← Product Code [BSO](/productcode/BSO) · K172777

# Catheter Clamping Adapter LUER 80369-7, Catheter Clamping Adapter NRFit 80369-7 (K172777)

_PAJUNK GmbH Medizintechnologie · BSO · May 3, 2018 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172777

## Device Facts

- **Applicant:** PAJUNK GmbH Medizintechnologie
- **Product Code:** [BSO](/productcode/BSO.md)
- **Decision Date:** May 3, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5120
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Catheter Clamping Adapter, a connection device, is used to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with catheters for continuous administration of anesthetic agents.

## Device Story

Catheter Clamping Adapter is a sterile, disposable connection device; provides single, common access point to 20G anesthesia conduction catheters (epidural or peripheral) for anesthetic/fluid delivery. Device clamps onto proximal end of catheter; features either NRFit (ISO 80369-6) or LUER (ISO 80369-7) connector. Used by clinicians in clinical settings for continuous administration of anesthetic agents. Device ensures secure, leak-tight connection between catheter and administration equipment; facilitates reliable drug delivery; minimizes risk of disconnection or leakage during anesthesia procedures.

## Clinical Evidence

Bench testing only. No clinical data. Performance validated via ISO 80369-6 and ISO 80369-7 testing (liquid/air leakage, stress cracking, axial load, unscrewing torque, overriding), biocompatibility (ISO 10993-1), and sterilization validation (ISO 11135).

## Technological Characteristics

Materials: Polycarbonate body, silicone clamping seal. Connectivity: NRFit (ISO 80369-6) or LUER (ISO 80369-7). Sterilization: Ethylene Oxide (SAL 10^-6). Form factor: Disposable catheter clamping adapter. Standards: ISO 11135, ISO 10993-1, ISO 10993-7, ISO 80369-6, ISO 80369-7.

## Regulatory Identification

An anesthesia conduction catheter is a flexible tubular device used to inject local anesthetics into a patient and to provide continuous regional anesthesia.

## Predicate Devices

- BBraun Perifix Catheter Connector ([K022019](/device/K022019.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized depiction of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 3, 2018

PAJUNK GmbH Medizintechnologie Christian Quass Director of Regulatory Affairs, Safety Officer Karl-Hall-Str. 1 Geisingen, 78187 Germany

Re: K172777

Trade/Device Name: Catheter Clamping Adapter NRFit (ISO80369-6) and Catheter Clamping Adapter LUER (ISO80369-7) Regulation Number: 21 CFR 868.5120 Regulation Name: Anesthesia Conduction Catheter Regulatory Class: Class II Product Code: BSO Dated: April 4, 2018 Received: April 6, 2018

Dear Christian Quass:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172777

Device Name

Catheter Clamping Adapter NRFit (ISO80369-6) and Catheter Clamping Adapter LUER (ISO80369-7)

Indications for Use (Describe)

The Catheter Clamping Adapter, a connection device, is used to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics.

The connector is used in conjunction with catheters for continuous administration of anesthetic agents.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# PAJUNK

Pioneering medical technology

# 510(k) Summary as required by 21 CFR 807.92(c).

Date of Preparation: 2018-04-25

# Document Control Number: K172777

510(k) owner:

PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen, Baden-Wuerttemberg, Germany

Establishment Registration Number: 9611612

### Submitter Information/ production site:

PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen Baden-Wuerttemberg, Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612

# USA Contact:

PAJUNK MEDICAL SYSTEMS 6611 Bay Circle, Suite 140 Norcross, GA 30071 Phone +1 (888) 9PAJUNK (72-5865) Fax +1 (678) 514-3388

Establishment Registration Number: 3004076349

## Contact:

Christian G. H. Quass Director Regulatory Affairs, Safety Officer Patricia Weisbrod Requlatory Affairs Manager Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com E-Mail: patricia.weisbrod@pajunk.com

## Contact

Marco Wohnig President PAJUNK® Medical Systems L.P. Phone: (770) 493-6832 Direct: (470) 448-4070 Cell: (770) 757-2449 (678) 514-3388 Fax: E-Mail: marco.wohnig@pajunk-usa.com

