The ECG Cable/Leadwires is intended to be used with ECG. The ECG Cable/Leadwires is used to connect electrodes placed at appropriate sites on the patient to ECG for general monitoring and/or diagnostic evaluation by health care professional.
Device Story
Device consists of plug, cable/leadwires, and connector. Transmits ECG signals from patient-affixed electrodes to ECG monitoring equipment. Used in clinical settings by healthcare professionals for general monitoring and diagnostic evaluation. Reusable device; connects to standard chest wall electrodes via clip or snap terminations.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical performance, safety, and biocompatibility assessments.
Technological Characteristics
Materials: TPU. Sensing: Passive electrical signal transmission. Connectivity: Wired connection between electrodes and ECG monitor. Standards: AAMI/ANSI EC53:2013, IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012, ISO 10993-1:2009. Non-sterile.
Indications for Use
Indicated for use with ECG monitoring devices to connect patient-placed electrodes for general monitoring and diagnostic evaluation by healthcare professionals.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2018
Shenzhen Coreray Technology Co., Ltd. % Field Fu Consultant Shenzhen Joyantech Consulting Co., Ltd. No. 55 Shizhou zhong Road Nanshan District, Shenzhen Guangdong, China 518100
Re: K172797
Trade/Device Name: ECG Cable/Leadwires Regulation Number: 21 CFR 870.2900 Regulation Name: Patient transducer and electrode cable (including connector) Regulatory Class: Class II Product Code: DSA Dated: January 24, 2018 Received: January 30, 2018
Dear Field Fu:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Wilhelmsen
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Druq Administration
### Indications for Use
510(k) Number (if known) K172797
Device Name ECG Cable/Leadwires
Indications for Use (Describe)
The ECG Cable/Leadwires is intended to be used with ECG. The ECG Cable/Leadwires is used to connect electrodes placed at appropriate sites on the patient to ECG for general monitoring and/or diagnostic evaluation by health care professional.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# VOL 05 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR § 807.92.
## 5.1 Administrative Information
| Date of Summary<br>prepared | Aug 17, 2017 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Company title: Shenzhen Coreray Technology<br>Co., Ltd |
| | Company address: |
| | ChuangYe Technology Park, 1th Dong Huan |
| | Road, Longhua New District, Shenzhen, 510000 |
| | P.R China. P.C.:518109 |
| | Contact person: Simon Fan |
| | Phone: +86-755-28839229 |
| | Fax: +86-755-28839229 |
| | E-mail: manager@core-ray.com |
| Submission<br>Correspondent | Shenzhen Joyantech Consulting Co., Ltd.<br>Address: Room 1122#, International Mayors<br>Communication Centre, NO. 55 Shizhou middle |
| | road , Nanshan District, Shenzhen |
| | Contact person: Mr. Field Fu |
| | E-Mail: field@cefda.com; cefda13485@163.com |
### Establishment registration number
#### 5.2 Device Information
Type of 510(k) Traditional submission: Trade Name: ECG Cable/Leadwires Model: Classification name: | Cable, Transducer And Electrode, Patient,
{5}------------------------------------------------
(Including Connector) Cardiovascular Review Panel: Product Code: DSA Device Class: -870.2900 Requlation Number:
#### 5.3 Predicate Device Information
SHENZHEN MED-LINK ELECTRONICS TECH CO., Sponsor: LTD. Device: CABLE/ LEAD-WIRE K120010 510(K) Number:
#### 5.4 Device Description
The ECG Cable/Leadwires is comprised of Plug, cable/leadwires and connector. The device is used to transmit ECG signals from electrodes which are affixed to the patient's body for both diagnostic and monitoring purposes. Each lead wire is attached to ECG patient electrodes. The leadwires connector plugs into one end of the external trunk cable which is plug into an ECG monitor.
#### 5.5 Intended Use/ Indications for Use
The ECG Cable/Leadwires is intended to be used with ECG. The ECG Cable/Leadwires is used to connect electrodes placed at appropriate sites on the patient to ECG for general monitoring and/or diagnostic evaluation by health care professional.
## 5.6 Technological characteristics of the subject device compared to the predicate device
#### Predicate Device Information:
| 510(K) No.: | K120010 |
|------------------------|-----------------------------------------------------------------|
| Common name: | Cable / lead-wire |
| Classification name: | Cable, Transducer and Electrode, Patient, (Including Connector) |
| Production regulation: | 21 CFR §870.2900 |
{6}------------------------------------------------
| Product code: | DSA |
|---------------|----------------|
| Panel: | Cardiovascular |
| Elements of<br>Comparison | Proposed device | Predicate device<br>K120010 | Remarks |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Product Code | DSA | DSA | Same |
| Regulation<br>Number | 870.2900 | 870.2900 | Same |
| Classification | II | II | Same |
| Indication for Use | Shenzhen Coreray ECG<br>Cable/Leadwires is intended to<br>be used with ECG. The ECG<br>Cable/Leadwires is used to<br>connect electrodes placed at<br>appropriate sites on the patient<br>to ECG for general monitoring<br>and/or diagnostic evaluation by<br>health care professional. | Shenzhen Med-link<br>Cable/lead-wire are intended<br>to be used with ECG, EKG,<br>SpO2 and invasive Blood<br>Pressure monitoring devices.<br>The Cable/lead-wire are used<br>to connect electrodes,<br>catheters, and/or sensors<br>placed at appropriate sites on<br>the patient to a monitoring<br>device for general monitoring<br>and/or diagnostic evaluation by<br>health care professional. | Same as<br>ECG/EK<br>G cable/<br>lead-<br>wire. |
| Patent Usage | Reusable | Reusable | Same |
| Anatomical sites | Attached to electrodes placed<br>at standard specified locations<br>on chest wall | Attached to electrodes placed<br>at standard specified locations<br>on chest wall | Same |
| Patient end<br>termination | Clip, Snap | Clip, Snap | same |
| Leadwire<br>material | TPU | Shielded & Unshielded Copper<br>with PVC or TPU Jacket | TPU is<br>same. |
| Sterile | Non sterile | Non sterile | same |
| Biocompatibility | No cytotoxicity,<br>No irritation,<br>No sensitization | No cytotoxicity,<br>No irritation,<br>No sensitization | SE |
| Electrical<br>Performance and<br>Safety | Comply with AAMI/ANSI<br>EC53: 2013 IEC 60601-<br>1:2005+CORR.1: 2006+CORR.<br>2:2007+AM1: 2012 | Comply with ANSI/AAMI EC<br>53:1995/(R)2001 IEC 60601-<br>1:1998: Am1: A2:1995 | SE |
## Comparison to predicate device:
## 5.7 Brief discussion of the nonclinical tests
ECG Cable/Leadwires conforms to the following standards:
{7}------------------------------------------------
IEC 60601-1:2005+CORR.1:2006+CORR.2007+A1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
AAMI / ANSI EC53:2013, ECG Trunk Cables and Patient Leadwires.
ISO 10993-1:2009, Biological Evaluation Of Medical Devices - Part 1: Evaluation And Testing Within A Risk Management Process.
## 5.8 Brief discussion of clinical tests
N/A.
5.9 Other information (such as required by FDA guidance/Test) N/A.
### 5.10 Conclusions
Based on the above information, the subject device and the predicate device have the same intended use and same technological characteristics; we conclude the subject device, ECG Cable/ Leadwires, is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.