K172999 · Shanghai United Imaging Healthcare Co., Ltd. · LLZ · Nov 7, 2018 · Radiology
Device Facts
Record ID
K172999
Device Name
uWS-MR
Applicant
Shanghai United Imaging Healthcare Co., Ltd.
Product Code
LLZ · Radiology
Decision Date
Nov 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
uWS-MR is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications: The MR Stitching is intended to create full-format images from overlapping MR volume data sets acquired at multiple stages. The Dynamic application is intended to provide a general post-processing tool for time course studies. The Image Fusion application is intended to combine two different image series so that the displayed anatomical structures match in both series. MRS (MR Spectroscopy) is intended to evaluate the molecule constitution and spatial distribution of cell metabolism. It provides a set of tools to view, process, and analyze the complex MRS data. This application supports the analysis for both SVS (Single Voxel Spectroscopy) and CSI (Chemical Shift Imaging) data. The MAPs application is intended to provide a number of arithmetic and statistical functions for evaluating dynamic processes and images. These functions are applied to the grayscale values of medical images. The MR Breast Evaluation application provides the user a tool to calculate parameter maps from contrast-enhanced time-course images. The Brain Perfusion application is intended to allow the visualization of temporal variations in the dynamic susceptibility time series of MR datasets. MR Vessel Analysis is intended to provide a tool for viewing, manipulating, and evaluating MR vascular images. The Inner view application is intended to perform a virtual camera view through hollow structures (cavities), such as vessels.
Device Story
uWS-MR is a standalone software solution for processing, reviewing, and analyzing MR images. Input: DICOM 3.0 images from network or external storage (CD/DVD/flash). Operation: Software performs 2D/3D visualization (MIP, VR, CPR, MPR), image fusion, and advanced post-processing (MRS, dynamic perfusion, vessel analysis, breast evaluation). Output: Processed images, parameter maps, and statistical data (mean/max/min values). Used in clinical settings by trained, licensed physicians to interpret images and statistics. Clinical decision-making is supported by providing quantitative and qualitative analysis of anatomical and functional MR data. Benefits include enhanced visualization and standardized analysis of complex MR datasets, aiding in diagnostic evaluation.
Clinical Evidence
No clinical studies were required. Evidence consists of bench testing and performance verification for all software modules, including MR Dynamic, Breast, Stitching, Brain Perfusion, Spectroscopy, Fusion, Inner View, Vessel, and MAPs. Software verification and validation included hazard analysis, requirements specification, and architecture description. The device was found to have a safety and effectiveness profile similar to the predicate device.
Technological Characteristics
Standalone software running on Windows 7. Supports DICOM 3.0 network protocols (TCP/IP). Features include 2D/3D visualization (MIP, VR, CPR, MPR), ROI/VOI analysis, and automated image registration. Complies with NEMA PS 3.1-3.20 (DICOM), ISO 14971 (risk management), and IEC 62304 (software life cycle).
Indications for Use
Indicated for viewing, manipulation, communication, and storage of medical images for interpretation and evaluation within healthcare institutions. Supports MR imaging applications including stitching, dynamic time-course studies, image fusion, spectroscopy (SVS/CSI), arithmetic/statistical mapping, breast evaluation, brain perfusion, vessel analysis, and virtual endoscopy (inner view).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
November 7, 2018
Shanghai United Imaging Healthcare Co., Ltd. % Shumei Wang QM&RA VP No.2258 Chengbei Road, Jiading District Shanghai, 201807 CHINA
Re: K172999
Trade/Device Name: uWS-MR Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: October 9, 2018 Received: October 10, 2018
Dear Shumei Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hol 2. Nils
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172999
Device Name uWS-MR
Indications for Use (Describe)
uWS-MR is a software solution intended to be used for viewing, manipulation, and storage of medical images. It supports interpretation and evaluations within healthcare institutions. It has the following additional indications:
The MR Stitching is intended to create full-format images from overlapping MR volume data sets acquired at multiple stages.
The Dynamic application is intended to provide a general post-processing tool for time course studies.
The Image Fusion application is intended to combine two different image series so that the displayed anatomical structures match in both series.
MRS (MR Spectroscopy) is intended to evaluate the molecule constitution and spatial distribution of cell metabolism. It provides a set of tools to view, process, and analyze the complex MRS data. This application supports the analysis for both SVS (Single Voxel Spectroscopy) and CSI (Chemical Shift Imaging) data.
