K172754 · STERIS Corporation · MLR · Feb 9, 2018 · General Hospital
Device Facts
Record ID
K172754
Device Name
V-PRO maX 2 Low Temperature Sterilization System
Applicant
STERIS Corporation
Product Code
MLR · General Hospital
Decision Date
Feb 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6860
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172754 · Feb 9, 2018
V-PRO maX 2 Low Temperature Sterilization System
STERIS Corporation
Clinically-cleaned, patient-soiled medical instruments
The 'In Use Test' was conducted to demonstrate the efficacy of the V-PRO maX 2 sterilization cycles on medical instruments that had been used in a clinical setting and subsequently cleaned via routine clinical procedures.
In Use Test; Performance validation using clinically-used instruments
Patient-soiled, clinically-cleaned medical instruments
Not applicable for this study
Sterilization efficacy on patient-soiled instruments
Indications for Use
The V-PRO maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues. Each Cycle can sterilize non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle and Fast Non Lumen Cycle. The Non Lumen Cycle can sterilize: Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes. The Fast Non Lumen Cycle can sterilize: Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes. The Flexible Cycle can sterilize: Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations: 1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. 2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. The Lumen Cycle can sterilize: Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations: Single channeled devices with a stainless lumen that is > 0.77 mm ID and ≤ 500 mm in length; Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length; Triple channeled devices with stainless lumens that are either: ≥ 1.2 mm ID and ≤ 275 mm in length, ≥ 1.8 mm ID and ≤ 310 mm in length, or ≥ 2.8 mm ID and ≤ 317 mm in length.
Device Story
V-PRO maX 2 is a vaporized hydrogen peroxide (VHP) sterilizer for healthcare facilities; uses VAPROX HC Sterilant (59% hydrogen peroxide). Operates via preprogrammed cycles (Lumen, Non Lumen, Flexible, Fast Non Lumen) involving conditioning, sterilization, and aeration phases. Input: cleaned/dried medical devices; Output: terminally sterilized devices ready for immediate use. Operated by healthcare staff. System uses RFID to read sterilant cups. Provides automated cycle control to ensure sporicidal efficacy; no cool-down required. Benefits include rapid processing of heat-sensitive instruments without toxic residues.
Clinical Evidence
Bench testing only. No clinical data. Efficacy demonstrated via AOAC sporicidal tests (720 carriers), D-value/total kill endpoint analysis (>12 log reduction), and 1/2 cycle verification for all four cycles under worst-case load conditions. Biocompatibility and material compatibility verified across 23 materials.
Technological Characteristics
Vaporized hydrogen peroxide sterilizer; uses VAPROX HC (59% H2O2). Proprietary microcomputer control board with memory backup. RFID reader for sterilant identification. Dimensions/form factor: floor-standing unit. Connectivity: data storage for up to 300 cycles. Standards: EN 61010-1:2010 (safety), EN 61326-1:2006 (EMC).
Indications for Use
Indicated for terminal sterilization of cleaned, rinsed, and dried medical devices in healthcare facilities. Suitable for heat/moisture-sensitive devices, including non-lumened instruments, rigid/semi-rigid/flexible endoscopes, and specific single/dual/triple lumen configurations. Contraindicated for materials incompatible with vaporized hydrogen peroxide.
Regulatory Classification
Identification
An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.
Predicate Devices
V-PRO maX Low Temperature Sterilization System (K162413)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure in profile, with three overlapping faces suggesting community and connection. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2018
STERIS Corporation Bill Brodbeck, Ph.D. Director, Regulatory Affairs 5960 Heisley Rd. Mentor, Ohio 44060
Re: K172754
Trade/Device Name: V-PRO® maX 2 Low Temperature Sterilization System Regulation Number: 21 CFR 880.6860 Regulation Name: Sterilizer, Ethylene Oxide Gas Regulatory Class: Class II Product Code: MLR Dated: January 9, 2018 Received: January 10, 2018
Dear Dr. Bill Brodbeck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172754
### Device Name
V-PRO® maX 2 Low Temperature Sterilization System
### Indications for Use (Describe)
The V-PRO maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.
Each Cycle can sterilize non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle and Fast Non Lumen Cycle.
The Non Lumen Cycle can sterilize: †
Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes.
t: The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22.7 kg).
