VITEK® 2 Gram Negative Ceftazidime/Avibactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ceftazidime/ Avibactam is a quantitative test. Ceftazidime/Avibactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active both in vitro and in clinical infections: Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa The following in vitro data are available, but clinical significance is unknown: Citrobacter koseri Enterobacter aerogenes Morganella morganii Providencia stuartii Serratia marcescens The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Streptococcus spp., Streptococcus spp. and clinically significant yeast.
Device Story
VITEK 2 AST-GN Ceftazidime/Avibactam is an automated, miniaturized, growth-based antimicrobial susceptibility test (AST) system. It utilizes 64-well test cards containing premeasured antimicrobial concentrations. Input consists of isolated bacterial colonies diluted in saline. The system automatically (VITEK 2) or manually (VITEK 2 Compact) fills, seals, and incubates cards. An optical scanner measures light attenuation every 15 minutes to monitor bacterial growth. The system calculates the Minimum Inhibitory Concentration (MIC) based on growth kinetics. Results are reported as MIC values and interpretive categories (Susceptible/Resistant). Used in clinical laboratories to guide antibiotic therapy for Gram-negative infections. The device provides rapid, standardized susceptibility data, aiding clinicians in selecting appropriate antimicrobial treatments and improving patient outcomes by identifying effective therapies.
Clinical Evidence
Performance evaluated via clinical and challenge studies (n=1073 total results). Compared to CLSI broth microdilution reference method. Overall Essential Agreement (EA) was 94.5% and Category Agreement (CA) was 98.7%. Major discrepancy rate was 1.4% (adjusted to 0.5% due to lack of intermediate category). P. aeruginosa showed a trend toward higher MICs; a limitation was added for repeat testing when MIC is ≥16 µg/mL.
Technological Characteristics
Miniaturized broth microdilution card containing antimicrobial concentrations. Automated incubation and optical growth monitoring system. Connectivity via VITEK 2/Compact instrument platforms. Software-based analysis of growth kinetics to determine MIC.
Indications for Use
Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli, including Citrobacter, Enterobacter, Escherichia, Klebsiella, Proteus, Pseudomonas, Morganella, Providencia, and Serratia species.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{0}
Page 1 of 15
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K172944
B. Purpose for Submission:
To obtain a substantial equivalence determination for Ceftazidime/Avibactam for testing of gram negative bacilli on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test (AST) Systems
C. Measurand:
The VITEK 2 AST-Gram Negative card contains the following concentrations of Ceftazidime/Avibactam: 0.06/4, 0.25/4, 1/4, 4/4, and 8/4 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The Ceftazidime/Avibactam MIC reporting ranges for the card are ≤ 0.12/4 – ≥16/4 µg/mL for Enterobacteriaceae and ≤ 0.25/4 – ≥16/4 µg/mL for Pseudomonas aeruginosa.
D. Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for Ceftazidime/Avibactam
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
VITEK 2 AST- GN Ceftazidime/Avibactam (≤ 0.12 – ≥16 µg/mL)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System
{1}
Page 2 of 15
2. Classification:
Class II
3. Product code(s):
LON – Fully automated short-term incubation cycle antimicrobial susceptibility system
LTW – Susceptibility Test Cards, Antimicrobial
LTT – Panels, Test, Susceptibility, Antimicrobial
4. Panel:
Microbiology, 83
H. Intended Use/Indications for Use:
1. Intended Use (s):
The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
2. Indications for Use:
VITEK 2 AST-Gram Negative Ceftazidime/Avibactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST-Gram Negative Ceftazidime/Avibactam is a quantitative test. Ceftazidime/Avibactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active both in vitro and in clinical infections
Citrobacter species
Enterobacter species
Escherichia coli
Klebsiella oxytoca
Klebsiella pneumoniae
Proteus mirabilis
Pseudomonas aeruginosa
The following in vitro data are available, but clinical significance is unknown:
Citrobacter koseri
Enterobacter aerogenes
{2}
Morganella morganii
Providencia stuartii
Serratia marcescens
The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
3. Special conditions for use statement(s):
For Prescription Use Only
Limitations:
- Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination(s):
- Ceftazidime/Avibactam: Providencia rettgeri
- Although within Essential Agreement when compared to the reference method, errors (major and/or very major) were observed due to the lack of intermediate category. Testing should be repeated using an alternative testing/reference method prior to reporting results for the following antibiotic/organism(s):
- Ceftazidime/Avibactam: Pseudomonas aeruginosa when the VITEK 2 MIC is ≥16μg/mL due to observed major errors (4.1%), with an adjusted categorical major error rate of 1.2%.
