Bencox Mirabo Cup System of Benox Total Hip System is intended for Cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions: a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis b. Inflammatory degenerative joint disease, such as rheumatoid arthritis c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques d. Patients with failed previous surgery where pain, deformity, or dysfunction persists e. Revision of previously failed total hip arthroplasty
Device Story
The Bencox Mirabo Cup System is a cementless acetabular hip prosthesis used in primary or revision hip arthroplasty. The system comprises an acetabular cup (titanium alloy with porous titanium coating) and a UHMWPE liner. It is implanted by orthopedic surgeons in a clinical/OR setting. The device replaces the natural acetabulum to restore joint function, reduce pain, and improve mobility. The cup is secured via press-fit or optional screw fixation; the liner locks into the cup via an internal taper mechanism. The system is used with dedicated non-sterile instrumentation that requires hospital-based sterilization. Clinical benefit is derived from the restoration of hip joint mechanics and stability.
Clinical Evidence
Bench testing only. Performance validated via ISO 14242-1 (Wear), ASTM F1820 (Torsion, Pull-out, Push-out), ASTM F2582 (Impingement), and ISO 21535 (Range of Motion). Sterility and shelf-life validated per ISO 11137-1/2 (Gamma), ISO 11135-1 (EtO), and ASTM F1980 (Accelerated Aging). Endotoxin testing performed per ANSI/AAMI ST72.
Technological Characteristics
Acetabular cup: Titanium alloy (ASTM F136) with porous titanium powder coating (ASTM F1580). Liner: UHMWPE (ASTM F648, Type 2, GUR 1050) cross-linked with 10 Mrad gamma radiation. Locking mechanism: Internal taper. Sterilization: Gamma (cup) and EtO (liner). Shelf life: 5 years.
Indications for Use
Indicated for patients requiring partial or total hip arthroplasty (primary or revision) due to non-inflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis, traumatic arthritis), inflammatory degenerative joint disease (e.g., rheumatoid arthritis), non-union, femoral neck/trochanteric fractures with head involvement, or failed previous hip surgery.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Pinnacle Acetabular System with ALTRX Liner (K000306)
Pinnacle Acetabular System with ALTRX Liner (K062148)
Submission Summary (Full Text)
{0}------------------------------------------------
October 17, 2017
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below it. The logo is simple and professional, and it is easily recognizable.
Corentec Co., Ltd. J.S. Daniel Associate Director- Global RA/QA 8F Chungho Tower. 483. Gangnam-daero. Seocho Gu, Seoul, South Korea-06541
Re: K172806
Trade/Device Name: Bencox Mirabo Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH Dated: September 15, 2017 Received: September 18, 2017
Dear J.S. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172806
Device Name
Bencox Mirabo Cup System
#### Indications for Use (Describe)
Bencox Mirabo Cup System of Benox Total Hip System is intended for Cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for "Corentec Implant Specialist". The logo features the word "Corentec" in gray letters, with a red circle in place of the "O". Below the word is a thin red line, and below that is the phrase "IMPLANT SPECIALIST" in smaller, gray letters. The logo is simple and modern, with a focus on the company's name and specialization.
# 510(K) SUMMARY Corentec Co., Ltd.
Bencox Mirabo Cup System 15th Sept., 2017
### ADMINISTRATIVE INFORMATION
| Manufacturer | Corentec Co., Ltd.<br>12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu<br>Cheonan-si, Chungchongnam-do, South Korea-31056<br>Telephone: +82-41-585-7114; Fax: +82-41-585-7113 |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | J.S. Daniel<br>Associate Director – Global RA/QA<br>Corentec Co., Ltd<br>8F Chungho Tower, 483, Gangnam-daero,<br>Seocho Gu, Seoul, South Korea-06541<br>Ph: +82 70 4393 3819; Fax: +82 2 3445 5467<br>Email: isdaniel@corentec.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Bencox Mirabo Cup System |
|-----------------------------|---------------------------|
| Common Name: | Acetabular Cup System |
| Classification Regulations: | 21 CFR 888.3358 |
| Regulatory Class: | II |
| Product Codes: | LPH |
| Classification Panel: | Orthopedic Products Panel |
#### INDICATIONS FOR USE
Bencox Mirabo Cup System of Bencox Hip Replacement System is intended for cementless use in partial or total hip arthroplasty in primary or revision surgery for the following conditions:
a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis
b. Inflammatory degenerative joint disease, such as rheumatoid arthritis
c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques
d. Patients with failed previous surgery where pain, deformity, or dysfunction persists
e. Revision of previously failed total hip arthroplasty
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "orentec IMPLANT SPECIALIST". The logo features the word "orentec" in gray letters, with the "o" in "orentec" being a red circle. Below the word "orentec" is a thin red line, and below the line is the phrase "IMPLANT SPECIALIST" in gray letters.
#### DEVICE DESCRIPTION
#### 1) Bencox Mirabo Cup System: (Sterile)
The Bencox Mirabo Cup System is a Cementless hip acetabular system (Metal on Poly Liner or Ceramic on Poly Liner) for hip arthroplasty. This submission consists of the following components:
- Acetabular Cup Bencox Mirabo Cup o
- Acetabular Liner Bencox Mirabo PE Liner ●
### Acetabular Cup: Bencox Mirabo Cup
The Bencox Mirabo Cup Spec. Inclusion is similar to Bencox Mirabo Cup cleared under K162127 & K120924 with respect to material - Titanium alloy (ASTM F136), coating with pure Titanium powder (ASTM F1580), design, locking system, manufacturing, packaging and sterilization. This submission is only for the inclusion of specification of acetabular cups with external diameter 44 mm.
