T2 Alpha Femur Antegrade GT/PF Nailing System: The T2 Alpha Femur Antegrade GT/PF Nailing System is intended for temporary stabilization of bone segments or fragments until bone consolidation has been achieved. IMN Screws System: The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization. IMN Instruments System: The IMN Instruments System is intended to enable the implantation and extraction of intramedullary nail and screw.
Device Story
System comprises intramedullary nails, compression screws, set screws, end caps, and locking screws for femoral fracture fixation; includes targeting instruments for implantation/extraction. Implants are titanium alloy (Ti6Al4V ELI); set screws include PEEK. Used in surgical settings by orthopedic surgeons for temporary stabilization of bone segments until consolidation. System provides mechanical support to stabilize nail-bone construct. Performance verified via bench testing (fatigue, mechanical properties per ASTM F1264/F543, cut-out testing). MR Conditional. Biocompatible materials.
Clinical Evidence
No clinical data. Bench testing only, including fatigue strength, mechanical properties (ASTM F1264, ASTM F543), cut-out testing, biocompatibility (ISO 10993-1), and bacterial endotoxin testing.
Technological Characteristics
Implants: Titanium alloy (Ti6Al4V ELI) per ASTM F136; set screws include PEEK. Instruments: Stainless steel, PEEK (unreinforced and carbon-fiber reinforced). Mechanical fixation via intramedullary nail and locking screws. MR Conditional. Non-sterile instruments; sterile implants.
Indications for Use
Indicated for fixation of subtrochanteric, intertrochanteric, ipsilateral neck/shaft, and comminuted proximal femoral shaft fractures; femoral fixation for pathological disease; temporary stabilization of femoral shaft fractures from femoral neck to supracondylar regions; open and closed femoral fractures; pseudoarthrosis and correction osteotomy; pathologic fractures, impending pathologic fractures, and tumor resections; ipsilateral femur fractures; fractures proximal to total knee arthroplasty; nonunions and malunions; and fractures involving osteopenic and osteoporotic bone.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration" with "Administration" written on the second line.
January 12, 2018
Stryker Trauma GmbH Heike Gustke Staff Regulatory Affairs Specialist Prof-Kuentscher-Str. 1-5 24232 Schoenkirchen GERMANY
Re: K172774
Trade/Device Name: T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System, IMN Instruments System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB, HWC Dated: December 8, 2017 Received: December 12, 2017
Dear Heike Gustke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known)
K172774
Device Name
T2 Alpha Femur Antegrade GT/PF Nailing System
Indications for Use (Describe)
The indications for use of these internal fixation devices include:
- Fixation of subtrochanteric, intertrochanteric, ipsilateral neck/shaft, comminuted proximal femoral shaft fractures
- Femoral fixation required as a result of pathological disease
- Temporary stabilization of fractures of the femoral shaft ranging from the femoral neck to the supracondylar regions of the femur
- Open and closed femoral fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures and tumor resections
- Ipsilateral femur fractures
- Fractures proximal to a total knee arthroplasty
- Nonunions and malunions
- Fractures involving osteopenic and osteoporotic bone
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
Device Name IMN Screws System
Indications for Use (Describe)
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name IMN Instruments System
Indications for Use (Describe)
The IMN Instruments System is intended to enable the implantation of intramedullary nail and screw.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{5}------------------------------------------------
## Section 9: 510(k) Summary
# I. SUBMITTER
| <b>I. SUBMITTER</b> | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Sponsor: | Stryker Trauma GmbH<br>Prof.-Kuentscher-Str. 1-5<br>24232 Schoenkirchen / Germany | |
| Contact Person: | Dr. Heike Gustke<br>Staff Regulatory Affairs Specialist<br>Prof.-Kuentscher-Str. 1-5<br>24232 Schoenkirchen / Germany<br>heike.gustke@stryker.com<br>Phone: +49 4348 702 637<br>Fax: +49 4348 702 8637 | |
| Date Prepared: | November 30, 2017 | |
| <b>II. DEVICE</b> | | |
| Name of Device: | T2 Alpha Femur Antegrade GT/PF Nailing System<br>IMN Screws System<br>IMN Instruments System | |
| Common Name: | T2 Alpha Femur Antegrade GT/PF Nailing System<br>Rod, fixation, intramedullary and accessories<br>IMN Screws System<br>Screw, fixation, bone<br>IMN Instruments System<br>Rod, fixation, intramedullary and accessories | |
| Regulation Number / Name: | T2 Alpha Femur Antegrade GT/PF Nailing System<br>21CFR 888.3020 (Intramedullary fixation rod)<br>IMN Screws System | |
| | | 21CFR 888.3040 (Smooth or threaded metallic bone fixation |
| | | fastener) |
| | | IMN Instruments System |
| | | 21CFR 888.3020 (Intramedullary fixation rod) |
| Product Code: | | T2 Alpha Femur Antegrade GT/PF Nailing System |
