K172959 · Perahealth, Inc. · MWI · May 1, 2018 · Cardiovascular
Device Facts
Record ID
K172959
Device Name
PeraServer and PeraTrend
Applicant
Perahealth, Inc.
Product Code
MWI · Cardiovascular
Decision Date
May 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172959 · May 1, 2018
PeraServer and PeraTrend
Perahealth, Inc.
Electronic Health Record (EHR) data; Routine clinical vital signs; Nursing assessments; Laboratory results
Retrospective clinical studies were used to validate the Rothman Index (RI) and pediatric RI (pRI) scores by analyzing historical patient data from routine clinical care settings.
Retrospective study; EHR data; Clinical validation; Rothman Index
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Rothman et al. (2013); Retrospective study
Adult patients
Not applicable for this study
Validation of adult RI score
Finlay et al. (2014); Retrospective study
—
MEWS (Modified Early Warning Score)
Comparison between RI scores and MEWS
Rothman et al. (2017); Retrospective study
Pediatric patients
Not applicable for this study
Validation of pediatric RI score
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Patient status index (Rothman Index)
—
—
—
—
—
Rothman et. al (2013): mechanics and validation of the adult RI score; Finlay et. al (2014): comparison between RI scores and MEWS; Rothman et. al (2017): mechanics and validation of the pediatric RI score.
—
Indications for Use
The Rothman Index uses commonly recorded vital sign, nursing assessment, and lab data to compute a patient status index. The Rothman Index is a single measure of a patient's physiologic condition based on the aggregate statistical mortality risk associated with the values of the patient's vital signs, nursing assessments, and selected lab values. PeraServer is indicated for use wherever there is interest in generating Rothman Index (RI) scores and/or associated configurable warnings. PeraTrend is indicated for use by healthcare providers whenever there is need for displaying and/or trending RI scores and displaying associated configurable warning states as an adjunct to clinical decision support. PeraServer/PeraTrend is intended for the care of patients throughout the hospital setting (e.g., in the emergency department, on the wards, in intensive care units). The Rothman Index score is validated for use with neonatal, pediatric, and adult patients. It is an adjunct-to and is not intended to replace vital signs monitoring and is not intended for use in the Neonatal Intensive Care Unit.
Device Story
PeraServer/PeraTrend system; software-only platform; imports EHR data (vital signs, nursing assessments, lab values). Computes Rothman Index (RI) score; represents aggregate statistical mortality risk; provides configurable warnings. PeraTrend provides web-based UI for trending/displaying scores; used by healthcare providers in hospital settings (ED, wards, ICU). Adjunct to clinical decision support; does not replace vital signs monitoring. Output assists clinicians in assessing patient physiologic condition; aids clinical decision-making; potentially benefits patients through early identification of status changes.
Clinical Evidence
Three retrospective studies provided: Rothman et al. (2013) (adult RI validation), Finlay et al. (2014) (RI vs. MEWS comparison), and Rothman et al. (2017) (pediatric RI validation). Studies demonstrate performance equivalence to predicate for both pediatric and adult populations.
Technological Characteristics
Software-only system; installed on user-provided hardware. Interfaces with hospital EHR systems. Web-based user interface (PeraTrend). HIPAA compliant. No arrhythmia detection performed; heart rhythm data imported from EHR.
Indications for Use
Indicated for neonatal, pediatric, and adult patients in hospital settings (ED, wards, ICU) to provide a patient status index (Rothman Index) as an adjunct to clinical decision support and vital signs monitoring. Not for use in the Neonatal Intensive Care Unit.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 1. 2018
PeraHealth, Inc. % Donna-Bea Tillman Senior Consultant Biologics Consulting 1555 Kingh Street, Suite 300 Alexandria, Virginia 22314
Re: K172959
Trade/Device Name: PeraServer™ and PeraTrend™ System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: March 29, 2018 Received: March 29, 2018
Dear Donna-Bea Tillman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
M.A. Wilhelmsen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172959
Device Name PeraServer and PeraTrend System
# Indications for Use (Describe)
The Rothman Index uses commonly recorded vital sign, nursing assessment, and lab data to compute a patient status index. The Rothman Index is a single measure of a patient's physiologic condition based on the aggregate statistical mortality risk associated with the values of the patient's vital signs, nursing assessments, and selected lab values.
