The ToFscan is a neuromuscular transmission monitor for monitoring the neuromuscular block of a patient in the operating theatre, recovery room or intensive care unit. The effect of neuromuscular blocking agents (NMBAs) is monitored by measuring the acceleration of the muscle movement (acceleromyography) or by visually observing muscle contractions consequent to electrical stimulation. The ToFscan has a three-dimensional acceleration sensor (accelerometer) to detect and quantify a patient's thumb movement (contracting adductor pollicis). The sensor is directly integrated into the finger's splint, making it possible to obtain its optimal and reproducible positioning.
Device Story
ToFscan is a neuromuscular transmission (NMT) monitor used in ORs, recovery rooms, and ICUs by clinicians (anesthesiologists, doctors, nurse anesthetists). It stimulates the ulnar nerve and measures resulting adductor pollicis thumb movement via a 3D accelerometer integrated into a finger splint. The device supports multiple stimulation modes (TOF, PTC, ATP, DBS, ST, TET) selected via a rotary knob. It processes electrical stimulation responses to quantify muscle contraction, providing objective and subjective monitoring of neuromuscular block and curare levels. Output is displayed on the monitor, allowing clinicians to assess NMBA effectiveness and guide clinical decision-making regarding patient paralysis levels. The system includes a monitor, hand sensor, and power supply.
Clinical Evidence
No clinical or animal testing was required. Substantial equivalence is supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-1), electrical safety, EMC, and software verification/validation.
Technological Characteristics
Constant current stimulator (0-60 mA, 200µsec pulse width, 300V max); 3D accelerometer sensor; surface-contacting (intact skin, <24h); rotary knob interface; IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 compliant; powered by AC supply.
Indications for Use
Indicated for neuromuscular transmission monitoring in adult and pediatric patients in operating theatres, recovery rooms, or intensive care units to assess the effect of neuromuscular blocking agents.
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized caduceus. To the right of it is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Mav 31, 2018
IDMED Frédéric Bernert President Hotel Technoptic 2 rue Marc Donadille 13013 Marseille, France
Re: K172690
Trade/Device Name: ToFscan Regulation Number: 21 CFR 868.2775 Regulation Name: Electrical Peripheral Nerve Stimulator Regulatory Class: Class II Product Code: KOI Dated: April 19, 2018 Received: April 19, 2018
Dear Frédéric Bernert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. | 0910-0120 |
|------------------------|--------------------------|
| Expiration Date: | January 31, 2017 |
| | See PRA Statement below. |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K172690 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | ToFscan |
| Indications for Use (Describe) | The ToFscan is a neuromuscular transmission monitor for monitoring the neuromuscular block of a patient in the<br>operating theatre, recovery room or intensive care unit.<br>The effect of neuromuscular blocking agents (NMBAs) is monitored by measuring the acceleration of the muscle<br>movement (acceleromyography) or by visually observing muscle contractions consequent to electrical stimulation. The<br>ToFscan has a three-dimensional acceleration sensor (accelerometer) to detect and quantify a patient's thumb movement<br>(contracting adductor pollicis). The sensor is directly integrated into the finger's splint, making it possible to obtain its<br>optimal and reproducible positioning. |
| Type of Use (Select one or both, as applicable) | <span> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </span> <span> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
프
1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for idmed. The word "idmed" is written in a sans-serif font, with "id" in green and "med" in blue. To the right of the word is a plus sign made up of four blue squares and one green square. Below the word "idmed" is the phrase "an eye on your patient" in a smaller sans-serif font.
5-1
# 5. 510 (k) Summary
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92.