4. Special instrument requirements:
VITEK 2 and VITEK 2 Compact Systems using VITEK 2 Systems 9.01 software
I. Device Description:
The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 24 hours for Streptococcus species). The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration
Page 3 of 15
{3}
or "MIC" values for the anti-microbial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK 2 AST-GN Ceftazidime/Avibactam has the following concentrations in the card: 0.06/4, 0.25/4, 1/4, 4/4, and 8/4μg/mL (equivalent standard method concentration by efficacy in μg/mL). The Ceftazidime/Avibactam MIC result ranges for the VITEK 2 are:
- ≤ 0.12 – ≥ 16 μg/mL for Enterobacteriaceae
- ≤ 0.25 – ≥ 16 μg/mL for Pseudomonas aeruginosa
J. Substantial Equivalence Information:
1. Predicate device name(s):
VITEK 2 AST-GN Ceftolozane/Tazobactam
2. Predicate 510(k) number(s):
K161510
3. Comparison with predicate:
Page 4 of 15
{4}
Table 1: Comparison with Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device: VITEK 2 AST-GN Ceftazidime/Avibactam (K172944) | Predicate Device: VITEK 2 AST-GN Ceftolozane/Tazobactam (K161510) |
| Intended Use | The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. | Same |
| Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of Gram negative bacilli | Same |
| Inoculum | Standardized saline suspension of test organism | Same |
| Test Card | VITEK 2 Gram Negative Susceptibility Test Card | Same |
| Instrument | VITEK 2 and VITEK 2 Compact Systems | Same |
| Analysis Algorithm | Growth pattern analysis | Same |
| Differences | | |
| Antimicrobial Agent | Ceftazidime/Avibactam | Ceftolozane/Tazobactam |
| Antimicrobial Concentrations | 0.06/4, 0.25/4, 1/4, 4/4, and 8/4 μg/mL | 0.5/4, 1/4, 4/4, 8/4, and 32/4 μg/mL |
| Reporting Range | Enterobacteriaceae: ≤0.12 - ≥16 μg/mL P. aeruginosa: ≤0.25 - ≥16 μg/mL | Enterobacteriaceae and P. aeruginosa: ≤0.25 - ≥32μg/mL |
# K. Standard/Guidance Documents Referenced (if applicable)
FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
- CLSI M07-A10, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard-Ninth Edition" Vol. 35 No. 2 (January 2015)
- CLSI M100-S27, "Performance Standards for Antimicrobial Susceptibility Testing"; Twenty-Fourth Informational Supplement, Vol. 37 No. 1 (January 2017)
{5}
# L. Test Principle:
The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems uses visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well measures organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
A reproducibility study for the VITEK 2 AST-GN card with Ceftazidime/Avibactam was conducted at three external clinical sites using ten isolates of gram-negative bacilli consistent with the Intended Use. Testing was performed on three separate days and in triplicate for a total of 270 data points. The isolates tested in the reproducibility study included one isolate each for Klebsiella pneumoniae pneumoniae and Escherichia coli, two isolates each for Citrobacter freundii complex, Enterobacter aerogenes, Pseudomonas aeruginosa, and Serratia marcescens. Inocula were prepared using both the auto-dilution and manual dilution methods for testing in the VITEK 2 System. Inocula were prepared by the manual dilution method only for use with the VITEK 2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values that fell within $+ / -$ one doubling dilution from the mode MIC value. The majority of data points were on-scale and within $\pm$ one doubling dilution agreement as compared to the mode MIC (Table 2). Two data points were off-scale when tested by the VITEK 2 and the VITEK 2 Compact. The data was analyzed taking into consideration best case and worst case scenarios as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems.