The MAPs application is intended to provide a number of arithmetic and statistical functions for evaluating dynamic processes and images. These functions are applied to the grayscale values of medical images.
The MR Breast Evaluation application provides the user a tool to calculate parameter maps from contrast-enhanced timecourse images.
The Brain Perfusion application is intended to allow the visualizations in the dynamic susceptibility time series of MR datasets.
MR Vessel Analysis is intended to provide a tool for viewing, manipulating MR vascular images. The Inner view application is intended to perform a virtual camera view through hollow structures (cavities), such as vessels.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol, which is dark gray and has a white horizontal line through the middle. The overall design is clean and modern.
K172999
## 510 (k) SUMMARY
- 1. Date of Preparation: March 26, 2018
### 2. Sponsor Identification
## Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Establishment Registration Number: Not yet registered or the Number
Contact Person: Shumei Wang Position: QM&RA VP Tel: +86-021-67076888-6776 Fax: +86-021-67076889 Email: shumei.wang@united-imaging.com
### 3. Identification of Proposed Device
Trade Name: uWS-MR Common Name: MR Image Post-Processing Software Model(s): uWS-MR
Regulatory Information Classification Name: Picture archiving and communications system Classification: II Product Code: LLZ Regulation Number: 21 CFR 892.2050 Review Panel: Radiology
### 4. Identification of Predicate Device(s)
Predicate Device 510(k) Number: K123920 Device Name: syngo.via
Reference Device#1 510(k) Number: K151353 Device Name: Syngo.MR Neurology, Syngo.MR Oncology
Reference Device#2 510(k) Number: K130749 Device Name: syngo.MR General, syngo.MR Cardiology and syngo.MR Vascular
## SECTION 2 - 2 of 11
{4}------------------------------------------------
Reference Device#3 510(k) Number: K120315 Device Name: syngo.MR Spectroscopy
## Reference Device#4
510(k) Number: K133401 Device Name: syngo.MR Neurology, syngo.MR Oncology, syngo.MR BreVis and syngo.mMR General
Reference Device#5 510(k) Number: K960265 Device Name: Advantage Windows with FuncTool Option
Reference Device#6 510(k) Number: K090546 Device Name: Nordic Image Control and Evaluation (nordicICE) Software
Reference Device#7 510(k) Number: K081985 Device Name: AW Server
#### 5. Device Description
uWS-MR is a comprehensive software solution designed to process, review and analyze MR (Magnetic Resonance Imaging) studies. It can transfer images in DICOM 3.0 format over a medical imaging network or import images from external storage devices such as CD/DVDs or flash drives. These images can be functional data, as well as anatomical datasets. It can be at one or more time-points or include one or more time-frames. Multiple display formats including MIP and volume rendering and multiple statistical analysis including mean, maximum and minimum over a user-defined region is supported. A trained, licensed physician can interpret these displayed images as well as the statistics as per standard practice.
### 6. Indications for use
uWS-MR is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:
- The MR Stitching is intended to create full-format images from overlapping MR volume data sets acquired at multiple stages.
- . The Dynamic application is intended to provide a general post-processing tool for time course studies.
- . The Image Fusion application is intended to combine two different image series so that the displayed anatomical structures match in both series.
## SECTION 2 - 3 of 11
{5}------------------------------------------------
- . MRS (MR Spectroscopy) is intended to evaluate the molecule constitution and spatial distribution of cell metabolism. It provides a set of tools to view, process, and analyze the complex MRS data. This application supports the analysis for both SVS (Single Voxel Spectroscopy) and CSI (Chemical Shift Imaging) data.
- The MAPs application is intended to provide a number of arithmetic and statistical functions for evaluating dynamic processes and images. These functions are applied to the grayscale values of medical images.
- . The MR Breast Evaluation application provides the user a tool to calculate parameter maps from contrast-enhanced time-course images.
- . The Brain Perfusion application is intended to allow the visualization of temporal variations in the dynamic susceptibility time series of MR datasets.
- . MR Vessel Analysis is intended to provide a tool for viewing, manipulating, and evaluating MR vascular images.
- . The Inner view application is intended to perform a virtual camera view through hollow structures (cavities), such as vessels.
### 7. Summary of Technological Characteristics
uWS-MR comprises the post-processing applications for viewing, manipulation, 3D visualization, post-processing and comparison of MR images. After importing the DICOM image data into the system, the operator is able to perform image browsing and processing and can further obtain advanced information. This is identical to the predicate device.
