The 2MP LCD Monitor (G22S+, G22SP+, C22S+, C22SP+) is intended to be used in displaying and viewing digital images for review, analysis and diagnosis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
Device Story
2MP LCD monitor (monochrome and color models) for medical image display; inputs digital video signals (DVI-D, DisplayPort, VGA) from imaging systems; transforms signals into visual output via LED-backlit TFT LCD panels; features built-in brightness stabilization and anti-glare coating; used in clinical settings by trained medical practitioners for review, analysis, and diagnosis; output allows clinicians to visualize medical images for diagnostic decision-making; benefits include stable brightness and persistent calibration throughout device life.
Clinical Evidence
Bench testing only. No clinical or animal data. Testing verified conformance to DICOM GSDF and AAPM TG18 guidelines, including luminance/chromaticity non-uniformity, spatial resolution (MTF), and pixel defect analysis. Results demonstrated equivalence to predicate devices.
Technological Characteristics
21.3" TFT LCD panel (monochrome or color), LED backlight, 1600x1200 resolution, 1400:1 contrast ratio, 500 cd/m² typical brightness. Connectivity: DVI-D, DisplayPort, VGA. Power: DC 12V/6.0A. Features: anti-glare hard coating, brightness stabilization circuit, ambient light sensor, backlight sensor. VESA DDC 2B plug-and-play. Operating temp: 5°C-35°C.
Indications for Use
Indicated for trained medical practitioners to display and view digital images for review, analysis, and diagnosis. Contraindicated for the display of mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
EIZO Corporation, 2MP Color LCD Monitor, RadiForce RX250 and RX250-AR (K160247)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services, which features a stylized human figure. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 27, 2017
Shenzhen Beacon Display Technology Co., Ltd. % Fu Ailing Document Engineer 12F, Block B1, Nanshan Zhiyuan, No.1001 Xueyuan Road Shenzhen Guangdong 518055 CHINA
### Re: K172815
Trade/Device Name: 2MP LCD Monitor (G22S+, G22SP+, C22S+, C22SP+) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: August 29, 2017 Received: September 18, 2017
Dear Fu Ailing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Robert Oels
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
#### K172815
Device Name
2MP LCD Monitor (G22S+, G22SP+, C22S+, C22SP+)
#### Indications for Use (Describe)
The 2MP LCD Monitor (G22S+, G22S+, C22S+, C22SP+) is intended to be used in displaying and viewing digital images for review, analysis and diagnosis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
Type of Use (Select one or both, as applicable)
V Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR807.92 (a) (1)]
August 29, 2017
## 2. Submitter's Information [21 CFR807.92 (a) (1)]
| Name of Sponsor: | Shenzhen Beacon Display Technology Co., Ltd. | |
|------------------|---------------------------------------------------------------------------------------------------|--|
| Address: | 12F, Block B1, Nanshan Zhiyuan, No.1001 Xueyuan Road, Nanshan<br>District, Shenzhen, 518055 China | |
| Contact Name: | Fu Ailing | |
| Telephone No.: | +86-0755-33985118-5150 | |
| Fax No.: | +86-0755-26995755 | |
| Email Address: | fuailing@beacon-display.cn | |
## 3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]
Trade Name/Model: 2MP LCD Monitor (G22S+, G22SP+, C22S+, C22SP+) Common Name: 2MP LCD Monitor Classification Name: Picture archiving and communications system Regulation Number: 21 CFR 892.2050 Product code: PGY Classification Panel: Radiology Device Class: II
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]
The identified predicates within this submission are as follows: EIZO NANAO Corporation, 2MP Monochrome LCD Monitor, RadiForce GX240 has been cleared by FDA through 510(k) No. K120407 (Decision Date - April 19, 2012).
{4}------------------------------------------------
EIZO Corporation, 2MP Color LCD Monitor, RadiForce RX250 and RX250-AR has been cleared by FDA through 510(k) No. K160247 (Decision Date - February 23, 2016).