The reproducibility performance is shown in Table 2.
Table 2: Reproducibility Performance
| | VITEK 2 | | VITEK 2 Compact |
| --- | --- | --- | --- |
| | Auto-Dilution | Manual Dilution | Manual Dilution |
| Best Case | 99.63% | 99.63% | 100% |
| Worst Case | 99.26% | 99.26% | 98.89% |
The combined reproducibility results for all three sites were acceptable.
{6}
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Inoculum Density Control**
The DensiCHEK Plus was used to standardize the inoculum to a 0.5 McFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.
**Purity Check**
A purity check of all organisms was performed on the dilution tube used to prepare the VITEK 2 card inoculum. Only those cultures that were pure were evaluated in the study.
**Growth Failure Rate**
There were ten isolates that failed to grow in the VITEK 2 card in the clinical study. Complete test results were available for 980 isolates from a total of 990 clinical isolates. The growth failure rate was 1.0% and was acceptable.
A challenge set of 93 isolates was evaluated at one site. All 93 challenge organisms grew in the VITEK 2 GN card with Ceftazidime/Avibactam using both the autodilution and manual dilution methods for the VITEK 2 and manual inoculation for the VITEK 2 Compact System.
A total of 1073 VITEK 2 AST results were available.
**Quality Control (QC) Testing**
The FDA/CLSI recommended QC organisms (E. coli ATCC 25922, E. coli ATCC 35218, Pseudomonas aeruginosa ATCC 27853, and K. pneumoniae pneumoniae ATCC 700603) were tested using both the VITEK 2 card and the reference method at each site. Both the automatic dilution and manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact.
As shown in Table 3, both the auto-dilution and the manual dilution methods for VITEK 2 and the manual dilution for VITEK 2 Compact were within the expected range >95% of the time.
Page 7 of 15
{7}
Table 3: Quality Control Summary Results for VITEK 2 (Auto-Dilution and Manual Dilution Methods) and VITEK 2 Compact (Manual Dilution Method)
| Ceftazidime/Avibactam | | VITEK 2 | | | | VITEK 2 Compact Manual Dilution | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Auto-Dilution | | Manual Dilution | | | |
| Organism | Conc. (μg/mL) | Reference | Test | Reference | Test | Reference | Test |
| E. coliATCC 25922FDA/CLSIExpected Range0.06 – 0.5μg/mL(VITEK 2:≤0.12-0.5μg/mL) | ≤0.015625 | | | | | | |
| | 0.03125 | | | | | | |
| | 0.0625 | 113 | | 62 | | 62 | |
| | ≤0.12** | 137 | 112 | 74 | 87 | 75 | 103 |
| | 0.25 | 7 | 146 | 4 | 23 | 4 | 38 |
| | 0.5 | | | | | | |
| | 1 | 1 | | | | | |
| | 2 | | | | | | |
| | ≥4 | | | | | | |
| E. coliATCC 35218FDA/CLSIExpected Range0.03-0.12μg/mL(VITEK 2:≤0.12μg/mL) | ≤0.015625 | | | | | | |
| | 0.03125 | 41 | | 13 | | 13 | |
| | 0.0625 | 204 | | 119 | | 119 | |
| | ≤0.12** | 8 | 252 | 4 | 135 | 4 | 136 |
| | 0.25 | | 1 | | 1 | | |
| | 0.5 | | | | | | |
| | 1 | | | | | | |
| | 2 | | | | | | |
| | ≥4 | | | | | | |
| K. pneumoniae pneumoniaeATCC 700603Expected Range0.025- 2μg/mL | ≤0.015625 | | | | | | |
| | 0.03125 | | | | | | |
| | 0.0625 | | | | | | |
| | 0.12 | | | | | | 1 |
| | 0.25 | 64 | | 18 | | 16 | |
| | 0.5 | 156 | 258 | 103 | 142 | 103 | 139 |
| | 1 | 39 | 1 | 21 | | 21 | |
| | 2 | | 1 | 1 | 1 | 1 | 1 |
| | ≥4 | | | | | | |
| | | | | | | | |
| P. aeruginosaATCC 27853Expected Range0.5 – 4 μg/mL | ≤0.0625 | | | | | | |
| | 0.12 | | | | | | |
| | 0.25 | | | | | | |
| | 0.5 | | | | | | |
| | 1 | 1 | 8 | 107 | 11 | 107 | 9 |
| | 2 | 196 | 244 | 26 | 124 | 26 | 124 |
| | 4 | 50 | | 3 | | 3 | |
| | 8 | 6 | | | | | |
| | 16 | | 1 | | 1 | | 3 |
| | 32 | | | | | | |
| | ≥64 | | | | | | |
* MIC for ceftazidime in the presence of a fixed concentration of 4 μg/mL of avibactam
** The lowest dilution of the VITEK 2 Ceftazidime/Avibactam MIC range is ≤0.12μg/mL. Obtaining this value was considered an indicator that the quality control test results were acceptable.
