The BW-X07HD Handheld VitalSigns Monitoring System is intended to be used in hospital, clinic, homecare, home and other medical settings where patient care is offered by qualified healthcare personnel who can use the device to monitor the patient's physical indexes, including ECG, SpO2, NIBP and TEMP as well as the pulse rate. For measuring ECG, SpO2 the intended patient population is adult, for measuring NIBP the intended patient population is both adult and pediatric aged more than 3 years old, whereas for measuring TEMP the intent population is all types of people including newborn, pediatric and adult. The device is used to collect, store, and transmit general patient health information and patient vital signs data between the patient and a health care professional.
Device Story
Handheld, wireless-connected vital signs monitor; measures ECG, SpO2, NIBP, TEMP, and pulse rate. Inputs: physiological sensors for ECG, SpO2, NIBP, and temperature. Device integrates embedded system with Android application; stores data locally; transmits via WiFi, 3G, or 4G. Includes HD video conferencing for communication between patient and healthcare professional. Used in hospitals, clinics, and home settings by healthcare personnel or patients. Output: vital signs data displayed on device and transmitted to healthcare professionals for clinical decision-making. Benefits: enables remote monitoring and data collection for patient health management.
Clinical Evidence
Clinical reports provided for NIBP, SpO2, and Thermometer functions. NIBP validation performed per ISO 81060-2:2013; SpO2 validation per ISO 80601-2-61:2011; Thermometer validation per ISO 80601-2-56:2009. Studies included adult and pediatric populations (NIBP >3 years; TEMP all ages). No adverse events or side effects reported.
Technological Characteristics
Handheld device with embedded system and Android application. Sensors: ECG, NIBP, SpO2, and infrared temperature. Power: 3.7V/2500mAh Li-polymer battery, AC-DC 100-240Vac. Connectivity: WiFi, 3G, 4G. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-49, IEC 60601-2-27, IEC 80601-2-30, ISO 81060-2, ISO 80601-2-61, ISO 80601-2-56, ISO 10993-5/10.
Indications for Use
Indicated for monitoring ECG, SpO2, NIBP, TEMP, and pulse rate. ECG/SpO2: adults. NIBP: adults and pediatric patients >3 years. TEMP: all populations including newborns, pediatric, and adults. Intended for use in hospital, clinic, and home settings by qualified healthcare personnel.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue and white.
December 17, 2018
Visiomed Technology Co., Ltd. % Filed Fu Consultant Shenzhen Joyantech Consulting Co., Ltd 1122#, International Mayor Communication Center, Baishizhong Road 55# Nanshan District, Shenzhen, 518000 CN
Re: K172965
Trade/Device Name: Handheld Vital Signs Monitoring System, Model: BW-X07HD Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, DRT, DXN, DOA, FLL Dated: January 2, 2018 Received: January 19, 2018
Dear Filed Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Stephen C. Browning -S5
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172965
Device Name
Handheld VitalSigns Monitoring System, Model: BW-X07HD
Indications for Use (Describe)
The BW-X07HD Handheld VitalSigns Monitoring System is intended to be used in hospital, clinic, homecare, home and other medical settings where patient care is offered by qualified healthcare personnel who can use the device to monitor the patient's physical indexes, including ECG, SpO2, NIBP and TEMP as well as the pulse rate.
For measuring ECG, SpO2 the intended patient population is adult, for measuring NIBP the intended patient population is both adult and pediatric aged more than 3 years old, whereas for measuring TEMP the intent population is all types of people including newborn, pediatric and adult.
The device is used to collect, store, and transmit general patient health information and patient vital signs data between the patient and a health care professional.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) information is submitted as required by requirements of SMDA and 21 CFR §807.92.
