The Electric Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Electric Breast Pump is intended for a single user.
Device Story
Personal-use electric breast pump; single-user; powered by internal batteries or AC adapter. Device uses diaphragm-type vacuum pump driven by DC motor; controlled by microcontroller. Two modes: Stimulation (fast cycles, low vacuum) to initiate flow; Expression (slow cycles, high vacuum) for milk collection. Components include motor, breast shields, tubing, bottle, and valve. Operated by lactating women in home environment. User adjusts suction levels via interface; device provides visual feedback via LCD. Benefits include efficient, gentle milk expression.
Clinical Evidence
Bench testing only. Testing included vacuum pressure range, maximum vacuum, backflow-leakage, and battery life. Compliance verified for AAMI/ANSI ES60601-1, IEC 60601-1-11, and IEC 60601-1-2. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization/irritation).
Technological Characteristics
Materials: silicone (breast shield), polypropylene (body/bottle). Diaphragm-type vacuum pump; DC-motor; microcontroller-based control. Power: 4 AA batteries or AC adapter (DC6.0V, 1000mA). Vacuum range: -60 to -250 mmHg. LCD interface. Non-sterile. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two distinct elements: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2017
Joytech Healthcare Co., Ltd. Ren Yunhua General manager No.365, Wuzhou Road, Yuhang Economic Development Zone Hangzhou City, Zhejiang 311100 China
K172989 Re: Trade/Device Name: Electric Breast Pump (Models LD-202 and LD-213) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: September 15, 2017 Received: September 27, 2017
Dear Ren Yunhua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the text "Charles Viviano -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. The letters are evenly spaced, and the overall impression is clean and legible.
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172989
Device Name
Electric Breast Pump (Models LD-202 and LD-213)
## Indications for Use (Describe)
The Electric Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Electric Breast Pump is intended for a single user.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# JOYTECH HEALTHCARE CO. LTD.
## 510(k) Summary
## K172989
## 1. Date Prepared: 2017.12.19
## 2. Submitter's Identification:
Name: JOYTECH HEALTHCARE CO., LTD.
- Address: No.365, Wuzhou Road, Yuhang Economic Development Zone,Hangzhou City, 311100 Zhejiang,China
Contact Person: Yunhua Ren
Phone: +86-571-81957767
Fax: +86-571-81957750
RENYH@SEJOY.COM Email:
## 3. Name of the Device:
Trade Name: Electric Breast Pump (Models LD-202, LD-213)
## 4. Classification Information:
| Common Name: | Electric Breast Pump |
|----------------------|--------------------------|
| Classification name: | Powered breast pump |
| Regulation number: | 884.5160 |
| Product Code: | HGX, Powered breast pump |
| Device Class: | II |
| Panel: | Obstetrics/Gynecology |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for JOYTECH. The logo is in blue and red. To the right of the word JOYTECH is Chinese characters.
## JOYTECH HEALTHCARE CO. LTD.
#### 5. Predicate Device Information:
Medela Swing Breast Pump (K053052); Medela AG
The predicate device has not been subject to a design-related recall.
### 6. Indications for Use:
The Electric Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The electric breast pump is intended for a single user.
## 7. Device Description:
The Electric Breast Pump is a personal use electric breast pump capable of single pumping. The device is electrically powered from either internal batteries or an external AC powered supply and is provided non-sterile.
The Electric Breast Pump consists of the motor, breast shields, tubing, bottle and valve. The materials used are silicone (breast shield) and polypropylene (Pump body and bottle).
The electric breast pump's drive unit employs a diaphragm-type vacuum pump, powered by a DC-motor, supervised by a microcontroller. The microcontroller provides control over motor speed (vacuum creation) and solenoid (vacuum release).
The device has two modes of operation:
- Stimulation mode: Suction patterns with fast cycles and low vacuum to start ● milk flowing
- . Expression mode: Suction patterns with slow cycles and high vacuum to express more milk gently and efficiently.
| | Subject Device<br>K172989 | Predicate<br>device<br>K053052 | Comparison |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|------------|
| Indication for use | A powered breast pump<br>to be used by lactating<br>women to express and<br>collect milk from<br>their breasts. | A powered breast pump<br>to be used by lactating<br>women to express and<br>collect milk from<br>their breasts. | Identical |
| Patient Population | Lactating women | Lactating women | Identical |
| Environment of Use | Home | Home or Hospital | Similar |
| Single/double pump | Single pump | Single pump | Identical |
| Re-usable | Yes | Yes | Identical |
| Direct user contact | Yes | Yes | Identical |
| Adjustable<br>suction<br>levels | Yes | Yes | Identical |
## 8. Predicate Device Comparison:
{5}------------------------------------------------
| | JOYTECH HEALTHCARE CO. LTD. | | |
|-----------------------------------------|-------------------------------------------------------------|------------------------------------------------|-----------|
| Visual indicator | LCD Screen | LED | Similar |
| Vacuum range<br>(mmHg) | -60~-250mmHg | -50~-250mmHg | Similar |
| Cycling/Suction<br>Control<br>Mechanism | Microprocessor | Microprocessor | Identical |
| Breastshield size | 24mm or 27mm | 21mm,24mm,27mm,30mm<br>or 36mm | Similar. |
| Cleaning method for<br>Accessories | Soap and warm water | Soap and warm water | Identical |
| Battery type | 4 AA batteries or Medical<br>AC Adapter (DC6.0V,<br>1000mA) | 4 AA batteries or<br>AC Adapter (DC4.8V, 1.2A) | Similar |
| Battery use life | Approx.1.5h for pumping time | Approx.1.5h for<br>pumping time | Identical |
| Low<br>battery<br>indicator | Yes | Yes | Identical |
| Automatic<br>Power-<br>Off | Yes | Yes | Identical |
The Electric Breast Pump has the same intended use but minor differences in technological characteristics compared to the predicate device. These minor differences in technological characteristics do not raise different questions of safety and effectiveness.
## 9. Performance data
The following performance data were provided in support of the substantial equivalence determination.
Performance testing was conducted to validate and verify that the Electric Breast Pump met all requirements of the applicable standards, including electrical safety, EMC, and biocompatibility. Results of these tests demonstrate compliance to the following standards:
Electrical Safety and performance requirements:
- AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 and A2:2010/(R)2012 Medical Electrical Equipment
Home-used medical equipment requirements and environmental test:
- · IEC 60601-1-11:2015 General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
Electromagnetic compatibility requirements:
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for . basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Biocompatibility Evaluation for patient contacting components:
- Cytotoxicity ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests . for in vitro cytotoxicity
- Sensitization ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: . Tests for irritation and skin sensitization
{6}------------------------------------------------
# JOXTEC-11116
## JOYTECH HEALTHCARE CO. LTD.
- Irritation ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for . irritation and skin sensitization
## Software Validation
The software/firmware verification and validation was provided in accordance with the FDA Guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.
## Bench Performance Testing
The following bench testing was completed on the subject device:
- . Vacuum pressure range (suction)
- Maximum vacuum pressure .
- Backflow-leakage testing .
- Battery life testing .
## 10. Conclusions:
The performance testing provided is sufficient to support substantial equivalence determination. Based on the information provided in this submission, the Joytech Electric Breast Pump is as safe and effective as the predicate Medela Swing Breast Pump.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.