The following tables compare the main features, principles of operation, fundamental scientific technology and intended use of uWS-MR when compared to the predicate devices.
### 8. Substantially Equivalent (SE) Comparison
The proposed device has similar indications for use and overall function and performs in a similar manner with respect to, display, review and processing applications with predicate device and reference devices.
## SECTION 2 - 4 of 11
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern.
| Item | Proposed Device<br>uWS-MR | Predicate Device<br>syngo.via(K123920) | Remark |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| General | | | |
| Device<br>Classification<br>Name | Picture Archiving and<br>Communications System | Picture Archiving and<br>Communications System | Same |
| Product Code | LLZ | LLZ | Same |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Device Class | II | II | Same |
| Classification<br>Panel | Radiology | Radiology | Same |
| Specification | | | |
| Image<br>communication | Standard network protocols like<br>TCP/IP and standard<br>communication protocol<br>DICOM. Additional fast image. | Standard network protocols<br>like TCP/IP and standard<br>communication protocol<br>DICOM. Additional fast<br>image. | Same |
| Hardware /OS | Windows 7 | Windows 7 | Same |
| Patient<br>Administration | Display and manage the image<br>data information of all patients<br>stored in the database. | Display and manage the image<br>data information of all patients<br>stored in the database. | Same |
| Review 2D | 2D Review consists of basic<br>processing of the 2D images,<br>e.g. rotation, scaling,<br>translation, windowing, and<br>measurements. | 2D Review consists of basic<br>processing of the 2D images,<br>e.g. rotation, scaling,<br>translation, windowing, and<br>measurements. | Same |
| Review 3D | 3D Review consist of<br>functionalities for displaying<br>and processing image in the 3D<br>form, e.g. VR, CPR, MPR,<br>MIP, etc. The module also<br>includes tools for VOI analysis. | 3D Review consist of<br>functionalities for displaying<br>and processing image in the 3D<br>form, e.g. VR, CPR, MPR,<br>MIP, etc. The module also<br>includes tools for VOI analysis. | Same |
| Filming | Filming is a module dedicated<br>for image printing. The print<br>tools provide layout editing for<br>both single images and series. | Filming is a module dedicated<br>for image printing. The print<br>tools provide layout editing for<br>both single images and series. | Same |
| Fusion | Including Auto registration,<br>Manual registration, Spot<br>registration | Including Auto registration,<br>Manual registration, Spot<br>registration | Same |
| Inner View | Inner view the vessel , colon ,<br>trachea | Inner view the vessel , colon ,<br>trachea | Same |
| Visibility | User-defined the display<br>property of fused image:<br>Adjustment of preset of T/B<br>value; Adjustment of the fused | User-defined the display<br>property of fused image:<br>Adjustment of preset of T/B<br>value; Adjustment of the fused | Same |
| Item | Proposed Device<br>uWS-MR | Predicate Device<br>syngo.via(K123920) | Remark |
| ROI/VOI | Plotting ROI or VOI, and<br>obtaining the maximum activity<br>value, the minimum activity<br>value, mean activity value, the<br>volume/area of region, and the<br>maximum diameter of volume,<br>peak activity value; | Plotting ROI or VOI, and<br>obtaining the maximum<br>activity value, the minimum<br>activity value, mean activity<br>value, the volume/area of<br>region, and the maximum<br>diameter of volume, peak<br>activity value; | Same |
| MIP Display | The image can be displayed as<br>MIP and rotating play | The image can be displayed as<br>MIP and rotating play | Same |
| Compare | Load two studies to compare. | Load two studies to compare. | Same |
# SECTION 2 - 5 of 11
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED" is stacked on top of the text "IMAGING" in a bold, sans-serif font. To the right of the text is a stylized, dark teal-colored symbol that resembles a shield or a stylized letter "U" with a vertical line running through the center.
