Arm Type Automatic Blood Pressure Monitor BP310A, BP313A, BP351A, BP362A, BP366A, BP369A, BP380A, BP389A, BP-JC312
K172896 · Shenzhen Jiacom Technology CO ., Ltd. · DXN · Jan 26, 2018 · Cardiovascular
Device Facts
Record ID
K172896
Device Name
Arm Type Automatic Blood Pressure Monitor BP310A, BP313A, BP351A, BP362A, BP366A, BP369A, BP380A, BP389A, BP-JC312
Applicant
Shenzhen Jiacom Technology CO ., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jan 26, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Arm Type Automatic Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the arm. It can be used at medical facilities or at home. The intended arm circumference is 22-32 cm.
Device Story
Battery-powered, non-invasive oscillometric blood pressure monitor; measures systolic/diastolic pressure and pulse rate via inflatable arm cuff. Used in clinical or home settings by adults. Device automatically inflates cuff, processes pressure oscillations via internal MCU, and displays results on LCD. Includes data storage for review and WHO blood pressure classification indicator. Provides irregular heartbeat detection. Healthcare providers or patients use output to monitor cardiovascular status; aids in hypertension management and general health assessment.
Clinical Evidence
Clinical accuracy evaluated per ISO 81060-2:2013. Results demonstrated compliance with standard requirements for automated non-invasive sphygmomanometers.
Technological Characteristics
Oscillometric measurement principle; battery-powered (6V, 4x AA). Components: LCD, MCU, pump, cuff (Terylene), tube (PVC). Dimensions vary by model. Software level of concern: Moderate. Biocompatibility per ISO 10993-5 and ISO 10993-10. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 80601-2-30.
Indications for Use
Indicated for adult patients requiring non-invasive blood pressure and pulse rate measurement via oscillometric technique using an upper arm cuff (22-32 cm circumference). Suitable for use in medical facilities or home environments.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2018
Shenzhen Jiacom Technology Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CN
Re: K172896
Trade/Device Name: Arm Type Automatic Blood Pressure Monitor BP310A, BP351A, BP362A, BP366A, BP369A, BP380A, BP389A, BP-JC312 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 25, 2017 Received: December 28, 2017
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Diana Hong
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
. A. Wilhelm
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172896
#### Device Name
Arm Type Automatic Blood Pressure Monitor BP310A, BP313A, BP351A, BP362A, BP366A, BP369A, BP380A, BP389A, BP-JC312
#### Indications for Use (Describe)
Arm Type Automatic Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the arm. It can be used at medical facilities or at home. The intended arm circumference is 22-32 cm.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #7 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K172896
- 1. Date of Preparation: 12/20/2017
- 2. Sponsor Identification
#### Shenzhen Jiacom Technology Co., Ltd
No.A6 Building, silicon valley power, Qinghu park, Longhua street, Bao'an district, Shenzhen, Guangdong, 518109, China
Establishment Registration Number: Not yet registered.
Contact Person: Luo Tianfeng Position: Registration specialist Tel: +86-755-29015600 Fax: +86-755-29575792 Email: zhongjingming@szjiakang.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Betty Xiao (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Arm Type Automatic Blood Pressure Monitor Common Name: Arm Blood Pressure Monitor Models: BP310A, BP313A, BP351A, BP362A, BP366A, BP369A, BP380A, BP389A, BP-JC312
Regulatory information: Classification Name: Noninvasive blood pressure measurement system Classification: 2 Product Code: DXN Regulation Number: 21 CFR 870.1130 Review Panel: Cardiovascular
#### Intended Use Statement:
Arm Type Automatic Blood Pressure Monitor is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population via non-invasive oscillometric technique in which an inflatable cuff is wrapped around the arm. It can be used at medical facilities or at home. The intended arm circumference is 22-32 cm.
#### Device Description
The proposed device, Arm Type Automatic Blood Pressure Monitor, is a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation and measurement, which can measure systolic and diastolic blood pressure and pulse rate of the adult person at upper arm within its claimed range and accuracy via the oscillometric technique. User can select the unit of the measurement: mmHg or KPa.
The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. It has a bar indicating function, which can indicate the WHO (World Health Organization) Blood Pressure Classification of the measured blood pressure by referencing Diastolic Blood Pressure issued at Journal of Hypertension 1999. Vol 17, No.2.
The proposed device is intended to be used in medical facilities or at home. And it is provided non-sterile, and not to be sterilized by the user prior to use.
The proposed blood pressure monitor includes 9 models, which are BP310A, BP351A, BP362A, BP366A, BP369A, BP380A, BP389A, and BP-JC312. All models follow the same software, measurement principle and NIBP algorithm. The main differences are product appearance and key numbers.
{5}------------------------------------------------
- న. Identification of Predicate Device
510(k) Number: K131558 Product Name: Electronic Blood Pressure Monitor
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012, Medical electrical equipment - Part 1: General requirements for basic safety, and essential performance.
- > IEC 60601-1-2 Edition 4.0 2014-02, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.
- A IEC 60601-1-11 Edition 2.0 2015-01, Medical Electrical Equipment - Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment
- A IEC 80601-2-30 Edition 1.1 2013-07, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.
- > ISO 10993-5:2009, Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.
- > ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
- 7. Clinical Test Conclusion
The clinical accuracy of the proposed device was evaluated per the following standard. The test results demonstrated that the proposed device comply with the standard requirements.
ISO 81060-2 Second Edition 2013-05-01, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type.
