The G-Beam™ Fusion Beaming System is intended to be inserted in the foot and ankle for alignment, stabilization and fixation of various fractures and osteotomies, fusions and reconstructions. It is indicated for fracture and osteotomy fixation, reconstruction procedures, non-unions and fusions of the foot and ankle including metatarsals, cuneiforms, cuboid, navicular, calcaneus and talus. Specific example: medial and lateral column fusions resulting from neuropathic osteoarthropathy (Charcot osteoarthropathy).
Device Story
G-Beam Fusion Beaming System; internal bone fixation system for foot and ankle. Device provides alignment, stabilization, and fixation for fractures, osteotomies, fusions, and reconstructions. Used by orthopedic surgeons in clinical settings. System components include metallic bone fixation fasteners (beams/bolts). Operates via mechanical fixation of bone segments to facilitate healing/fusion. Output is physical stabilization of bone structures. Benefits include restoration of alignment and structural integrity in complex foot/ankle pathologies like Charcot osteoarthropathy.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including mechanical performance testing per ASTM F1264-16 and ASTM F-543-13, and pyrogenicity testing per USP 38 and ANSI/AAMI ST72:2011.
Indicated for patients requiring fracture/osteotomy fixation, reconstruction, non-union repair, or fusion of foot/ankle bones (metatarsals, cuneiforms, cuboid, navicular, calcaneus, talus), including medial/lateral column fusions for Charcot osteoarthropathy.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 7, 2018
Orthofix Srl Gianluca Ricadona Sr. Quality & Regulatory Affairs Manager Via delle Nazioni, 9 37012 Bussolengo (VR) Italy
Re: K172698
Trade/Device Name: G-Beam Fusion Beaming System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: January 5, 2018 Received: January 8, 2018
Dear Gianluca Ricadona:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K172698
Device Name G-Beam™ Fusion Beaming System
#### Indications for Use (Describe)
The G-Beam™ Fusion Beaming System is intended to be inserted in the foot and ankle for alignment,
stabilization and fixation of various fractures and osteotomies, fusions and reconstructions.
It is indicated for fracture and osteotomy fixation, reconstruction procedures, non-unions and fusions of the foot and ankle including metatarsals, cuneiforms, cuboid, navicular, calcaneus and talus.
Specific example: medial and lateral column fusions resulting from neuropathic osteoarthropathy (Charcot osteoarthropathy).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Orthofix, a medical device company. Below the logo is the text "510(k) Summary" and the text "(as required by 21 CFR 807.92)". The logo consists of the word "ORTHOFIX" in a bold, sans-serif font, with a registered trademark symbol next to it. Above the word is a blue, abstract shape.
| Submitter Name | Orthofix Srl |
|----------------|------------------------------------------------------|
| Address | Via delle Nazioni, 9 37012 Bussolengo (VR) - Italy |
| Telephone | + 39 045 6719.000 |
| Fax | + 39 045 6719.380 |
| Contact Person | Gianluca Ricadona<br>Sr. Quality & Regulatory Affairs Manager |
|----------------|---------------------------------------------------------------|
| Address | Via delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy |
| Telephone | + 39 045 6719 000 |
| Fax | + 39 045 6719 380 |
| email | GianlucaRicadona@orthofix.it |
| Date Prepared | February 5, 2018 |
| Trade Name | G-Beam™ Fusion Beaming System |
|-------------------|----------------------------------------------------|
| Common Name | Screw, fixation, bone |
| Panel Code | Orthopedic |
| Classification | Smooth or threaded metallic bone fixation fastener |
| Name | |
| Class | Class II |
| Regulation Number | 21 CFR 888.3040 |
| Product Code | HWC |
| Predicate Device Name | 510(k) Number | Manufacturer |
|---------------------------------------------------------------------------------------------|---------------|---------------------------------|
| SALVATION™ Beams and Bolts<br>System | K140741 | Wright Medical Technology, Inc. |
| Axis Charcot Fixation System | K171018 | Extremity Medical LLC |
| Orthofix External Fixation Screw<br>(Pin) with hydroxyapatite coating<br>(Reference device) | K974186 | Orthofix SRL |
| Orthofix Titanium Nailing Systems<br>(Reference device) | K053261 | Orthofix SRL |
{4}------------------------------------------------
| procedures, non-unions and fusions of bones of the foot and ankle |
|---------------------------------------------------------------------|
| including metatarsals, cuneiforms, cuboid, navicular, calcaneus and |
| talus. |
| Specific example: medial and lateral column fusions resulting from |
| neuropathic osteoarthropathy (Charcot osteoarthropathy). |
{5}------------------------------------------------
| Technological<br>Characteristics and<br>Substantial<br>Equivalence | Documentation was provided to demonstrate that the G-Beam™<br>Fusion Beaming System is substantially equivalent to the legally<br>marketed predicates.<br>Components and instrumentation included in the G-Beam™ Fusion<br>Beaming System and the predicate device are both internal fracture<br>fixation systems, as defined in 21 CFR 888.3040.<br>The G-Beam™ Fusion Beaming System is substantially equivalent<br>to the predicate devices in: intended use, site of application, patient<br>population, conditions of use, mechanical performances, operating<br>principles.<br>Mechanical testing show how the mechanical properties of the<br>subject device are equivalent or better than the predicate device. |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data | The potential hazards have been evaluated and controlled through<br>a Risk Management Plan.<br>All testing met or exceeded the requirements as established by the<br>test protocols and applicable standards. A review of the mechanical<br>data indicates that the components of the Subject device are<br>capable of withstanding expected loads without failure. The Subject<br>device was therefore found to be substantially equivalent to the<br>predicate device. Clinical data was not needed to support the<br>safety and effectiveness of the Subject Device.<br>Mechanical testing was performed according to the following<br>standards:<br>• ASTM F1264-16 "Standard Specification and Test Methods<br>for Intramedullary Fixation Devices".<br>• ASTM F-543-13 "Standard Specification and Test Methods<br>for metallic bone screws" |
| Pyrogenicity<br>data | In order to establish the Subject device non-pyrogenicity, tests were<br>performed according to the following international standards:<br>• USP 38: 2014 < 85 > "Bacterial endotoxin test (LAL)".<br>• USP 38: 2014 < 161 > "Medical devices – bacterial endotoxin<br>and pyrogen tests".<br>• ANSI / AAMI ST72: 2011 "Bacterial endotoxins – Test<br>methodologies, routine monitoring and alternative batch testing". |
| Conclusion | Based upon similarities in: intended use, site of application, patient<br>population, conditions of use, mechanical performances, operating<br>principles, and according to the results of mechanical testing, G-Beam™ Fusion Beaming System has been shown to be<br>substantially equivalent to the legally marketed predicate device and<br>to be as safe and effective as the predicate for its intended use. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.