K172807 · Changzhou Rongxin Medicine Minimal Invasion Technology · EZL · Jun 4, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K172807
Device Name
Silicone Foley Catheter for single use
Applicant
Changzhou Rongxin Medicine Minimal Invasion Technology
Product Code
EZL · Gastroenterology, Urology
Decision Date
Jun 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Two-way Disposable Silicone Foley Catheterization for bladder drainage and urological use only; the indwelling time is no more than 30 days. Three-way Disposable Silicone Foley Catheterization for bladder drainage and bladder irrigation for urological use only; the indwelling time is no more than 30 days.
Device Story
Silicone Foley catheter designed for intermittent or indwelling bladder drainage and irrigation; available in two-way and three-way configurations. Device consists of a silicone shaft with an inflatable balloon for retention; includes pediatric models with internal steel wire. Used in clinical settings for urological procedures; operated by healthcare professionals. Device provides a pathway for fluid drainage or irrigation; balloon secures position within bladder. Benefits include effective bladder management and reduced risk of trauma due to silicone material. Sterile, single-use device.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including performance testing per ASTM F623-99 (2013) and EN 1616:1997, radiopacity testing per ASTM F640-2012, and comprehensive biocompatibility testing per ISO 10993-1.
Technological Characteristics
Material: Silicone. Design: Balloon-retention catheter, two-way or three-way. Sizes: 6Fr-24Fr. Sterilization: Ethylene Oxide (EO). Radiopaque. Complies with ASTM F623-99, EN 1616:1997, and ASTM F640-2012.
Indications for Use
Indicated for pediatric, male, and female patients requiring urethral catheterization for bladder drainage (two-way) or bladder drainage and irrigation (three-way). Indwelling time limited to 30 days.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 4, 2018
Changzhou Rongxin Medicine Minimal Invasion Technology Co., Ltd. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd. 8-9th Floor, R&D Building, No. 26 Qinglan Street Panyu District Guangzhou, 510006 CHINA
Re: K172807
Trade/Device Name: Silicone Foley Catheter for Single Use Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: Class II Product Code: EZL Dated: May 4, 2018 Received: May 4, 2018
# Dear Mike Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K172807
Device Name Silicone Foley Catheter for single use
### Indications for Use (Describe)
Two-way Disposable Silicone Foley Catheterization for bladder drainage and urological use only; the indwelling time is no more than 30 days.
Three-way Disposable Silicone Foley Catheterization for bladder drainage and bladder irrigation for urological use only; the indwelling time is no more than 30 days.
### Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with a green leaf on the left and a blue wave on the right. The leaf and wave are arranged in a circular shape. The leaf is a bright green color, and the wave is a deep blue color. The logo is simple and modern, and it likely represents a company or organization that is focused on environmental sustainability or water conservation.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
# 1. SUBMITTER
CHANGZHOU RONGXIN MEDICINE MINIMAL INVASION TECHNOLOGY CO., LTD. No. 8, Jianerkang Road, Industrial Concentration Area, Zhixi Town, Jintan City, Jiangsu Province, China Tel: +86-0519-82446628 Fax: +86-0519-82446610
| Primary Contact | Mike Gu |
|-------------------|----------------------------------------------------|
| Person: | Regulatory Affairs Manager |
| | Guangzhou Osmunda Medical Device Technical Service |
| | Co., Ltd. |
| | Tel: (+86)-20-6231 6262 |
| | Fax: (+86) -20-8633 0253 |
| Secondary Contact | Ping Liu |
| Person: | Quality Assurance Manager |
| | CHANGZHOU RONGXIN MEDICINE MINIMAL INVASION |
| | TECHNOLOGY CO., LTD. |
| | Tel: +86-0519-82446628 |
| | Fax: +86-0519-82446610 |
| Date prepared | May 3, 2018 |
| 2. DEVICE | |
| Device Name: | Silicone Foley Catheter for single use |
| Common/Usual Name: | Silicone Foley Catheter for single use |
|---------------------|----------------------------------------|
| Regulation number | 21 CFR 876.5130 |
| Regulation Name | Urological catheter and accessories |
| Regulation Class: | II |
| Product Code: | EZL |
| Classification Name | Catheter, Retention Type, Balloon |
| | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with a green leaf-like shape on the upper left and a blue wave-like shape on the lower right. The two shapes are arranged in a circular fashion, creating a sense of balance and harmony. The green color of the leaf suggests nature and growth, while the blue color of the wave evokes water and fluidity. The logo is simple, clean, and visually appealing.
#### 3.
ന് PREDICATE DEVICES
Predicate device: K130908, Disposable Balloon-retention Catheter These predicates have not been subject to a design-related recall.
## 4. DEVICE DESCRIPTION
The Silicone Foley Catheter for single use is designed as an intermittent pathway for drainage and irrigation of the bladder. The device is made of silicone. The catheter is provided sterile in a variety of lengths and sizes.