## 5. Description of the Device [21 CFR 807.92(a) (4)]
Beacon LCD monitor (G22S+, G22SP+, C22S+, C22SP+) is for viewing medical images other than those of mamography. Its using the latest generation of LED back light panel, with the resolution 1600 x 1200 (landscape), 1200 x 1600 (portrait), built-in brightness stabilization circuit, the monitor can guarantee the stable brightness and persistent calibration throughout its life. The anti-glare screen can prevent display from reflection under highlight conditions, making the image and display clearer. The monitor supports DVI-D, DP and VGA.
Model variations are distinguished by characters. G means the Monochrome monitor, C means the Color monitor, and P means the monitor with an additional glass screen. For example, G22S+ is a monochrome LCD monitor; C22SP+ is a color LCD monitor with the additional glass screen.
## 6. Intended Use [21 CFR 807.92(a)(5)]
The 2MP LCD Monitor (G22S+, G22SP+, C22S+, C22SP+) is intended to be used in displaying and viewing digital images for review, analysis and diagnosis by trained medical practitioners. It does not support the display of mammography images for diagnosis.
## 7. Technological Characteristics [21 CFR 807.92(a)(6)]
| Panel | 21.3", TFT, monochrome, LCD screen, anti-glare,<br>hard coating |
|-------------------|-----------------------------------------------------------------|
| Brightness (Typ.) | 500 cd/m² |
| CR (Typ.) | 1400:1 |
| Viewing angle | R/L 176°, U/D 176° Typ. (CR > 10) |
| Pixel Pitch | (H) 0. 270 × (V) 0.270 mm |
### G22S+/G22SP+ Monochrome LCD monitor
{5}------------------------------------------------
| Native resolution | 1,200 x 1,600 | |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------|--|
| Display area | 324 mm(H) x 432 mm (V) | |
| Aspect ratio | 3:4 | |
| Screen size | 21.3" real diagonal | |
| Power | DC 12 V/6.0 A | |
| Power consumption | Max. 40 W | |
| Input signals | DVI-D, Display Port, VGA | |
| Plug and play | VESA DDC 2B | |
| Dimension | 369 mm (W) x 506 ~ 626 mm (H) x 235 mm (D) (with stand)<br>369 mm (W) x 483 mm (H) x 65 mm (D) (without stand) | |
| Weight | 6.0 kg (without Stand)<br>9.2 kg (with Stand) | |
| Operating<br>temperature and<br>humidity | Temperature: 5°C ~ 35°C<br>Humidity: 15% ~ 85% | |
| Storage temperature<br>and humidity | Temperature: -20°C ~ 60°C<br>Humidity:10% ~ 90% | |
### C22S+/C22SP+ Color LCD monitor
| Panel | 21.3", TFT, color, LCD screen, anti-glare, hard coating |
|-------------------|---------------------------------------------------------|
| Brightness (Typ.) | 500 cd/m² |
| CR (Typ.) | 1400:1 |
| Viewing angle | R/L 176°, U/D 176° Typ. (CR > 10) |
| Pixel Pitch | (H) 0. 270 × (V) 0.270 mm |
| Native resolution | 1,200 x 1,600 |
| Display area | 324 mm(H) x 432 mm (V) |
| Aspect ratio | 3:4 |
| Screen size | 21.3" real diagonal |
{6}------------------------------------------------
| Power | DC 12 V/6.0 A | |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--|
| Power consumption | Max. 60 W | |
| Input signals | DVI-D, Display Port, VGA | |
| Plug and play | VESA DDC 2B | |
| Dimension | 369 mm (W) x 506 ~ 626 mm (H) x 235 mm (D) (with<br>stand)<br>369 mm (W) x 483 mm (H) x 65 mm (D) (without<br>stand) | |
| Weight | 6.0 kg (without Stand)<br>9.2 kg (with Stand) | |
| Operating<br>temperature and<br>humidity | Temperature: 5°C ~ 35°C<br>Humidity: 15% ~ 85% | |
