← Product Code [MUU](/productcode/MUU) · K172714

# LipiVage (K172714)

_Genesis Biosystems, Inc. · MUU · Dec 8, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172714

## Device Facts

- **Applicant:** Genesis Biosystems, Inc.
- **Product Code:** [MUU](/productcode/MUU.md)
- **Decision Date:** Dec 8, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted. the harvested fat is only to be used without any additional manipulation. LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.

## Device Story

LipiVage® is a 60cc hollow barrel harvesting syringe with an integrated in-line polyester filter; designed for aspiration, harvesting, filtering, and transferring autologous adipose tissue. Device connects to a user-supplied vacuum/aspirator source via universal tubing and to a cannula via a male nozzle. During lipoplasty, suction draws adipose tissue into the syringe; the in-line filter concentrates and filters the tissue, eliminating the need for centrifugation or decanting. The device is single-use, sterile, and intended for professional use by surgeons in clinical settings. By avoiding additional manipulation of harvested fat, the device maintains tissue viability for immediate re-implantation into the same patient. The system facilitates efficient fat grafting by streamlining the harvesting and processing workflow.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical performance testing, including biocompatibility (ISO 10993-1), sterilization (ISO 11137), packaging/shelf-life (ASTM F1980, F1140, F1929), and mechanical performance (ISO 80369). Adipose viability testing confirmed tissue viability was maintained.

## Technological Characteristics

60cc hollow barrel syringe; materials: polycarbonate barrel, polyolefin plunger tip, silicone O-rings. Filter: Estene® polyester thread, 180 μm mesh. Energy: external vacuum source. Connectivity: universal tubing connection. Sterilization: gamma radiation. Single-use, sterile.

## Regulatory Identification

A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.

## Special Controls

*Classification.* Class II (special controls). Consensus standards and labeling restrictions.

## Predicate Devices

- GID 700™ ([K120902](/device/K120902.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other and are connected by a series of lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 8, 2017

Genesis Biosystems, Incorporated % Mr. Stuart R. Goldman Senior Consultant, Regulatory Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin. Texas 78746

Re: K172714

Trade/Device Name: LipiVage® Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: August 24, 2017 Received: September 8, 2017

Dear Mr. Goldman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172714

Device Name LipiVage®

### Indications for Use (Describe)

LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted. the harvested fat is only to be used without any additional manipulation.

LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the word "EMERGO" in black font. To the left of the word is a red circular logo with three white curved lines inside. The logo is simple and modern. The text is clean and easy to read.

## 510(k) Summary

LipiVage®

к 172714

#### Submission Sponsor 5.1

Genesis Biosystems, Inc.

1500 Eagle Court

Lewisville, TX 75057

Phone Number: 972.315.7888

Contact: Jim Lafferty

Email: JLafferty@genesisbiosystems.com>

#### 5.2 Submission Correspondent

Emergo Global Consulting, LLC

2500 Bee Cave Road, Bldg. 1, Suite 300

Austin, TX 78746

Office Phone: 512.327.9997

Contact: Stuart R. Goldman, Senior Consultant, RA/QA

Email: project.management@emergogroup.com

#### 5.3 Date Prepared

August 24, 2017

#### Device Identification 5.4

Trade/Proprietary Name: LipiVage® Common/Usual Name: Harvesting syringe Classification Name: Suction lipoplasty system Regulation Number: MUU Product Code: 878.5040 Device Class: Class II Classification Panel: General & Plastic Surgery

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for EMERGO. The logo consists of a red circular symbol on the left and the word "EMERGO" in black letters on the right. The circular symbol has three curved white lines inside it, creating a stylized "E" shape.

#### Legally Marketed Predicate Device 5.5

GID 700™ (K120902)

#### 5.6 Device Description

LipiVage® is a 60cc hollow barrel harvesting syringe with plunger and an in-line polyester filter that is intended to receive, concentrate and filter autologous adipose tissue from lipoplasty procedures, for reimplantation of the extracted tissue back into the same patient at other anatomical locations in their body using a piston syringe and injection cannula. At one end of the LipiVage® harvesting syringe there is a male connector (nozzle) for fitting the female connector (hub) of a cannula for the harvesting of fat cells, while at the other end of the device there is a universal tubing connection to attach to a tubing set (not supplied by Genesis) that is connected to a controlled vacuum source, such as an aspirator or wall suction in the user facility. The LipiVage® harvesting syringe incorporates an in-line filter so there is no need to centrifuge, decant or expose fat cells to unnecessary handling. The device is for professional use only, made from medical grade polymers, single use, and provided sterile to the user.

#### 5.7 Indication for Use

LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.

LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, and laparoscopic surgery.

