K172700 · Ge Hualun Medical Systems Co. , Ltd. · OWB · Nov 9, 2017 · Radiology
Device Facts
Record ID
K172700
Device Name
OEC One
Applicant
Ge Hualun Medical Systems Co. , Ltd.
Product Code
OWB · Radiology
Decision Date
Nov 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The OEC One™ mobile C-arm system is intended to provide fluoroscopic and digital spot/film images of the patient anatomy, interventional tools/devices, and/or contrast agents during diagnostic, interventional, and surgical procedures.
Device Story
Mobile C-arm X-ray system; provides fluoroscopic and digital spot/film images for visualization of anatomy, interventional tools, and contrast agents. System comprises mobile stand, articulating arm, widescreen monitor, and TechView tablet. Used in OR, clinic, or emergency settings by physicians/technicians. Integrates mainframe and workstation into compact footprint for improved ergonomics and workflow. Features include Adaptive Dynamic Range Optimization (ARDO) and motion artifact reduction. Output displayed on monitor/tablet to assist in localization of clinical regions of interest and guidance during procedures. Benefits include improved maneuverability and space efficiency in clinical environments.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical verification, validation, and engineering bench testing using anthropomorphic phantoms to assess image quality and dose performance per FDA guidance.
Technological Characteristics
Mobile C-arm X-ray system. Components: mobile stand, articulating arm, widescreen monitor, TechView tablet. Energy: X-ray. Connectivity: Integrated workstation. Standards: IEC 60601-1 Ed. 3, IEC 60601-2-54, IEC 60601-2-43, NEMA XR-27. Software: Moderate level of concern.
Indications for Use
Indicated for adult and pediatric patient populations during diagnostic, interventional, and surgical procedures, including orthopedic, gastrointestinal, endoscopic, urologic, neurologic, critical care, and emergency procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a caduceus, which is a traditional symbol of medicine. The text on the right is in a blue sans-serif font. The word "FDA" is in a larger font than the rest of the text.
November 9, 2017
GE Hualun Medical Systems Co., Ltd. Lifeng Wang Regulatory Affairs Manager No 1 Yong Chang North Road Beijing Economic Technological Development Zone Beijing, 100176, China
Re: K172700
Trade/Device Name: OEC One™ Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB. OXO. JAA Dated: September 7, 2017 Received: September 7, 2017
Dear Lifeng Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Oolo
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K172700
Device Name OEC One™
#### Indications for Use (Describe)
The OEC One mobile C-arm system is designed to provide fluoroscopic and digital spot/film images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, critical care, and emergency procedures.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/6 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling, wave-like shapes, also in blue, which give the logo a dynamic and fluid appearance.
## GE Healthcare
510(k) Premarket Notification Submission- OEC One™
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the requirements of 21 CFR 807.92 the following summary of information is provided:
| Date: | Sep 7, 2017 |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE HUALUN MEDICAL SYSTEMS CO. Ltd.<br>No.1 Yong Chang North Road, Beijing Economic<br>Technological Development Zone<br>Beijing 100176, China |
| Manufacturer/<br>Manufacturing Location | GE HUALUN MEDICAL SYSTEMS CO. Ltd.<br>No.1 Yong Chang North Road, Beijing Economic<br>Technological Development Zone<br>Beijing 100176, China |
| Primary Contact Person: | Lifeng Wang<br>Regulatory Affairs Manager<br>GE Healthcare<br>+86 10 57083145<br>Lifeng.Wang@ge.com |
| Secondary Contact Person: | Helen Peng<br>Regulatory Affairs Director<br>GE Healthcare<br>+1 262 424-8222<br>Hong.Peng@ge.com |
| Device Trade Name: | OEC One™ |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Device Class | Class II |
| Classification regulation: | 21CFR 892.1650 |
| Primary Product Code: | OXO |
| Secondary Product Code: | JAA, OWB |
| Predicate Device(s): | K123603 OEC Brivo, 892.1650, OXO |
| Device Description: | The OEC One™ is a mobile C-arm x-ray system to provide<br>fluoroscopic images of the patient during diagnostic,<br>interventional, and surgical procedures such as orthopedic,<br>gastrointestinal, endoscopic, urologic, neurologic, critical care,<br>and emergency procedures. These images help the physician<br>visualize the patient's anatomy and localize clinical regions of<br>interest. The system consists of a mobile stand with an<br>articulating arm attached to it to support an image display monitor<br>(widescreen monitor) and a TechView tablet, and a "C" shaped<br>Page 1 of 3 |
