← Product Code [LON](/productcode/LON) · K172944

# VITEK 2 AST-GN Ceftazidime/Avibactam <=0.12 – >=16 µg/mL (K172944)

_bioMerieux, Inc. · LON · Dec 19, 2017 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172944

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/productcode/LON.md)
- **Decision Date:** Dec 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

VITEK® 2 Gram Negative Ceftazidime/Avibactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ceftazidime/ Avibactam is a quantitative test. Ceftazidime/Avibactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active both in vitro and in clinical infections: Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa The following in vitro data are available, but clinical significance is unknown: Citrobacter koseri Enterobacter aerogenes Morganella morganii Providencia stuartii Serratia marcescens The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Streptococcus spp., Streptococcus spp. and clinically significant yeast.

## Device Story

VITEK 2 AST-GN Ceftazidime/Avibactam is an automated, miniaturized, growth-based antimicrobial susceptibility test (AST) system. It utilizes 64-well test cards containing premeasured antimicrobial concentrations. Input consists of isolated bacterial colonies diluted in saline. The system automatically (VITEK 2) or manually (VITEK 2 Compact) fills, seals, and incubates cards. An optical scanner measures light attenuation every 15 minutes to monitor bacterial growth. The system calculates the Minimum Inhibitory Concentration (MIC) based on growth kinetics. Results are reported as MIC values and interpretive categories (Susceptible/Resistant). Used in clinical laboratories to guide antibiotic therapy for Gram-negative infections. The device provides rapid, standardized susceptibility data, aiding clinicians in selecting appropriate antimicrobial treatments and improving patient outcomes by identifying effective therapies.

## Clinical Evidence

Performance evaluated via clinical and challenge studies (n=1073 total results). Compared to CLSI broth microdilution reference method. Overall Essential Agreement (EA) was 94.5% and Category Agreement (CA) was 98.7%. Major discrepancy rate was 1.4% (adjusted to 0.5% due to lack of intermediate category). P. aeruginosa showed a trend toward higher MICs; a limitation was added for repeat testing when MIC is ≥16 µg/mL.

## Technological Characteristics

Miniaturized broth microdilution card containing antimicrobial concentrations. Automated incubation and optical growth monitoring system. Connectivity via VITEK 2/Compact instrument platforms. Software-based analysis of growth kinetics to determine MIC.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK® 2 AST-GN Ceftolozane/Tazobactam ([K161510](/device/K161510.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 15

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K172944

B. Purpose for Submission:
To obtain a substantial equivalence determination for Ceftazidime/Avibactam for testing of gram negative bacilli on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test (AST) Systems

C. Measurand:
The VITEK 2 AST-Gram Negative card contains the following concentrations of Ceftazidime/Avibactam: 0.06/4, 0.25/4, 1/4, 4/4, and 8/4 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The Ceftazidime/Avibactam MIC reporting ranges for the card are ≤ 0.12/4 – ≥16/4 µg/mL for Enterobacteriaceae and ≤ 0.25/4 – ≥16/4 µg/mL for Pseudomonas aeruginosa.

D. Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for Ceftazidime/Avibactam

E. Applicant:
bioMérieux, Inc.

F. Proprietary and Established Names:
VITEK 2 AST- GN Ceftazidime/Avibactam (≤ 0.12 – ≥16 µg/mL)

G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System

{1}

Page 2 of 15

2. Classification:
Class II

3. Product code(s):
LON – Fully automated short-term incubation cycle antimicrobial susceptibility system
LTW – Susceptibility Test Cards, Antimicrobial
LTT – Panels, Test, Susceptibility, Antimicrobial

4. Panel:
Microbiology, 83

H. Intended Use/Indications for Use:

1. Intended Use (s):
The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

2. Indications for Use:
VITEK 2 AST-Gram Negative Ceftazidime/Avibactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST-Gram Negative Ceftazidime/Avibactam is a quantitative test. Ceftazidime/Avibactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.