The Fast Non Lumen Cycle can sterilize: *
Non-lumened instruments including non-luments, non-luments, non-lumened rigid, semi-rigid and flexible endoscopes.
* The validation studies were conducted using a validation load consisting of one pouched instrument tray for a total weight of 11 lbs (5 kg).
The Flexible Cycle can sterilize:
Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations:
1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. * The flexible endoscopes may contain either:
- · A single lumen that is ≥ 1 mm internal diameter (ID) and ≤ 1050 mm in length
- · Or two lumens with:
- One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
- And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
* The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. ++ The flexible endoscope may contain either:
• A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length
• Or two lumens with:
- One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
- And the other lumen is ≥ 1 mm ID and ≤ 850 mm in length.
** The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).
{3}------------------------------------------------
The Lumen Cycle can sterilize: †
Medical devices (including single, dual and triple channeled rigid endoscopes) with the following configurations:
- Single channeled devices with a stainless lumen that is > 0.77 mm ID and ≤ 500 mm in length
- Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
- · Triple channeled devices with stainless lumens that are either:
- ≥ 1.2 mm ID and ≤ 275 mm in length
- > 1.8 mm ID and < 310 mm in length
or
- ≥ 2.8 mm ID and ≤ 317 mm in length
* Validation testing for all lumen sizes was conducted using a maximum of 20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs (8.9 kg).
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
മ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in a bold, sans-serif font, with the registered trademark symbol to the right of the word. Below the word "STERIS" is a graphic of several horizontal, wavy lines in blue, resembling water waves.
## 510(k) Summary For
# V-PRO® maX 2 Low Temperature Sterilization System
## K172754
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Contact:
Bill Brodbeck Director, Regulatory Affairs Tel: 440-392-7690 Fax: 440-357-9198
Submission Date: February 6, 2018
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
### 1. Device Name
| Trade Name: | V-PRO® maX 2 Low Temperature Sterilization<br>System |
|------------------------|------------------------------------------------------|
| Device Class: | Class II |
| Common/usual Name: | Vapor Phase Hydrogen Peroxide Sterilizer |
| Classification Name: | Sterilizer, Ethylene Oxide Gas |
| Classification Number: | 21 CFR 880.6860 |
| Product Code: | MLR |
### 2. Predicate Device
The claimed primary predicate device is the V-PRO maX Low Temperature Sterilization System, cleared most recently under K162413.
The V-PRO maX Low Temperature Sterilization System prior clearances include Premarket Submissions K102330, K112760, K112813, K120632, K131120 and K162413.
| System to the predicate device is summarized in the table below. | | |
|------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | V-PRO maX 2 Low Temperature<br>Sterlization System<br>(Proposed Device) | V-PRO maX Low Temperature Sterlization<br>System<br>(Predicate Device/K162413) |
| Indications<br>for Use | The V-PRO maX 2 Low Temperature<br>Sterilization Systems using VAPROX HC<br>Sterilant are intended for use in the<br>terminal sterilization of properly prepared<br>(cleaned, rinsed and dried) medical<br>devices in Healthcare Facilities. The<br>preprogrammed sterilization cycles<br>operate at low pressure and temperature,<br>suitable for processing medical devices.<br><br>Each Cycle can sterilize non-lumened<br>instruments with diffusion-restricted<br>spaces such as the hinged portion of<br>forceps and scissors. Only stainless steel<br>or titanium diffusion-restricted spaces<br>should be processed in the Non Lumen<br>Cycle and the Fast Non Lumen Cycle.<br><br>The Fast Non Lumen Cycle can<br>sterilize: * | The V-PRO 1, V-PRO 1 Plus and V-PRO<br>maX Low Temperature Sterilization Systems,<br>with VAPROX® HC Sterilant, are vaporized<br>hydrogen peroxide sterilizers intended for use<br>in the terminal sterilization of cleaned, rinsed<br>and dried metal and nonmetal medical devices<br>used in healthcare facilities. The three pre-<br>programmed sterilization cycles (Lumen<br>Cycle, Non Lumen Cycle, and Flexible Cycle)<br>operate at low pressure and low temperature<br>and are thus suitable for processing medical<br>devices sensitive to heat and moisture.<br><br>The V-PRO 1, V-PRO 1 Plus and V-PRO<br>maX Low Temperature Sterilization System's<br>Lumen Cycle, can sterilize:a<br>Lumened and non-lumened instruments<br>with diffusion-restricted spaces such as the<br>hinged portion of forceps and scissors Medical devices, including single, dual and<br>triple channeled rigid and semi-rigid |
