The iVitri®Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Device Story
iVitri® Straw is a cryopreservation storage device for human embryos; used in clinical embryology laboratories. Device consists of a polystyrene stick with a curved tip for embryo loading and an ABS cap. Embryos are loaded onto the tip, then the cap is applied to create a hermetic seal for storage in liquid nitrogen. The device facilitates rapid cooling (-3,022°C/min) and warming (+40,976°C/min) during vitrification. The closed system protects embryos during handling and storage. Healthcare providers (embryologists) use the device to preserve embryos for future use, potentially improving clinical outcomes in assisted reproduction.
Clinical Evidence
Bench testing only. Performance validated via cooling/warming rate analysis, durability/closure integrity testing in liquid nitrogen, endotoxin testing (USP <85>), and Mouse Embryo Assay (MEA) (≥80% expanded blastocyst formation at 96 hours). Sterilization validated per ISO 11137. Package integrity and shelf-life confirmed via ASTM F1980, F929, and F88 standards.
Indicated for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 9, 2018
Reprobiotech Corp. Huai L. Feng, Ph.D. Director 42-31 Colden Street. Suite 202 Flushing, NY 11355
Re: K172751
Trade/Device Name: iVitri® Straw Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MOK Dated: January 8, 2018 Received: January 16, 2018
Dear Huai L. Feng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172751
Device Name iVitri® Straw
#### Indications for Use (Describe)
The iVitri®Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------|--|
| <div> <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | |
| <div> <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary K172751 - iVitri® Straw
#### 1. Submission Sponsor:
#### Reprobitech Corp.
42-31 Colden St., Suite 202 Flushing, New York 11355 Tel: (516) 301-6171 Email: Reprobiotech@gmail.com
# 2. Contact Person:
Dr. Huai L. Feng Director of Reprobiotech Corp. 42-31 Colden Street, Suite 202 Flushing, NY 11355 Doctorf99@gmail.com Tel: (516) 301-6171
## 3. Date Prepared: February 8, 2018
## 4. Device Information:
| Name of Device | iVitri® Straw |
|---------------------|--------------------------------------|
| Common Name | Cryopreservation Storage Device |
| Classification Name | Assisted Reproduction Labware |
| Regulation | 21 CFR 884.6160 |
| Product Code | MQK (Labware, Assisted Reproduction) |
| Predicate | K162640-iVitri® EZ |
| Regulatory Class | II |
#### 5. Predicate Device Information:
iVitri® EZ (K162640) manufactured by Reprobiotech Corp. This device has not been subject to a design-related recall.
#### Device Description: 6.
The iVitri® Straw device is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. The iVitri® Straw device consists of a square-shaped polystyrene stick and acrylonitrile butadiene styrene (ABS) cap. As part of the vitrification procedure, the embryos are loaded on the tip of the stick, and capped for subsequent storage following vitrification. The tip of the storage device is curved to aid in maintaining and securing the embryos during handling procedures. The stick and cap include a tapered design that creates a hermetic seal, forming a closed storage system. Markings on the stick and tip of the device are used to aid in the proper orientation during embryo loading procedures. The device is provided sterile and is for single use only. Product specifications are listed in the table below:
{4}------------------------------------------------
| Parameter | Specification |
|---------------------------------|-----------------------------------------------------------|
| Cooling Rate | -3,022°C/min |
| Warming Rate | +40,976°C/min |
| Sterility Assurance Level (SAL) | 10-6 |
| Endotoxin | ≤0.5 EU/device |
| Mouse Embryo Assay (MEA) | 1-Cell MEA ≥80% expanded blastocyst formation at 96 hours |
## 7. Indication for Use:
The iVitn® Straw is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
## 8. Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices:
| Parameter | K172751 (subject device) | K162640 (predicate device) |
|-------------------------------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Same as the predicate device | The iVitri® EZ is a cryopreservation storage device that is<br>intended for use in vitrification procedures to contain and<br>maintain human 4-8 cell and blastocyst stage embryos. |
| Fundamental design | Same as the predicate device | Consists of two components: a body with a tip, where the<br>embryos are loaded, and a cap. The device is closed by<br>surface friction of the body and cap. |
| Body | Same as the predicate device | Consists of a single molded polystyrene stick that<br>incorporates a curved sample loading area. |
| Cap | Weighted, straw-like cap | Non-weighted, stick-like cap |
| Dimensions | Overall length: 130 mm<br>Body length: 105 mm<br>Cap length: 90 mm<br>Tip length:16 mm | Overall length: 130 mm<br>Body length: 116 mm<br>Cap length: 45.5 mm<br>Tip length:16 mm |
| Loading capacity | Same as the predicate device | Maximum of 3 embryos in ≤1 μl of medium |
| Mode of action | Same as the predicate device | The device is closed prior to freezing in liquid nitrogen<br>and opened during warming (i.e., the embryos have direct<br>contact with warming media). |
| Cooling rate | -3,022°C/min | -3,020°C/min |
| Warming rate | +40,976°C/min | +40,694°C/min |
| Embryo contacting<br>material | Same as the predicate device | Polystyrene including colorant (blue, orange, yellow,<br>green, pink) |
The subject and predicate devices have Indications for Use, fundamental design, mode of action, material used in the embryo-contacting component, and embryo loading capacity. They also have comparable cooling and warming rates. There are differences in dimensions and design of the cap between the subject and predicate devices. However, these differences do not raise different questions of safety and effectiveness.
#### 9. Non-Clinical Characteristics Performance Data:
As part of demonstrating substantial equivalence to the following non-clinical performance tests were conducted. The iVitri® Straw passed all of the testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device:
- . Cooling/warming rate testing: The purpose of this testing was to evaluate the cooling and warming rates of the subject device. Temperature recording equipment was used to document the temperature profile of devices when used as prescribed in the Instructions for Use throughout the cooling and warming portions of the vitrification process. Results showed a cooling rate of -3,022°C/min, and a warming rate of +40,976°C/min.
{5}------------------------------------------------
- . Durability and closure integrity testing: The purpose of this testing was to assess the integrity of the device following exposure to liquid nitrogen. The samples were assembled as described in the Instructions for Use and submerged in liquid nitrogen. The samples were then exposed to 37℃ temperatures. Samples were then assessed for signs of leakage, damage (breaks, ctc.), deformation or discoloration. The acceptance criteria were that devices must not exhibit any of the failure modes described above. Test samples met the acceptance criteria.
- . Endotoxin testing per USP<85>: < 0.5 EU/device
- Mouse Embryo Assay (MEA): 1-Cell mouse embryos were incubated in extracts of the subject device at ● 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage at 96 hours were assessed in comparison with the control group. The acceptance criterion was 1-Cell MEA: ≥80% expanded blastocyst formation at 96 hours.
- . Sterilization validation per ISO 11137-1:2006(R)2011 and ISO 11137-2:2013.
- Package integrity testing following accelerated aging per ASTM F1980-16:
- Dye penetration testing per ASTM F929-15 o
- Seal strength testing per ASTM F88/F88M-15 o
- Shelf-life studies (real-time and accelerated) were conducted to ensure that the following product specifications were met:
- MEA O
- Durability and closure integrity testing o
- Device appearance (discoloration, deformation, damage) o
- Dimensional assessments o
## 10. Conclusion:
The results of the testing described above demonstrate that the iVitri® Straw device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.