# 1. Administrative Information
| Submission Date | 2017-12-24 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Submitter's Name: Visiomed Technology Co.,Ltd<br>Address: 2 Floor of No.1 Building, Jia An Technological<br>Industial Park, 67 Distric,<br>Bao An, Shenzhen,<br>Guangdong,<br>China<br>Contact person: Chen XiaoFeng<br>TEL: 0755-29481701-103<br>FAX: 0755-29481705<br>E-Mail: jay@visiomed-tech.com |
| | Submission<br>Correspondent |
| Image: Logo | |
| | |
| Establishment<br>registration number | 3012429388 |
## 2. Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|----------------------------------------------------------------------|
| Trade Name: | Handheld VitalSigns Monitoring System |
| Model: | BW-X07HD |
| Classification name: | Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) |
| Review Panel: | Cardiovascular |
| Product Code: | Product Code: MWI<br>Subsequent Product Codes: DRT, DXN, DQA, FLL |
| Device Class: | 2 |
| Regulation Number: | 870.2300 |
{4}------------------------------------------------
### 3. Predicate Device Information
Sotera Wireless, Inc. Sponsor: Device: ViSi Mobile Monitoring System K150361 510(K) Number:
#### 4. Device Description
The BW-X07HD Handheld VitalSigns Monitoring System is a handheld, smart and wireless connected medical device. The device can be easily used by healthcare professional personnel, ordinary users or patients, to monitor multiple physiological indexes, including TEMP, ECG, NIBP and SpO2 as well as the pulse rate. Medical data can be stored locally and can be sent through WiFi, 3G or 4G network.
The device incorporates an embedded system and an Android application. The embedded system integrates the Temp module, the ECG, NIBP, SpO2 module and the wireless module together. The device allows users to communicate through an HD video conferencing system but cannot be used as phone (only IP connection, Micro SIM card for data transmission).
## 5. Intended Use/Indications for Use
The BW-X07HD Handheld VitalSigns Monitoring System is intended to be used in hospital, clinic, homecare, home and other medical settings where patient care is offered by qualified healthcare personnel who can use the device to monitor the patient's physical indexes, including ECG, SpO2, NIBP and TEMP as well as the pulse rate.
For measuring ECG, SPO2 the intended patient population is adult, for measuring SpO2 the intended patient population is adult, for measuring NIBP the intended patient population is both adult and pediatric aged more than 3 years old, whereas for measuring TEMP the intended patient population is all types of people including newborn, pediatric and adult.
The device is used to collect, store, and transmit general patient health information and patient vital signs data between the patient and a health care professional.
| Item | Subject Device<br>BW-X07HD | Predicate Device<br>K150361 | Remark | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-------------------|
| Intended<br>Use/Indication<br>for Use | The<br>BW-X07HD Handheld VitalSigns<br>Monitoring System is intended to be used in<br>hospital, clinic, homecare, home and other<br>medical settings where patient care is<br>offered by qualified healthcare personnel<br>who can use the device to monitor the<br>patient's physical indexes, including ECG,<br>SpO2, NIBP and TEMP as well as the pulse<br>rate.