K172932 · C.R. Bard, Inc. · OZT · Dec 13, 2017 · Cardiovascular
Device Facts
Record ID
K172932
Device Name
True Flow Valvuloplasty Perfusion Catheter
Applicant
C.R. Bard, Inc.
Product Code
OZT · Cardiovascular
Decision Date
Dec 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1255
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172932 · Dec 13, 2017
True Flow Valvuloplasty Perfusion Catheter
C.R. Bard, Inc.
Post-market observational clinical study (TRUE-FLOW)
The sponsor included results from a post-market observational study to update the Instructions for Use (IFU) with clinical experience data, demonstrating successful valve dilation and safety outcomes in patients undergoing TAVI.
TRUE-FLOW study; Prospective, observational post-market study; Follow-up/Duration: Through hospital discharge
Adult male and non-pregnant females scheduled to undergo Transcatheter Aortic Valve Implantation (TAVI) for aortic stenosis; Sample Size: 24 (Analysis Population)
Not applicable for this study
Primary performance: successful dilatation of the aortic valve. Primary safety: freedom from device-related or procedure-related death, stroke, annulus rupture, coronary occlusion, ventricular perforation.
Indications for Use
The True® Flow Valvuloplasty Perfusion Catheter is indicated for balloon aortic valvuloplasty.
Device Story
The True Flow Valvuloplasty Perfusion Catheter is an over-the-wire co-axial catheter featuring a fixed, non-compliant balloon at the tip. Designed for use in clinical settings by physicians, the device allows continuous hemodynamic flow through a central orifice during balloon inflation. The catheter is positioned across the aortic valve using fluoroscopic guidance via two radiopaque markers. Radiopaque contrast media is delivered through an inflation lumen to dilate the stenotic aortic valve. The device enables complete valve dilation without requiring rapid pacing. The catheter is compatible with specific introducer sheaths and guidewires. Clinical benefit includes effective valve dilation prior to TAVI procedures.
Clinical Evidence
Prospective, observational study (TRUE-FLOW) of 24 subjects (adults with aortic stenosis undergoing TAVI). Primary performance endpoint: successful aortic valve dilatation. Primary safety endpoint: freedom from device/procedure-related death, stroke, annulus rupture, coronary occlusion, or ventricular perforation. Results: 87.5% (21/24) achieved successful dilatation without clinically significant movement; no device-related serious adverse events reported. Study demonstrated the device allows adequate blood flow for complete dilation without rapid pacing.
Technological Characteristics
Over-the-wire co-axial catheter; 110 cm length; non-compliant balloon; two-lumen design (inflation and guidewire); radiopaque markers for fluoroscopic positioning. Compatible with 11F-16F introducer sheaths depending on balloon diameter (18-26 mm). Sterilization method same as predicate.
Indications for Use
Indicated for balloon aortic valvuloplasty in adult patients scheduled for Transcatheter Aortic Valve Implantation (TAVI) for treatment of aortic stenosis.
Regulatory Classification
Identification
A balloon aortic valvuloplasty catheter is a catheter with a balloon at the distal end of the shaft, which is intended to treat stenosis in the aortic valve when the balloon is expanded.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components.
(3) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use, including device delivery, inflation, deflation, and removal.
(4) In vivo evaluation of the device must demonstrate device performance, including the ability of the device to treat aortic stenosis.
(5) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to the use of the device.
In combination with the general controls of the FD&C Act, the NuMED NuCLEUS and NuCLEUS-X BAV Catheters are subject to the following special controls:
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" above the words "U.S. FOOD & DRUG" and the word "ADMINISTRATION" below.