# 5.1. Submitter
| Submitter: | IDMED |
|------------|------------------------|
| | Hotel Technoptic |
| | 2 rue Marc Donadille |
| | 13013 Marseille France |
| | +33 (0)491 118 784 |
Establishment Registration Number: Nº DUNS 260233256
| Contact person: | Frédéric BERNERT |
|------------------------|-----------------------------|
| | President |
| | E-mail : f.bernert@idmed.fr |
| Date Summary Prepared: | 2017/12/15 |
K172690 510(k) number:
# 5.2. Device
| Trade Name: | ToFscan |
|----------------------------|------------------------------------------|
| Common/Usual Name: | NeuroMuscular Transmission (NMT) monitor |
| Regulation Name: | Electrical peripheral nerve stimulator |
| Classification Regulation: | 21 CFR 868.2775 |
| Product Code: | KOI |
| Regulatory Class: | Class II |
# 5.3. Predicate device
| Predicate Device: | TOF-Watch, manufactured by Organon Teknika, K972698 |
|-----------------------------------|-----------------------------------------------------|
| 510(k) — ToFscan | CONFIDENTIAL |
| 001_510(k) Summary rev E 20180522 | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Idmed. The logo has the word "idmed" in blue and green letters. Below the word is the phrase "an eye on your patient".
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
# 5.4. Device description
The ToFscan is a neuromuscular transmission (NMT) monitoring the neuromuscular block of a patient in the operating theatre, recovery room or intensive care unit. It is used by health professionals (anaesthesiologists, doctors, or fully qualified nurse anaesthetists) for:
- 1. Objective neuromuscular transmission monitoring
- 2. Subjective neuromuscular transmission monitoring
The continuous monitoring of NMT blocking involves the stimulation of the ulnar nerve and the corresponding measurement of acceleration in the adductor muscle of the thumb (with a three-dimensional acceleration sensor). Based on the force of contraction resulting from the stimulation, it is possible to draw conclusions about the effectiveness of an injected neuromuscular blocking agents or the patient's curare level.
The apparatus and all of the settings associated with it are designed for use on adult and paediatric patients in hospital or health institutions so that the patient's curare level can be monitored.
The ToFscan system is composed of the monitor, the hand sensor, and the US AC power supply unit. The hand sensor itself consists of the thumb splint/3D accelerometer, the electrode clamps, and the cable which connects to the monitor.
The ToFscan can perform several different modes of electrical stimulation in accordance with usual clinical practice:
- -TOF (Train Of Four)
- PTC (Post Tetanic Count) -
- -ATP (Automated TOF PTC)
- DBS (Double Burst) (3,3) (3,2) -
- ST (Single Twitch) 0.1 Hz and 1 Hz -
- -TET (Tetanus) 50 Hz
The monitor displays the various stimulation settings, the time elansed between stimulations, and the results for the measurements made in response to TOF, PTC, and ATP stimulations. The responses to other types of stimulation are gaged visually by the health professional. The device and stimulation settings are chosen using a rotary knob.
The accessories that can be used with the device include:
- Extension Cable 1.9 m -
- Optic-serial (RS232) cable to transfer data -
- -Fixation clamp - regular size (10-40 mm / 0.4-1.6 inches)
- Fixation clamp large size (20-60 mm / 0.8-2.4 inches) -
- -Cable Support
#### CONFIDENTIAL
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for idmed. The logo has the letters "idmed" in green and blue. Below the letters is the phrase "an eye on your patient" in a smaller font.
# 5.5. Indications for use
The TOFscan is a neuromuscular transmission monitoring the neuromuscular block of a patient in the operating theatre, recovery room or intensive care unit.
The effect of neuromuscular blocking agents (NMBAs) is monitored by measuring the acceleration of the muscle movement (acceleromyography) or by visually observing muscle contractions consequent to electrical stimulation. The TOFscan has a three-dimensional acceleration sensor (accelerometer) to detect and quantify a patient's thumb movement (contracting adductor pollicis). The sensor is directly integrated into the finger's splint, making it possible to obtain its optimal and reproducible positioning.
The Indications for use statement for the ToFscan is not identical to the predicate device, however, the differences do not alter the intended diagnostic use of the device nor do they raise different questions of safety and effectiveness of the device relative to the predicate. Both devices rely on the use of acceleromyography to measure muscle response to electrical stimulation as an indication of a patient's curare level. The predicate device has an additional intended use, nerve location, which is not included in the ToFscan.