Active both in vitro and in clinical infections
Citrobacter species
Enterobacter species
Escherichia coli
Klebsiella oxytoca
Klebsiella pneumoniae
Proteus mirabilis
Pseudomonas aeruginosa

The following in vitro data are available, but clinical significance is unknown:
Citrobacter koseri
Enterobacter aerogenes

{2}

Morganella morganii
Providencia stuartii
Serratia marcescens

The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

3. Special conditions for use statement(s):

For Prescription Use Only

Limitations:

- Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination(s):
- Ceftazidime/Avibactam: Providencia rettgeri
- Although within Essential Agreement when compared to the reference method, errors (major and/or very major) were observed due to the lack of intermediate category. Testing should be repeated using an alternative testing/reference method prior to reporting results for the following antibiotic/organism(s):
- Ceftazidime/Avibactam: Pseudomonas aeruginosa when the VITEK 2 MIC is ≥16μg/mL due to observed major errors (4.1%), with an adjusted categorical major error rate of 1.2%.

4. Special instrument requirements:

VITEK 2 and VITEK 2 Compact Systems using VITEK 2 Systems 9.01 software

I. Device Description:

The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 24 hours for Streptococcus species). The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration

Page 3 of 15

{3}

or "MIC" values for the anti-microbial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK 2 AST-GN Ceftazidime/Avibactam has the following concentrations in the card: 0.06/4, 0.25/4, 1/4, 4/4, and 8/4μg/mL (equivalent standard method concentration by efficacy in μg/mL). The Ceftazidime/Avibactam MIC result ranges for the VITEK 2 are:
- ≤ 0.12 – ≥ 16 μg/mL for Enterobacteriaceae
- ≤ 0.25 – ≥ 16 μg/mL for Pseudomonas aeruginosa

J. Substantial Equivalence Information:

1. Predicate device name(s):
VITEK 2 AST-GN Ceftolozane/Tazobactam

2. Predicate 510(k) number(s):
K161510

3. Comparison with predicate:

Page 4 of 15

{4}

Table 1: Comparison with Predicate Device

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device: VITEK 2 AST-GN Ceftazidime/Avibactam (K172944) | Predicate Device: VITEK 2 AST-GN Ceftolozane/Tazobactam (K161510)  |
|  Intended Use | The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. | Same  |
|  Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of Gram negative bacilli | Same  |
|  Inoculum | Standardized saline suspension of test organism | Same  |
|  Test Card | VITEK 2 Gram Negative Susceptibility Test Card | Same  |
|  Instrument | VITEK 2 and VITEK 2 Compact Systems | Same  |
|  Analysis Algorithm | Growth pattern analysis | Same  |
|  Differences  |   |   |
|  Antimicrobial Agent | Ceftazidime/Avibactam | Ceftolozane/Tazobactam  |
|  Antimicrobial Concentrations | 0.06/4, 0.25/4, 1/4, 4/4, and 8/4 μg/mL | 0.5/4, 1/4, 4/4, 8/4, and 32/4 μg/mL  |
|  Reporting Range | Enterobacteriaceae: ≤0.12 - ≥16 μg/mL P. aeruginosa: ≤0.25 - ≥16 μg/mL | Enterobacteriaceae and P. aeruginosa: ≤0.25 - ≥32μg/mL  |

# K. Standard/Guidance Documents Referenced (if applicable)

FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
- CLSI M07-A10, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard-Ninth Edition" Vol. 35 No. 2 (January 2015)
- CLSI M100-S27, "Performance Standards for Antimicrobial Susceptibility Testing"; Twenty-Fourth Informational Supplement, Vol. 37 No. 1 (January 2017)

{5}

# L. Test Principle:

The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems uses visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well measures organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

A reproducibility study for the VITEK 2 AST-GN card with Ceftazidime/Avibactam was conducted at three external clinical sites using ten isolates of gram-negative bacilli consistent with the Intended Use. Testing was performed on three separate days and in triplicate for a total of 270 data points. The isolates tested in the reproducibility study included one isolate each for Klebsiella pneumoniae pneumoniae and Escherichia coli, two isolates each for Citrobacter freundii complex, Enterobacter aerogenes, Pseudomonas aeruginosa, and Serratia marcescens. Inocula were prepared using both the auto-dilution and manual dilution methods for testing in the VITEK 2 System. Inocula were prepared by the manual dilution method only for use with the VITEK 2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values that fell within  $+ / -$  one doubling dilution from the mode MIC value. The majority of data points were on-scale and within  $\pm$  one doubling dilution agreement as compared to the mode MIC (Table 2). Two data points were off-scale when tested by the VITEK 2 and the VITEK 2 Compact. The data was analyzed taking into consideration best case and worst case scenarios as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems.

