K172966 · Advanced Instumentations, Inc. · MHX · Dec 8, 2017 · Cardiovascular
Device Facts
Record ID
K172966
Device Name
CMS-2000 Central Monitoring System
Applicant
Advanced Instumentations, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Dec 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The CMS-2000 Central Monitoring System provides centralized monitoring and critical care management for patients monitored by bedside monitors. From the CMS-2000, clinicians can gain access to patient information for patients on the Network. The CMS-2000 displays waveforms, parameters and alarm status of bedside monitors for up to 32 patients on a single screen or up to 64 patients using two screens.
Device Story
CMS-2000 Central Monitoring System; software-based platform running on Windows XP/7. Inputs: physiological waveforms, parameters, and alarm data from bedside monitors via network protocol. Operation: collects, displays, and archives patient data; enables bidirectional control of bedside monitors (alarm limits, NIBP measurements). Output: centralized display of up to 32-64 patients; printed reports; historical data review. Used in clinical settings by healthcare providers for critical care management. Benefits: enables remote oversight of multiple patients, facilitating timely clinical response to alarm status and parameter changes.
Clinical Evidence
No clinical data; bench testing only. Quality assurance included software testing, risk analysis, safety testing, and performance testing.
Technological Characteristics
Software-based system; PC platform (Windows XP/7). Connectivity: hospital LAN, HL7 support. Features: bidirectional control, multi-patient display (up to 64), trend/waveform/alarm storage. No specific hardware materials or sterilization required as it is a software-based monitoring system.
Indications for Use
Indicated for centralized monitoring and critical care management of patients connected to bedside monitors in a clinical network environment.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
EDAN Central Monitoring System, model MFM-CMS (K120727)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 8, 2017
Advanced Instrumentations, Inc. Jorge Millan Regulatory Affairs Manager 6800 NW 77th. Ct. Miami. Florida 33166
Re: K172966
Trade/Device Name: CMS-2000 Central Monitoring System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX Dated: August 22, 2017 Received: September 26, 2017
Dear Jorge Millan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
M.A. Wilhelm
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K172966
Device Name
CMS-2000 Central Monitoring System
Indications for Use (Describe)
The CMS-2000 Central Monitoring System provides centralized monitoring and critical care management for patients monitored by bedside monitors. From the CMS-2000, clinicians can gain access to patient information for patients on the Network. The CMS-2000 displays waveforms, parameters and alarm status of bedside monitors for up to 32 patients on a single screen or up to 64 patients using two screens.
본 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510K SUMMARY
CMS-2000 Central Monitoring System
ADVANCED INSTRUMENTATIONS, INC SUBMITTER 6800 NW 77th Ct, Miami, FL 33166 Phone: (305) 477-6331
US AGENT JORGE MILLAN, PHD REGULATORY AFFAIRS CORRESPONDENT Email: regulatory@advanced-inst.com Web: https://www.sigmabiomedical.com
#### DEVICE NAME AND CLASSIFICATION
| TRADE NAME: | CMS-2000 Central Monitoring System |
|-------------------------------------------|-----------------------------------------------------------------------------------|
| CLASSIFICATION NAME: | 870.1025 monitor, physiological, patient (with arrhythmia<br>detection or alarms) |
| | Product Code: MHX |
| REGULATORY CLASS:<br>PANEL IDENTIFICATION | Class II<br>Cardiovascular |
#### DEVICE DESCRIPTION
The CMS-2000 Central Monitoring System is a software production, which runs on a PC platform running under the Microsoft Windows XP or Windows 7 operating system. Through specified protocol, one CMS-2000 can connect with multi-monitors from ADVANCED INSTRUMENTATIONS to collect patients' information and monitoring data such as physiological waveforms, physiological parameters and alarms. The CMS-2000 can also send bidirectional control instruction to bedside monitors to change patients' information, alarm limits and conduct NIBP measurements. The bedside Patient Physiological Monitors have been cleared by the FDA under K123048 separately. The monitoring information collected by the CMS-2000 can be saved and printed. At the same time, the old records can be searched conveniently and quickly.
Indications for Use: The CMS-2000 Central Monitoring System provides centralized monitoring and critical care management for patients monitored by bedside monitors. From the CMS-2000, clinicians can gain access to patient information for patients on the Network. The CMS-2000 displays waveforms, parameters and alarm status of bedside monitors for up to 32 patients on a single screen or up to 64 patients using two screens.
{4}------------------------------------------------
Predicate Devices: The CMS-2000 Central Monitoring System is equivalent to the EDAN Central Monitoring System, model MFM-CMS cleared under K120727.
#### Comparison with the Predicate Devices [21 CFR 807.92(a) (6)]
The CMS-2000 Central Monitoring System is comparable with and substantially equivalent to the EDAN Central Monitoring System, model MFM-CMS cleared under K120727. Compared with the MFM-CMS the Subject Device CMS-2000 has the same intended uses and functionality.
