Non-Sterile Powder Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs
K172864 · Maxter Glove Manufacturing Sdn Bhd · LZC · May 2, 2018 · General Hospital
Device Facts
Record ID
K172864
Device Name
Non-Sterile Powder Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs
Applicant
Maxter Glove Manufacturing Sdn Bhd
Product Code
LZC · General Hospital
Decision Date
May 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A powder free nitrile examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands and finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable nitrile examination glove; worn on hands/fingers by medical examiners to prevent cross-contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Provides barrier protection in clinical settings. Performance verified via bench testing for physical properties, dimensions, and chemical permeation breakthrough times. Not for use with certain drugs (e.g., Carmustine, Thiotepa) due to low permeation times.
Clinical Evidence
Bench testing only. No clinical data required. Testing included dimensions, physical properties, thickness, freedom from holes (ASTM D5151), powder residues (ASTM D6124), chemotherapy drug permeation (ASTM D6978-05), and biocompatibility (ISO 10993: primary skin irritation, dermal sensitization, and in vitro cytotoxicity).
Indicated for use as a disposable medical glove worn by examiners to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile. Tested for Use with Chemotherapy Drugs (K151750)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 2, 2018
Maxter Glove Manufacturing Sdn Bhd Yap Peak Geeh QA & Regulatory Affairs Manager Lot 6070, Jalan Haji Abdul Manah 6th Miles off Jalan Meru 41050 Klang, Selangor Malaysia
Re: K172864
Trade/Device Name: Non-Sterile Powder Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: March 30. 2018 Received: April 6, 2018
Dear Yap Peak Geeh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Geeta K.
Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known) K172864
R1/2001
#### Device Name
Non-Sterile Powder Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs
#### Indications for Use (Describe)
A powder free nitrile examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands and finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
| Tested Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time (Minutes) |
|-------------------------------------------------|-----------------------------------------------|
| Carmustine (BCNU). 3.3mg/ml (3,300ppm) | 8.3 |
| Cisplatin. 1.0 mg/ml (1,000ppm) | >240 |
| Cyclophosphamide (Cytoxan). 20mg/ml (20,000ppm) | >240 |
| Cytarabine 100mg/ml (100,000ppm) | >240 |
| Dacarbazine (DTIC), 10.0mg/ml (10,000ppm) | >240 |
| Doxorubicin Hydrochloride. 2.0mg/ml (2,000ppm) | >240 |
| Etoposide (Toposar), 20.0mg/ml (20,000ppm) | >240 |
| Fluorouracil 50.0mg/ml (50,000ppm) | >240 |
| Ifosfamide. 50.0mg/ml (50,000ppm) | >240 |
| Methotrexate 25mg/ml (25,000ppm) | >240 |
| Mitomycin C. 0.5mg/ml (500ppm) | >240 |
| Mitoxantrone 2.0mg/ml (2,000ppm) | >240 |
| Paclitaxel (Taxol). 6.0mg/ml (6,000ppm) | >240 |
| Thiotepa. 10.0mg/ml (10,000ppm) | 38.0 |
| Vincristine Sulfate 1.0mg/ml (1,000ppm) | >240 |
{3}------------------------------------------------
# 510(K) SUMMARY
## Date Prepared: 30th April 2018
### 1.0 Submitter:
| Name | : | Maxter Glove Manufacturing Sdn Bhd |
|-----------|---|---------------------------------------------------------------------------------------------------|
| Address | : | Lot 6070 Jalan Haji Abdul Manan,<br>6th Miles off Jalan Meru,<br>41050 Klang, Selangor, Malaysia. |
| Phone No. | : | 603-33929888 |
| Fax No. | : | 603-33923328 |
## 2.0 Name of the device:
Non-Sterile Powder Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Common Name Patient Examination Gloves : Brand Name Maxter Classification Name: Polymer Patient Examination Glove (21 CFR 880.6250 Product Code LZA)
# 3.0 Identification Of The Legally Marketed Devices that equivalency is claimed:
| | Predicate |
|-------------------|------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Kossan International Sdn Bhd |
| Device Name | Powder Free Nitrile Patient Examination<br>Glove, Blue Colored, Non-Sterile. Tested<br>for Use with Chemotherapy Drugs |
| 510(k) Number | K151750 |
| Regulation Number | 21 CFR 880.6250 |
| Regulatory Name | Patient Examination Glove |
| Regulatory Class | I |
Patient Examination Gloves Specialty (21 CFR 880.6250 Product Code LZC)
### 4.0 Description of the Device:
These patient examination gloves are formulated using nitrile, non-sterile, powder free, meet all the requirements of ASTM D6319 and tested for use with chemotherapy drugs.