<br>For measuring ECG, SpO2 the intended<br>patient population is adult, for measuring<br>SpO2 the intended patient population is<br>adult, for measuring NIBP the intended<br>patient population is both adult and<br>pediatric aged more than 3 years old,<br>whereas for measuring TEMP the intended<br>patient population is all types of people<br>including newborn, pediatric and adult. The<br>device is used to collect, store, and transmit<br>general patient health information and | The ViSi Mobile Monitoring System is<br>lintended for use by clinicians and<br>medically qualified personnel for single<br>or multi-parameter vital<br>sians<br>monitoring of adult patients (18 years or<br>older). It is indicated for ECG (3 or 5<br>lead-wire), respiration rate (RESP),<br>heart rate (HR), noninvasive blood<br>(NIBP),<br>pressure<br>noninvasive blood pressure (cNIBP),<br>Inoninvasive monitoring of functional<br>oxygen saturation<br>of arterial<br>hemoglobin (SpO2), pulse rate (PR),<br>and skin temperature (TEMP) in<br>hospital-based facilities; including,<br>medical-surgical<br>general<br>floors.<br>lintermediate care floors,<br>and<br>emergency departments. The ViSi<br>Mobile Monitoring System may be used<br>as standalone devices or networked to | Similar (Note 01) | |
| Item | Subject Device<br>BW-X07HD | Predicate Device<br>K150361 | Remark | |
| | patient vital signs data between the patient<br>and a health care professional. | ViSi Mobile Remote Viewers through<br>wireless 802.11 communications. | | |
| Product Code | MWI, DRT, DXN, FLL, DQA | MWI, DRT, DXN, FLL, DQA | Same | |
| Patient<br>population | The intended patient population for<br>measuring Temp is all types of people<br>including newborn, pediatric and adult,<br>whereas, for measuring NIBP is both<br>pediatric and adult, for measuring ECG is<br>adult and SpO2 is adult. | Adults ≥ 18 years | Different (Note<br>02) | |
| Usage<br>Environment | home and hospital | Hospital-based facility including<br>emergency departments, general<br>medical-surgical and intermediate care<br>floors. | Similar (Note 03) | |
| Power Supply | | | | |
| Internal<br>supply | Type<br>Rechargeable Li-polymer<br>battery | Rechargeable Li-polymer battery | | |
| | Capacity<br>3.7V/2500mAh | 3.7 V/2000 mAh | Similar (Note 04) | |
| | Performance<br>capacity<br>4 hours | > 12 hours | | |
| AC-DC Power<br>supply | Line Voltage<br>100Vac to 240Vac | 100Vac to 240Vac | | |
| | Current<br>less than 0.5A rms. at 90Vac<br>input and maximum load | / | SE | |
| | Frequency<br>50Hz or 60Hz | 50Hz or 60Hz | | |
| ECG Function | | | | |
| | | 3-wire: II | | |
| | Measuring Mode<br>5 Electrodes, 7-Lead ECG | 5-wire: I, II, III, AVL, AVR, AVF, V | | |
| | DC offset Voltage Range<br>±300mV | ±300mV | SE | |
| | Differential Voltage<br>Measurement Range<br>±5mV | ±5mV | | |
| | Resolution<br>1bpm | 1bpm | Same | |
| NIBP Function: | | | | |
| | Principle of<br>Operation | oscillation | Same | |
| Blood<br>Pressure | Measurement<br>Range<br>Adult & Pediatric:<br>0-295mmHg (0-39.3kPa) | Systolic Range: 60 to 240 mmHg<br>Diastolic Range: 40 to 160 mmHg | | |
| | Accuracy<br>±3mmHg (±0.4kPa) | Mean error of less than ± 5 mmHg<br>and a std. dev. of ≤ 8 mmHg | Similar (Note 05) | |
| | Pressure<br>Resolution<br>1mmHg | 1mmHg | | |
| Pulse Rate | Measuring<br>Range<br>40 bpm to 180 bpm | 30 to 240 BPM | | |
| | Accuracy<br>±3 bpm | ±3 BPM | | |
| | Resolution<br>1 bpm | 1 BPM | | |
| SpO2 Function | | | | |
| | Principle of Operation<br>Lambert Beer Law | | Same | |
| SpO2 | Measuremen<br>t Range<br>0~100% | 49 to 100% | Similar (Note 06) | |
| Item | Subject Device<br>BW-X07HD | Predicate Device<br>K150361 | Remark | |
| Pulse Rate | Accuracy | Adult ±2 %<br>(70% to 100% SpO2),<br>Undefined (0 to 69% SpO2); | ≤ 2% from 70-100% (no motion)<br>Unspecified from 49-69% | |
| | Resolution | 1% | 1% | |
| | Measuring Range | 25 bpm to 250 bpm | 30 to 240 BPM | |
| | Accuracy | ±3 bpm | ≤ 3 BPM | |
| | Resolution | 1 bpm | 1 BPM | |
| Temperature Function | | | | |
| Principle of Operation |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.