The devices are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive oximeter equipment. The sensors are reusable or disposable depending on models, and are intended for adult, pediatric or infant, depending on models. They shall be used in healthcare settings.
Device Story
SpO2 sensors function as accessories to non-invasive oximeters; measure arterial oxygen saturation and pulse rate. Input: red and infrared light absorption through patient finger; output: electrical signal forwarded to host oximeter. Operation: two LEDs (red/infrared) emit light through tissue; photodiode detects transmitted light; host oximeter amplifies signal and calculates ratio of oxygenated hemoglobin to total hemoglobin. Used in healthcare settings by clinicians for patient monitoring. Benefits: provides non-invasive, real-time physiological data to support clinical decision-making regarding patient oxygenation status.
Clinical Evidence
Clinical hypoxia testing performed on human adult volunteers. Accuracy of the proposed device validated against arterial oxygen saturation (SaO2) measured via co-oximetry. Results support accuracy claims within the specified saturation range.
Technological Characteristics
Surface-contact device (limited contact <24h). Sensing principle: 2-wavelength optical absorption (Red 660nm, IR 940nm). Components: two LEDs and photodiode. Biocompatibility per ISO 10993-5/10. Electrical safety/EMC per IEC 60601-1, IEC 60601-1-2, and ISO 80601-2-61. Non-sterile.
Indications for Use
Indicated for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in adult, pediatric, or infant patients in healthcare settings using compatible non-invasive oximeter equipment.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
June 8, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
APK Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K172981
Trade/Device Name: Reusable SpO2 Sensors, Disposable SpO2 Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: April 26, 2018 Received: April 30, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K172981
Device Name
Reusable SpO2 Sensors, Disposable SpO2 Sensors
Indications for Use (Describe)
The devices are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive oximeter equipment. The sensors are reusable depending on models, and are intended for adult, pediatric or infant, depending on models. They shall be used in healthcare settings.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right: 20px;"> <span style="border: 1px solid black; padding: 0px 2px;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="border: 1px solid black; padding: 0px 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
{3}------------------------------------------------
## Exhibit #2 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K172981
- 1. Date of Preparation: 03/26/2018
- 2. Sponsor
### APK Technology Co., Ltd.
6Floor, B2 Building, Industry of Hengfeng, Hezhou, Xixiang, Bao'an Distric, Shenzhen, China
Establishment Registration Number: 3007699081
Contact Person: Caifang Wang Position: Management Representative Tel: +86-755-27325581 Fax: +86-755-27325585 Email: caifang@apk-technology.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
#### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
- 4. Identification of Proposed Device
Trade Name: Reusable SpO2 Sensors Disposable SpO2 Sensors
{4}------------------------------------------------
Common Name: Oximeter (Accessory-sensor)
Regulatory Information Classification Name: Oximeter Classification: Class II Product Code: DQA Regulation Number: 21 CFR 870.2700 Review Panel: Anesthesiology
Intended Use Statement:
The devices are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in non-invasive oximeter equipment. The sensors are reusable or disposable depending on models, and are intended for adult, pediatric or infant, depending on models. They shall be used in healthcare settings.
#### Device Description
The proposed device, SpO2 Sensors are accessories to the oximeters, which are intended for spot checking or continuous monitoring of functional arterial oxygen saturation and pulse rate in noninvasive U.S. legally marketed oximeters or patient monitors. The sensors are reusable or disposable depending on models, and are intended for adult, pediatric or infant, depending on models.
The proposed devices measure, along with the corresponding oximeters, non-invasively, the arterial oxygen saturation of blood. The measurement method is based on the red and infrared light absorption of hemoglobin and oxyhemoglobin. Each sensor has two LEDs, emitting both red and infrared light, and a photodiode. The light is emitted through human finger and received by a photodiode. Then the received signal is forwarded to the corresponding oximeter that amplifies the signal and an algorithm that calculates the ratio. The saturation value is determined by the percentage ratio of the oxygenated hemoglobin (HbO2) to the total amount of hemoglobin (Hb).
The patient contact components include cable and sensor. The contact level is surface device and contact duration is limited contact (less than 24h).
#### న్. Identification of Predicate Device
510(k) Number: K153184 Device Name: Caremed Reusable & Disposable SpO2 Sensors
- 6. Non-Clinical Test Conclusion
{5}------------------------------------------------
Non clinical tests were conducted to verify that the proposed device met all design specifications and to demonstrate Substantially Equivalence(SE) to the predicate device. The test include following items
The performance tests performed on proposed devices include following items
- > Pulse and SpO2 Accuracy test
- > Pulse and SpO2 Accuracy test after High and Low Temperature
- > Pulse and SpO2 Accuracy test after Maximum Times Disinfection
Biocompatibility tests were conducted per ISO 10993-1, the test items include
- ▲ Cytotoxicity Test
- > Skin Irritation Test
- > Skin Sensitization Test
Electrical safety and electromagnetic compatibility test were conducted on proposed devices per IEC 60601-1, IEC 60601-1-2 and ISO 80601-2-61.
- 7. Clinical Test Conclusion
Clinical testing has been performed under an approved protocol with subject informed consent. Clinical hypoxia test results were obtained in human adult volunteers to validate the accuracy of proposed device versus arterial oxygen saturation (SaO2) as determined by co-oximetry. Clinical test results support device accuracy claims for the specified saturation range.
- 8. Substantially Equivalent (SE) Comparison
| ITEM | Proposed device | Predicate Device |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K153184 |
| Product Code | DQA | DQA |
| Regulation<br>Number | 21 CFR 880.2910 | 21 CFR 880.2910 |
| Intended Use | The devices are intended for spot<br>checking or continuous monitoring of<br>functional arterial oxygen saturation<br>and pulse rate in non-invasive<br>oximeter equipment. The sensors are<br>reusable or disposable depending on<br>models, and are intended for adult,<br>pediatric or infant, depending on | Caremed Reusable & Disposable<br>SPO2 Sensors are indicated for<br>continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate(PR)<br>for adult patients weighing greater than<br>40 kg and pediatric patients weighing<br>10-50 kg at hospital facilities |
### Table 1 Comparison of Technology Characteristics
{6}------------------------------------------------
| | models. They shall be used<br>in healthcare settings. | |
|----------------------------|-------------------------------------------------------|---------------------------------------------|
| Principle of<br>Operation | 2-wavelength Relative Optical<br>Absorption | 2-wavelength Relative Optical<br>Absorption |
| Light Emitting | Red: 660nm<br>Ired: 940nm | Red:660-666nm<br>Ired:880-950nm |
| Signal Detection<br>Method | Photodetector | Photodetector |
| SpO2 Range | 70-100% | 70-100% |
| SpO2 Accuracy | $\pm$ 3% @ 70-100% | $\pm$ 3% @ 70-100% |
| PR Range | 30-245 bpm | 30-250 bpm |
| PR Accuracy | $\pm$ 3 bpm | $\pm$ 3 bpm |
| Sterile | No | No |
| Usage | Reusable & Disposable | Reusable & Disposable |
| EMC | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 |
| Performance | Complied with ISO 80601-2-61 | Complied with ISO 80601-2-61 |
| Biocompatibility | | |
| Cytotoxicity | Complied with ISO 10993-5 | Complied with ISO 10993-5 |
| Skin Irritation | Complied with ISO 10993-10 | Complied with ISO 10993-10 |
| Sensitization | Complied with ISO 10993-10 | Complied with ISO 10993-10 |
- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.