| Storage temperature<br>and humidity | Temperature: -20°C ~ 60°C<br>Humidity:10% ~ 90% | |
## 8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]
### 8.1 Intended uses:
| ID | Comparison<br>Item | Proposed Device<br>2MP Monochrome LCD<br>Monitor (G22S+, G22SP+) | Predicate Device<br>2MP Monochrome LCD<br>Monitor (RadiForce GX240) |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Intended Use | The 2MP LCD Monitor<br>(G22S+, G22SP+) is intended<br>to be used in displaying and<br>viewing digital images for<br>review, analysis and diagnosis<br>by trained medical<br>practitioners. It does not<br>support the display of<br>mammography images for<br>diagnosis. | The RadiForce GX240 is<br>intended to be used in displaying<br>and viewing digital images by<br>trained medical practitioners. The<br>RadiForce GX240 does not<br>support the display of<br>mammography images for<br>diagnosis. |
| | | | | | Table 1 Intended Use Comparison of G22S+/G22SP+ |
|--|--|--|--|--|--------------------------------------------------|
|--|--|--|--|--|--------------------------------------------------|
{7}------------------------------------------------
| ID | Comparison Item | Proposed Device<br>2MP Color LCD Monitor<br>(C22S+, C22SP+) | Predicate Device<br>2MP Color LCD Monitor<br>(RX250, RX250-AR) |
|----|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Intended Use | The 2MP LCD Monitor (C22S+,<br>C22SP+) is intended to be<br>used in displaying and viewing<br>digital images for review,<br>analysis and diagnosis by<br>trained medical practitioners. It<br>does not support the display of<br>mammography images for<br>diagnosis. | This product is intended to be<br>used in displaying and viewing<br>digital images for review, analysis<br>and diagnosis by trained<br>medical practitioners. It does not<br>support the display of<br>mammography images for<br>diagnosis. |
### Table 2 Intended Use Comparison of C22S+/C22SP+
## 8.2 Comparison table
| ID | Comparison<br>Item | Proposed Device<br>2MP Monochrome<br>LCD Monitor (G22S+,<br>G22SP+) | Predicate Device<br>2MP Monochrome<br>LCD Monitor<br>(RadiForce GX240) | Explanation of<br>Difference |
|------|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| 2 | Display Performance/Specifications | | | |
| 2.1 | Screen<br>Technology | Monochrome TFT<br>LCD Panel (IPS) | Monochrome TFT<br>LCD Panel (IPS) | - |
| 2.2 | Viewing angle<br>(H, V) | H: 176°, V: 176° | H: 176°, V: 176° | - |
| 2.3 | Resolution | 2MP (1,200 x 1,600) | 2MP (1,200 x 1,600) | - |
| 2.4 | Aspect ratio | 3 : 4 | 3 : 4 | - |
| 2.5 | Active screen<br>size | 324.0 mm x 432.0 mm | 324.0 mm x 432.0 mm | - |
| 2.6 | Pixel pitch | 0.270 mm x 0.270 mm | 0.270 mm x 0.270 mm | - |
| 2.7 | Typical<br>luminance | 1900 cd/m² | 1200 cd/m² | Different screen |
| | | | | |
| 2.8 | DICOM calibrated luminance | 500 cd/m² | 500 cd/m² | - |
| 2.9 | Contrast ratio | 1400:1 | 1400 : 1 | - |
| 2.10 | Backlighting | LED | LED | - |
| 2.11 | Grayscale Tones | 10-bit (DisplayPort):<br>1,024 from a palette of 16,369 tones<br>8-bit: 256 from a palette of 16,369 tones | 10-bit (DisplayPort):<br>1,024 from a palette of 16,369 tones<br>8-bit: 256 from a palette of 16,369 tones | - |
| 2.12 | Luminance non-uniformity compensation | - | Digital Uniformity Equalizer | Different design scheme |
| 3 | Video Signals | | | |