PeraServer is indicated for use wherever there is interest in generating Rothman Index (RI) scores and/or associated configurable warnings.
PeraTrend is indicated for use by healthcare providers whenever there is need for displaying and/or trending RI scores and displaying associated configurable warning states as an adjunct to clinical decision support.
PeraServer/PeraTrend is intended for the care of patients throughout the hospital setting (e.g., in the emergency department, on the wards, in intensive care units).
The Rothman Index score is validated for use with neonatal, pediatric, and adult patients. It is an adjunct-to and is not intended to replace vital signs monitoring and is not intended for use in the Neonatal Intensive Care Unit.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary for the PeraServer and PeraTrend System is provided below:
| Device Common Name: | Adjunct to Multiparameter Patient Monitor |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | PeraServerTM and PeraTrendTM System |
| Applicant: | PeraHealth, Inc.<br>6302 Fairview Road, Suite 310<br>Charlotte, NC 28210 |
| Contact: | Joseph Beals, Ph.D.<br>Vice President, Informatics<br>jbeals@perahealth.com |
| Prepared by: | Donna-Bea Tillman, Ph.D.<br>Senior Consultant<br>Biologics Consulting Group, Inc.<br>dtillman@biologicsconsulting.com |
| Date Prepared: | April 26, 2018 |
| Classification Regulation: | 21 CFR 870.2300 (Class II) |
| Regulation Description: | Cardiac Monitor (including Cardiotachometer and Rate Alarm) |
| Panel: | Cardiovascular |
| Product Code: | MWI |
| Product Code Description: | Physiological Patient Monitor (without arrhythmia<br>detection or alarms) |
| Predicate Device: | BedsidePEWS (K124038) |
#### 1. INDICATIONS FOR USE:
The indications for use for the PeraServer and PeraTrend System are the following:
"The Rothman Index uses commonly recorded vital sign, nursing assessment, and lab data to compute a patient status index. The Rothman Index is a single measure of a patient's physiologic condition based on the aggregate statistical mortality risk associated with the values of the patient's vital signs, nursing assessments, and selected lab values.
{4}------------------------------------------------
PeraServer is indicated for use wherever there is interest in generating Rothman Index (RI) scores and/or associated configurable warnings.
PeraTrend is indicated for use by healthcare providers whenever there is need for displaying and/or trending RI scores and displaying associated configurable warning states as an adjunct to clinical decision support.
PeraServer/PeraTrend is intended for the care of patients throughout the hospital setting (e.g. in the emergency department, on the wards, in intensive care units).
The Rothman Index score is validated for use with neonatal, pediatric, and adult patients. It is an adjunct-to and is not intended to replace vital signs monitoring and is not intended for use in the Neonatal Intensive Care Unit."
### 2. DEVICE DESCRIPTION:
The subject of this submission is the PeraServer and PeraTrend System, consisting of:
- PeraServer a backend hardware/software platform that imports electronic health record ● (EHR) data, which are used to compute and store Rothman Index (RI) scores and related measures
- PeraTrend – the frontend web-based user interfaces for displaying RI scores and related measures, as well as simple manipulations of the display (i.e., graphical representations over time, comparison of Rothman Index scores and input measures)
PeraServer and PeraTrend are software-only devices that are installed on user-provided hardware.
## SOFTWARE: 3.
Documentation has been provided in the submission commensurate with a Moderate Level of Concern.
## SUBSTANTIAL EQUIVALENCE: 4.
### 4.1. Cited Predicate Device:
The cited predicate device is the Bedside Clinical Systems Bedside Paediatric Early Warning System (BedsidePEWS), cleared under K124038.