### 5.0 Indications for Use statement for the subject device:
{4}------------------------------------------------
A Non-Sterile Powder Free Blue Nitrile Examination Glove Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands and finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05.
| Tested Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time (Minutes) |
|-------------------------------------------------|-----------------------------------------------|
| Carmustine (BCNU). 3.3mg/ml (3,300ppm) | 8.3 |
| Cisplatin. 1.0 mg/ml (1,000ppm) | >240 |
| Cyclophosphamide (Cytoxan). 20mg/ml (20,000ppm) | >240 |
| Cytarabine 100mg/ml (100,000ppm) | >240 |
| Dacarbazine (DTIC), 10.0mg/ml (10,000ppm) | >240 |
| Doxorubicin Hydrochloride. 2.0mg/ml (2,000ppm) | >240 |
| Etoposide (Toposar), 20.0mg/ml (20,000ppm) | >240 |
| Fluorouracil 50.0 mg/ml (50,000ppm) | >240 |
| Ifosfamide 50.0 mg/ml (50,000ppm) | >240 |
| Methotrexate 25mg/ml (25,000ppm) | >240 |
| Mitomycin C. 0.5mg/ml (500ppm) | >240 |
| Mitoxantrone 2.0mg/ml (2,000ppm) | >240 |
| Paclitaxel (Taxol) 6.0mg/ml (6,000ppm) | >240 |
| Thiotepa 10.0mg/ml (10,000ppm) | 38 |
| Vincristine Sulfate 1.0mg/ml (1,000ppm) | >240 |
Please note that the following drugs have low permeation time of less than 240minutes: Carmustine (BCNU), 3.3mg/ml: 8.3 Minutes Thiotepa. 10.0mg/ml: 38.0 Minutes
# 6.0 Summary of the Technological Characteristics Comparison:
Below is the summary of the technological characteristics of the Non-Sterile Powder Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs compare to ASTM D6319 or equivalent standards. Chemotherapy claim is similar to predicate device, which has a glove thickness of minimum 0.05mm and is shorter than 270mm but in compliance with the ASTM standards.
| Characteristics | Standards | Device Performance | |
|----------------------------------------------------------|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Predicate | Current |
| Manufacturer | | Kossan<br>International Sdn<br>Bhd | Maxter Gloves<br>Manufacturing Sdn<br>Bhd |
| 510 (K) Number | | K151750 | K172864 |
| Dimensions | ASTM D6319-10 | $\geq$ 230mm | $\geq$ 230mm |
| Physical Properties | ASTM D6319-10 | Meet | Meet |
| Thickness- Finger<br>-palm | ASTM D6319-10 | ≥ 0.05mm | ≥ 0.05mm |
| Freedom From<br>Holes | ASTM D6319-10<br>and ASTM D5151 | Pass | Pass |
| Powder Free<br>Residue | ASTM D6319-10<br>and ASTM D6124 | Meet | Meet |
| Resistance to<br>Permeation by<br>Chemotherapy<br>Drugs: | ASTM D6978-05 | | |
| Test Chemotherapy<br>Drug | Concentration | Minimum Breakthrough Detection Time<br>(min) | |
| Carmustine (BCNU) | 3.3mg/ml | 10.1 | 8.3 |
| Cisplatin | 1.0mg/ml | >240 | >240 |
| Cyclophosphamide<br>(Cytoxan) | 20mg/ml | >240 | >240 |
| Cytarabine | 100mg/ml | >240 | >240 |
| Dacarbazine (DTIC) | 10.0mg/ml | >240 | >240 |
| Doxorubicin<br>Hydrochloride | 2.0mg/ml | >240 | >240 |
| Etoposide (Toposar) | 20.0mg/ml | >240 | >240 |
| Ifosfamide | 50.0 mg/ml | >240 | >240 |
| Methotrexate | 25mg/ml | >240 | >240 |
| Mitomycin C. | 0.5mg/ml | >240 | >240 |
| Mitoxantrone | 2.0mg/ml | >240 | >240 |
| Paclitaxel (Taxol) | 6.0mg/ml | >240 | >240 |
