V2K Rinspiration System

K172716 · V2k Medical, Inc. · QEW · Jan 26, 2018 · Cardiovascular

Device Facts

Record IDK172716
Device NameV2K Rinspiration System
ApplicantV2k Medical, Inc.
Product CodeQEW · Cardiovascular
Decision DateJan 26, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The V2K Rinspiration System is intended to infuse physician specified fluids and remove/aspirate fluid, fresh, soft emboli and thrombi from the peripheral vasculature.

Device Story

V2K Rinspiration System is a multi-lumen catheter system for fluid mechanical thrombectomy in peripheral vasculature. Input: physician-specified fluids (via infusion syringe) and vacuum source (via hospital line/pump). Operation: catheter placed over 0.014" guidewire; infusion holes dispense fluid while central lumen aspirates emboli/thrombi. Accessory Filter Line connects to vacuum source, featuring in-line flow control switch and filter for debris collection/visualization. Used in clinical settings by physicians. Output: removal of thrombi/emboli; visual confirmation of aspirated material in filter. Benefits: enables simultaneous infusion and aspiration to clear peripheral vessels; provides simple flow management and debris inspection.

Clinical Evidence

Bench testing only. No clinical data. Performance verified via bench and lab testing per ISO 10555-1 and V2K methods, including simulated use, flow rate verification, and fatigue testing. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, and hemocompatibility (PTT, complement activation, in vivo thrombogenicity in dogs). All tests passed.

Technological Characteristics

Multi-lumen catheter; 7.0 French; 135cm length; 0.014" guidewire compatible. Materials: medical grade plastics, stainless steel, silicone oil coating. Radiopaque markers at tip and infusion site. E-Beam sterilization (SAL 10^-6). Accessories: 12cc infusion syringe with check valve, filter line with flow control switch. Connectivity: manual connection to external vacuum source.

