Blood Collecting Needle, Safety Blood Collecting Needle, Blood Collecting Needle with Holder, Safety Blood Collecting Needle with Holder, Blood Collecting Set, Safety Blood Collecting Set, Blood Collecting Set with Holder, Safety Blood Collecting Set with Holder
K172763 · Zhejiang Kindly Medical Devices Co., Ltd. · JKA · Jun 4, 2018 · Clinical Chemistry
Device Facts
Record ID
K172763
Device Name
Blood Collecting Needle, Safety Blood Collecting Needle, Blood Collecting Needle with Holder, Safety Blood Collecting Needle with Holder, Blood Collecting Set, Safety Blood Collecting Set, Blood Collecting Set with Holder, Safety Blood Collecting Set with Holder
Applicant
Zhejiang Kindly Medical Devices Co., Ltd.
Product Code
JKA · Clinical Chemistry
Decision Date
Jun 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The Safety Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury. The Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The Safety Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury. The Blood Collecting Set is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The Safety Blood Collecting Set is intended to be used with vacuum blood collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury. The Blood Collecting Set with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The Safety Blood Collecting Set with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
Device Story
Venous blood collection devices; channel between patient vein and vacuum tube. Patient-end needle punctures vein; non-patient end needle penetrates vacuum tube; blood flows via differential pressure. Available as needles, sets, or with integrated holders. Safety versions feature manual safety shield to cover needle post-venipuncture. Used in clinical settings by healthcare professionals. Output is collected blood sample in vacuum tube. Benefits include standardized blood collection and reduced risk of sharps injury to clinicians.
Clinical Evidence
No clinical study included. Bench testing only. Simulated clinical study performed per FDA guidance for sharps injury prevention features; results met pre-established criteria. Biocompatibility testing (ISO 10993 series, ASTM F756-17) and physical/mechanical testing (ISO 7864, ISO 9626) confirm device safety and performance.
Technological Characteristics
Materials: Stainless steel needles, PP (hubs, shields, caps), ABS (fittings), PVC (tubing), natural rubber (sleeves). Manual operation. EO sterilized (SAL 10^-6). Dimensions: 18G-25G gauges; 5/8"-2" lengths. Complies with ISO 7864, ISO 9626, ISO 10993, ASTM F88, ASTM F1929, ASTM F756.
Indications for Use
Indicated for multiple venous blood collections using vacuum tubes. Safety versions include a shield to prevent accidental needle stick injuries. Intended for prescription use.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
VACUTAINER® Brand Safety-Lok Blood Collection Set (K965202)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
August 26, 2020
Zhejiang Kindly Medical Devices Co.,Ltd % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.Box 120-119 Shanghai, Zhejiang 325025 China
Re: K172763
Trade/Device Name: Blood Collecting Needle, Safety Blood Collecting Needle, Blood Collecting Needle with Holder, Safety Blood Collecting Needle with Holder, Blood Collecting Set, Safety Blood Collecting Set, Blood Collecting Set with Holder, Safety Blood Collecting Set with Holder Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA
Dear Diana Hong:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 4, 2018. Specifically, FDA is updating this SE Letter to update the product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Vacant, OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices, Paval Patel, (240) 402-6029, payal.patel(@fda.hhs.gov.
Sincerely,
Sapana Patel -S
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices
{1}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size.
