Optilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein Controls
Device Facts
| Record ID | K172868 |
|---|---|
| Device Name | Optilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein Controls |
| Applicant | The Binding Site Group , Ltd. |
| Product Code | DCN · Immunology |
| Decision Date | Feb 28, 2018 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5270 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K172868 · Feb 28, 2018 | Optilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein Controls | The Binding Site Group , Ltd. | Clinical serum samples | Clinical serum samples were used in a method comparison study to demonstrate substantial equivalence between the Optilite C-Reactive Protein Kit and the predicate device. | Method comparison; Clinical samples; Substantial equivalence |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Method comparison study | 110 clinical samples and 83 normal donor samples; Sample Size: 193 | Roche Diagnostics Tina-Quant C-Reactive Protein Gen. 3 assay | Method comparison (regression statistics and clinical concordance) |
Indications for Use
The Optilite C-Reactive Protein Reagent is intended for the quantitative in vitro determination of C-reactive protein (CRP) concentration in serum using the Binding Site Optilite analyser. Measurement of C-Reactive Protein aids in evaluation of the amount of injurv to body tissues and for evaluation of infection, tissue injurv, and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings. The Optilite C-Reactive Protein Calibrator is intended for the calibration of the Optilite C-Reactive Protein Reagent on the Optilite analyser. The Optilite C-Reactive Protein Controls are intended for use in quality control by monitoring accuracy and precision for the Optilite C-Reactive Protein Reagent.
Device Story
The Optilite C-Reactive Protein (CRP) system is an in vitro diagnostic test used in clinical laboratories. It utilizes serum samples to measure CRP concentration via immunoturbidimetry. The device consists of a reagent (antiserum and reaction buffer), calibrator, and controls, all designed for use on the Binding Site Optilite analyser. During operation, the sample reacts with specific anti-CRP antiserum to form insoluble complexes. The analyser passes light through the suspension and measures transmitted light via a photodiode; the intensity of transmitted light is inversely proportional to the CRP concentration. The system automatically calculates concentrations using a stored calibration curve. Healthcare providers use these results to assess tissue injury, infection, and inflammatory states in conjunction with other clinical findings. The system benefits patients by providing quantitative data to support diagnostic and monitoring decisions for inflammatory conditions.
Clinical Evidence
Bench testing only. Performance evaluated per CLSI guidelines (EP05-A3, EP06-A, EP17-A2, EP07-A2). Precision: total CV% 2.6–6.2%. Linearity: R=1.00 over 2.70–316.03 mg/L. LoQ: 5.00 mg/L. No interference observed for common endogenous substances or 14 therapeutic drugs. Method comparison (n=113) against predicate showed Passing-Bablok regression y = 1.00x + 5.56 (r=1.00). Reference interval verified against literature consensus (<5 mg/L).
Technological Characteristics
Immunoturbidimetric assay; goat anti-human CRP antiserum; reaction buffer (TRIS pH 7.5, 0.099% sodium azide). Photometric detection at 340 nm. Traceable to ERM-DA474. Automated processing on Optilite analyzer. Reagent stability: 5 months shelf-life, 3 months open-vial, 30 days on-board.
Indications for Use
Indicated for the quantitative determination of C-reactive protein (CRP) in serum to aid in the evaluation of tissue injury, infection, and inflammatory disorders in patients. Prescription use only.
Regulatory Classification
Identification
A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.
Predicate Devices
- Roche Diagnostics Tina-Quant C-Reactive Protein Gen. 3 (K083444)