Indications for Use

Indicated for patients requiring infusion of physician-specified fluids and aspiration of fluid, fresh, soft emboli, and thrombi from the peripheral vasculature.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below. January 26, 2018 V2K Medical Inc. Jake Wolenberg Quality and Regulatory Consultant 1221 Innsbruck Drive Sunnyvale, California 94089 Re: K172716 Trade/Device Name: V2K Rinspiration System, P7 Catheter - 9mm Rinse x 135cm Working Leneth, V2K Rinspiration System, P7 Catheter - 20mm Rinse x 135cm Working Length, V2K Rinspiration System, Filter Line - 108" Length Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE, KRA Dated: December 13, 2017 Received: December 18, 2017 Dear Jake Wolenberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good {1}------------------------------------------------ manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K172716 Device Name V2K Rinspiration System Indications for Use (Describe) The V2K Rinspiration System is intended to infuse physician specified fluids and remove/aspirate fluid, fresh, soft emboli and thrombi from the peripheral vasculature. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020. See PRA Statement below. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is a logo featuring a large letter 'V' split into two colors, green on the left and blue on the right. Below the 'V' is the text 'V2K Medical' in a smaller font size. The entire logo is encircled by a green ring, creating a circular emblem. The logo appears to be for a medical company. # 510(k) Summary V2K Rinspiration System # A. Submitter Information Submitter's Name: Address: Telephone: Fax: Email: Contact Person: Date of Preparation: ## B. Subject Device Proprietary Name: 510(k) #: Common/Usual Name: Classification Name: Product Code: ## C. Predicate Device Proprietary Name: 510(k) #'s: Common/Usual Name: Classification Name: Product Code: V2K Medical Inc. 1221 Innsbruck Drive Sunnyvale, CA 94089 408-940-5587 408-726-2977 jwolenbergV2K(@gmail.com Jake Wolenberg January 25, 2018 V2K Rinspiration System K172716 Embolectomy Catheter and Infusion Catheter Catheter, Embolectomy and; Catheter, Continuous Flush DXE per 21 C.F.R. 870.5150 and; KRA per 21 C.F.C 870.1210 Kerberos Proximal Solutions Rinspiration System K062275 Embolectomy Catheter Catheter, Embolectomy DXE per 21 C.F.R. 870.5150 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a logo for V2K Medical. The logo consists of a green circle with a white background. Inside the circle is a large letter V, with the left half of the V being green and the right half being blue. Below the V is the text "V2K Medical" in black font. # D. Device Description: The V2K Rinspiration System is comprised of the V2K Rinspiration Catheter and two accessory components; an Infusion Syringe and a Filter Line. The V2K Rinspiration Catheter is a multi-lumen catheter that has infusion holes located near the distal end of the catheter to dispense an infusible fluid. The central lumen of the catheter is used for aspiration. Simultaneous infusion and aspiration can be performed through these lumens to perform fluid mechanical thrombectomy to remove fluid, fresh, soft emboli and thrombi from peripheral vasculature. The Rinspiration Catheter will be placed in the target vasculature of a patient over an 0.014" guide wire. The V2K Rinspiration Catheter uses a rapid exchange (a.k.a. rail) configuration for the guidewire lumen. The Rinspiration catheter distal tip will be positioned at the intended treatment site. Infusion occurs approximately 1 to 2cm proximal to the distal tip. The catheter includes a smaller radiopaque marker band at the distal tip and two larger radiopaque marker bands designating the infusion portion of the catheter. The catheter has a V-hub on the proximal end that allows access to the infusion and aspiration lumens. Included in the same packaging as the V2K Rinspiration Catheter is a 12cc Infusion Syringe with an attached check valve. The Infusion Syringe is intended to be used for infusing physician specified fluids through the V2K Rinspiration Catheter. The V2K Rinspiration Catheter will also be supplied with an accessory Filter Line which is intended to be used to connect the catheter to a continuous vacuum source such as an aspiration pump or hospital vacuum line. The Filter Line has two connections, one male Luer and one slip-fit connector. The male Luer is connected to the aspiration lumen of the V2K Rinspiration Catheter. The slip-fit connector is connected to the continuous vacuum source. The Filter Line also has an in-line flow control