Insight BD

K172832 · Siemens Medi Cal Solutions, Inc. · LLZ · Feb 6, 2018 · Radiology

Device Facts

Record IDK172832
Device NameInsight BD
ApplicantSiemens Medi Cal Solutions, Inc.
Product CodeLLZ · Radiology
Decision DateFeb 6, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Insight BD is a software application for the volumetric density assessment of digital x-ray images of the breast to aid health care professionals in the assessment of breast tissue composition. Insight BD is not a diagnostic aid and should be used only as adjunctive information when the final assessment of breast density category is made visually by medical professionals.

Device Story

Insight BD is a software application for volumetric breast density assessment. It processes digital mammograms and raw tomosynthesis projection images to generate density maps where pixel values represent fibroglandular tissue thickness. The software calculates fibroglandular tissue volume, total breast volume, volumetric breast density percentage, and a BI-RADS 5th Edition breast density grade. Used in clinical settings by healthcare professionals, the software operates on Siemens PC hardware or integrated into Siemens digital mammography systems. Results are stored in DICOM headers for documentation. The output serves as adjunctive information to assist radiologists in visual breast density categorization, potentially improving consistency in tissue composition assessment.

Clinical Evidence

Bench testing only. Performance verified through phantom measurements for accuracy, and large dataset analysis for reproducibility across different views (CC/MLO), imaging modes (FFDM, tomosynthesis, PRIME), and patient age cohorts. Consistency validated against known physiological trends (age-related density decrease). Equivalence to predicate established via comparative analysis of results on identical datasets and correlation with radiologist-assigned BI-RADS 5th Edition categories.

Technological Characteristics

Software-based volumetric density analysis. Operates on Windows-based PC or integrated mammography system hardware. Compliant with DICOM (NEMA PS 3.1-3.20) for image handling and reporting. Software lifecycle follows ISO 62304; risk management follows ISO 14971. Moderate level of concern software.