$^b$ K. pneumoniae ATCC 700603 should be tested against Ceftazidime/Avibactam and Ceftazidime alone to confirm the activity of Avibactam in the combination and to ensure that the plasmid encoding the beta-lactamase has not been lost in this strain. The acceptable range for Ceftazidime alone is greater than 16 μg/mL.
{8}
The FDA/CLSI Ceftazidime/Avibactam expected ranges for E. coli ATCC 25922 and E. coli ATCC 35218 are 0.06 – 0.5 and 0.03 – 0.12μg/mL respectively. However, the VITEK 2 MIC reporting range is ≤0.12 – ≥16 μg/mL (MIC results: ≤0.12, 0.25, 0.5, 1, 2, 4, 8, ≥16 μg/mL). The VITEK 2 systems do not provide results lower than 0.12 μg/mL; therefore, all results for E. coli ATCC 35218 and the lower range value for E. coli ATCC 25922 are off scale. The quality control test results were considered acceptable for E. coli ATCC 25922 and E. coli ATCC 35218 when the MIC value of ≤0.12 μg/mL was obtained for these organisms by the VITEK 2 system.
bioMérieux includes the FDA/CLSI broth microdilution expected range in the labeling when the VITEK 2 system range is not aligned with that of the FDA/CLSI range. The additional information is acceptable.
In addition, K. pneumoniae ATCC 700603 is recommended to test against ceftazidime alone to ensure the presence of the plasmid encoding beta-lactamase when testing this beta-lactam/beta-lactamase inhibitor combination, ceftazidime/avibactam. The acceptable result for ceftazidime alone is greater than 16μg/mL. The footnote below is added to the QC section of the Package Insert for VITEK 2 AST-GN Ceftazidime/Avibactam:
K. pneumoniae ATCC 700603 should be tested against Ceftazidime/Avibactam and Ceftazidime alone to confirm the activity of Avibactam in the combination and to ensure that the plasmid encoding the beta-lactamase has not been lost in this strain. The acceptable range for Ceftazidime alone is greater than 16 μg/mL.
d. Detection limit:
Not applicable
e. Analytical Specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Results obtained with the bioMérieux VITEK 2 AST-Gram Negative card with Ceftazidime/Avibactam were compared to results obtained with the CLSI broth microdilution reference panel. The following concentrations of Ceftazidime/Avibactam are contained in the VITEK 2 AST-GN test card: 0.06/4,
Page 9 of 15
{9}
0.25/4, 1/4, 4/4, and 8/4μg/mL (equivalent standard method concentration by efficacy in μg/mL). However, the reporting range is ≤ 0.12 – ≥16 μg/mL for Enterobacteriaceae and ≤ 0.25 – ≥16 μg/mL for Pseudomonas aeruginosa. The reference panel contained two-fold serial dilutions with a range of ≤ 0.0156 to ≥128 μg/mL. The testing conditions for the reference method consisted of the following:
- Medium – Mueller Hinton broth with appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension
- Incubation – 35°C ambient air incubator; 16-20 hours
All test inocula used for the evaluation of VITEK 2 AST-GN Ceftazidime/Avibactam and the reference method were standardized using the DensiCHEK Plus instrument. The cards were inoculated with each test organism by auto-dilution for reading by the VITEK 2 System and by manual dilution for reading on the VITEK 2 and VITEK 2 Compact Systems. Reference broth microdilution panels were inoculated in adherence with CLSI document, M07-A10.