| Intended Use: | The OEC One™ mobile C-arm system is intended to provide fluoroscopic and digital spot/film images of the patient anatomy, interventional tools/devices, and/or contrast agents during diagnostic, interventional, and surgical procedures. |
| Indications for Use | The OEC One™ mobile C-arm system is designed to provide fluoroscopic and digital spot/film images of adult and pediatric patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, neurologic, critical care, and emergency procedures. |
| Technology: | The modified OEC One™ employs the same fundamental scientific technology as that of the unmodified predicate device(K123603). The primary change for the subject device was to integrate the separate mainframe and workstation of the predicate device into a compact configuration to save the space of the operation room with smaller foot print and to offer better workflow, ergonomics and usability.<br><br>Additionally, a separate articulating arm attached to the mobile stand was added to support the single widescreen image display monitor and a TechView tablet and to offer more degrees of freedom for easier positioning and maneuverability.<br><br>The software was updated primarily to support changing to a single image display monitor, adding a new TechView tablet, consolidation of user interface, and modifying existing features for Adaptive Dynamic Range Optimization(ARDO) and motion artifact reduction. Additionally, the software was also modified to add an Alternating Current (AC) recovery feature and to comply to NEMA XR-27(User Quality Mode) requirement. |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Testing:<br>Verification and validation including hazard mitigation has been<br>executed with results demonstrating the OEC One™ system met<br>design input and user needs. |
| | The system has been tested by an NRTL and certified compliant<br>with the IEC 60601-1 Ed. 3 series, including IEC60601-2-54 and<br>IEC 60601-2-43. All applicable 21CFR Subchapter J<br>performance standards are met. |
| | The OEC One™ system was developed under the GE<br>Healthcare's Quality Management System, including design<br>controls, risk management and software development life cycle<br>processes. The following quality assurance measures were<br>applied to the development of the system:<br>Risk AnalysisRequired ReviewsDesign ReviewsUnit Testing (Sub System verification)Integration testing (System verification)Performance testing (Verification)Safety testing (Verification)Simulated use testing (Validation) |
| | Additional engineering bench testing on image performance using<br>anthropomorphic phantoms was also performed. All the image<br>quality/performance testing identified for fluoroscopy found in<br>FDA's "Information for Industry: X-ray Imaging Devices-<br>Laboratory Image Quality and Dose Assessment, Tests and<br>Standards" was performed with acceptable results. |
| | The substantial equivalence was also based on software<br>documentation for a "Moderate" level of concern device. |
| | Clinical Testing:<br>Because OEC One's modification based on the predicate device<br>does not change the system's intended use and represent<br>equivalent technological characteristics, clinical studies are not<br>required to support substantial equivalence. |
| Conclusion: | The differences discussed in this section do not introduce any<br>adverse effects nor raise new questions of safety and<br>effectiveness. Based on the successful verification and validation<br>testing, additional engineering bench testing, conformance to<br>standards, and development under GE Healthcare's Quality<br>Management System, we believe that the OEC One is of<br>comparable type and substantially equivalent to the predicate<br>device OEC Brivo Series (K123603). |
{4}------------------------------------------------
# GE Healthcare
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white, stylized script in the center. The circle is surrounded by a swirling, decorative pattern, also in white, that gives the impression of movement or energy.
## 510(k) Premarket Notification Submission- OEC One™
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are decorative swirls around the letters. The logo is simple and recognizable.
# GE Healthcare
510(k) Premarket Notification Submission- OEC One™
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.