### 5. Indications for USE
Two-way Disposable Silicone Foley Catheter: Urethral catheterization for bladder drainage for urological use only; the indwelling time is no more than 30 days. Three-way Disposable Silicone Foley Catheter: Urethral catheterization for bladder drainage and bladder irrigation for urological use only; the indwelling time is no more than 30 days.
| Specification | Predicate Device | Proposed Device | Discussion of<br>Differences |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Device name | Disposable Balloon-retention<br>Catheter | Silicone Foley Catheter for single<br>use | similar |
| K number | K130908 | -- | |
| Indications for<br>use | Two-way Disposable Silicone<br>Foley Catheter: Urethral<br>catheterization for bladder<br>drainage and urological use only;<br>the indwelling time is no more<br>than 30 days.<br>Three-way Disposable Silicone<br>Foley Catheter: Urethral<br>catheterization for bladder<br>drainage and bladder irrigation<br>for urological use only; the<br>indwelling time is no more than<br>30 days. | Two-way Disposable Silicone<br>Foley Catheter: Urethral<br>catheterization for bladder<br>drainage and urological use only;<br>the indwelling time is no more<br>than 30 days.<br>Three-way Disposable Silicone<br>Foley Catheter: Urethral<br>catheterization for bladder<br>drainage and bladder irrigation<br>for urological use only; the<br>indwelling time is no more than<br>30 days. | Identical |
| Specification | Predicate Device | Proposed Device | Discussion of<br>Differences |
| Intended<br>population | Pediatric ,male and female | Pediatric ,male and female | Identical |
| Anatomical<br>Sites | Urethra, bladder | Urethra, bladder | Identical |
| Lumen | Two way, three way | Two way, three way | Identical |
| Size range | 6Fr, 8Fr, 10Fr, 12Fr, 14Fr, 16Fr,<br>18Fr, 20Fr, 22Fr, 24Fr, 26Fr | 6Fr, 8Fr, 10Fr, 12Fr, 14Fr, 16Fr,<br>18Fr, 20Fr, 22Fr, 24Fr | Identical |
| Tube Length | 6Fr, 8Fr ,10Fr: 310mm<br>12Fr-26Fr: 400mm | 6Fr, 8Fr ,10Fr: 310mm<br>12Fr-24Fr: 400mm | Identical |
| Balloon | Yes | Yes | Identical |
| Balloon size | 1.5ml, 3ml, 5ml, 10ml, 15ml,<br>20ml, 30ml | 3ml, 5ml, 10ml, 15ml, 30ml | Identical |
| Eyes in tip | Two-way: no<br>Three-way: yes | Two-way: no<br>Three-way: yes | Identical |
| Tip shape | circular | circular | Identical |
| Standard funnel | Yes | Yes | Identical |
| Steel wire in<br>pediatric models | yes | yes | Identical |
| Material of main<br>shaft | silicone | silicone | Identical |
| Single use | Yes | Yes | Identical |
| Shelf life | 3 years | 3 years | Identical |
| Sterility | EO | EO | Identical |
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a green leaf-like shape on the left and a blue wave-like shape on the right. The two shapes are intertwined, creating a circular design. The green color suggests nature or growth, while the blue color suggests water or sky. There is a trademark symbol in the upper right corner of the logo.
K172807
Silicone Foley Catheter for single use is substantially equivalent to the cleared predicate device, Disposable Balloon-retention Catheter (K130908) because it has same indications for use, is composed of the same materials, and has similar technological characteristics.
# 7. NON-CLINICAL DATA
The following non-clinical data were provided in support of the substantial equivalence determination.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with a green leaf-like shape on the left and a blue wave-like shape on the right. The two shapes are intertwined. Below the logo, the words "Performance testing" are written in black text. The text is underlined.
The device were evaluated according to ASTM F623-99 (2013) and EN 1616:1997. Also, the device was evaluated to be radiopaque in accordance with ASTM F640-2012.
# Biocompatibility
According to the guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", the following tests have been conducted. Sample preparation and reference materials were conducted based on ISO 10993-12:2012.
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Subchronic Toxicity
- Genotoxicity
- Implantation
# Sterility
The device is designed for EO sterilization.
# Animal Study
The subject of this premarket submission, Silicone Foley Catheter for single use, does not require animal studies to support substantial equivalence.
# 8. CLINICAL DATA
The subject of this premarket submission, Silicone Foley Catheter for single use, did not require clinical studies to support substantial equivalence.
# 9. CONCLUSION
The differences between the Silicone Foley Catheter for single use and its predicate devices do not raise different questions of safety and effectiveness. The non-clinical data support the safety of the device and the performance testing report demonstrate that the Silicone Foley Catheter for single use should perform as intended in the specified use conditions.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with a green leaf on the left and a blue wave on the right. The leaf and wave are connected, forming a circular shape. The logo is simple and modern, and it likely represents a company or organization that is focused on environmental sustainability or water resources.
From the results of non-clinical data including the performance testing and comparative performance testing described, CHANGZHOU RONGXIN MEDICINE MINIMAL INVASION TECHNOLOGY CO., LTD. concludes that the Silicone Foley Catheter for single use is as safe and as effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.