A total of 990 clinical isolates were evaluated at three sites with VITEK 2 AST – Gram Negative cards inoculated by automatic dilution and interpreted using the VITEK 2 instrument. There were ten isolates that did not grow in the studies. The no growth rate was 0.1% (10/990). Most isolates were recently isolated from clinical specimen (885 isolates, 89.4%). The remainder were stock isolates (105 isolates, 10.6%).
A total of 93 challenge organisms (47 Enterobacteriaceae and 46 P. aeruginosa) were evaluated at one site. The challenge set was tested with the auto-dilution and manual dilution options of the VITEK 2 System and with the manual dilution method on the VITEK 2 Compact System. Overall performance of the clinical and challenge organisms by auto-dilution is shown in Table 4.1.
Page 10 of 15
{10}
Table 4.1: Performance ${}^{ \ddagger }$ of Clinical and Challenge Isolates,VITEK 2 Auto-Dilution Method
| Ceftazidime/Avibactam | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Enterobacteriaceae ≤8 (Susceptible), --, ≥16 (Resistant) | | | | | | | | | | | |
| Clinical | 819 | 770 | 94.0 | 265 | 216 | 81.5 | 813 | 99.3 | 10 | 6 | 0 |
| Challenge | 47 | 46 | 97.9 | 20 | 19 | 95.0 | 46 | 97.9 | 27 | 1 | 0 |
| Combined | 866 | 816 | 94.2 | 285 | 235 | 82.5 | 859 | 99.2 | 37 | 7 (0.8%) | 0 |
| Pseudomonas aeruginosa ≤8 (Susceptible), --, ≥16 (Resistant) | | | | | | | | | | | |
| Clinical | 161 | 152 | 94.4 | 145 | 136 | 93.8 | 154 | 95.7 | 10 | 7 | 0 |
| Challenge | 46 | 46 | 100 | 18 | 18 | 100 | 46 | 100 | 28 | 0 | 0 |
| Combined | 207 | 198 | 95.7 | 163 | 154 | 94.5 | 200 | 96.6 | 38 | 7 (4.1%) | 0 |
| Enterobacteriaceae + Pseudomonas aeruginosa | | | | | | | | | | | |
| Clinical | 980 | 922 | 94.1 | 410 | 352 | 85.9 | 967 | 98.7 | 20 | 13 | 0 |
| Challenge | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0 |
| Combined | 1073 | 1014 | 94.5 | 448 | 389 | 86.8 | 1059 | 98.7 | 75 | 14† (1.4%) | 0 |
The overall categorical major error rate for Enterobacteriaceae and Pseudomonas aeruginosa combined was $1.4\%$ (14/998). Nine (9) major errors were one dilution apart from the reference method and as such fall within essential agreement. Based on the essential agreement, and the lack of an intermediate breakpoint for Ceftazidime/Avibactam, the adjusted categorical major error rate is $0.5\%$ (5/998).
$\ddagger$ EA - Essential Agreement (+/- 1 doubling dilution)
CA - Category Agreement
EVAL - Evaluable isolates
R - Resistant isolates
Min - Minor discrepancies
Maj - Major discrepancies
Vmj - Very major discrepancies
Essential Agreement (EA) occurs when there is agreement between the result of the reference method and that of VITEK 2 test card within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the VITEK 2 test card and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the VITEK 2 test card.
The overall performance using the auto-dilution method for the VITEK 2 System demonstrated an essential agreement of $94.5\%$ and an overall category agreement of $98.7\%$ . The major discrepancy rate was acceptable at $1.4\%$ . However, due to the lack of intermediate category, the maj discrepancy rate was adjusted and was described in a footnote:
The overall categorical major error rate for Enterobacteriaceae and Pseudomonas aeruginosa combined was $1.4\%$ (14/998). Nine (9) major errors were one dilution apart from the reference method and as such fall within essential agreement. Based on the essential agreement, and the lack of an intermediate breakpoint for Ceftazidime/Avibactam, the adjusted categorical major error rate is $0.5\%$ (5/998).