Indications for Use

Indicated for health care professionals to aid in the assessment of breast tissue composition using digital mammography or tomosynthesis images. Not for diagnostic use; intended as adjunctive information for visual assessment of breast density.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" underneath. Siemens Medical Solutions, Inc Denise Adams Regulatory Affairs Specialist 40 Liberty Boulevard 65-1A MALVERN, PA 19355 February 6, 2018 Re: K172832 Trade/Device Name: Insight BD Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: February 1, 2018 Received: September 19, 2017 Dear Denise Adams: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mechat D. O'Hara For Robert A. Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172832 Device Name Insight BD Indications for Use (Describe) Insight BD is a software application for the volumetric density assessment of digital x-ray images of the breast to aid health care professionals in the assessment of breast tissue composition. Insight BD is not a diagnostic aid and should be used only as adjunctive information when the final assessment of breast density category is made visually by medical professionals. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is an orange graphic of several dots arranged in a circular pattern. # 510(k) Summary: Insight BD Company Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Date Prepared December 15, 2017 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. ### 1. General Information Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869 ### Location of Manufacturing Site Siemens Healthcare GmbH Siemensstr. 1 91031 Forchheim, Germany Establishment Registration Number: 3004977335 ### 2. Contact Person Denise Adams, RAC Requlatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 adams.denise@siemens-healthineers.com ### Alternate Contact Person Patricia D. Jones, RAC Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 patricia.d.jones@siemens-healthineers.com {4}------------------------------------------------ - 3. Device Name and Classification | Trade Name | Insight BD | |---------------------------|----------------------------------------| | Classification Name | System, Image Processing, Radiological | | Classification Panel | Radiology | | Classification Regulation | 21 CFR §892.2050 | | Device Class | Class 2 | | Product Code | LLZ | ### 4. Legally Marketed Predicate Device | Trade Name | Volpara Imaging Software | |---------------------------|----------------------------------------------------------------| | 510(k) # | K152028 | | Clearance Date | 10/26/2015 | | Classification Name | System, Image Processing, Radiological | | Classification Panel | Radiology | | Classification Regulation | 21 CFR §892.2050 | | Device Class | Class 2 | | Product Code | LLZ | | Recall Information | This device has not been subject to a design<br>related recall | ### 5. Device Description The Siemens volumetric breast density analysis application software Insight BD provides volumetric assessment of digital x-ray images of the breast, including both digital mammograms and raw tomosynthesis projection images. The assessment consists of generating and evaluating density maps where the value at each pixel represents the thickness of fibroglandular tissue above that pixel. From those density maps the software calculates quantitative values, namely - . volume of fibroglandular tissue in the breast in cm³ - breast volume in cm³ . - . volumetric breast density in % - . breast density grade The breast density grade is computed by thresholding of the volumetric breast density and conforms to BI-RADS® 5th Edition breast composition categories. The masking effect of dense fibroglandular tissue is automatically included in the results. {5}------------------------------------------------ Insight BD is a software option for the Siemens digital mammography systems. Insight BD may be installed either on Siemens released PC hardware or on the Siemens digital mammography systems. Breast Density results are stored in the DICOM header and can be used for further documentation. ### 6. Indications for Use Insight BD is a software application for the volumetric density assessment of digital x-ray images of the breast to aid health care professionals in the assessment of breast tissue composition. Insight BD is not a diagnostic aid and should be used only as adjunctive information when the final assessment of breast density category is made visually by medical professionals. ### 7. Substantial Equivalence The Siemens Insight BD volumetric breast density application software is substantially equivalent to the commercially available Volpara Imaging Software (K152028). The Insight BD has the same intended use, fundamental scientific technologies and performance characteristics as the predicate device Volpara Imaging Software and the performance of the Insight BD application software has been verified with similar tests as the predicate. The substantial equivalence comparison table below compares the main performance data of the subject device Siemens Insight BD to the cleared predicate Volpara Imaging Software (K152028). | Features /<br>Technology | Subject Device<br>Insight BD | Predicate Device<br>Volpara Imaging Software<br>K152028 | Remarks | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------| | Indications for Use | Insight BD is a software<br>application for the volumetric<br>density assessment of digital<br>x-ray images of the breast to<br>aid health care professionals<br>in the assessment of breast<br>tissue composition. Insight<br>BD is not a diagnostic aid and<br>should be used only as<br>adjunctive information when<br>the final assessment of<br>breast density category is<br>made visually by medical<br>professionals. | Volpara is a software<br>application intended for use<br>with the raw data from digital<br>breast x-ray systems,<br>including tomosynthesis.<br>Volpara calculates a density<br>map and from that<br>determines volumetric breast<br>density as a ratio of<br>fibroglandular tissue and total<br>breast volume estimates.<br>Volpara provides these<br>numerical values along with a<br>BI-RADS breast density 4th<br>or 5th Edition category to aid<br>health care professionals in<br>the assessment of breast<br>tissue composition. Volpara<br>produces adjunctive | Same meaning,<br>only wording is<br>different | {6}------------------------------------------------ | Features /<br>Technology | Subject Device<br>Insight BD | Predicate Device<br>Volpara Imaging Software<br>K152028 | Remarks | |------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | | | information. It is not an<br>interpretive or diagnostic aid. | | | Intended Users | Health Care Professionals | Health Care Professionals | Same | | Image Source | Digital mammograms from<br>• mammography or<br>• tomosynthesis systems | Digital mammograms from<br>• mammography or<br>• tomosynthesis systems | Same | | Anatomical Area | Breast | Breast | Same | | Assessment Scope | Volumetric | Volumetric | Same | | Operating<br>Environment | Windows | Windows | Same | | Image Storage and<br>Report Generation | Yes<br>Output to the console | Yes<br>Output to the console | Same | | Numeric Output | • Volume of Fibroglandular<br>tissue<br>• Volume of Breast<br>• Volumetric Breast Density<br>• Siemens Breast Density<br>Grade corresponding to<br>BI-RADS® 5th Edition<br>Breast Density Category | • Volume of Fibroglandular<br>tissue<br>• Volume of Breast<br>• Volumetric Breast Density<br>• Volpara Density Grade<br>corresponding to BI-RADS®<br>4th and 5th Edition Breast<br>Density Category | Siemens<br>volumetric breast<br>density<br>information<br>corresponds to BI-<br>RADS® 5th Edition<br>breast density<br>category only | | Storage of Breast<br>Density Information<br>in Image | Yes | No | Siemens<br>volumetric breast<br>density<br>information is<br>stored in the<br>DICOM header | | Classification | 21 CFR 892.2050; LLZ | 21 CFR 892.2050; LLZ | Same | | Software Level of<br>Concern | Moderate | Moderate | Same | ## 8. Summary of Technological Characteristics of Subject Device as Compared with the Predicate Device The Insight BD software is based on similar technology and algorithms to calculate the volumetric density as the predicate Volpara Imaging Software. ### 9. Performance Data The Siemens Insight BD has been verified and validated according to the company's design control process. A risk analysis compliant with ISO 14971 has been provided and incorporated into the development effort. Software testing included both unit level and integrated system level testing. {7}------------------------------------------------ The device was tested and determined to be compliant to the following standards: | ISO 14971 | Medical devices – Application of risk management to medical devices | |------------------|---------------------------------------------------------------------| | ISO 62304 | Medical device software – Software life cycle processes | | NEMA PS 3.1-3.20 | Digital Imaging and Communications in Medicine (DICOM) Set | Density parameters were evaluated using various image data sets and results were compared against predicate performance, when appropriate. - . Measurement accuracy was assessed by comparing Insight BD estimates with known values from breast phantoms. - . Reproducibility was assessed by running Insight BD over substantial data sets and the results for left and right breasts and CC and MLO views were compared to confirm that the results were similar for each view and each breast. - . Reproducibility was assessed by running Insight BD over substantial data sets where the same woman had been imaged on Siemens mammography systems in both tomosynthesis (DBT) and mammography (FFDM) modes, and the results were compared to confirm they were similar. - . Reproducibility was assessed by running Insight BD over substantial datasets where the same women had been imaged on Siemens units running in both mammography and PRIME modes and the results were compared to confirm they were similar. - . Consistency was assessed by running Insight BD over substantial datasets where the women's age was known and results were compared with the expected and known decrease in breast density with age. - . Relationship to visual assessment was assessed by running Insight BD over x-ray images for which a BI-RADS® 5th Edition density category was available from MQSA qualified radiologists followed by a comparison of the two sets of data. - . Equivalence was assessed by running both the predicate device Volpara and Insight BD over substantial datasets and the results were compared to confirm they were similar. {8}------------------------------------------------ ### 10. General Safety and Effectiveness Concerns The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification and validation testing to ensure that the product meets its intended uses. ## 11.Conclusion as to Substantial Equivalence The Siemens volumetric breast density analysis application software Insight BD has the same intended use, fundamental scientific technology and performance characteristics as the predicate device. The 510(k) premarket notification for Insight BD contains adequate information and data to demonstrate substantial equivalence to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...