{4}------------------------------------------------

### Device Name and Classification Details

| Device Information:       |                                                                                                                                                                                 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:              | Catheter Clamping Adapter NRFit™ (ISO80369-6)<br>Catheter Clamping Adapter LUER (ISO80369-7)                                                                                    |
| Sterilization method:     | Ethylene Oxide<br>disposable device, supplied sterile to the end user and<br>non-sterile intended to be sterilized prior to use to<br>repackagers/ medical device manufacturers |
| Contract Sterilizer:      | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number: 3002807090                                           |
| Document Control Number   | K172777                                                                                                                                                                         |
| Trade Name:               | Clamping Adapter                                                                                                                                                                |
| Common name               | Catheter Connector                                                                                                                                                              |
| Classification Name:      | Anaesthesia Conduction Catheter                                                                                                                                                 |
| Classification Reference: | 21 CFR 868.5120                                                                                                                                                                 |
| Product Code:             | BSO                                                                                                                                                                             |
| Regulatory Class:         | II                                                                                                                                                                              |
| Panel:                    | Anesthesiology                                                                                                                                                                  |

PAJUNK® GmbH Medizintechnologie is submitting this 510(k) for the Catheter Clamping Adapter with either NRFit™ Connector according to ISO80369-6 or LUER-Connector according to ISO80369-7.

The device is intended to provide a specified connection (either NRFit™ Connector according to ISO80369-6 or LUER-Connector according to ISO80369-7) to an unspecified proximal end of a 20G anesthesia conduction catheter (epidural or peripheral). Therefore it may also be considered an accessory for epidural and peripheral anesthesia conduction catheters.

The Catheter Clamping Adapter is considered a Class II medical device according to the predicate device as defined in 21 CFR §868.5120, product code BSO.

The device is a sterile finished disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/ medical device manufacturers.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image features the word "PAJUNK" in bold, white letters against a teal background. The font is sans-serif and appears to be slightly blurred, giving it a soft, glowing effect. A registered trademark symbol is located to the right of the letter "K".

#### Indications for use

The Catheter Clamping Adapter, a connection device, is used to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics.

The connector is used in conjunction with catheters for continuous administration of anesthetic agents.

#### Determination methods and results of Substantial Equivalence Determination:

Intended Use Design Performance

Result: Substantially Equivalent

Result: Substantially Equivalent

Result: Substantially Equivalent

#### Equivalence in materials used

| Characteristics        | Predicate device<br>K022019<br>BBraun Perifix Catheter<br>Connector                                                                                                                                                                                                                                                                                  | Subject Device<br>Catheter Clamping Adapter                                                                                                                                                                                                                                                                                              | Result of comparison,<br>if necessary with<br>rationale                                                                                          |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility       | Both devices are classified as externally communicating. Therefore and<br>based upon ISO10993-1 biocompatibility applies.                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                          | Substantially equivalent                                                                                                                         |
| Packaging              | Individually sterile in soft blister<br>pouch or as part of a kit                                                                                                                                                                                                                                                                                    | Individually sterile in soft blister<br>pouch or as part of a kit                                                                                                                                                                                                                                                                        | Same packaging<br>The PAJUNK®<br>individual packaging<br>was selected according<br>to a validated<br>sterilization and<br>transportation process |
| Indications for<br>Use | A connection device used to provide<br>various anesthetic and fluid<br>administration devices with a single,<br>common access point to an 18 or 20<br>gauge Perifix catheter for delivery of<br>anesthetics. The connector is used in<br>conjunction with 18 or 20 gauge Perifix<br>catheters for continuous administration<br>of anesthetic agents. | The Catheter Clamping Adapter, a<br>connection device, is used to provide<br>various anesthetic and fluid<br>administration devices with a single,<br>common access point to a catheter for<br>delivery of anesthetics.<br>The connector is used in conjunction<br>with catheters for continuous<br>administration of anesthetic agents. | Substantially equivalent<br>Size limitation in<br>predicate device has<br>been altered by<br>K032144 and K033952<br>of predicate's sponsor       |
| Overall design:        | Adapter clamps catheter                                                                                                                                                                                                                                                                                                                              | Adapter clamps catheter                                                                                                                                                                                                                                                                                                                  | Substantially equivalent                                                                                                                         |