#### 5.8 Substantial Equivalence Discussion

Genesis Biosystems has chosen the GID 700™ harvesting canister system as the predicate device for its LipiVage® harvesting syringe. The GID 700™ was cleared by the FDA under K120902. The following table compares the subject device to the predicate device with respect to their indications for use, technology and performance testing, thus demonstrating the basis for determination of substantial equivalence between these devices.

| Regulatory Information             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                        |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                        | LipiVage®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | GID 700™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similarities/Differences                                                                                                                                                                                                                                               |
| Manufacturer                       | Genesis Biosystems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Gid Group                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | NA                                                                                                                                                                                                                                                                     |
| 510(k)                             | Pending                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K120902                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | NA                                                                                                                                                                                                                                                                     |
| Product Code                       | MUU                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | MUU                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                   |
| Regulation                         | §878.5040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | §872.5040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                   |
| Class                              | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                   |
| Indications for Use                | LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.<br><br>The LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery. | The GID 700™ is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.<br><br>The GID 700™ is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery. | Same                                                                                                                                                                                                                                                                   |
| Prescription use                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                   |
| Fundamental Scientific Technology  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                        |
| System Concept                     | Closed loop                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Closed loop                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                   |
| Source of Energy                   | User supplied vacuum.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | User supplied vacuum.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                   |
| Device Size<br>(fill volume)       | Syringe: 60 cc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Canister: 700 ml                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | The differences in the size of the subject and predicate devices are a function of their overall differences in design geometries (syringe vs. canister), and not considered to have any significant difference in the overall quality of the                          |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | harvested fat.                                                                                                                                                                                                                                                         |
| Device Materials                   | Syringe: Molded, polycarbonate<br>- Plunger Tip: Polyolefin<br>- O-Rings: Silicone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Canister: Polymer (unknown)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | The subject and predicate<br>device are made of similar<br>medical grade polymer<br>materials.                                                                                                                                                                         |
| Device Filter                      | Image: Estene Polyester thread                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Image: Similar woven fabric material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The subject device uses<br>Estene® Polyester thread<br>(natural color) for its woven<br>filter fabric with a calculated<br>mesh opening of 180 μm,<br>while the predicate device<br>uses a similar woven fabric<br>material with a reported<br>mesh opening of 200 μm. |
|                                    | Estene® Polyester thread (natural<br>color) woven filter fabric with a<br>calculated mesh opening of<br>180 μm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar woven fabric material<br>with a reported mesh opening<br>of 200 µm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                        |
| Provided Sterile                   | Yes (gamma radiation)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes (gamma radiation)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                   |
| Single Patient<br>Use              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                   |
| Implantable                        | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                   |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Relevant Testing Performed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                        |
| Biocompatibility<br>Testing        | Per ISO 10993-1:<br>- Part 5 Cytotoxicity<br>- Part 10 Intracutaneous<br>Reactivity<br>- Part 10 Irritation-Sensitization<br>- Part 11 Systemic Toxicity<br>- Part 11 Pyrogenicity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                   |
| Device<br>Sterilization            | ISO 11137-1,-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | ISO 11137                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                   |
| Device Packaging<br>and Shelf-Life | ASTM F1980<br>ASTM F1140-07<br>ASTM F1929-15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Not known.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Device packaging and shelf<br>have been validated for the<br>subject device in accordance<br>with the referenced<br>ASTM standards.                                                                                                                                    |
| Device<br>Performance<br>Testing   | ISO 80369-1<br>ISO 80369-7<br>ISO 80369-20                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Implosion<br>Tubing Connection<br>Tensile Strength<br>Mechanical Plugging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                        |
| Animal Testing                     | NA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | NA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                   |
| Clinical Testing                   | NA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | NA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                   |
| Risk Analysis                      | ISO 14971                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Not Known.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | An appropriate Risk Analysis                                                                                                                                                                                                                                           |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | has been performed on the                                                                                                                                                                                                                                              |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | subject device.                                                                                                                                                                                                                                                        |

## Table 5-1 – Device Comparison: LipiVage® vs. GID 700™

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for EMERGO. The logo consists of a red circular icon with three curved lines inside, followed by the word "EMERGO" in black, sans-serif font. The logo is simple and modern, with a focus on the company name.

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the Emergo logo. The logo consists of a red circular icon with three curved white lines inside, followed by the word "EMERGO" in black, sans-serif font. The logo is simple and modern, with a focus on the company name.

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the logo for Emergo. The logo consists of a red circular symbol with white curved lines inside, followed by the word "EMERGO" in black, sans-serif font. The logo is simple and modern, with a focus on the company name.

#### Non-Clinical Performance Data 5.9

Genesis Biosystems has submitted LipiVage® and its packaging system for the appropriate biocompatibility testing per ISO 10993-1 and its applicable parts, sterilization per ISO 11137, accelerated aging per ASTM F1980, and performance testing per ISO 80369-1/-7. Adipose viability testing demonstrated that fat tissue viability was maintained. The subject device meets all internal Genesis Biosystems specification requirements to support substantial equivalence to the predicate device.

## 5.10 Clinical Performance Data

The performance and use of Suction Lipoplasty Systems classified under product code MUU and regulated within CFR 878.5040 has been well established in the clinical environment since the first such devices were cleared by the FDA. Therefore, there was no clinical testing required to support LipiVage®, as the indications for use is equivalent to the predicate device, which was not subjected to any clinical testing. The substantial equivalence of LipiVage® to the GID 700™ is supported by the non-clinical testing that was performed, along with other supporting documentation, and presented in this submission.

## 5.11 Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared device, or has the same intended use and different technological characteristics, but it can be demonstrated that the new device is substantially equivalent to the predicate device, and that the new device does not raise any questions regarding its safety and effectiveness when compared to the predicate device.

LipiVage® has the same intended use and indications for use as the predicate device. The conclusions drawn from the data included in this submission, demonstrate that LipiVage® is as safe, as effective, and are substantially equivalent to the predicate device in intended use, indications for use, technological characteristics, mode of operation, mechanical performance, materials, biocompatibility and is therefore substantially equivalent to the GID 700™.

---

**Source:** [https://fda-staging.innolitics.com/device/K172714](https://fda-staging.innolitics.com/device/K172714)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).