## Table 2 SE Discussion for advanced applications
| Application | Function name | Proposed<br>device<br>uWS-MR | Reference<br>Device#1<br>Syngo.MR<br>Neurology,<br>Syngo.MR<br>Oncology<br>(K151353) | Reference<br>Device#6<br>Nordic Image<br>Control and<br>Evaluation<br>(nordicICE)<br>Software<br>(K090546) | Remark |
|-----------------------|-------------------------------------------------------------------------------|-------------------------------|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------|
| MR Brain<br>Perfusion | Type of imaging<br>scans | MRI | MRI | / | Same |
| | Intended Body part | MR<br>neurological<br>images. | MR<br>neurological<br>images. | / | Same |
| | Rectify motion | Yes | Yes | / | Same |
| | Remove background | Yes | Yes | / | Same |
| | Arterial Select:<br>Automatically or<br>manually select<br>arterial location. | Yes | Yes | / | Same |
| | Parametric Mapping<br>Calculation | Yes | / | Yes | Same |
| | TIC Analysis | Yes | Yes | / | Same |
| | Save Image Data | Yes | Yes | / | Same |
| | Print Image Data | Yes | Yes | / | Same |
| | Report | Yes | Yes | / | Same |
# SECTION 2 - 6 of 11
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, with a white line running vertically through the center. The logo is simple and modern, and the use of bold font and a strong color makes it easily recognizable.
| Application | Function name | Proposed device | Reference Device#4 | Reference Device#5 | Remark |
|----------------------|-------------------------------------|-----------------|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------|
| MR Breast Evaluation | Type of imaging scans | MRI | MRI | MRI | Same |
| | Intended Body part | Breast | Breast | Breast | Same |
| | Automatic Subtraction | Yes | Yes | Yes | Same |
| | Motion Correction | Yes | Yes | Yes | Same |
| | TIC Analysis | Yes | Yes | Yes | Same |
| | Background Removal | Yes | Yes | Yes | Same |
| | TIC Analysis | Yes | Yes | Yes | Same |
| | Automatic Parameter Map Calculation | Yes | Yes | Yes | Same |
| | Save Image Data | Yes | Yes | Yes | Same |
| | Print Image Data | Yes | Yes | Yes | Same |
| | Report | Yes | Yes | Yes | Same |
| | | | syngo.MR<br>Neurology,<br>syngo.MR<br>Oncology,<br>syngo.MR<br>BreVis and<br>syngo.mMR<br>General<br>(K133401) | Advantage<br>Windows<br>with<br>FuncTool<br>Option<br>(K960265) | |
| Application | Function name | Proposed<br>device<br>uWS-MR | Reference Device#2<br>syngo.MR General,<br>syngo.MR Cardiology and<br>syngo.MR Vascular(K130749) | Remark |
|--------------|----------------------------|------------------------------|--------------------------------------------------------------------------------------------------|--------|
| MR Stitching | Type of imaging<br>scans | MRI | MRI | Same |
| | Intended Body<br>part | WholeBody | WholeBody | Same |
| | Automatic Stitching | Yes | Yes | Same |
| | Manual Stitching | Yes | Yes | Same |
| | Homogenous<br>Modification | Yes | Yes | Same |
| | Sharp/Smooth | Yes | Yes | Same |
| | Save Image Data | Yes | Yes | Same |
| | Print Image Data | Yes | Yes | Same |
| | Report | Yes | Yes | Same |
# SECTION 2 - 7 of 11
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and it is likely used to represent the company's brand.
| Application | Function name | Proposed<br>device<br>uWS-MR | Reference Device#2<br>syngo.MR General,<br>syngo.MR<br>Cardiology and<br>syngo.MR Vascular<br>(K130749) | Reference<br>Device#7<br>AW<br>SERVER(K<br>081985) | Remark |
|-----------------------|------------------------------------------------------|------------------------------|---------------------------------------------------------------------------------------------------------|----------------------------------------------------|--------|
| MR Vessel<br>Analysis | Type of imaging<br>scans | MRI | MRI | / | Same |
| | Intended Body<br>part | WholeBody | WholeBody | / | Same |
| | Automatic blood<br>vessel center lines<br>extraction | Yes | Yes | / | Same |
| | Vascular stenosis<br>analysis: | Yes | / | Yes | Same |
| | Vascular display<br>with<br>MPR\CPR\VR\MIP | Yes | Yes | / | Same |
| | Save Image Data | Yes | Yes | / | Same |
| | Print Image Data | Yes | Yes | / | Same |
| | Report | Yes | Yes | / | Same |
| Application | Function name | Proposed<br>device | Reference Device#3<br>syngo.MR Spectroscopy | Remark |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-----------------------------------------------------------------|--------|
| | | uWS-MR | (K120315) | |
| MRS | Type of imaging scans | MRI | MRI | Same |
| | Intended Body part | Brain | Brain | Same |
| | Single-voxel Spectrum Data Analysis | Yes | Yes | Same |
| | Chemical Shift Imaging Data | Yes | Yes | Same |
| | Protocol Management: | Yes | Yes | Same |