## Comparison of Intended Use: 4.2.
The PeraServer and PeraTrend System and the predicate BedsidePEWS have the same intended use, namely to provide healthcare providers with a patient status index that reflects the underlying patient condition. Both devices acquire their inputs from the hospital Electronic Medical Record System and are intended to be an adjunct to vital signs monitoring. Both devices are intended to be used by healthcare professionals in the clinical care environment. The intended patient population of BedsidePEWS is the pediatric population, while the PeraServer
{5}------------------------------------------------
and PeraTrend System is intended for both pediatric and adult populations. This difference in indications does not raise different questions of safety and effectiveness.
### Comparison of Technological Characteristics: 4.3.
| | Subject Device | Predicate Device |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | PeraHealth, Inc. | Bedside Clinical Systems, Inc. |
| Device Name | PeraServerTM and PeraTrendTM System | Bedside Paediatric Early Warning<br>System (BedsidePEWSTM) |
| Classification<br>Regulation | 21 CFR 870.2300, Class II | 21 CFR 870.2300, Class II |
| Product Code | MWI: Physiological Patient Monitor<br>(without arrhythmia detection or<br>alarms) | MWI: Physiological Patient Monitor<br>(without arrhythmia detection or<br>alarms) |
| Device Description | PeraServer collects specified<br>demographic and clinical data from the<br>EHR and ancillary systems. PeraServer<br>computes an Early Warning Score<br>called the Rothman Index (RI),<br>determines if RI scores and/or clinical<br>criteria meet configurable EWS<br>criteria; specified input data, and RI<br>score and warning states are stored for<br>analysis, remote viewing, and<br>interfacing.<br>PeraTrend provides a web-based user-<br>interface for trending and review of RI<br>scores and underlying data and<br>warning status.<br>PeraServer/PeraTrend also interface to<br>other medical automation systems such<br>as the Hospital Electronic Health<br>Record (EHR). | BedsidePEWS is a web-based clinical<br>decision-support tool that can also be<br>run as a standalone application for use<br>on computers running Microsoft<br>Software Operating features with a<br>local network connection, which access<br>medical data. It is also interfaces to<br>other medical automation systems such<br>as Hospital Information System (HIS),<br>Electronic Medical Record (EMR), and<br>Practice Management Software. |
| | Subject Device | Predicate Device |
| Indications for Use | The Rothman Index uses commonly<br>recorded vital sign, nursing<br>assessment, and lab data to compute a<br>patient status index. The Rothman<br>Index is a single measure of a patient's<br>physiologic condition based on the<br>aggregate statistical mortality risk<br>associated with the values of the<br>patient's vital signs, nursing<br>assessments, and selected lab values.<br>PeraServer is indicated for use<br>wherever there is interest in generating<br>Rothman Index (RI) scores and/or<br>associated configurable warnings.<br>PeraTrend is indicated for use by<br>healthcare providers whenever there is<br>need for displaying and/or trending RI<br>scores and displaying associated<br>configurable warning states as an<br>adjunct to clinical decision support.<br>PeraServer/PeraTrend is intended for<br>the care of patients throughout the<br>hospital setting (e.g. in the emergency<br>department, on the wards, in intensive<br>care units).<br>The Rothman Index score is validated<br>for use with neonatal, pediatric, and<br>adult patients. It is an adjunct-to and is<br>not intended to replace vital signs<br>monitoring and is not intended for use<br>in the Neonatal Intensive Care Unit. | The Bedside Paediatric Early Warning<br>System (BedsidePEWS) is an<br>electronic documentation tool that is<br>designed to be used in conjunction with<br>multi-parameter patient monitoring. It<br>is indicated for use by healthcare<br>professionals with paediatric patients<br>between the ages of term newborn (>37<br>weeks gestational age) and 18 years,<br>who are hospitalized with any medical<br>or surgical condition.