# Pseudomonas aeruginosa (Reporting range: $\leq 0.25 / 4 - \geq 16 / 4\mu \mathrm{g} / \mathrm{mL}$ )
When $P.$ aeruginosa data was analyzed separately, it was observed that the major discrepancy rate for this organism was $4.1\%$ (7/169) and did not meet the acceptance criterion. Due to the lack of "intermediate" interpretation category, adjusted major discrepancy rate was calculated taking into consideration the MIC values were within essential agreement compared to reference MIC. Of the seven maj discrepancies, five
{11}
were within essential agreement [i.e. VITEK 2: $\geq 16$ (R), Reference: 8 (S)] which would have been considered minor discrepancies if there was an "intermediate" interpretative category for $P$ . aeruginosa. The adjusted major discrepancy rate was $1.2\%$ (2/169) which met the acceptance criterion.
The adjusted $P$ . aeruginosa performance is noted in Table 4.2 below.
Table 4.2: Adjusted Major Discrepancy Performance of $P$ . aeruginosa Due to the Lack of an Intermediate Category (VITEK 2 Auto-Dilution Method)
| Ceftazidime/Avibactam | Susceptible# | Major Discrepancies | |
| --- | --- | --- | --- |
| | | # | Rate |
| Before Adjustment | 169 | 7 | 4.1% |
| After Adjustment | 169 | 2† | 1.2% |
† Five (5) P. aeruginosa isolates were within essential agreement when compared to the reference method
The original $P$ aeruginosa performance with seven major discrepancies was mitigated by adding a limitation to address the need for repeat testing $P$ aeruginosa by alternative/reference methods when the VITEK 2 Systems MIC is $\geq 16\mu \mathrm{g / mL}$ . The limitation is noted below:
Although within Essential Agreement when compared to the reference method, errors (major and/or very major) were observed due to the lack of intermediate category. Testing should be repeated using an alternative testing/reference method prior to reporting results for the following antibiotic/organism(s):
- Ceftazidime/Avibactam: Pseudomonas aeruginosa when the VITEK 2 MIC is $\geq 16\mu \mathrm{g} / \mathrm{mL}$ due to observed major discrepancies $(4.1\%)$ , with an adjusted categorical major error rate of $1.2\%$ .
For implementation of the repeat testing limitation, bioMerieux indicated that they will issue a customer communication requesting VITEK 2 users to create a custom BioART rule in the VITEK 2 System software. Implementation of an indication for use BioART rule for a U.S. customer will prevent the user from generating a result when using the AST-GN Ceftazidime/Avibactam test for Pseudomonas aeruginosa.
The indication for use and limitation BioART rule will be pre-defined in the VITEK 2 System software in the next software update.
# MIC Trends:
The claimed Pseudomonas aeruginosa and Enterobacteriaceae organisms were also evaluated for trending. This trending calculation takes into account MIC values that are determined to be $\leq 1$ and $\geq 1$ doubling dilutions compared to the reference method irrespective whether the device MIC values are on-scale or not. The trending analysis is shown in Table 4.3 below.
{12}
Table 4.3: Trending Analysis of Evaluable Clinical and Challenge Results for P. aeruginosa and Enterobacteriaceae
| Ceftazidime/ Avibactam ≤ 1.25 – ≥ 16μg/mL | Total a (Total # Tested) | Difference in MIC as Compared to the CLSI Reference Method | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | ≥2 dil. lower | 1 dil. lower | Exact | 1 dil. higher | ≥2 dil. higher |
| P. aeruginosa + Enterobacteriaceae | 508/1073 | 15 | 75 | 155 (30.51%) | 209 | 54 |
| | | 90 (17.72%) b 95% CI (14.64% to 21.28%) | | | 263 (51.77%) b 95% CI (47.43% to 56.09%) | |
| P. aeruginosa | 168 (207) | 0 | 12 | 69 (41.07%) | 78 | 9 |
| | | 12 (7.14%) c 95% CI (4.13% to 12.07%) | | | 87(51.79%) c 95% CI (44.27% to 59.22%) | |
a Total number of evaluable results for trending analysis
b Difference between isolates trending higher and isolates trending lower for Enterobacteriaceae and P. aeruginosa: 34.06%; 95% CI (28.44% to 39.55%)
c Difference between isolates trending higher and isolates trending lower for P. aeruginosa: 44.64%; 95% CI (35.66% to 52.66%)
A trend towards higher MIC was observed with Enterobacteriaceae and P. aeruginosa isolates when compared to the CLSI broth microdilution reference method. This trending and the potential for occurrence of major discrepancies for Ceftazidime/Avibactam when testing clinical and challenge isolate results with the VITEK 2 GN Ceftazidime/Avibactam was addressed by adding the following footnote to the performance table for Ceftazidime/Avibactam in the device labeling (Table 127: Performance Characteristics for Gram-Negative Antimicrobial Susceptibility Testing, VITEK 2 Product Information Manual).