{6}------------------------------------------------

# PAJUNK®

# K172777

| Pioneering medical technology |  |  |
|-------------------------------|--|--|
|                               |  |  |

| Characteristics | Predicate device<br>K022019<br>BBraun Perifix Catheter<br>Connector                  | Subject Device<br>Catheter Clamping Adapter                                          | Result of comparison, if necessary with rationale                                                                                                                                                                                                              |
|-----------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Picture         | Image: BBraun Perifix Catheter Connector                                             | Image: Catheter Clamping Adapter                                                     | Both systems allow<br>easy handling and<br>connecting with one<br>hand                                                                                                                                                                                         |
| Materials used  | Plastics, rubber                                                                     | Body Polycarbonate<br>Clamping seal Silicone                                         | Similar<br>Each of the materials<br>used either in the<br>Predicate Devices or<br>the Subject Device are<br>established materials<br>used for manufacturing<br>medical devices.                                                                                |
| Connectivity    | ISO 80369-7                                                                          | ISO 80369-6 NRFitTM<br>ISO 80369-7 LUER                                              | Both systems do have<br>identical connectivity<br>towards syringe and<br>needle; 2017's state of<br>the art connection type<br>is NRFitTM which slightly<br>differs in dimensions of<br>the male and female<br>connectors in order to<br>avoid misconnections. |
| Tightness       | Leak tight under normal conditions<br>when used according to<br>instructions for use | Leak tight under normal<br>conditions when used according<br>to instructions for use | identical                                                                                                                                                                                                                                                      |
| Tensile Force   | > 8N                                                                                 | > 8N                                                                                 | identical                                                                                                                                                                                                                                                      |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image features the word "PAJUNK" in a bold, sans-serif font. The letters are white with a subtle gradient effect, giving them a slightly three-dimensional appearance. The background is a solid teal color, providing a strong contrast that makes the text stand out. A small registered trademark symbol is visible to the right of the letter "K".

### Equivalence in the Indications for use

#### Subiect Device:

The Catheter Clamping Adapter, a connection device, is used to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics.

The connector is used in conjunction with catheters for continuous administration of anesthetic agents.

#### Predicate Devices:

A connection device used to provide various anesthetic and fluid administration devices with a single, common access point to an 18 or 20 gauge Perifix catheter for delivery of anesthetics. The connector is used in conjunction with 18 or 20 gauge Perifix catheters for continuous administration of anesthetic agents.

#### Discussion

The intended use of the predicate devices and of the subject device are identical. The indications for use differ slightly but the differences do not raise different questions of safety or effectiveness.

Conclusion: Substantially Equivalent

#### Sterilization

#### Subject Device:

Sterilized with Ethylene Oxide. Shelf Life: 5 years.

#### Predicate Devices:

Sterilized with Ethylene Oxide. Shelf Life: 5 years.

#### Discussion

Both devices are sterilized using Ethylene Oxide and have a shelf life of 5 years.

Conclusion: Substantially Equivalent

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows a sequence of alphanumeric characters. The sequence is "K172777". The characters are in a bold, sans-serif font and are black against a white background.

## Technology Characteristics/ Performance Testing

#### Subject Device:

Connectivity: LUER and NRFit™ (ISO 80369-6 and ISO 80369-7).

Leak Tightness: Complies with internal protocol

Security of connection, tensile strength: Complies with internal protocol

#### Predicate Devices:

Leak Tightness: Complies with internal protocol of the sponsor

Security of connection, tensile strength: Complies with internal protocol of the sponsor

#### Sterilization

The contract sterilizer and the sterilizing process are identical to the process and sterilizer used for all PAJUNK® - manufactured devices which are already cleared for market or exempt.

Sterilization parameters are

| SAL             | 10-6                                                                             |
|-----------------|----------------------------------------------------------------------------------|
| Type of gas     | Ethylene Oxide 99,99%                                                            |
| Exposure time   | 300 min.                                                                         |
| Aeration method | evacuation<br>2 airwashes                                                        |
| Aeration period | residual EtO-gas is removed in circulating<br>air at 40° C (±5) for at least 48h |

Sterilization has been validated according to ISO 11135-1 Overkill Approach (1 sublethal cycle, 2 half cycle, 1 full cycle)

Residuals of EO and ECH are in compliance with ISO 10993-7.