# Technical Characteristics Comparison:
The basic and main technical features of the subject device CMS-2000 are the same as the predicated device MFM-CMS including Design, Operation Controls, Display Modes and performance results. Table 1 provides the comparison of features and functionality:
| ITEM | PROPOSED DEVICE | PREDICATE DEVICE |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CMS-2000<br>K# TBD | MFM-CMS<br>K120727 |
| PRODUCT CODE | MHX | MHX |
| REGULATION NO. | 870.1025 | 870.1025 |
| CLASS | II | II |
| INTENDED USE | The CMS-2000 Central<br>Monitoring System provides<br>centralized monitoring and<br>critical care management for<br>patients monitored by<br>bedside monitors. From the<br>CMS-2000, clinicians can gain<br>access to patient information<br>for patients on the Network.<br>The CMS-2000 displays<br>waveforms, parameters and<br>alarm status of bedside<br>monitors for up to 32 patients<br>on a single screen or up to 64<br>patients using two screens. | The MFM-CMS Central<br>Monitoring System provides<br>centralized monitoring and<br>critical care management for<br>patients monitored by<br>bedside monitors. From the<br>MFM-CMS clinicians can gain<br>access to patient information<br>for patients on the Network.<br>The MFM-CMS displays<br>waveforms, parameters and<br>alarm status of bedside<br>monitors for up to 32<br>patients on a single screen<br>or up to 64 patients using<br>two screens. |
| WAVEFORMS | 2 ECG Waveforms<br>1 RESP waveform<br>1 PLETH waveform | 2 ECG Waveforms<br>1 RESP waveform<br>1 PLETH waveform |
{5}------------------------------------------------
| | 8 IBP waveform | 8 IBP waveform | | | |
|-----------------|----------------------------------|--------------------------------|--|--|--|
| | 1 C02 waveform | 1 CO2 waveform | | | |
| | 4 AG waveforms for CO2, 02, | 4 AG waveforms for CO2, 02, | | | |
| | N20 and AA | N20 and AA | | | |
| PARAMETERS | ECG: HR, ST value, PVCs | ECG: HR, ST value, PVCs | | | |
| | RESP: RR | RESP: RR | | | |
| | NIBP: SYS, DIA, MAP | NIBP: SYS, DIA, MAP | | | |
| | SP02: SP02, PR | SP02: SP02, PR | | | |
| | IBP: ART, PA, CVP, RAP, ICP, | IBP: ART, PA, CVP, RAP, ICP, | | | |
| | LAP, P1, P2 (only IBP | LAP, P1, P2 (only IBP | | | |
| | supported by the monitor will | supported by the monitor | | | |
| | be displayed) | will be displayed) | | | |
| | CO2: EtCO2, FiCO2, AwRR | CO2: EtCO2, FiCO2, AwRR | | | |
| | TEMP: T1, T2, TD | TEMP: T1, T2, TD | | | |
| | AG: EtCO2, FiCO2, AwRR, | AG: EtCO2, FiCO2, AwRR, | | | |
| | Et02, FiO2, EtN20, FuN20, | Et02, Fi02, EtN20, FuN20, | | | |
| | HAL/ISO/ENF/SEV/DES: Et, | HAL/ISO/ENF/SEV/DES: Et, | | | |
| | Fi, MAC | Fi, MAC | | | |
| | C.O: C.O., TB | C.O. C.O., TB | | | |
| DISPLAY | The<br>Central<br>Monitoring | The Central<br>Monitoring | | | |
| | supports one or two displays. | supports one or two | | | |
| | It can display up to 32 | displays. It can display up to | | | |
| | bedside monitors on one | 32 bedside monitors on one | | | |
| | and 64<br>bedside<br>display - | display and 64 bedside | | | |
| | displays<br>monitors on two | monitors on two displays | | | |
| | simultaneously | simultaneously | | | |
| RECORD CAPACITY | 240-hour trend data, 72 hour | 240-hour trend<br>data,<br>72 | | | |
| | waveform, 720 alarm events, | hour waveform, 720 alarm | | | |
| | 1~2 hours short trend, 720 | events, 1~2 hours<br>short | | | |
| | group<br>NIBP measurement | 720 group<br>trend,<br>NIBP | | | |
| | for each bedside<br>review | measurement review<br>for | | | |
| | monitor | each bedside monitor | | | |
| CALCULATION | Drug calculation and titration | calculation<br>Drug<br>and | | | |
| | table | titration table | | | |
| REVIEW | information,<br>Print<br>patient | Print patient information, | | | |
| | wave review, alarm review, | wave review, alarm review, | | | |
| | trend review, NIBP review, | trend review, NIBP review, | | | |
| | and drug calculation result | and drug calculation result | | | |
| ALARMS | Audible and visible alarms | Audible and visible alarms | | | |
{6}------------------------------------------------
| NETWORK/CONNECTIVITY | Web observation in the<br>hospital local area network<br>Bidirectional control<br>HL7 | Web observation in the<br>hospital local area network<br>Bidirectional control<br>HL7 |
|----------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
|----------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
## Clinical Test:
Clinical testing is not required.
## Non-clinical Test:
The following quality assurance measures were applied to the development of the CMS-2000 System:
- . Software testing
- Risk analysis
- . Safety testing
- Performance test
# Substantially Equivalent Determination
The subject device has similar technology characteristics and has the same intended use, same design principle and same functionality as the predicate device. There are no differences between the devices. In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, Advanced Instrumentations concludes that the CMS-2000 Central Monitoring System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.