| | Statistics Analysis | Yes | Yes | Same |
| | Metabolite Pseudo-color Map: | Yes | Yes | Same |
| | Curve Peak Setting | Yes | Yes | Same |
| | Save Image Data | Yes | Yes | Same |
| | Print Image Data | Yes | Yes | Same |
| | Report | Yes | Yes | Same |
| | | uWS-MR | Advantage<br>Windows<br>with<br>FuncTool<br>Option<br>(K960265) | |
| MR | ADC and eADC Calculate | Yes | Yes | Same |
| Dynamic | T1, T2, T2*/R2* Calculate | Yes | Yes | Same |
| MR MAPS | Background Removal | Yes | Yes | Same |
| | ROI Curves Analysis: Plot interested region,<br>including correlated or non-correlated,<br>symmetric or asymmetric, and get mean time-<br>intensity curve | Yes | Yes | Same |
| | Statistic Table: Produce statistical table for<br>parameters, and freely select to display items<br>in demand. | Yes | Yes | Same |
| | DCE (Dynamic Contrast Enhance)<br>customizes background pixel range<br>calculation (Support to modify manually), and<br>automatically provides parameter value, end-<br>stage scanning value (Support to modify by<br>need). The relative Parameter Maps includes<br>PEI (Positive Enhanced Integration), TTP<br>(Time to Peak), MSI (Maximum Slope of<br>Increasing) and MTE (Mean Time to<br>Enhance). | Yes | Yes | Same |
| | DSC (Dynamic Susceptibility Contrast-<br>enhance) customizes background pixel range<br>calculation (Support to modify manually), and<br>automatically provides parameter value, end-<br>stage scanning value (Support to modify by<br>need). The relative Parameter Maps includes<br>NEI (Negative Enhanced Integration), TTP<br>(Time to Peak), MSD (Maximum Slop of<br>Descending) and MTE (Mean Time to<br>Enhance) | Yes | Yes | Same |
| | Save Image Data | Yes | Yes | Same |
| | Print Image Data | Yes | Yes | Same |
| | Report | Yes | Yes | Same |
| Application | Function name | Proposed device | Reference Device#5 | Remark |
|-------------|---------------|-----------------|--------------------|--------|
|-------------|---------------|-----------------|--------------------|--------|
# SECTION 2 - 8 of 11
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED IMAGING" are stacked on top of each other in bold, sans-serif font. To the right of the words is a stylized "U" symbol, which is made up of two vertical bars and a horizontal bar in the middle.
### 9. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
# SECTION 2 - 9 of 11
{11}------------------------------------------------
## Biocompatibility
Not Applicable to the proposed device, because the device is stand-alone software.
## Electrical Safety and Electromagnetic Compatibility (EMC)
Not Applicable to the proposed device, because the device is stand-alone software.
## Software Verification and Validation
Software verification and validation testing was provided to demonstrate safety and efficacy of the proposed device. This includes a hazard analysis, and the potential hazards have been classified as a moderate level of concern (LOC). These documentations include:
- Software description
- · Hazard Analysis
- · Software requirements specification (SRS)
- · Software Architecture Description
- · Software Development Environment Description
- · Software Verification and Validation
- · Cyber security Documents
## Animal Study
No animal study was required.
### Clinical Studies
No clinical study was required.
### Performance Verification
- . Performance Evaluation For MR Dynamic
- Performance Evaluation For MR Breast
- Performance Evaluation For MR Stitching
- Performance Evaluation For MR Brain Perfusion
- . Performance Evaluation For MR Spectroscopy
- . Performance Evaluation For MR Fusion
- . Performance Evaluation For MR Inner View
- Performance Evaluation For MR Vessel
- . Performance Evaluation For MR Maps
### Product Standards and Guidance
- NEMA PS 3.1 - 3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2016).
- ISO 14971 Medical devices - Application of risk management to medical devices (Edition 2.0, corrected version, 2007).
## SECTION 2 - 10 of 11
{12}------------------------------------------------
- IEC 62304 Medical device software Software life cycle processes (Edition 1.1, ● 2015).
## Summary
The features described in this premarket submission are supported with the results of the testing mentioned above, the uWS-MR was found to have a safety and effectiveness profile that is similar to the predicate device.
## 10. Substantially Equivalent (SE) Conclusion
The proposed device is equivalent to the predicate device with regard to safety and efficacy. This conclusion is based upon a comparison of intended use, technological characteristics, performance specification, device hazards as well as verification and validation results.
In summary, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
## SECTION 2 - 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.