| 3.1 | Input video signals | DVI-D x 1,<br>DisplayPort x 1,<br>VGA x 1 | DVI-D x 1,<br>DisplayPort x 1 | Different design scheme |
| 3.2 | Ouput video signals | - | DisplayPort x 1 (daisy chain) | Different design scheme |
| 3.3 | Scanning Frequency (H, V) | 31 - 100 kHz / 59 - 61 Hz<br>Frame synchronous mode: 59 - 61 Hz | 31 - 100 kHz / 59 - 61 Hz<br>Frame synchronous mode: 59 - 61 Hz | - |
| 4 | Power Related Specifications | | | |
| 4.1 | Power Requirements | DC 12 V/6.0 A | AC 100 -120 V, 200 - 240 V: 50 / 60 Hz | Difference between Built-in power supply and Built-out power supply |
| 4.2 | Power Consumption / Save Mode | 40 W/Less than 5 W | 76 W / Less than 1.6 W | Different design scheme |
| 4.3 | Power Management | DVI DMPM,<br>DisplayPort 1.1a | DVI DMPM,<br>DisplayPort 1.1a | - |
| 5 | Miscellaneous Features/Specifications | | | |
| 5.1 | QC software | Beacon Monitor<br>Manage | RadiCS | Different design<br>scheme |
| 5.2 | Sensors | G22S+:<br>Backlight Sensor,<br>Ambient Light Sensor<br><br>G22SP+:<br>Backlight Sensor,<br>Ambient Light Sensor<br>Integrated Front<br>Sensor | Backlight Sensor,<br>Presence Sensor,<br>Integrated Front<br>Sensor<br>Ambient Light<br>Sensor | Different design<br>scheme |
| 5.3 | USB<br>Ports/Standard | 1 upstream,<br>2 downstream | 1 upstream,<br>2 downstream | - |
| 5.4 | Dimensions<br>w/o stand (W x<br>H x D) | 369 x 483 x 65 mm | 376 x 505 x 98 mm | Different housing<br>design due to the<br>different panel<br>size |
#### Table 3 General Comparison of G22S+/G22SP+
{8}------------------------------------------------
{9}------------------------------------------------
## Table 4 General Comparison of C22S+/C22SP+
| ID | Comparison<br>Item | Proposed Device<br>2MP Color LCD<br>Monitor (C22S+,<br>C22SP+) | Predicate Device<br>2MP Color LCD<br>Monitor<br>(RX250, RX250-AR) | Explanation of<br>Difference |
|------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| 2 | Display Performance/Specifications | | | |
| 2.1 | Screen<br>Technology | IPS TFT Color LCD<br>Panel | IPS TFT Color LCD<br>Panel | - |
| 2.2 | Viewing angle<br>(H, V) | H: 176°, V: 176° | H: 176°, V: 176° | - |
| 2.3 | Resolution | 2MP (1,200 x 1,600) | 2MP (1,200 x 1,600) | - |
| 2.4 | Aspect ratio | 3 : 4 | 3 : 4 | - |
| 2.5 | Active screen<br>size | 324.0 mm x 432.0 mm | 324.0 mm x 432.0 mm | - |
| 2.6 | Pixel pitch | 0.270 mm x 0.270 mm | 0.270 mm x 0.270 mm | - |
| | | | | |
| 2.7 | Typical<br>luminance | 900 cd/m² | 800 cd/m² | Different screen |
| 2.8 | DICOM<br>calibrated<br>luminance | 500 cd/m² | 400 cd/m² | Different screen |
| 2.9 | Contrast ratio | 1400 : 1 | 1400 : 1 | - |
| 2.10 | Backlighting | LED | LED | - |
| 2.11 | Display Colors | From a palette of 68<br>billion colors:<br>- 10-bit input<br>(DisplayPort): 1.07<br>billion colors<br>(maximum)<br>- 8-bit input: 16.77<br>million colors | From a palette of 68<br>billion colors:<br>- 10-bit input<br>(DisplayPort): 1.07<br>billion colors<br>(maximum)<br>- 8-bit input: 16.77<br>million colors | - |
| 2.12 | Luminance<br>non-uniformity<br>compensation | - | Digital Uniformity<br>Equalizer | Different design<br>scheme |
| 3 | Video Signals | | | |
| 3.1 | Input video<br>signals | DVI-D x 1,<br>DisplayPort x 1,<br>VGA x 1 | DVI-D x 1,<br>DisplayPort x 1 | Different design<br>scheme |
| 3.2 | Ouput video<br>signals | - | - | - |