| Feature | V-PRO maX 2 Low Temperature<br>Sterilization System<br>(Proposed Device) | V-PRO maX Low Temperature Sterilization<br>System<br>(Predicate Device/K162413) |
| | Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.<br>* The validation studies were conducted using a validation load consisting of one pouched instrument tray for a total weight of 11 lbs (5 kg). | endoscopes, with the following configurations: a single channeled devices with a stainless steel lumen that is ≥ 0.77 mm internal diameter (ID) and ≤ 500 mm in length dual lumen devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length triple lumen devices with stainless steel lumens that are ≥ 1.2 mm ID and ≤ 275 mm in length ≥ 1.8 mm ID and ≤ 310 mm in length or ≥ 2.8 mm ID and ≤ 317 mm in length |
| | The Non Lumen Cycle can sterilize: ‡<br>Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.<br>‡ The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22.7 kg). | a The validation studies for all channel/ lumen configurations were conducted using a maximum of twenty (20) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs. |
| | The Flexible Cycle can sterilize:<br>Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two load configurations:<br><br>1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. * The flexible endoscopes may contain either: A single lumen that is ≥ 1 mm internal diameter (ID) and ≤ 1050 mm in length Or two lumens with: One lumen that is ≥ 1 mm ID and ≤ 990 mm in length And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length * The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).<br><br>2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. †† The flexible endoscope may contain either: A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length | The V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems' Non Lumen Cycle can sterilize:b Non-lumened instruments including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened instruments with stainless steel or titanium diffusion-restricted areas such as the hinged portion of forceps or scissors. |
| | | b The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50.0 lbs. |
| | | The V-PRO maX Low Temperature Sterilization System's Flexible Cycle can sterilize single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and |
| Feature | V-PRO maX 2 Low Temperature<br>Sterilization System<br>(Proposed Device) | V-PRO maX Low Temperature Sterilization<br>System<br>(Predicate Device/K162413) |
| | One lumen that is ≥ 1 mm ID and ≤ 990 mm in length And the other lumen is ≥ 1 mm ID and ≤ 850 mm in length. †† The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).<br>The Lumen Cycle can sterilize: † Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations: Single channeled devices with a stainless lumen that is ≥ 0.77 mm ID and ≤ 500 mm in length Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length Triple channeled devices with stainless lumens that are either: ≥ 1.2 mm ID and ≤ 275 mm in length ≥ 1.8 mm ID and ≤ 310 mm in length or ≥ 2.8 mm ID and ≤ 317 mm in length † Validation testing for all lumen sizes was conducted using a maximum of 20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs (8.9 kg). | bronchoscopes in either of two load configurations: Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load.c<br>The flexible endoscopes may contain either: a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length or two lumens with: one lumen that is ≥ 1 mm ID and ≤ 990 mm in length and the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length c The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope). One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments including instruments with diffusion-restricted areas such as the hinged portion of forceps or scissors.d<br>The flexible endoscope can contain either: a single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length or two lumens with: a single lumen that is ≥ 1 mm ID and ≤ 990 mm in length and the other lumen that is ≥ 1 mm and ≤ 850 mm in length d The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total load weight of 24.0 lbs. |
| Process<br>Parameters | The critical process parameters are:<br>Time Chamber Temperature Vaporizer Temperature Chamber Pressure Prior to Injection Sterilant Injection Weight | The critical process parameters are:<br>Time Chamber Temperature Vaporizer Temperature Chamber Pressure Prior to Injection Sterilant Injection Weight |
| Feature | V-PRO maX 2 Low Temperature<br>Sterlization System<br>(Proposed Device) | V-PRO maX Low Temperature Sterlization<br>System<br>(Predicate Device/K162413) |