The reproducibility performance is shown in Table 2.

Table 2: Reproducibility Performance

|   | VITEK 2 |   | VITEK 2 Compact  |
| --- | --- | --- | --- |
|   |  Auto-Dilution | Manual Dilution | Manual Dilution  |
|  Best Case | 99.63% | 99.63% | 100%  |
|  Worst Case | 99.26% | 99.26% | 98.89%  |

The combined reproducibility results for all three sites were acceptable.

{6}

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

**Inoculum Density Control**

The DensiCHEK Plus was used to standardize the inoculum to a 0.5 McFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.

**Purity Check**

A purity check of all organisms was performed on the dilution tube used to prepare the VITEK 2 card inoculum. Only those cultures that were pure were evaluated in the study.

**Growth Failure Rate**

There were ten isolates that failed to grow in the VITEK 2 card in the clinical study. Complete test results were available for 980 isolates from a total of 990 clinical isolates. The growth failure rate was 1.0% and was acceptable.

A challenge set of 93 isolates was evaluated at one site. All 93 challenge organisms grew in the VITEK 2 GN card with Ceftazidime/Avibactam using both the autodilution and manual dilution methods for the VITEK 2 and manual inoculation for the VITEK 2 Compact System.

A total of 1073 VITEK 2 AST results were available.

**Quality Control (QC) Testing**

The FDA/CLSI recommended QC organisms (E. coli ATCC 25922, E. coli ATCC 35218, Pseudomonas aeruginosa ATCC 27853, and K. pneumoniae pneumoniae ATCC 700603) were tested using both the VITEK 2 card and the reference method at each site. Both the automatic dilution and manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact.

As shown in Table 3, both the auto-dilution and the manual dilution methods for VITEK 2 and the manual dilution for VITEK 2 Compact were within the expected range &gt;95% of the time.

Page 7 of 15

{7}

Table 3: Quality Control Summary Results for VITEK 2 (Auto-Dilution and Manual Dilution Methods) and VITEK 2 Compact (Manual Dilution Method)

|  Ceftazidime/Avibactam |  | VITEK 2 |   |   |   | VITEK 2 Compact Manual Dilution  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Auto-Dilution |   | Manual Dilution  |   |   |   |
|  Organism | Conc. (μg/mL) | Reference | Test | Reference | Test | Reference | Test  |
|  E. coliATCC 25922FDA/CLSIExpected Range0.06 – 0.5μg/mL(VITEK 2:≤0.12-0.5μg/mL) | ≤0.015625 |  |  |  |  |  |   |
|   |  0.03125 |  |  |  |  |  |   |
|   |  0.0625 | 113 |  | 62 |  | 62 |   |
|   |  ≤0.12** | 137 | 112 | 74 | 87 | 75 | 103  |
|   |  0.25 | 7 | 146 | 4 | 23 | 4 | 38  |
|   |  0.5 |  |  |  |  |  |   |
|   |  1 | 1 |  |  |  |  |   |
|   |  2 |  |  |  |  |  |   |
|   |  ≥4 |  |  |  |  |  |   |
|  E. coliATCC 35218FDA/CLSIExpected Range0.03-0.12μg/mL(VITEK 2:≤0.12μg/mL) | ≤0.015625 |  |  |  |  |  |   |
|   |  0.03125 | 41 |  | 13 |  | 13 |   |
|   |  0.0625 | 204 |  | 119 |  | 119 |   |
|   |  ≤0.12** | 8 | 252 | 4 | 135 | 4 | 136  |
|   |  0.25 |  | 1 |  | 1 |  |   |
|   |  0.5 |  |  |  |  |  |   |
|   |  1 |  |  |  |  |  |   |
|   |  2 |  |  |  |  |  |   |
|   |  ≥4 |  |  |  |  |  |   |
|  K. pneumoniae pneumoniaeATCC 700603Expected Range0.025- 2μg/mL | ≤0.015625 |  |  |  |  |  |   |
|   |  0.03125 |  |  |  |  |  |   |
|   |  0.0625 |  |  |  |  |  |   |
|   |  0.12 |  |  |  |  |  | 1  |
|   |  0.25 | 64 |  | 18 |  | 16 |   |
|   |  0.5 | 156 | 258 | 103 | 142 | 103 | 139  |
|   |  1 | 39 | 1 | 21 |  | 21 |   |
|   |  2 |  | 1 | 1 | 1 | 1 | 1  |
|   |  ≥4 |  |  |  |  |  |   |
|  |   |   |   |   |   |   |   |
|  P. aeruginosaATCC 27853Expected Range0.5 – 4 μg/mL | ≤0.0625 |  |  |  |  |  |   |
|   |  0.12 |  |  |  |  |  |   |
|   |  0.25 |  |  |  |  |  |   |
|   |  0.5 |  |  |  |  |  |   |
|   |  1 | 1 | 8 | 107 | 11 | 107 | 9  |
|   |  2 | 196 | 244 | 26 | 124 | 26 | 124  |
|   |  4 | 50 |  | 3 |  | 3 |   |
|   |  8 | 6 |  |  |  |  |   |
|   |  16 |  | 1 |  | 1 |  | 3  |
|   |  32 |  |  |  |  |  |   |
|   |  ≥64 |  |  |  |  |  |   |