| 3.3 | Scanning<br>Frequency (H,<br>V) | 31 - 100 kHz / 59 - 61<br>Hz<br>(VGA Text: 69 - 71 Hz)<br>Frame synchronous<br>mode: 59 - 61 Hz | 31 - 100 kHz / 59 - 61<br>Hz<br>(VGA Text: 69 - 71<br>Hz)<br>Frame synchronous<br>mode: 59 - 61 Hz | - |
| 4 | Power Related Specifications | | | |
| 4.1 | Power<br>Requirements | DC 12 V/6.0 A | AC 100 - 240 V: 50 /<br>60Hz | Difference<br>between Built-in<br>power supply<br>and Built-out<br>power supply |
| | | | | |
| 4.2 | Power<br>Consumption /<br>Save Mode | 60 W/Less than 5 W | 79 W / Less than 1.6<br>W | Different design<br>scheme |
| 4.3 | Power<br>Management | DVI DMPM,<br>DisplayPort 1.1a | DVI DMPM,<br>DisplayPort 1.1a | - |
| 5 | Miscellaneous Features/Specifications | | | |
| 5.1 | QC software | Beacon Monitor<br>Manage | RadiCS | Different design<br>scheme. |
| 5.2 | Sensors | C22S+:<br>Backlight Sensor,<br>Ambient Light Sensor<br>C22SP+:<br>Backlight Sensor,<br>Ambient Light Sensor<br>Integrated Front<br>Sensor | Backlight Sensor,<br>Presence Sensor | Different design<br>scheme |
| 5.3 | USB<br>Ports/Standard | 1 upstream,<br>2 downstream / Rev.<br>2.0 | 1 upstream,<br>2 downstream / Rev.<br>2.0 | - |
| 5.4 | Dimensions<br>w/o stand (W x<br>H x D) | 369 x 483 x 65 mm | 361 x 465 x 78 mm | Different housing<br>design due to the<br>different panel<br>size |
{10}------------------------------------------------
{11}------------------------------------------------
It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the G22S+, G22SP+, C22S+, C22SP+.
### 8.3 Performance Testing
The bench tests were performed on G22S+, G22SP+, C22S+, C22SP+ as below.
- Verify the conformance to DICOM GSDF in accordance with Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline).
- Measure the luminance non-uniformity characteristics of the display screen in accordance with TG18 guideline.
{12}------------------------------------------------
- Measure the chromaticity non-uniformity characteristics of the display screen in accordance with TG18 guideline.
- Measure the chromaticity at the center of the display screen at 5%, 50% and 95% of the maximum luminance.
- Visually check the presence or absence of miscellaneous artifacts on the display screen in accordance with TG18 guideline.
- Measure the spatial resolution expressed as modulation transfer function (MTF)
- Maximum number allowed for each type of pixel defects/faults
The test results showed that G22S+, G22SP+, C22S+, C22SP+ are with display characteristics equivalent to those of the predicate device, RadiForce GX240, RadiForce RX250 and RX250-AR except some items, each of which was determined that it would not affect observer's performance.
No animal or clinical testing is needed for G22S+, G22SP+, C22S+, C22SP+.
## 9. Conclusion [21 CFR 807.92(b) (3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Shenzhen Beacon Display Technology Co., Ltd. concludes that:
- The intended use of G22S+, G22SP+, C22S+, C22SP+ is totally same as that of the ● predicate device.
- The technological characteristics differences between G22S+, G22SP+, C22S+, C22SP+ and RadiForce GX240, RadiForce RX250 and RX250-AR do not affect the safety and effectiveness, no new risk is raised.
- Demonstrated by the bench tests, the display characteristics of G22S+, G22SP+, C22S+, C22SP+ are equivalent to those of the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.