# 5.6. Comparison of technological characteristics with the predicate device
Measuring and/or observing the movement of the thumb in response to well-defined electrical stimulation patterns in order to monitor the neuromuscular blocking of a patient is the technological principle for both the subject and predicate devices.
At a high level, the subject and predicate devices are based on the following same technological elements as presented in the table below. The information regarding the nerve location function of the TOF-Watch have not been included as this function is not available on the ToFscan.
| Features | ToFscan (subject device) | TOF-watch (predicate device) |
|--------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|
| Stimulation Current<br>Output | Constant current, 0-60 mA,<br>monophasic 200µsec pulse width | Constant current, 0-60 mA,<br>monophasic 200µsec pulse width |
| Maximum stimulation<br>voltage | 300V (60 mA, 5kΩ) | 300V (60 mA, 5kΩ) |
| Stimulation | | |
| TOF | Yes | Yes |
| TOF auto | Yes | Yes |
| PTC | Yes | Yes |
| 0.1Hz and 1Hz ST | Yes | Yes |
| TET 50 Hz | Yes | Yes |
| TET 100 Hz | No | Yes |
| 3.3 and 3.2 DBS | Yes | Yes |
| ATP | Yes | No |
#### Calculations and measurements
510(k) — ToFscan
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Idmed. The word "idmed" is written in a sans-serif font, with "id" in green and "med" in blue. To the right of the word "med" is a blue plus sign made up of four squares. Below the word "idmed" is the tagline "an eye on your patient" in a smaller, sans-serif font.
IDMED Hotel Technoptic 2 rue Marc Donadille 13013 Marseille - France 2 +33 491 118 784 □ info@idmed.fr www.idmed.fr
| Features | ToFscan (subject device) | TOF-watch (predicate device) |
|-------------------------------------------|--------------------------------------------|--------------------------------------------|
| TOF calculation | Yes | Yes |
| PTC count | Yes | Yes |
| T4/T1 | Yes | Yes |
| T4/Tref | Yes | No |
| Displays amplitude of<br>responses / Bars | Yes | No |
| Type of sensor | Thumb - accelerometer | Thumb - accelerometer |
| General safety | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 | IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10 |
The following technological differences exist between the subject and predicate device:
- The ToFscan does not perform 100Hz TET stimulation -
- -The ToFscan allows the user to select a repeat frequency of TOF stimulation from 15s to 15min instead of 15s only
- -The ToFscan includes the ATP mode, a combination of the TOF and PTC stimulation modes
- -The ToFscan can measure a patient's reference response to TOF stimulation and calculate T4Tet
- -The ToFscan monitor displays the amplitude of responses/bars
The differences presented here are supported by following performance data. These differences do not raise different questions of safety and effectiveness.
## 5.7. Performance data
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility testing
The biocompatibility evaluation for the ToFscan device was conducted in accordance with the FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:
- -Cytotoxicity
- -Sensitization
- -Irritation
The thumb sensor is the only part of the ToFscan that is in contact with the patient. The ToFscan thumb sensor is a surface-contacting device in contact with intact skin for a duration of less than 24 hours.
#### Electrical safety and electromagnetic compatibility (EMC)
#### CONFIDENTIAL
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for idmed. The logo has the word "idmed" in blue and green letters. Below the word is the phrase "an eye on your patient" in a smaller font.
Electrical safety and EMC testing were conducted on the ToFscan device. The system complies with the IEC 60601-1, IEC 60601-1-6 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
#### Animal and Clinical Studies
No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate.
## 5.8. Conclusions
Based on its intended use, design principles, and technological characteristics, the ToFscan device was found to be as safe, as effective, and performs comparably to the predicate device.
The technological differences identified do not raise different questions of safety and effectiveness as the non-clinical data and the hardware and software verification demonstrate that the ToFscan device should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.