<br>BedsidePEWS allows input by the<br>healthcare professional of commonly<br>recorded vital sign data and provides<br>clinician with a patient status index (the<br>"BedsidePEWS score") based on a<br>weighted average of seven vital signs<br>when entered by the clinician, namely<br>Heart Rate, Respiratory Rate, Blood<br>Pressure, Oxygen Saturation, Oxygen<br>Therapy, Respiratory Effort, and<br>Capillary Refill. The "BedsidePEWS<br>score" is a single measure of a patient's<br>condition and indicates the variation in<br>the patient's vital signs with respect to<br>normality. BedsidePEWS is an adjunct-<br>to and is not intended to replace vital<br>signs monitoring.<br>BedsidePEWS is intended for use in<br>wards and emergency rooms in<br>hospitals that provide care for children<br>between the ages of term newborn (>37<br>weeks gestational age) and 18 years. It<br>is not intended for use in the Neonatal<br>Intensive Care Unit. |
| Intended users | Healthcare Professionals | Healthcare Professionals |
| Patient Population | Adult and pediatric patients | Pediatric patients between the ages of<br>term newborn (>37 weeks gestational<br>age) and 18 years |
| Use environment | Hospital | Hospital |
| Parameters | Vital signs (Temp, Sys. BP, Dias. BP,<br>HR, RR) Pulse-ox, heart rhythm*,<br>nursing assessments, Braden score,<br>selected labs (potassium, sodium,<br>creatinine, chloride, WBC, BUN,<br>HGB) | Vital signs (Heart Rate, Respiratory<br>Rate, Blood Pressure, Oxygen<br>Saturation, Oxygen Therapy,<br>Respiratory Effort, and Capillary<br>Refill) |
| | Subject Device | Predicate Device |
| Score | Rothman Index (RI) and Pediatric RI<br>(pRI) | BedsidePEWS score |
| Material | Software only | Software only |
| Hardware Platform | PeraServer/PeraTrend require<br>dedicated application and database<br>servers. | Windows PC or Tablet |
| Data Source | Hospital EHR systems | Hospital EHR systems or manually<br>input |
| Data Display | PeraTrend Web – a browser based<br>display<br>PeraTrend Kiosk – a browser based<br>display configured for dedicated<br>monitors providing an always-up,<br>auto-refreshed display<br>PeraTrend EHR – a browser based<br>display accessed and viewed from<br>within the EHR | Browser based display |
| Ability to trend | Yes | Yes |
| Real-time display | Yes | Yes |
| Privacy | HIPPA compliant | HIPPA compliant |
{6}------------------------------------------------
{7}------------------------------------------------
*Note: Although the PeraServer and PeraTrend System uses heart rhythm as one of its inputs, this information is obtained from the EHR and therefore the PeraServer and PeraTrend System is not performing arrhythmia detection.
### PERFORMANCE TESTING: 5.
To demonstrate the validity of the RI score, three retrospective studies are described in the submission:
- Rothman et. al (2013), to describe the mechanics and validation of the adult RI score ●
- Finlay et. al (2014), to provide a comparison between RI scores and MEWS ●
- Rothman et. al (2017), to describe the mechanics and validation of the pediatric RI score ●
These studies demonstrate that the performance of the PeraServer and PeraTrend System is equivalent to that of the predicate device for both pediatric and adult populations.
### CONCLUSION: 6.
The subject PeraServer and PeraTrend System and the predicate BedsidePEWS have the same intended use, which is to provide healthcare providers with a patient status index that reflects the
{8}------------------------------------------------
underlying patient condition. The differences in technological characteristics do not raise different questions of safety and effectiveness, and the results of performance testing demonstrate that the PeraServer and PeraTrend System performs in accordance with specifications and meets user needs and intended uses. The software testing provided in the submission verifies the functionality of the software, and the studies provided in the submission validate the Rothman Index score. Therefore, the PeraServer and PeraTrend System has been demonstrated to be substantially equivalent to the predicate BedsidePEWS.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.