| Thiotepa | 10.0mg/ml | 30.2 | 38.0 |
| Vincristine Sulfate | 1.0mg/ml | >240 | >240 |
| Warning Statement | | Warning: Please note<br>that the following<br>drugs have extremely<br>low permeation times:<br>Carmustine(BCNU):<br>10.1 minutes and<br>Thiotepa: 30.2<br>minutes | Warning : Please<br>note that the<br>following drugs have<br>low permeation time<br>of less than 240<br>minutes: Carmustine<br>(BCNU): 8.3<br>minutes and<br>Thiotepa: 38.0<br>minutes. Do not use<br>with Carmustine or<br>Thiotepa. |
| Biocompatibility | Primary Skin<br>Irritation-ISO 10993 | Under the conditions<br>of the study, this is a<br>non-irritant. | Under the conditions<br>of the study, this is a<br>non-irritant. |
| | Dermal<br>Sensitization-ISO<br>10993 | Under the conditions<br>of the study, this is a<br>non-sensitizer. | Under the conditions<br>of the study, this is a<br>non-sensitizer. |
| | In vitro Cytotoxicity-<br>ISO 10993 | Not available. | Under the conditions<br>of the study, the<br>device extract was<br>not cytotoxic. |
| Intended Use | - | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that<br>is worn on the<br>examiner's hand to<br>prevent<br>contamination<br>between patient and<br>examiner. These<br>gloves were tested for<br>use with<br>chemotherapy drugs<br>per ASTM D6978-05<br>(Reapproved 2013)<br>Standard Practice for<br>Assessment of<br>Medical Gloves to<br>Permeation by<br>Chemotherapy Drugs. | Non-Sterile Powder<br>Free Blue Nitrile<br>Examination Gloves<br>Tested for Use with<br>Chemotherapy Drugs<br>is a disposable<br>device intended for<br>medical purposes<br>that is worn on the<br>examiner's hands<br>and finger to prevent<br>contamination<br>between patient and<br>examiner. In<br>addition, these<br>gloves were tested<br>for use with<br>chemotherapy drugs<br>in accordance with<br>ASTMD6978-05. |
| Material | ASTM D6319-10 | Nitrile | Nitrile |
| Color | - | Blue, and white | Blue |
| Texture | - | Finger Textured | Finger Textured |
| Size | Medical Glove<br>Guidance Manual-<br>Labeling | Extra Small | Extra Small |
| | | Small | Small |
| | | Medium | Medium |
| | | Large | Large |
| | | Extra Large | Extra Large |
| Single Use | Medical Glove<br>Guidance Manual-<br>Labeling | Single use | Single use |
Table 1
{5}------------------------------------------------
{6}------------------------------------------------
# 7.0 Summary of Non-Clinical Performance Data
The non-clinical performance test results for the subject device met the acceptance criteria in the standard or test method for dimension, physical properties, thickness,
{7}------------------------------------------------
freedom from holes, powder residues, chemotherapy permeation, and biocompatibility testing.
# 8.0 Summary of Clinical Performance Data
Not applicable- clinical data is not needed for the intended use of this subject device.
# 9.0 Conclusion
Based on intended use, technological characteristics and non-clinical performance data, the Non-Sterile Powder Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs is substantially equivalent to the predicate device K151750.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.