switch and an in-line filter. The in-line flow control switch is used to open/close the connection between the continuous vacuum source and the aspiration lumen of the catheter. This allows the user a simple method to manage the aspiration flow. The in-line filter is used to collect any debris that is aspirated through the catheter and allows the user to clearly visualize what has been aspirated. This also allows for temporary storage, inspection, analysis, transport and / or disposal of the aspirated materials. # E. Intended Use: The V2K Rinspiration System is intended to infuse physician specified fluid and remove/aspirate fluid, fresh, soft emboli and thrombi from the peripheral vasculature. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image is a logo for V2K Medical. The logo features a large letter "V" with the left side in dark green and the right side in blue. Below the "V" is the text "V2K Medical" in a smaller font. The entire logo is encircled by a thick, dark green ring, creating a circular emblem. # F. Predicate Comparison: The subject V2K Rinspiration Catheter has the same basic catheter design, intended use, and operating principles as the predicate Kerberos Rinspiration System. The primary update made was to replace the handheld pump (Rinspirator) that was used as part of the Kerberos Rinspiration System to simultaneously operate both an infusion syringe and an aspiration syringe. In the V2K Rinspiration System, the Rinspirator was replaced with the 12cc Infusion Syringe and the Rinspiration Filter Line. These accessories still allow for simultaneous infusion and aspiration. The catheter hub design was also updated to help make it easier to identify how to connect the catheter to the accessory devices. Additional updates can be found in the table below which compares the subject device models included in the 510(k) to the most similar predicate device models. | | Predicate Device | Subject Device | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Kerberos Proximal Solutions<br>Rinspiration System | V2K Rinspiration System | | Product Code | DXE per 21 C.F.R. 870.5150 | DXE per 21 C.F.R. 870.5150<br>KRA per 21 C.F.C 870.1210 | | Class | Class II | Same | | Name | Embolectomy Catheter<br>Infusion Catheter | Same | | Indications for Use | The Kerberos Proximal Solutions<br>Rinspiration Catheter System is<br>intended to infuse physician<br>specified fluid and remove/aspirate<br>fresh, soft emboli and thrombi from<br>the coronary and peripheral<br>vasculature. | The V2K Rinspiration System is<br>intended to infuse physician<br>specified fluid and remove/aspirate<br>fluid, fresh, soft emboli and thrombi<br>from the peripheral vasculature. | | Catheter &<br>Accessory Materials | Commonly used medical grade<br>plastics & stainless steel. | Same | | Lubricious Coating | Silicone Oil Solution | Same | | Catheter Visibility | Radiopaque Tip and Infusion<br>Markers | Same | | Outer Diameter | 7.0 French (<0.093") | Same | | Aspiration Lumen<br>ID | ID ≥ 0.050" | ID ≥ 0.048" | | Infusion Hole<br>Spacing | 9mm | 9mm & 20mm | | Effective Length | 135cm | Same | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image is a logo for V2K Medical. The logo consists of a green circle with a large letter "V" inside. The left half of the "V" is dark green, while the right half is blue. Below the "V" is the text "V2K Medical" in black font. | | Predicate Device | Subject Device | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------| | | Kerberos Proximal Solutions<br>Rinspiration System | V2K Rinspiration System | | Guidewire<br>Compatibility | 0.014” | Same | | Maximum Infusion<br>Pressure | 3.1 psi | Same | | Catheter Accessories | 3cc Infusion syringe<br>Infusion Supply Line<br>Rinspirator Handle<br>5cc aspiration syringe<br>In-Line Filter<br>Aspiration bag | 12cc Infusion Syringe with Check<br>Valve<br>Filter Line | | Sterile Barrier | Tyvek with polymer backing | Same | | Sterilization | EtO, SAL 10-6 | E-Beam, SAL 10-6 | | Condition Supplied | Sterile, Single Use, Disposable | Same | ## G. Performance Data Supporting Substantial Equivalence: Bench and Lab testing was conducted in order to evaluate the differences between the proposed V2K Rinspiration System and the predicate Kerberos Rinspiration System. As V2K has access to the original design verification and validation documentation for the predicate Kerberos Rinspiration System, the data from this testing was leveraged, where possible, for comparison to the proposed V2K Rinspiration System. The test results were reviewed and found to demonstrate that the differences between the proposed and predicate device do not significantly impact any catheter performance parameters that would affect the safety or efficacy of the proposed V2K Rinspiration Catheter. A summary of the tests and performance specifications that were evaluated is presented on the following pages. These tests were performed per V2K approved methods based on prior Kerberos test methods and catheter performance standard ISO 10555-1. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image is a logo for V2K Medical. The logo consists of a green circle with a large letter V inside. The left half of the V is dark green, and the right half is blue. Below the V, the text "V2K Medical" is written in black. V2K Medical Inc. 1221 Innsbruck Drive Sunnyvale, CA 94089 Sunnyvale, CA 94089 Phone: 408-940-5587 | Test Performed | Specification Summary | |-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Visual Inspection | The product and packaging shall be free of significant defects or<br>assembly issues. | | Dimensional<br>Verification | All product dimensions critical to device performance shall meet the<br>product specifications with the required tolerances. | | Catheter Bond<br>Strength | All bonds/joints made throughout the catheter and accessories shall<br>be strong enough to prevent separation during normal use. | | Fatigue Testing -<br>Infusion | The catheter and accessories shall be able to function and maintain<br>integrity following infusion of a worst-case fluid volume through the<br>infusion lumen. | | Fatigue Testing -<br>Filter Line | The Filter Line shall function and maintain integrity following<br>continuous aspiration for a clinically relevant time. | | Fatigue Testing -<br>Flow Control Switch | The flow control switch shall be able to function and maintain<br>integrity following a clinically relevant number of on/off cycles and<br>following continuous aspiration for a clinically relevant time. | | Freedom from<br>Leakage | The catheter and accessories shall remain free of leaks or damage<br>following exposure to clinically relevant positive pressure and<br>vacuum. | | Simulated Use | The catheter shall be able to be delivered through a simulated use<br>model and successfully used to infuse fluids, aspirate and be<br>retracted per the IFU without incurring any damage. | | Infusion Flow Rate<br>& Height | The user shall be able to inject physician specified fluids at a rate of<br>1.5cc/sec without generating excessive pressure/fluid velocity at the<br>distal tip of the catheter. | | Aspiration Flow<br>Rate | When connected to an aspiration pump, the user should be able to<br>aspirate at an average flow rate >= 1.5cc/sec. | | Flexibility and Kink<br>Resistance | The catheter shall not kink when passed through clinically relevant<br>bend radii. | | Strain Relief<br>Flexibility | The strain relief shall protect the catheter shaft when deflected with<br>a force of 1 lbf. | | Test Performed | Specification Summary | | Guide Sheath<br>Compatibility /<br>Coating Durability | After a clinically relevant number of insertions and removals<br>through a 7 French guide sheath, the catheter shall be intact with no<br>deformation or kinks and no detectable change in resistance. | | Guidewire<br>Compatibility | The user shall be able to advance the catheter with little resistance<br>over a 0.014" diameter wire while infusing at a rate of 1.5cc/sec. | | Radiopacity | The radiopaque markers on the distal and infusion site of catheter<br>shall be visible under typical fluoroscopic methods. | | Corrosion Test | The metallic components of the catheter shall show no signs of<br>corrosion following worst case exposure conditions to a heated<br>saline solution. | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image is a logo for V2K Medical. The logo features a large letter "V" split into two colors, green on the left and blue on the right. Below the "V" is the text "V2K Medical" in a smaller font. The entire design is encircled by a green ring, creating a circular emblem. # H. Biocompatibility Biocompatibility testing was performed on all direct and indirect patient contacting components of the V2K Rinspiration System. This testing was conducted to ensure that the components and raw materials used in the proposed V2K Rinspiration System, as well as the manufacturing processes and sterilization processes result in a biocompatible product. All biocompatibility tests were conducted pursuant to 21 CFR Part 58, Good Laboratory Practices, ISO 10993-1, and per the recommendations in FDA guidance document titled. Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". A summary of the selected biocompatibility tests performed for the direct and indirect patient contacting components of the V2K Rinspiration System is presented for reference on the following pages. All test results passed, indicating that the V2K Rinspiration is biocompatible for its intended use. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image is a logo for V2K Medical. The logo features a large letter "V" with the left half in dark green and the right half in blue. Below the "V" is the text "V2K Medical" in a smaller font. The entire logo is encircled by a green ring, creating a circular emblem. | Test | Test Method | Extraction<br>Methods/Conditions | Acceptance Criteria | Results | |--------------------------------------------------------------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| | Cytotoxicity: ISO<br>MEM Elution<br>Method | ISO 10993-5 | Test device extracted in<br>MEM with 5% serum at<br>37+/-1°C for 24-25 hours. | Sample extracts must<br>yield cell lysis grade 2<br>or lower. | Pass,<br>Non-cytotoxic | | Sensitization:<br>Magnusson-<br>Kligman Method | ISO 10993-10 | Guinea pigs exposed to test<br>device extracts. Challenged<br>sites observed for skin<br>sensitization 24 +/- 2 and<br>48 +/- 2 hours after<br>removal of extracts.<br><br>Extracts were prepared at<br>37+/-1°C for 72 +/- 2<br>hours. | Test Group shall yield<br>Grade < 1 score on<br>Magnusson and<br>Kligman scale<br>(provided control<br>Grade < 1). | Pass,<br>Non-Sensitizing | | Irritation: ISO<br>Intracutaneous<br>Toxicity | ISO 10993-10 | Rabbits are injected with<br>extracts of test device.<br>Injection sites<br>examined/scored at 24 +/-<br>2, 48 +/- 2 and 72 +/- 2<br>hours after injections.<br><br>Extracts were prepared at<br>37+/-1°C for 72 +/- 2<br>hours. | The difference in the<br>mean test article and<br>mean control score<br>must be grade 1.0 or<br>lower. | Pass,<br>Non-Irritant | | Systemic Toxicity | | | | | | Systemic Toxicity:<br>ISO Systemic<br>Injection | ISO 10993-11 | Mice are injected with<br>extracts of test device.<br>Animals are observed for<br>signs of toxicity<br>immediately after<br>injection, 4+/- 0.75, 24 +/-<br>2, 48 +/- 2 and 72 +/- 2<br>hours after injections.<br>Body weights are<br>measured prior to injection<br>and on all 3 days.<br><br>Extracts were prepared at<br>37+/-1°C for 72 +/- 2<br>hours. | Sample extracts must<br>not cause the<br>following:<br>• > 10% weight loss in<br>3 or more test animals<br>• Mortality of 2 or<br>more test animals<br>• Abnormal behavior<br>in 2 or more test<br>animals | Pass,<br>Non-Toxic | | Test | Test Method | Extraction<br>Methods/Conditions | Acceptance Criteria | Results | | Systemic Toxicity:<br>ISO Material<br>Mediated Pyrogen | ISO 10993-11 | Test device extracted in<br>0.9% saline solution and<br>injected in the marginal ear<br>vein. Rectal temperatures<br>recorded prior to injection<br>and every 30 min until 1-3<br>hours post injection.<br><br>Extracts were prepared at<br>37+/- 1°C for 72 +/- 2<br>hours. | Sample Extracts must<br>not cause a total rise in<br>body temperature of<br>$\geq$ 0.5°C. | <b>Pass</b> ,<br><b>Non-pyrogenic</b> | | Hemocompatibility:<br>ASTM Hemolysis,<br>indirect (human<br>blood) | ISO 10993-4<br>ASTM F756 | Extract exposed to blood<br>cell suspension.<br>Hemoglobin releases is<br>measured. | Sample extracts must<br>be nonhemolytic ( $\leq$<br>2% hemolytic index). | <b>Pass</b> ,<br><b>Non-hemolytic</b> | | Hemocompatibility:<br>ASTM Hemolysis,<br>direct contact<br>(human blood) | ISO 10993-4<br>ASTM F756 | Test device exposed to<br>blood cell suspension.<br>Hemoglobin releases is<br>measured. | Sample extracts must<br>be nonhemolytic ( $\leq$<br>2% hemolytic index). | <b>Pass</b> ,<br><b>Non-hemolytic</b> | | Hemocompatibility:<br>Partial<br>Thromboplastin<br>Time (PTT) | ISO 10993-4<br>ASTM F2382 | Test device and controls<br>are directly exposed to<br>human blood plasma with<br>anticoagulants at 37+/- 2°C<br>for 60 +/- 5 minutes.<br><br>Following exposure, the<br>anticoagulant is neutralized<br>and the time for the serum<br>to start clotting is recorded. | The clotting time for<br>the test device should<br>be similar to or less<br>than either a predicate<br>device, a negative<br>control, or the plasma<br>control. | <b>Pass. Minimal<br/>Activator</b> | | Hemocompatibility:<br>Complement<br>Activation Test | ISO 10993-4<br>ASTM F756 | Preformed in vitro using a<br>prepared Enzyme<br>Immunoassay (EIA) kit.<br>This kit will detect the<br>presence of specific<br>complement enzymes.