Cleaning and Sterilization method, which ensures an SAL of 10-6 as well as compliance with limits for chemical burden, bioburden, pyroburden (i.e. LAL) and Et0-residuals as well as shelf life have been validated.

The limits listed below are met by each device:

Limits for Residuals of Ehyleneoxide and Ethylene chlorhydrine are met.

Limit for Pyroburden/ endotoxine is met.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the alphanumeric string "K172777" in a bold, sans-serif font. The string is horizontally oriented and appears to be a code or identifier. The characters are uniformly sized and spaced, creating a clear and legible sequence.

The clamping adapters are also available in bulk non sterile. If appropriately packed and sterilized with Ethylene oxide according to the parameters above the technological parameters remain unchanged. However, final responsibility for sterilization remains with the customer of catheter clamping adapters purchased bulk non-sterile.

#### Shelf Life

Efficacy of sterile product's lifecycle has been validated.

Sterility tests have been performed. Performance of the essential performance of the device (NRFit and LUER connection, stability of connections) has been tested with real time aged devices (1 year) and devices subject to accelerated aging (1 year, 3 years). There is no decrease in performance after 5 years.

Shelf-life is set to 5 years.

#### Biocompatibility:

All products comply with ISO 10993-1, 2nd and 3rd edition.

The Subject Device is an external communicating device with no direct patient contact. Therefore based upon ISO10993-1 biocompatibility is evaluated.

Therefore and based upon sterilization and residuals validation the devices are considered to be biocompatible.

#### Standards/ Requirements

The Subject Device has been tested to comply with the state-of-the-art standards listed below. For connector standards both, the male and female connectors have been tested:

| Test Detail       | Standard         | FDA-Rec.-No. | Result |
|-------------------|------------------|--------------|--------|
| Sterilization     | ISO 11135        | 14-452       | Pass   |
| Residuals         | ISO 10993-7      | 14-408       | Pass   |
| Biocompatibility  | ISO 10993-1      | 2-220        | Pass   |
| Liquid Leakage    | ISO 80369-7, 6.1 | 5-115        | Pass   |
| Air Leakage       | ISO 80369-7, 6.2 | 5-115        | Pass   |
| Stress Cracking   | ISO 80369-7, 6.3 | 5-115        | Pass   |
| Axial Load        | ISO 80369-7, 6.4 | 5-115        | Pass   |
| Unscrewing torque | ISO 80369-7, 6.5 | 5-115        | Pass   |
| Overriding        | ISO 80369-7, 6.6 | 5-115        | Pass   |

{10}------------------------------------------------

# PAJUNK®

# K172777

Pioneering medical technology

| Test Detail                            | Standard          | FDA-Rec.-No. | Result |
|----------------------------------------|-------------------|--------------|--------|
| Liquid Leakage                         | ISO 80369-6, 6.1  | 5-108        | Pass   |
| Air Leakage                            | ISO 80369-6, 6.2  | 5-108        | Pass   |
| Stress Cracking                        | ISO 80369-6, 6.3  | 5-108        | Pass   |
| Axial Load                             | ISO 80369-6, 6.4  | 5-108        | Pass   |
| Unscrewing torque                      | ISO 80369-6, 6.5  | 5-108        | Pass   |
| Overriding                             | ISO 80369-6, 6.6  | 5-108        | Pass   |
| Leak Tightness                         | Internal Protocol | n.a.         | Pass   |
| Security of connection                 | Internal Protocol | n.a.         | Pass   |
| ISTA Procedure 2A (ASTM D 6653 et al.) |                   |              |        |
| ASTM F1886 - 09                        |                   |              |        |
| ASTM F 1929 – 98                       |                   |              |        |
| ASTM E 515 – 05 (4a))                  |                   |              |        |
| DIN EN 868-5:2009-09                   |                   |              |        |
| DIN EN 868-10:2009-09                  |                   |              |        |

## Conclusion:

The comparison between the predicate devices and the subject device of this submission as well as the validated sterilization process and the results of the standard testing and performance testing demonstrates that the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K172777](https://fda-staging.innolitics.com/device/K172777)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