June 4, 2018
Zhejiang kindly medical devices Co., Ltd % Diana Hong Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 CHINA
Re: K172763
Trade/Device Name: Blood Collecting Needle, Safety Blood Collecting Needle, Blood Collecting Needle with Holder, Safety Blood Collecting Needle with Holder, Blood Collecting Set, Safety Blood Collecting Set, Blood Collecting Set with Holder, Safety Blood Collecting Set with Holder
Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: April 28, 2018 Received: May 4, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Tina Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172763
## Device Name
Blood Collecting Needle, Safety Blood Collecting Needle, With Holder, Safety Blood Collecting Needle with Holder, Blood Collecting Set, Safety Blood Collecting Set with Holder, Safety Blood Collecting Set with Holder
### Indications for Use (Describe)
The Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
The Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
The Blood Collecting Set is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety Blood Collecting Set is intended to be used with vacuum blood collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
The Blood Collecting Set with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety Blood Collecting Set with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
Type of Use *(Select one or both, as applicable)*
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Requisition Use (Part 21 CFR 601 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 601 Subpart G) |
|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
# Exhibit #2 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K172763
- 1. Date of Preparation: 04/28/2018
- 2. Sponsor Identification
# Zhejiang kindly medical devices Co., Ltd.
No.758, 5th Binhai Road, Binhai Industrial Park, Longwan District, 325025 Wenzhou, Zhejiang Province, PRC.
Establishment Registration Number: Not yet registered
Contact Person: Yong Zhang Position: General Manager Tel: +86-577-86960616 Fax: +86-577-86374972 Email: zjkdl@kdlchina.com
- Designated Submission Correspondent 3.
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
# Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{7}------------------------------------------------
## K172763
#### Identification of Proposed Device 4.
Trade Name: Blood Collecting Needle
- Safety Blood Collecting Needle Blood Collecting Needle with Holder Safety Blood Collecting Needle with Holder Blood Collecting Set Safety Blood Collecting Set Blood Collecting Set with Holder Safety Blood Collecting Set with Holder
Common Name: Blood Collecting Needle and Set
### Regulatory Information
Classification Name: Needle, Hypodermic, Single Lumen Classification: II; Product Code: FMI: Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital;
Indications for Use:
The Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
The Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
The Blood Collecting Set is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
{8}------------------------------------------------
The Safety Blood Collecting Set is intended to be used with vacuumblood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
The Blood Collecting Set with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety Blood Collecting Set with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
# Device Description
The proposed devices are blood collection devices form a channel between patient's vein and the vacuum blood collection tube. The patient-end needle is punctured into vein, and non-patient end needle will penetrate into vacuum blood collection tube, blood will flow into blood collection tube under the action of differential pressure. The proposed devices are divided into several types and available in different specifications, see following table
| Device Name | Ref No. | Needle<br>Gauge | Needle Length |
|--------------------------------------------|---------|-----------------|--------------------------------------------|
| Blood Collecting Needle | 04-11 | 18G~23G | 1", 1 1/4", 1 1/2" |
| Blood Collecting Needle | 04-13 | 18G~23G | 1", 1 1/4", 1 1/2" |
| Blood Collecting Needle | 04-31 | 19G, 25G | 5/8", 3/4", 1", 1 1/4", 1 1/2", 1 3/4", 2" |
| Safety Blood Collecting Needle | 04-12 | 18G~23G | 1", 1 1/4", 1 1/2" |
| Safety Blood Collecting Needle | 04-32 | 19G, 25G | 5/8", 3/4", 1", 1 1/4", 1 1/2", 1 3/4", 2" |
| Blood Collecting Needle with Holder | 04-14 | 18G~23G | 1", 1 1/4", 1 1/2" |
| Blood Collecting Needle with Holder | 04-16 | 18G~23G | 1", 1 1/4", 1 1/2" |
| Blood Collecting Needle with Holder | 04-33 | 19G, 25G | 5/8", 3/4", 1", 1 1/4", 1 1/2", 1 3/4", 2" |
| Safety Blood Collecting Needle with Holder | 04-15 | 18G~23G | 1", 1 1/4", 1 1/2" |
| Safety Blood Collecting Needle with Holder | 04-34 | 19G, 25G | 5/8", 3/4", 1", 1 1/4", 1 1/2", 1 3/4", 2" |
| Blood Collecting Set | 04-21 | 19G, 25G | 5/8", 3/4", 1" |
| Safety Blood Collecting Set | 04-22 | 19G, 25G | 5/8", 3/4", 1" |
| Blood Collecting Set with Holder | 04-23 | 19G, 25G | 5/8", 3/4", 1" |
| Safety Blood Collecting Set with Holder | 04-24 | 19G, 25G | 5/8", 3/4", 1" |