* MIC for ceftazidime in the presence of a fixed concentration of 4 μg/mL of avibactam
** The lowest dilution of the VITEK 2 Ceftazidime/Avibactam MIC range is ≤0.12μg/mL. Obtaining this value was considered an indicator that the quality control test results were acceptable.
$^b$  K. pneumoniae ATCC 700603 should be tested against Ceftazidime/Avibactam and Ceftazidime alone to confirm the activity of Avibactam in the combination and to ensure that the plasmid encoding the beta-lactamase has not been lost in this strain. The acceptable range for Ceftazidime alone is greater than 16 μg/mL.

{8}

The FDA/CLSI Ceftazidime/Avibactam expected ranges for E. coli ATCC 25922 and E. coli ATCC 35218 are 0.06 – 0.5 and 0.03 – 0.12μg/mL respectively. However, the VITEK 2 MIC reporting range is ≤0.12 – ≥16 μg/mL (MIC results: ≤0.12, 0.25, 0.5, 1, 2, 4, 8, ≥16 μg/mL). The VITEK 2 systems do not provide results lower than 0.12 μg/mL; therefore, all results for E. coli ATCC 35218 and the lower range value for E. coli ATCC 25922 are off scale. The quality control test results were considered acceptable for E. coli ATCC 25922 and E. coli ATCC 35218 when the MIC value of ≤0.12 μg/mL was obtained for these organisms by the VITEK 2 system.

bioMérieux includes the FDA/CLSI broth microdilution expected range in the labeling when the VITEK 2 system range is not aligned with that of the FDA/CLSI range. The additional information is acceptable.

In addition, K. pneumoniae ATCC 700603 is recommended to test against ceftazidime alone to ensure the presence of the plasmid encoding beta-lactamase when testing this beta-lactam/beta-lactamase inhibitor combination, ceftazidime/avibactam. The acceptable result for ceftazidime alone is greater than 16μg/mL. The footnote below is added to the QC section of the Package Insert for VITEK 2 AST-GN Ceftazidime/Avibactam:

K. pneumoniae ATCC 700603 should be tested against Ceftazidime/Avibactam and Ceftazidime alone to confirm the activity of Avibactam in the combination and to ensure that the plasmid encoding the beta-lactamase has not been lost in this strain. The acceptable range for Ceftazidime alone is greater than 16 μg/mL.

d. Detection limit:

Not applicable

e. Analytical Specificity:

Not applicable

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Results obtained with the bioMérieux VITEK 2 AST-Gram Negative card with Ceftazidime/Avibactam were compared to results obtained with the CLSI broth microdilution reference panel. The following concentrations of Ceftazidime/Avibactam are contained in the VITEK 2 AST-GN test card: 0.06/4,

Page 9 of 15

{9}

0.25/4, 1/4, 4/4, and 8/4μg/mL (equivalent standard method concentration by efficacy in μg/mL). However, the reporting range is ≤ 0.12 – ≥16 μg/mL for Enterobacteriaceae and ≤ 0.25 – ≥16 μg/mL for Pseudomonas aeruginosa. The reference panel contained two-fold serial dilutions with a range of ≤ 0.0156 to ≥128 μg/mL. The testing conditions for the reference method consisted of the following:

- Medium – Mueller Hinton broth with appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension
- Incubation – 35°C ambient air incubator; 16-20 hours

All test inocula used for the evaluation of VITEK 2 AST-GN Ceftazidime/Avibactam and the reference method were standardized using the DensiCHEK Plus instrument. The cards were inoculated with each test organism by auto-dilution for reading by the VITEK 2 System and by manual dilution for reading on the VITEK 2 and VITEK 2 Compact Systems. Reference broth microdilution panels were inoculated in adherence with CLSI document, M07-A10.