| | | HSB (Rod, fixation, intramedullary and accessories) |
| | | IMN Screws System |
| | | HWC (Screw, fixation, bone) |
| | | IMN Instruments System |
| | | HSB (Rod, fixation, intramedullary and accessories) |
| Regulatory Class: | | Class II |
{6}------------------------------------------------
## III. PREDICATE DEVICE
| T2 Alpha Femur Antegrade GT/PF Nailing System | | |
|-----------------------------------------------|---------------------------------------------------------|--|
| Primary predicate: | T2 Recon Nail System (K102992) | |
| Additional predicate: | T2 Greater Trochanter Nail (GTN) (K101438) | |
| | Zimmer Natural Nail System Piriformis Fossa and Greater | |
| | Trochanter Antegrade Femoral Nails (K083497) | |
| IMN Screws System | | |
| Primary predicate: | Titan Tibial Nail (K003018) | |
| IMN Instruments System | | |
| Primary predicate: | Gamma3 and T2 Recon Target Devices (K123401) | |
| Reference device: | T2 Recon Nail System (K102992) | |
## IV. DEVICE DESCRIPTION
This Traditional 510(k) submission is being supplied to the U.S. FDA to gain clearance to market the T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System, and IMN
{7}------------------------------------------------
Instruments System. This submission encompasses multiple systems that have similar intended use and will be used together during the surgical procedure.
### T2 Alpha Femur Antegrade GT/PF Nailing System
The T2 Alpha Femur Antegrade GT/PF Nailing System is a fracture fixation system and includes sterile implants (intramedullary nails in various diameter and sizes, compression screw, set screws and end caps) as well as non-sterile instruments (targeting devices). The femoral nails, compression screw and end caps are made of titanium alloy (Ti6Al4V ELI) as per ASTM F136. The set screws are manufactured from titanium alloy (Ti6Al4V ELI) as per ASTM F136 and PEEK.
Additionally, the T2 Alpha Femur Antegrade GT/PF Nailing System will be used with the existing locking screws previously cleared in K003018 (Titan Tibial Nail System), the T2 Lag Screw Recon previously cleared in K032898 (T2 Recon Nail System) as well as existing instruments previously cleared in K123401 (Gamma 3 and T2 Recon Targeting Devices), K131365 (T2 Tibial Nailing System) and 510(k) exempt devices. The T2 Alpha Femur Antegrade GT/PF Nailing System is intended for use with IMN Screws System and IMN Instruments System.
### IMN Screws System
The IMN Screws System includes locking screws that are inserted through the intramedullary nail to stabilize the nail-bone construct. The locking screws are made of titanium alloy (Ti6Al4V ELI) as per ASTM F136. The IMN Screws System is intended for use with T2 Alpha Femur Antegrade GT/PF Nailing System and IMN Instruments System.
### IMN Instruments System
The IMN Instruments System includes instruments that support the implantation and extraction of intramedullary nails and screws. The IMN Instruments System is intended for use with T2 Alpha Femur Antegrade GT/PF Nailing System and IMN Screws System.
{8}------------------------------------------------
### V. INTENDED USE
### T2 Alpha Femur Antegrade GT/PF Nailing System
The T2 Alpha Femur Antegrade GT/PF Nailing System is intended for temporary stabilization of bone segments or fragments until bone consolidation has been achieved.
#### IMN Screws System
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
#### IMN Instruments System
The IMN Instruments System is intended to enable the implantation and extraction of intramedullary nail and screw.
### VI. INDICATION FOR USE
#### T2 Alpha Femur Antegrade GT/PF Nailing System
The indications for use of these internal fixation devices include:
- Fixation of subtrochanteric, intertrochanteric, ipsilateral neck/shaft, comminuted proximal . femoral shaft fractures
- o Femoral fixation required as a result of pathological disease
- Temporary stabilization of fractures of the femoral shaft ranging from the femoral neck to ● the supracondylar regions of the femur
- Open and closed femoral fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures and tumor resections ●
- Ipsilateral femur fractures ●
- . Fractures proximal to a total knee arthroplasty
- o Nonunions and malunions
- Fractures involving osteopenic and osteoporotic bone ●
{9}------------------------------------------------
### IMN Screws System
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
#### IMN Instruments System
The IMN Instruments System is intended to enable the implantation and extraction of intramedullary nail and screw.
# VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
#### T2 Alpha Femur Antegrade GT/PF Nailing System
Device comparison demonstrated that the T2 Alpha Femur Antegrade GT/PF Nailing System is substantially equivalent to the previously cleared T2 Recon Nail System (K102992), T2 GTN System (K101438) and Zimmer Natural Nail System (K083497) in regards to intended use, indications for use, technological characteristics (design features, material and performance) as well as operating principle.