{9}------------------------------------------------
#### 5. Identification of Predicate Devices
Predicate Device 1 510(k) Number: K982541 Product Name: VACUTAINER® Brand ECLIPSE™ Blood Collection Needle
Predicate Device 2 510(k) Number: K965202 Product Name: VACUTAINER® Brand Safety-Lok Blood Collection Set
{10}------------------------------------------------
# 6. Technological Characteristics
| ITEM | Proposed Device | Predicate Device 1 | Predicate Device 2 |
|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K982541 | K965202 |
| Class | II | II | II |
| Indications for Use | The Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood.<br>The Safety Blood Collecting Needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.<br>The Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The Safety Blood Collecting Needle with Holder is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.<br>The Blood Collecting Set is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The Safety Blood Collecting Set is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The safety | The VACUTAINER® Brand ECLIPSETM Blood Collection Needle is a sterile, multiple sample, single-use device for blood collection. The needle is designed with an attached safety shield, which can be activated to cover the needle immediately after venipuncture to provide protection from accidental needle sticks. | The VACUTAINER® Brand Safety-Lok™ Blood Collection Set is a winged blood collection needle and flexible tubing for venipuncture to collect blood specimens from patients. The Safety-Lok™ Blood Collection Set also contains a needle safety shield which minimizes the possibility of needlesticks if manually activated following blood collection. The VACUTAINER® Brand Safety-Lok™ Blood Collection Set is also recommended for use in patients with small veins |
| shield is intended to aid in the protection against<br>accidental needle stick injury. | | | |
| The Blood Collecting Set with Holder is intended<br>to be used with vacuum blood collection tube for<br>multiple collections of venous blood. | | | |
| The Safety Blood Collecting Set with Holder is<br>intended to be used with vacuum blood collection<br>tube for multiple collections of venous blood. | | | |
| The safety shield is intended to aid in the<br>protection against accidental needle stick injury. | | | |
| Configuration<br>and material | Patient-end Needle (Stainless Steel) | Protective Cover of Needle | Protective Cover of Patient-end Needle |
| | Protective Cover of Patient-end Needle (PP) | Needle | Patient-end Needle |
| | Patient-end Needle Hub (PP) | Needle Hub | Needle Handle |
| | Non-patient end Needle Hub (ABS) | Rubber Sleeve | Double Wing |
| | Non-patient end Needle (Stainless Steel) | Safety Shield | Flexible Tube |
| | Rubber Sleeve (Natural Rubber) | | Conical Fitting Connector |
| | Protective Cover of Non-patient end Needle (PP) | | Conical Fitting |
| | Double Wing (PVC) | | Non-patient end Needle |
| | Flexible Tube (PVC) | | Rubber Sleeve |
| | Conical Fitting Connector (ABS) | | Protective Cover of Non Patient end<br>Needle |
| | Conical Fitting (ABS) | | Safety Shield |
| | Safety Shield (PP) | | |
| | Needle Holder (PP) | | |
| Operate mode | Manual | Same | Same |
| Specification | 18G, 19G, 20G, 21G, 22G, 23G, 25G | Similar | Similar |
| | 5/8", 3/4", 1", 1 1/4", 1 1/2", 1 3/4", 2" | | |
| Safety<br>Mechanism | The safety shield is intended to prevent needle<br>sticks | Same | Same |
{11}------------------------------------------------
{12}------------------------------------------------
| K172763 | |
|---------|--|
| Label/Labeling | Conform with Part 801 | Same | Same |
|------------------|----------------------------------------------------------------------|----------------------------------|---------------------------------|
| Sterilization | EO sterilized<br>SAL: $10^{-6}$<br>Endotoxin Limit: 20 EU per device | Same | Same |
| Single Use | Single Use | Same | Same |
| Biocompatibility | In Vitro Cytotoxicity | Conform with ISO 10993 Standards | Conform with ISO 10993 Standard |
| | Skin Sensitization | | |
| | Intracutaneous Reactivity | | |
| | Acute Systemic Toxicity | | |
| | Hemolytic Properties | | |
| | Pyrogen | | |
The proposed devices Blood Collecting Needle, Blood Collecting Needle with Holder, Safety Blood Collecting Neede with Holder are similar to the predicate device design, Indications for use, sterilization, method of operation and technological characteristics. The proposed devices Blood Collecting Set, Blood Collecting Set with Holder, Safety Blood Collecting Set with Holder are similar to the predicate device design, Indications for use, sterilization, method of operation and technological characteristics. The difference in configuration between proposed devices is that the proposed devices have the additional configuration of the needle holder. In addition the proposed devices are available in a series of specification will be selected by physician per patients' condition. These two differ intended use, therefore, the proposed devices can be determined substantially equivalence to predicate devices.