A total of 990 clinical isolates were evaluated at three sites with VITEK 2 AST – Gram Negative cards inoculated by automatic dilution and interpreted using the VITEK 2 instrument. There were ten isolates that did not grow in the studies. The no growth rate was 0.1% (10/990). Most isolates were recently isolated from clinical specimen (885 isolates, 89.4%). The remainder were stock isolates (105 isolates, 10.6%).

A total of 93 challenge organisms (47 Enterobacteriaceae and 46 P. aeruginosa) were evaluated at one site. The challenge set was tested with the auto-dilution and manual dilution options of the VITEK 2 System and with the manual dilution method on the VITEK 2 Compact System. Overall performance of the clinical and challenge organisms by auto-dilution is shown in Table 4.1.

Page 10 of 15

{10}

Table 4.1: Performance  ${}^{ \ddagger  }$  of Clinical and Challenge Isolates,VITEK 2 Auto-Dilution Method

|  Ceftazidime/Avibactam | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Enterobacteriaceae ≤8 (Susceptible), --, ≥16 (Resistant)  |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 819 | 770 | 94.0 | 265 | 216 | 81.5 | 813 | 99.3 | 10 | 6 | 0  |
|  Challenge | 47 | 46 | 97.9 | 20 | 19 | 95.0 | 46 | 97.9 | 27 | 1 | 0  |
|  Combined | 866 | 816 | 94.2 | 285 | 235 | 82.5 | 859 | 99.2 | 37 | 7 (0.8%) | 0  |
|  Pseudomonas aeruginosa ≤8 (Susceptible), --, ≥16 (Resistant)  |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 161 | 152 | 94.4 | 145 | 136 | 93.8 | 154 | 95.7 | 10 | 7 | 0  |
|  Challenge | 46 | 46 | 100 | 18 | 18 | 100 | 46 | 100 | 28 | 0 | 0  |
|  Combined | 207 | 198 | 95.7 | 163 | 154 | 94.5 | 200 | 96.6 | 38 | 7 (4.1%) | 0  |
|  Enterobacteriaceae + Pseudomonas aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 980 | 922 | 94.1 | 410 | 352 | 85.9 | 967 | 98.7 | 20 | 13 | 0  |
|  Challenge | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0  |
|  Combined | 1073 | 1014 | 94.5 | 448 | 389 | 86.8 | 1059 | 98.7 | 75 | 14† (1.4%) | 0  |

The overall categorical major error rate for Enterobacteriaceae and Pseudomonas aeruginosa combined was  $1.4\%$  (14/998). Nine (9) major errors were one dilution apart from the reference method and as such fall within essential agreement. Based on the essential agreement, and the lack of an intermediate breakpoint for Ceftazidime/Avibactam, the adjusted categorical major error rate is  $0.5\%$  (5/998).
$\ddagger$ EA - Essential Agreement (+/- 1 doubling dilution)
CA - Category Agreement
EVAL - Evaluable isolates
R - Resistant isolates

Min - Minor discrepancies
Maj - Major discrepancies
Vmj - Very major discrepancies

Essential Agreement (EA) occurs when there is agreement between the result of the reference method and that of VITEK 2 test card within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the VITEK 2 test card and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the VITEK 2 test card.

The overall performance using the auto-dilution method for the VITEK 2 System demonstrated an essential agreement of  $94.5\%$  and an overall category agreement of  $98.7\%$ . The major discrepancy rate was acceptable at  $1.4\%$ . However, due to the lack of intermediate category, the maj discrepancy rate was adjusted and was described in a footnote:

The overall categorical major error rate for Enterobacteriaceae and Pseudomonas aeruginosa combined was  $1.4\%$  (14/998). Nine (9) major errors were one dilution apart from the reference method and as such fall within essential agreement. Based on the essential agreement, and the lack of an intermediate breakpoint for Ceftazidime/Avibactam, the adjusted categorical major error rate is  $0.5\%$  (5/998).