| Software/<br>Firmware<br>Controlled | Control system consists of a proprietary<br>microcomputer control board and<br>peripheral function circuit boards, located<br>within the control housing. A memory<br>backup system maintains user settings<br>and calibration data indefinitely. Up to<br>300 cycle data files can be stored for<br>review or downloading by the user.<br><br>The software allows user selection of<br>either the Lumen, Non Lumen, Flexible<br>or Fast Non Lumen pre-programmed<br>cycle. | Programmable Logic Control (PLC). The<br>software allows user selection of either the<br>Lumen, Non Lumen, or Flexible pre-<br>programmed cycle. |
| Total<br>Cycle<br>Time | Lumen Cycle - 52 minutes<br>Non Lumen Cycle - 28 minutes<br>Flexible Cycle - 35 minutes<br>Fast Non Lumen Cycle – 16 minutes | Lumen Cycle - 55 minutes<br>Non Lumen Cycle - 28 minutes<br>Flexible Cycle - 35 minutes |
| Sterilant | VAPROX HC Sterilant (59% Hydrogen<br>Peroxide).<br>The same amount of sterilant is injected<br>for each of the sterilization pulses for all<br>three cycles. | VAPROX HC Sterilant (59% Hydrogen<br>Peroxide).<br>The same amount of sterilant is injected for<br>each of the sterilization pulses for all three<br>cycles. |
| | Sterilant Cup is read by an RFID reader. | The sterilant cup is read by a 2-D matrix code |
| Accessories | Accessories were submitted under<br>separate, individual, concurrent 510(k)s<br>and cover the following:<br>Self-contained biological indicator Biological indicator challenge pack Fast Acting Biological Indicator Chemical indicator Trays & Tray Accessories Pouches | The following accessories are available for the<br>V-PRO maX Low Temperature Sterilization<br>System<br>Self-contained biological indicator Biological indicator challenge pack Fast Acting Biological Indicator Chemical indicator Trays & Tray Accessories Pouches |
A comparison between the proposed V-PRO maX 2 Low Temperature Sterilization System to the predicate device is summarized in the table helow
{6}------------------------------------------------
# TRADITIONAL PREMARKET NOTIFICATION
V-PRO® maX 2 Low Temperature Sterilization System
{7}------------------------------------------------
{8}------------------------------------------------
### TRADITIONAL PREMARKET NOTIFICATION V-PRO® maX 2 Low Temperature Sterilization System
The proposed device has an intended use similar to the predicate with the differences being the addition of the Fast Non Lumen Cycle and slight modification to the text to allow conciseness and clarity. Both the predicate and proposed devices have the same technological characteristics. Although, the devices slightly differ, the provided descriptive characteristics and performance data demonstrate equivalence. Therefore, the proposed V-PRO maX 2 Low Temperature Sterilization System is substantially equivalent to the predicate device, the V-PRO maX Low Temperature Sterilization System.
### 3. Description of Device
The V-PRO maX 2 Low Temperature Sterilization System is a new vaporized hydrogen peroxide sterilizer model to be added to the STERIS V-PRO family of
{9}------------------------------------------------
sterilizers. The V-PRO product line currently consists of the Amsco V-PRO 1, Amsco V-PRO 1 Plus, V-PRO maX and V-PRO 60 Sterilizers.
As with the predicate device (K162413), the V-PRO maX 2 Sterilizer has three preprogrammed cycles (the Lumen Cycle, the Non Lumen Cycle and the Flexible Cycle) with a fourth added (the Fast Non Lumen Cycle). The V-PRO maX 2 Low Temperature Sterilization System is intended for terminal sterilization of cleaned, rinsed, dried and packaged reusable surgical instruments used in healthcare facilities.
The V-PRO maX 2 Sterilizer uses VAPROX® HC Sterilant to sterilize the intended devices through exposure to vaporized hydrogen peroxide (VHP). Its four preprogrammed cycles all utilize a conditioning phase, a sterilize phase and an aeration phase. The packaged sterilized devices are ready for use at the completion of the cycle, no cool down or aeration period is required following completion of the cycle.
### 4. Intended Use
The V-PRO maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.
Each Cycle can sterilize non-luments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle and Fast Non Lumen Cycle.
## The Non Lumen Cycle can sterilize: +
Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.
* The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22.7 kg).
## The Fast Non Lumen Cycle can sterilize: *
Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.
* The validation studies were conducted using a validation load consisting of one pouched instrument tray for a total weight of 11 lbs (5 kg).
## The Flexible Cycle can sterilize:
Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two load configurations:
{10}------------------------------------------------
1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. * The flexible endoscopes may contain either:
- A single lumen that is > 1 mm internal diameter (ID) and < 1050 mm in length
- · Or two lumens with:
- -One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
- And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length
* The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).