### Acetabular Liner: Bencox Mirabo PE Liner
The Bencox Mirabo PE Liner specification inclusion is similar to Bencox Mirabo PE Liner cleared under K162127 & K150007, with respect to material, conforming to ASTM F648, Type 2 (GUR 1050), and irradiated with average dose of 10.0 Mrad of gamma radiation, design, locking system, manufacturing, packaging and sterilization. This submission is only for the inclusion of specification of Liners with head size, 28/35 & 32/39 mm.
#### 2) Bencox Hip Replacement System Instrumentation (Non Sterile)
Bencox Hip Replacement System Instrumentation is a set of accessories to be used with Bencox Hip Implants. The parts of the instruments are made of stainless steel and/or polymers and/or silicon and cleared for use in previous premarket notifications of Corentec. Bencox Hip System Instrumentation to be used with Bencox Mirabo Cup System is same with Instrumentation cleared under K162127 and K150007 & K120924 along with the inclusion of instruments specific to Bencox Mirabo Cup System components included in this submission.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "orentec IMPLANT SPECIALIST". The logo features a red circle with a gray arc around it, followed by the gray text "rentec". Below the text is a thin red line, and below that is the text "IMPLANT SPECIALIST" in gray.
# SUBSTANTIAL EQUIVALENCE
Bencox Mirabo Cup System is similar to the 510(k) cleared devices as mentioned below with respect to indications, design, specifications, operating principles and material.
| Predicate<br>Category | Manufacturer | Trade Name | 510(k) |
|-----------------------|--------------|-------------------------------------------------------|----------------------|
| Primary | Corentec | Bencox Mirabo Cup System | K162127 |
| | Corentec | Bencox Mirabo PE Liner | |
| Additional | Corentec | Bencox Mirabo PE Liner | K150007 |
| | | Bencox Mirabo Cup System | K120924 |
| | DePuy | Pinnacle Acetabular System with<br><i>ALTRX Liner</i> | K000306 &<br>K062148 |
The subject devices Bencox Mirabo Cup System and the Bencox Mirabo Cup (K162127 & K120924) has acetabular cups made of the same material and both have plasmasprayed porous coatings with similar pore size, porosity and coating thickness. The subject and predicate acetabular cups also incorporate holes through the cup in similar positions and angulations for optional screw fixation, and both an internal taper mechanism for locking the UHMWPE liner to the cup.
The subject devices Bencox Mirabo Cup System and the Bencox Mirabo PE Liner (K162127 & K150007) is made of the same material, UHMWPE and cross-linked to 10 Mrad. Both incorporate a taper mechanism for locking to the acetabular cup and have standard and elevated types of Inserts Type A with angles 10°, 15°. The Bencox Mirabo Cup system has similar dimensional specification with additional predicate device, Pinnacle Acetabular System, K000306 & AltrX Liner K062148.
# PERFORMANCE DATA
Bencox Mirabo Cup System was subjected to a series of testing protocols to document the performance of the components and to demonstrate substantial equivalence. Nonclinical testing and analysis included Wear Testing (ISO 14242-1), Liner Torsion (ASTM F1820 Torque Out Disassembly) & Pull out Testing (ASTM F1820 Offset Pull out disassembly), Push out (ASTM F1820 Axial Disassembly), Impingement Testing (ASTM F2582), Range of Motion Testing (ISO 21535) & Endotoxin Testing (ANSI/AAMI ST72) for Acetabular Cup and Liner and Risk Management (ISO 14971). The results of this testing/analysis showed that the subject devices are expected to be safe and effective for the proposed indications and are substantially equivalent to the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for Corentec Implant Specialist. The logo features the word "Corentec" in gray letters, with a red circle in the upper left corner. Below the word is a thin red line, and below that are the words "IMPLANT SPECIALIST" in smaller gray letters.
### STERILIZATION & PACKAGING
For sterile metal component Bencox Mirabo Cup, following to gamma sterilization, packaging was subjected to sterile barrier testing to validate a shelf life of 5 years as per ISO & ASTM standards which confirms the stability and effectiveness of packaging of the sterilized product during the shelf-life, by evaluating changes by accelerated aging, as per ASTM F1980. Sterilization validation as per ISO 11137-1 & 2 Sterilization of health care products – Radiation ensures sterility of the components for a SAL of 10-8.
For the sterile UHMWPE components - Bencox Mirabo PE Insert, following to EtO sterilization process, packaging was subjected to sterile barrier testing to validate a shelf life of 5 years as per ISO & ASTM standards which confirms the stability and effectiveness of packaging of the sterilized product during the shelf-life, by evaluating changes by accelerated aging, as per ASTM F1980. Sterilization validation as per ISO 11135-1, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices ensures sterility of the components for a SAL of 106. EtO Residuals were determined according to ISO 10993-7, Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals. The test results are within the limits specified in ISO 10993-7.
Sterile Bencox Mirabo Cup System components have same shelf life of 5 years, similar to other cleared sterile devices of Corentec. The non sterile Bencox hip replacement system instrumentation used in the surgery must be sterilized by the hospital, prior to use, as mentioned in the IFU.
# CONCLUSION
Overall Bencox Mirabo Cup System components included in this PMN is similar to the identified primary predicate device and additional predicates. The identified minor differences between the subject devices and predicate(s) do not constitute a new intended use and the minor differences in the technological characteristics do not affect or raise new queries of safety and effectiveness based on successful performance testing conforming to recognized performance standards for hip replacement devices and has been adequately addressed in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.