At a high level, the subject device and predicate devices are based on the following same technological elements:
- . Intramedullary nailing systems to provide a fracture fixation of the long bone
- Nail and screw design (length, diameter, and shape)
- Nails, compression screw and end caps manufactured from titanium alloy (Ti6Al4V ELI) . as per ASTM F136
- Set screws are manufactured from titanium alloy (Ti6A14V ELI) as per ASTM F136 and . PEEK.
The following technological differences exist between the subject and predicate devices:
- . Hole configurations in proximal and distal part of nail
- . Locking configurations.
The differences have been addressed through performance testing which demonstrates that the subject device is as safe and effective as the predicate devices.
#### IMN Screws System
{10}------------------------------------------------
Device comparison demonstrated that the IMN Screws System is substantially equivalent to the locking screws previously cleared in Titan Tibial Nail System (K003018) in regards to intended use, technological characteristics (design features, material and performance) as well as operating principle.
At a high level, the subject device and predicate devices are based on the following same technological elements:
- Stabilization of intramedullary nail-bone construct .
- Used for proximally and distally locking of nail-bone construct ●
- o Screw design (length, diameter, and thread design)
- Manufactured from titanium alloy (Ti6Al4V ELI) as per ASTM F136. .
### IMN Instruments System
Device comparison demonstrated that the IMN Instruments System is substantially equivalent to the previously cleared Gamma3 and T2 Recon Target Devices (K123401) in regards to intended use, technological characteristics (design features, material and performance) as well as operating principle.
At a high level, the subject device and predicate devices are based on the following same technological elements:
- Used for guided distal locking of nails .
- Design consisting of distal targeting arm and adjusting device o
- . Manufactured from Stainless Steel, PEEK unreinforced as well as PEEK with carbon fibers.
The following technological differences exist between the subject and predicate devices:
- A single adjusting device to cover all potential locking options. .
The difference has been addressed through performance testing which demonstrates that the subject device is as safe and effective as the predicate devices.
### VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### T2 Alpha Femur Antegrade GT/PF Nailing System
{11}------------------------------------------------
Non-clinical testing was performed for the T2 Alpha Femur Antegrade GT/PF Nailing System under worst case scenario to determine substantial equivalence.
The following bench testing was conducted:
- Fatigue strength testing .
- Testing of mechanical properties as per ASTM F1264 .
- Cut-out testing .
Testing demonstrated that the T2 Alpha Femur Antegrade GT/PF Nailing System is equivalent in mechanical performance to the predicate devices (K102992, K101438, and K083497).
Non-clinical testing has demonstrated that the T2 Alpha Femur Antegrade GT/PF Nailing System is MR Conditional.
Biocompatibility testing was performed to evaluate the biological safety of the T2 Alpha Femur Antegrade GT/PF Nailing System according to FDA Guidance 'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'. Testing demonstrated that the T2 Alpha Femur Antegrade GT/PF Nailing System is biocompatible.
The Bacterial Endotoxin Testing demonstrated that the sterile implants of T2 Alpha Femur Antegrade GT/PF Nailing System meet the specified endotoxin limit.
### IMN Screws System
Non-clinical testing was performed for the IMN Screws System. The following bench testing was conducted:
- Testing of mechanical properties as per ASTM F543 ●
Testing demonstrated that the IMN Screws System is equivalent in mechanical performance to the predicate device (K003018).
Non-clinical testing has demonstrated that the IMN Screws System is MR Conditional.
Biocompatibility assessment was performed according to FDA Guidance 'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'. It has been demonstrated that the IMN Screws System is biocompatible.
The Bacterial Endotoxin Testing demonstrated that the IMN Screws System meets the specified endotoxin limit.
{12}------------------------------------------------
### IMN Instruments System
Biocompatibility testing was performed to evaluate the biological safety of the IMN Instruments System according to FDA Guidance 'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'. Testing demonstrated that the IMN Instruments System is biocompatible.
### IX. CLINICAL TESTING
#### T2 Alpha Femur Antegrade GT/PF Nailing System
No clinical testing of the T2 Alpha Femur Antegrade GT/PF Nailing System has been conducted.
#### IMN Screws System
No clinical testing of the IMN Screws System has been conducted.
#### IMN Instruments System
No clinical testing of the IMN Instruments System has been conducted.
### X. CONCLUSION
### T2 Alpha Femur Antegrade GT/PF Nailing System
The T2 Alpha Femur Antegrade GT/PF Nailing System is substantially equivalent to the predicate devices (K102992, K101438, and K083497) identified in this premarket notification.
#### IMN Screws System
The IMN Screws System is substantially equivalent to the predicate device (K003018) identified in this premarket notification.
#### IMN Instruments System
The IMN Instruments System is substantially equivalent to the predicate device (K123401) identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.