# Pseudomonas aeruginosa (Reporting range:  $\leq 0.25 / 4 - \geq 16 / 4\mu \mathrm{g} / \mathrm{mL}$ )

When  $P.$  aeruginosa data was analyzed separately, it was observed that the major discrepancy rate for this organism was  $4.1\%$  (7/169) and did not meet the acceptance criterion. Due to the lack of "intermediate" interpretation category, adjusted major discrepancy rate was calculated taking into consideration the MIC values were within essential agreement compared to reference MIC. Of the seven maj discrepancies, five

{11}

were within essential agreement [i.e. VITEK 2:  $\geq 16$  (R), Reference: 8 (S)] which would have been considered minor discrepancies if there was an "intermediate" interpretative category for  $P$ . aeruginosa. The adjusted major discrepancy rate was  $1.2\%$  (2/169) which met the acceptance criterion.

The adjusted  $P$ . aeruginosa performance is noted in Table 4.2 below.

Table 4.2: Adjusted Major Discrepancy Performance of  $P$  . aeruginosa Due to the Lack of an Intermediate Category (VITEK 2 Auto-Dilution Method)

|  Ceftazidime/Avibactam | Susceptible# | Major Discrepancies  |   |
| --- | --- | --- | --- |
|   |   |  # | Rate  |
|  Before Adjustment | 169 | 7 | 4.1%  |
|  After Adjustment | 169 | 2† | 1.2%  |

† Five (5) P. aeruginosa isolates were within essential agreement when compared to the reference method

The original  $P$  aeruginosa performance with seven major discrepancies was mitigated by adding a limitation to address the need for repeat testing  $P$  aeruginosa by alternative/reference methods when the VITEK 2 Systems MIC is  $\geq 16\mu \mathrm{g / mL}$ . The limitation is noted below:

Although within Essential Agreement when compared to the reference method, errors (major and/or very major) were observed due to the lack of intermediate category. Testing should be repeated using an alternative testing/reference method prior to reporting results for the following antibiotic/organism(s):

- Ceftazidime/Avibactam: Pseudomonas aeruginosa when the VITEK 2 MIC is  $\geq 16\mu \mathrm{g} / \mathrm{mL}$  due to observed major discrepancies  $(4.1\%)$ , with an adjusted categorical major error rate of  $1.2\%$ .

For implementation of the repeat testing limitation, bioMerieux indicated that they will issue a customer communication requesting VITEK 2 users to create a custom BioART rule in the VITEK 2 System software. Implementation of an indication for use BioART rule for a U.S. customer will prevent the user from generating a result when using the AST-GN Ceftazidime/Avibactam test for Pseudomonas aeruginosa.

The indication for use and limitation BioART rule will be pre-defined in the VITEK 2 System software in the next software update.

# MIC Trends:

The claimed Pseudomonas aeruginosa and Enterobacteriaceae organisms were also evaluated for trending. This trending calculation takes into account MIC values that are determined to be  $\leq 1$  and  $\geq 1$  doubling dilutions compared to the reference method irrespective whether the device MIC values are on-scale or not. The trending analysis is shown in Table 4.3 below.

{12}

Table 4.3: Trending Analysis of Evaluable Clinical and Challenge Results for P. aeruginosa and Enterobacteriaceae

|  Ceftazidime/ Avibactam ≤ 1.25 – ≥ 16μg/mL | Total a (Total # Tested) | Difference in MIC as Compared to the CLSI Reference Method  |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |  ≥2 dil. lower | 1 dil. lower | Exact | 1 dil. higher | ≥2 dil. higher  |
|  P. aeruginosa + Enterobacteriaceae | 508/1073 | 15 | 75 | 155 (30.51%) | 209 | 54  |
|   |   |  90 (17.72%) b 95% CI (14.64% to 21.28%) |   |   | 263 (51.77%) b 95% CI (47.43% to 56.09%)  |   |
|  P. aeruginosa | 168 (207) | 0 | 12 | 69 (41.07%) | 78 | 9  |
|   |   |  12 (7.14%) c 95% CI (4.13% to 12.07%) |   |   | 87(51.79%) c 95% CI (44.27% to 59.22%)  |   |

a Total number of evaluable results for trending analysis
b Difference between isolates trending higher and isolates trending lower for Enterobacteriaceae and P. aeruginosa: 34.06%; 95% CI (28.44% to 39.55%)
c Difference between isolates trending higher and isolates trending lower for P. aeruginosa: 44.64%; 95% CI (35.66% to 52.66%)