{13}------------------------------------------------
#### 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed devices met all design specifications. The test include following items
Physical, Mechanical, Chemical testing performed on the proposed device:
| Cleanliness | Clause 4.3 of ISO 7864:2016 |
|-----------------------------------|------------------------------|
| Limits for acidity or alkalinity | Clause 4.4 of ISO 7864:2016 |
| Limits for extractable metals | Clause 4.5 of ISO 7864:2016 |
| Size designation | Clause 4.6 of ISO 7864:2016 |
| Colour coding | Clause 4.7 of ISO 7864:2016 |
| Needle hub | Clause 4.8 of ISO 7864:2016 |
| Needle Cap | Clause 4.9 of ISO 7864:2016 |
| Needle tube | Clause 4.10 of ISO 7864:2016 |
| Needle point | Clause 4.11 of ISO 7864:2016 |
| Bond between hub and needle tube | Clause 4.12 of ISO 7864:2016 |
| Patency of lumen | Clause 4.13 of ISO 7864:2016 |
| Surface finish | Clause 5.2 of ISO 9626:2016 |
| Cleanliness | Clause 5.3 of ISO 9626:2016 |
| Limits for acidity and alkalinity | Clause 5.4 of ISO 9626:2016 |
| Size designation | Clause 5.5 of ISO 9626:2016 |
| Dimensions | Clause 5.6 of ISO 9626:2016 |
| Stiffness | Clause 5.8 of ISO 9626:2016 |
| Resistance to breakage | Clause 5.9 of ISO 9626:2016 |
| Resistance to corrosion | Clause 5.10 of ISO 9626:2016 |
Sterile Barrier Packaging Testing performed on the proposed device:
| Seal strength | ASTM F88/F88-15 |
|-----------------|-----------------|
| Dye penetration | ASTM F1929-15 |
Sterilization and Shelf Life Testing performed on the proposed device:
| EO residue | ISO 10993-7:2008 |
|--------------------------|-----------------------------------------------|
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin Limit | USP 38-NF 33 <85> |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package Tests |
{14}------------------------------------------------
# were performed on aging samples to verify the claimed shelf life of the device
Biocompatibility testing include in vitro cytotoxicity test (ISO 10993-5), skin sensitization test (ISO 10993-10), irritation sensitivity (ISO 10993-10), Acute Toxicity Test (ISO 10993-11), pyrogen test (ISO 10993-11) and Hemolysis Test (ASTM F756-17).
# Simulated Clinical Study
A simulated clinical study was performed on proposed device according to FDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 to evaluate the safety mechanism of the proposed devices. The results demonstrated that the proposed device met the pre-established criteria.
# Safety Feature Test
The safety feature test was performed on both proposed device and predicate device to determine safety feature. The results demonstrated that the proposed devices did not show a significant difference from predicate devices.
- 8. Clinical Test Conclusion
No clinical study is included in this submission.
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.