2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-luments. ** The flexible endoscope may contain either:
- · A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length
- · Or two lumens with:
- . One lumen that is ≥ 1 mm ID and ≤ 990 mm in length
- . And the other lumen is > 1 mm ID and < 850 mm in length.
ff The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).
### The Lumen Cycle can sterilize: أ
Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:
• Single channeled devices with a stainless lumen that is ≥ 0.77 mm ID and ≤ 500 mm in length
• Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length
· Triple channeled devices with stainless lumens that are either:
≥ 1.2 mm ID and ≤ 275 mm in length
- ≥ 1.8 mm ID and ≤ 310 mm in length
or
≥ 2.8 mm ID and ≤ 317 mm in length
* Validation testing for all lumen sizes was conducted using a maximum of 20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs (8.9 kg).
#### Description of Safety and Substantial Equivalence ട്.
The V-PRO maX 2 Low Temperature Sterilization System has the same intended use and different technological characteristics that do not raise different questions of safety and effectiveness as compared to the predicate device. Performance testing to assess and demonstrate substantial equivalence to the predicate is summarized below.
| Test | Result | Conclusion |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| AOAC Sporicidal<br>Test | All 720 carriers processed using 3 lots of EOSL<br>sterilant were sterile. | PASS |
| Test | Result | Conclusion |
| Determination of D-<br>value and Total Kill<br>Endpoint | Greater than a 12 log reduction of the most<br>resistant organism is achieved within all cycles<br>(Lumen Cycle, Non Lumen Cycle, Flexible Cycle<br>and Fast Non Lumen) of the V-PRO maX 2<br>Sterilizer. | PASS |
| 1/2 Cycle Modified<br>Total Kill Endpoint<br>Verification | Modified total kill end point analysis was<br>demonstrated for all four V-PRO maX 2<br>Sterilizer cycles. The standard injection weight<br>of 2.1 g and at least one lower injection weight<br>resulted in all sterile results within the validation<br>load used to qualify each sterilizer cycle. Partial<br>positives or all survive results were seen at lower<br>injection weights. | PASS |
| 1/2 Cycle<br>Sterilization<br>Verification of<br>Cycle Claims | • The Lumen Cycle reproducibly sterilizes<br>single, dual and triple lumen devices under<br>worst case conditions in 1/2 Cycle<br>• The Flexible Cycle reproducibly sterilizes 1 x<br>990 mm flexible endoscope lumens under<br>worst case conditions in 1/2 Cycle<br>• The Non Lumen Cycle reproducibly sterilizes<br>non-lumened devices under worst case<br>conditions in 1/2 Cycle<br>• The Fast Non Lumen Cycle reproducibly<br>sterilizes non-lumened devices under worst<br>case conditions in 1/2 Cycle | PASS |
| ½ Cycle Verification<br>of Mated Surfaces | Sterile efficacy was demonstrated for mated<br>surfaces under worst case conditions | PASS |
| Simulated Use Test | Simulated use testing verified the ability of the<br>V-PRO maX 2 Sterilizer cycles to sterilize<br>medical devices under worst case processing<br>conditions. | PASS |
| In Use Test | The in use investigation demonstrated the ability<br>of the V-PRO maX 2 Sterilizer cycles to sterilize<br>patient-soiled, clinically-cleaned, medical<br>instruments. | PASS |
| Biocompatibility | Cytotoxicity and residue analysis of 23 materials<br>have demonstrated biocompatibility after<br>processing in the V-PRO 60 Sterilizer. | PASS |
| Medical Device<br>Material<br>Compatibility | Evaluation of medical devices after multiple<br>cycles in the V-PRO maX 2 Sterilizer has<br>demonstrated compatibility with 23 materials of<br>construction. | PASS |
| Final Process<br>Qualification | The V-PRO maX 2 Sterilizer final process<br>qualification was successful for all four (4)<br>sterilizer cycles. | PASS |
{11}------------------------------------------------
# TRADITIONAL PREMARKET NOTIFICATION
V-PRO® maX 2 Low Temperature Sterilization System
{12}------------------------------------------------
The V-PRO maX 2 Low Temperature Sterilization System has been tested for conformity and is certified to the following standards:
- EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control and laboratory use. General requirements; Part 1: General Requirements
- . EN 61326-1:2006 Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
### 6. Conclusion
Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs at least as well as the legally marketed predicate device K162413, Class II (21 CFR 880.6860), product code MLR.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.