VITEK 2 Ceftazidime/Avibactam MIC values tended to be in exact agreement or at least one doubling dilution higher when testing Enterobacteriaceae and Pseudomonas aeruginosa compared to the reference broth micro-dilution.
## Challenge Data – Auto and Manual Dilution:
The challenge set of 93 isolates (47 Enterobacteriaceae and 46 Pseudomonas aeruginosa) was evaluated at one site. The challenge set was tested with both the autodilution and manual dilution options of the VITEK 2 System and with the manual dilution method on the VITEK 2 Compact System.
Overall performance is shown in Table 5.
Page 13 of 15
{13}
Table 5: Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact
| Ceftazidime/Avibactam | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| VITEK 2 | | | | | | | | | | | |
| Auto-dilution | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0 |
| Manual Dilution | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0 |
| VITEK 2 Compact | | | | | | | | | | | |
| Manual Dilution | 93 | 92 | 98.9 | 38 | 37 | 97.4 | 92 | 98.9 | 55 | 1 | 0 |
The performance of the challenge isolates using the VITEK 2 Ceftazidime/Avibactam test on both the VITEK 2 and VITEK 2 Compact Systems was acceptable.
# Enzyme Group Characterization/Resistance Markers Information
Enterobacteriaceae with beta-lactamases were included in the ceftazidime/avibactam comparative studies. They were KPC (6), CTX-M (4), TEM (11), SHV (12) and plasmid mediated AmpC (4). The Enterobacteriaceae were Klebsiella pneumoniae, E. coli, and Enterobacter cloacae. The study showed susceptibility to Ceftazidime/Avibactam with VITEK2 MIC values range from 0.5/4 to $2/4 \mu \mathrm{g} / \mathrm{mL}$ .
Additionally, seven $P_{\cdot}$ aeruginosa with beta-lactamases AmpC variants PDC (7), KPC (1), GES (1), VEB (1), and outer membrane porin loss OprD (7) were tested. Two isolates demonstrated resistance of $\geq 16 / 4~\mu \mathrm{g / mL}$ and five isolates showed susceptibility ranging from $2 / 4$ to $8 / 4~\mu \mathrm{g / mL}$ . The following footnote was added to the ceftazidime/avibactam labeling regarding other resistance markers and enzyme characterization:
> Enzyme group characterization was not available for the following organisms at the time of comparative testing, and therefore the performance of the AST card with Ceftazidime/Avibactam is unknown: Enterobacteriaceae (OXA, metallo-beta-lactamases); Pseudomonas aeruginosa (metallo-beta-lactamases).
> Ceftazidime/Avibactam is not active against bacteria that produce metallo beta-lactamase enzymes and may not have activity against Gram-negative bacteria that overexpress efflux pumps or have porin mutations.
# b. Matrix comparison:
Not applicable
{14}
3. Clinical studies:
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Table 6: Interpretive Criteria for Ceftazidime/Avibactam (FDA Drug Label)
| Organism | FDA Interpretive Criteria for Ceftazidime/Avibactam MIC (μg/mL) | | |
| --- | --- | --- | --- |
| | S | I | R |
| Enterobacteriaceae
Pseudomonas aeruginosa | ≤8/4 | -- | ≥16/4 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision
Page 15 of 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.