A trend towards higher MIC was observed with Enterobacteriaceae and P. aeruginosa isolates when compared to the CLSI broth microdilution reference method. This trending and the potential for occurrence of major discrepancies for Ceftazidime/Avibactam when testing clinical and challenge isolate results with the VITEK 2 GN Ceftazidime/Avibactam was addressed by adding the following footnote to the performance table for Ceftazidime/Avibactam in the device labeling (Table 127: Performance Characteristics for Gram-Negative Antimicrobial Susceptibility Testing, VITEK 2 Product Information Manual).

VITEK 2 Ceftazidime/Avibactam MIC values tended to be in exact agreement or at least one doubling dilution higher when testing Enterobacteriaceae and Pseudomonas aeruginosa compared to the reference broth micro-dilution.

## Challenge Data – Auto and Manual Dilution:

The challenge set of 93 isolates (47 Enterobacteriaceae and 46 Pseudomonas aeruginosa) was evaluated at one site. The challenge set was tested with both the autodilution and manual dilution options of the VITEK 2 System and with the manual dilution method on the VITEK 2 Compact System.

Overall performance is shown in Table 5.

Page 13 of 15

{13}

Table 5: Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact

|  Ceftazidime/Avibactam | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  VITEK 2  |   |   |   |   |   |   |   |   |   |   |   |
|  Auto-dilution | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0  |
|  Manual Dilution | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0  |
|  VITEK 2 Compact  |   |   |   |   |   |   |   |   |   |   |   |
|  Manual Dilution | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0  |

The performance of the challenge isolates using the VITEK 2 Ceftazidime/Avibactam test on both the VITEK 2 and VITEK 2 Compact Systems was acceptable.

# Enzyme Group Characterization/Resistance Markers Information

Enterobacteriaceae with beta-lactamases were included in the ceftazidime/avibactam comparative studies. They were KPC (6), CTX-M (4), TEM (11), SHV (12) and plasmid mediated AmpC (4). The Enterobacteriaceae were Klebsiella pneumoniae, E. coli, and Enterobacter cloacae. The study showed susceptibility to Ceftazidime/Avibactam with VITEK2 MIC values range from 0.5/4 to  $2/4 \mu \mathrm{g} / \mathrm{mL}$ .

Additionally, seven  $P_{\cdot}$  aeruginosa with beta-lactamases AmpC variants PDC (7), KPC (1), GES (1), VEB (1), and outer membrane porin loss OprD (7) were tested. Two isolates demonstrated resistance of  $\geq 16 / 4~\mu \mathrm{g / mL}$  and five isolates showed susceptibility ranging from  $2 / 4$  to  $8 / 4~\mu \mathrm{g / mL}$ . The following footnote was added to the ceftazidime/avibactam labeling regarding other resistance markers and enzyme characterization:

&gt; Enzyme group characterization was not available for the following organisms at the time of comparative testing, and therefore the performance of the AST card with Ceftazidime/Avibactam is unknown: Enterobacteriaceae (OXA, metallo-beta-lactamases); Pseudomonas aeruginosa (metallo-beta-lactamases).
&gt; Ceftazidime/Avibactam is not active against bacteria that produce metallo beta-lactamase enzymes and may not have activity against Gram-negative bacteria that overexpress efflux pumps or have porin mutations.

# b. Matrix comparison:

Not applicable

{14}

3. Clinical studies:

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

Table 6: Interpretive Criteria for Ceftazidime/Avibactam (FDA Drug Label)

|  Organism | FDA Interpretive Criteria for Ceftazidime/Avibactam MIC (μg/mL)  |   |   |
| --- | --- | --- | --- |
|   |  S | I | R  |
|  Enterobacteriaceae
Pseudomonas aeruginosa | ≤8/4 | -- | ≥16/4  |

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision

Page 15 of 15

---

**Source:** [https://fda-staging.innolitics.com/device/K172944](https://fda-staging.innolitics.com/device/K172944)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).