<br><br>Test device extracted in<br>Normal Human Serum<br>(NHS) at 37+/- 2°C for 0.5,<br>1, and 1.5 hours. | The concentrations of<br>C3a and SC5b-9 in the<br>test samples are<br>statistically similar to<br>the predicate<br>(Exposure Control &<br>Ref Material) control<br>and statistically lower<br>than the positive<br>control for all<br>exposure times. | <b>No Significant<br/>Differences</b> | | In Vivo<br>Thrombogenicity<br>(Dog) | ISO 10993-4<br>ASTM F756 | Performed in duplicate.<br>Placed in one side of the<br>jugular vein. Vasculature<br>closed up and remains in<br>place for 4 hours. | The device must be<br>nonthrombogenic after<br>4 hours in vivo when<br>compared to a control<br>device (Merit Fountain<br>Infusion Catheter). | <b>Pass</b> .<br><b>Minimal<br/>Thrombus<br/>Formation</b> | {10}------------------------------------------------ Image /page/10/Picture/0 description: The image is a logo for V2K Medical. The logo consists of a large letter "V" that is split into two colors, green on the left and blue on the right. Below the "V" is the text "V2K Medical" in a smaller font. The entire logo is enclosed in a green circle. V2K Medical Inc. 1221 Innsbruck Drive Sunnyvale, CA 94089 Sunnyvale, CA 94089 Phone: 408-940-5587 {11}------------------------------------------------ Image /page/11/Picture/0 description: The image is a logo for V2K Medical. The logo features a green circle with the letters 'V' in green and blue inside the circle. Below the letters, the words 'V2K Medical' are written in black. The logo is simple and professional, and it is likely used to represent the company's brand. # I. Sterilization The V2K Rinspiration Catheter and Rinspiration Filter Line are both sterilized using an electron beam process with a sterility assurance level of 1x106. The sterilization process was validated per the VDmax 25 method described in ANSI/AAMI/ISO 11137-1, "Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices". A summary of the completed testing is presented below. | Test Group | Catheter<br>Results | Filter Line<br>Results | |-------------------------------------------|---------------------|------------------------| | Sterilization Dose Map | Pass | Pass | | Bioburden Recovery Correction Factor | Pass | Pass | | Device Bioburden Assessment | Pass | Pass | | Sterile Verification Dose | Pass | Pass | | Bacteriostasis/ Fungistasis (B/F) Testing | Pass | Pass | ## J. Shelf Life and Packaging Accelerated aging testing based on ASTM F1980 was conducted to verify device, accessory, and packaging performance. A real time aging equivalent of 25 months was used for this testing and will allow for labeling of product with a 2-year shelf life. Device and accessory performance was verified by repeating the functional tests previously discussed in Section G. Packaging and sterile barrier integrity through transportation and aging testing was verified per the testing summarized below. | Test | Test Method | Time<br>Points | Catheter<br>Results | Filter Line<br>Results | |------------------------------------------------|-----------------------------|----------------|---------------------|------------------------| | Packaging Visual<br>Inspection | ASTM F1886 /<br>V2K TM00294 | 25 months | Pass | Pass | | Pouch Integrity Test -<br>Gross Leak Detection | ASTM F2096 | 25 months | Pass | Pass | | Pouch Seal Strength<br>– Peel Strength | ASTM F88 | 25 months | Pass | Pass | | Label Integrity | V2K TM00294 | 25 months | Pass | Pass | {12}------------------------------------------------ Image /page/12/Picture/0 description: The image is a logo for V2K Medical. The logo features a large letter V, with the left half in dark green and the right half in blue. Below the V, the text "V2K Medical" is written in a smaller font. The entire logo is encircled by a green ring. # K. Conclusions Where differences were identified between the subject V2K Rinspiration Catheter and the predicate Kerberos Rinspiration Catheter, an assessment was conducted to determine if the difference would result in new safety or efficacy concerns regarding the use of the device. As appropriate, bench and lab testing was conducted to support this assessment. Based on the results of the conducted assessments and testing, it is concluded that V2K Rinspiration System is substantially equivalent to the predicate Kerberos Rinspiration System and that there are no new safety or efficacy concerns associated with the identified differences. This conclusion is based on both devices sharing the same intended use, basic technology characteristics, and performance characteristics, as demonstrated through welldesigned bench and lab testing.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...