- 8. Substantially Equivalent (SE) Comparison
{6}------------------------------------------------
| Item | Proposed Device | Predicate Device |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K131558 |
| Product Code | DXN | DXN |
| Regulation Number | 21 CFR 870.1130 | 21 CFR 870.1130 |
| Class | II | II |
| Intended Use | Arm Type Automatic Blood Pressure<br>Monitor is a digital monitor intended for<br>use in measuring blood pressure and pulse<br>rate in adult patient population via<br>non-invasive oscillometric technique in<br>which an inflatable cuff is wrapped around<br>the arm. It can be used at medical facilities<br>or at home. The intended arm<br>circumference is 22-32 cm. | Electronic Blood Pressure Monitor is<br>intended to measure the systolic and<br>diastolic blood pressure as well as the<br>pulse rate of adult person via<br>non-invasive oscillometric technique<br>in which an inflatable cuff is wrapped<br>around the arm. It can be used at<br>medical facilities or at home. The<br>intended arm circumference is 22-32<br>cm. |
| Measurement Site | Upper arm | Upper arm |
| Patient Population | Adult | Adult |
| Measurement Item | Systolic Pressure, Diastolic Pressure, Pulse<br>Rate | Systolic Pressure, Diastolic Pressure,<br>Pulse Rate |
| Measurement<br>Method | Oscillometric | Oscillometric |
| Component | LCD / Key / Cuff / MCU / Pump /<br>Batteries | LCD / Key / Cuff / MCU / Pump /<br>Batteries |
| Size | BP310A:<br>L133.2mm*W102.6mm*H65mm | PG-800B6D:<br>L138mm*W121mm*H83mm |
| | BP313A: L137mm*W121mm*H90mm | |
| | BP351A: L136mm*W95mm*H65mm | |
| | BP362A: L133mm*W103mm*H63mm | |
| | BP366A: L145mm*W105mm*H53mm | |
| | BP369A: L130mm*W108mm*H53mm | |
| | BP380A: L125mm*W71mm*H61mm | |
| | BP389A: L141.7mm*W96.5mm*H53mm | |
| | BP-JC312: L137mm*W121mm*H90mm | |
| | | |
| Weight | BP310A: Approx. 390g(with batteries) | PG-800B6D:<br>About 430g excluding batteries |
| | BP313A: Approx. 313g(with batteries) | |
| | BP351A: Approx.412g(with batteries) | |
| | BP362A: Approx.362(with batteries) | |
| | BP366A: Approx 405g(with batteries) | |
| Table 1 Comparison of Technology Characteristics | | | | |
|--------------------------------------------------|--|--|--|--|
|--------------------------------------------------|--|--|--|--|
{7}------------------------------------------------
| | BP369A: Approx.415g(with batteries)<br>BP380A: Approx.368g(with batteries)<br>BP389A: Approx.368g(with batteries)<br>BP-JC312: Approx.417g(with batteries) | Systolic pressure, diastolic pressure,<br>unit of pressure, pulse rate, pulse<br>indicator, date, time, WHO BP<br>classification indicating bar, low<br>battery indicator, memory |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display | Systolic pressure, diastolic pressure, unit<br>of pressure, pulse rate, irregular heartbeat,<br>pulse indicator, date, time, WHO BP<br>classification indicating bar, low battery<br>indicator, memory | |
| Blood Pressure<br>Range | 30 ~ 280 mmHg | 30 ~ 280 mmHg |
| Blood Pressure<br>Accuracy | 3 mmHg | 3 mmHg |
| Pulse Rate Range | 40 ~ 200 bpm | 40 ~ 199 bpm |
| Pulse Rate<br>Accuracy | $\pm$ 5% | $\pm$ 5% |
| WHO BP<br>Classification | Yes<br>Per WHO Classification | Yes<br>Per WHO Classification |
| Arm circumference | 22-32mm | 22-32mm |
| Particular<br>Performance | Comply with<br>IEC 80601-2-30:2009 and ISO<br>81060-2:2013 | Comply with ANSI/AAMI SP10 |
| Power Supply | four AA batteries | four AA or LR6 batteries |
| Voltage | 6V | 6V |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Patient-contact<br>component and<br>material | Cuff: Terylene<br>Tube: Polyvinyl chloride (PVC) | Cuff – Nylon<br>Enclosure - ABS<br>Key - ABS |
| Biocompatibility | No Cytotoxicity, sensitization, or irritation | Comply with ISO 10993<br>requirements |
| Software Level<br>Concern | Moderate | Moderate |
| Labeling | Conforms to 21CFR part 801 | Conforms to 21CFR part 801 |
The size, weight, and Pulse Rate Range of proposed device and predicate device are different. But the difference is very slight, and size, weight and Pulse Rate Range of proposed device is clearly indicated in user manual. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
The proposed device displays the irregular heartbeat, while the predicate device does not display. The
{8}------------------------------------------------
function of irregular heartbeat display has been validated per Software testing. Therefore, will not result in any safety and effectiveness issue of the proposed device.
The SP10 standard is no longer an FDA recognized consensus standard for evaluating the safety and performance of NIBP system, therefore the particular performance proposed device was conducted according to IEC 80601-2-30, the test results demonstrated that the proposed device meet the requirements of IEC 80601-2-30. In addition, the clinical investigation of the proposed device has been conducted according to ISO 81060-2:2013 to demonstrate the clinical accuracy. Therefore this difference on the particular performance will not affect the safety and effectiveness of the proposed device.
The patient contact components and materials of the proposed and predicate device are different. But both of them comply with ISO 10993 requirements. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.
#### Substantially Equivalent (SE) Conclusion 9.
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.