December 13, 2017
C. R. Bard, Inc. Aaron Conovaloff Regulatory Affairs Specialist 1625 West 3rd Street Tempe, Arizona 85281
Re: K172932
Trade/Device Name: True Flow Valvuloplasty Perfusion Catheter Regulation Number: 21 CFR 870.1255 Regulation Name: Balloon Aortic Valvuloplasty Catheter Regulatory Class: Class II Product Code: OZT Dated: September 25, 2017 Received: September 26, 2017
Dear Mr. Conovaloff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172932
#### Device Name
True® Flow Valvuloplasty Perfusion Catheter
Indications for Use (Describe)
The True® Flow Valvuloplasty Perfusion Catheter is indicated for balloon aortic valvuloplasty.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### True® Flow Valvuloplasty Perfusion Catheter
#### Submitter Information:
Bard Peripheral Vascular, Inc. 1625 West Third St. Tempe, AZ 85281
Phone: (480) 350-6012 Fax: (480) 449-2546
Contact Person: Aaron Conovaloff, Regulatory Affairs Specialist Date of Submission: September 25, 2017
### Subject Device Name:
| Name of Device: | True® Flow Valvuloplasty Perfusion Catheter |
|-----------------------|---------------------------------------------|
| Common or Usual Name: | Balloon Aortic Valvuloplasty |
| Classification Name: | Balloon Aortic Valvuloplasty |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR 870.1255 |
### Predicate Device:
| Name of Device: | True® Flow Valvuloplasty Perfusion Catheter (K152613 |
|-----------------------|------------------------------------------------------|
| Common or Usual Name: | Balloon Aortic Valvuloplasty |
| Classification Name: | Balloon Aortic Valvuloplasty |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR 870.1255 |
### Device Description:
The True® Flow Valvuloplasty Perfusion Catheter is an over-the-wire co-axial catheter with a balloon fixed at the tip. The balloon enables continuous hemodynamic flow through its central orifice. The catheter is 110 cm long and has two lumen is used to inflate and deflate the balloon and the other permits the use of a guidewire to position the catheter. The balloon inflation luer-lock hub (angled) connects to a syringe inflation device to deliver radiopaque contrast media for inflation. The guidewire luer-lock hub (straight) connects to the guidewire lumen. The balloon is non-compliant and is designed to reach a known diameter and length when inflated within the specified pressure range. Two radiopaque markers are located
{4}------------------------------------------------
on the guidewire lumen. These bands are positioned at the proximal and distal balloon shoulders. These markers are provided for fluoroscopic positioning of the device across the aortic valve. Balloon catheter dimensions, nominal pressure, maximum inflation pressure, recommended introducer size, and maximum guidewire size are indicated on the package label.
| Attribute | True® Flow Valvuloplasty Perfusion Catheter<br>Product Offering |
|-------------------------------------------------------------------|-----------------------------------------------------------------|
| Balloon Diameter<br>(mm) | 18, 20, 22, 24, 26 |
| Balloon Length<br>(cm) | 3.5 |
| Catheter Shaft<br>Length (cm) | 110 |
| Introducer Sheath<br>Compatibility by<br>Balloon Diameter<br>(mm) | 11F: 18 mm, 20 mm<br>12F: 22 mm<br>14F: 24 mm<br>16F: 26 mm |
## Indications for Use of Device:
The True® Flow Valvuloplasty Perfusion Catheter is indicated for balloon aortic valvuloplasty.
### Technological Comparison to Predicate Devices:
The True® Flow Valvuloplasty Perfusion Catheter has the following similarities to the predicate device, the True® Flow Valvuloplasty Perfusion Catheter (clearance to market via K152613 on January 19, 2016):
- . Same intended use
- . Same indications for use
- . Same target population
- Same operating principle .
- . Same fundamental scientific technology
- Same sterility assurance level and method of sterilization .
It should be noted that the subject True® Flow Valvuloplasty Perfusion Catheter is identical to the predicate device with respect to manufacturing and design. The only difference between the subject and predicate devices is an update to the Instructions for Use to provide results from clinical experience with the device.
{5}------------------------------------------------
#### Performance Data:
Subsequent to clearance of K152613, a post-market clinical study (the TRUE-FLOW study) was conducted on the predicate True® Flow Valvuloplasty Perfusion Catheter. Twenty- five subjects were enrolled, of which all were successfully treated with the study device. One subject withdrew consent following all study procedures and hospital discharge, and is thus not included in the Analysis Population (n =24). Subjects eligible to be enrolled in this study were adult male and non-pregnant females scheduled to undergo Transcatheter Aortic Valve Implantation (TAVI) for the treatment of aortic stenosis who had an annulus diameter that could be treated with the available sizes of the study device, in accordance with the IFU. The primary performance endpoint was successful dilatation of the aortic valve using the True® Flow Valvuloplasty Perfusion Catheter. The primary safety endpoint was freedom from device-related or procedure-related death, stroke, annulus rupture, coronary occlusion, ventricular perforation, during the pre-dilatation procedure. In 21 (87.5%) cases the True® Flow catheter successfully dilated the aortic annulus without clinically significant movement. For one subject these data were not recorded. No device related serious adverse events were reported during the study. The results of the TRUE-FLOW prospective, observational study indicated that the design of the True® Flow Valvuloplasty Perfusion Catheter allowed adequate blood flow through the device's central orifice to perform complete dilation of a stenotic aortic valve prior to TAVI without the need for rapid pacing.
#### Conclusions:
As the only difference between the subject and predicate devices is the addition of the above clinical experience to the IFU, there is no change to the intended use or indications for use, or any design specifications or manufacturing. Therefore the subject True® Flow Valvuloplasty Perfusion Catheter is substantially equivalent to the legally marketed predicate device, the True® Flow Valvuloplasty Perfusion Catheter.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.