K172689 · Merz North America, Inc. · PKO · Nov 29, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172689
Device Name
DESCRIBE PFD Patch
Applicant
Merz North America, Inc.
Product Code
PKO · General, Plastic Surgery
Decision Date
Nov 29, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Indications for Use
The DESCRIBE PFD Patch is indicated for use as an accessory to laser tattoo removal procedures using 532, 694, 755, and 1064 nm standard Q-Switched (QS) lasers and 532, 755, 785, and 1064 nm standard picosecond lasers in Fitzpatrick Skin Type I-III patients.
Device Story
The DESCRIBE PFD Patch is a sterile, dual-layer silicone film containing perfluorodecalin (PFD) optical clearing agent; used as an accessory during laser tattoo removal. The patch is applied to the patient's skin; the PFD liquid fills the space between the skin and the patch, providing optical clearing. The device is transparent to laser light, allowing the laser beam to pass through to the tattoo pigment without attenuation. It is used in a clinical setting by a physician or trained professional. The patch facilitates multiple laser passes by reducing the scattering of laser light, potentially allowing for faster treatment and improved clinical outcomes. The device is disposable and intended for single use.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included UV/Vis/NIR spectroscopy to verify optical transparency and stability of the patch and PFD after exposure to Q-switched (532, 694, 755, 1064 nm) and picosecond (532, 755, 785, 1064 nm) lasers. GC/MS testing confirmed no chemical changes to the PFD after laser exposure. Three-year real-time aging studies were conducted per ISO 11607, including package burst and bubble emission leak testing.
Technological Characteristics
Dual-layer medical-grade silicone film (low-friction upper layer, tacky lower layer) and sterile, high-purity perfluorodecalin (PFD) optical clearing agent. Optically transparent to dermatological lasers. Shelf life validated to 3 years via real-time aging and ISO 11607 package integrity testing.
Indications for Use
Indicated for Fitzpatrick Skin Type I-III patients undergoing laser tattoo removal using specified Q-Switched (532, 694, 755, 1064 nm) or picosecond (532, 755, 785, 1064 nm) lasers.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized caduceus. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 29, 2017
Merz North America, Inc. Misty D'Ottavio Director, Regulatory Affairs 6501 Six Forks Road Raleigh, North Carolina 27615
Re: K172689
Trade/Device Name: DESCRIBE PFD Patch Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: PKO Dated: September 5, 2017 Received: September 6, 2017
Dear Misty D'Ottavio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K172689
Device Name
DESCRIBE PFD PATCH
Indications for Use (Describe)
The DESCRIBE PFD Patch is indicated for use as an accessory to laser tattoo removal procedures using 532, 694, 755, and 1064 nm standard Q-Switched (QS) lasers and 532, 755, 785, and 1064 nm standard picosecond lasers in Fitzpatrick Skin Type I-III patients.
Type of Use (Select one or both, as applicable)
X | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
# Merz North America's DESCRIBE PFD Patch
#### Submitted by:
Merz North America, Inc. 6501 Six Forks Road Raleigh, NC 27615 Phone: (919) 582-8000
# Contact Person:
Misty D'Ottavio Director of Regulatory Affairs 6501 Six Forks Road Raleigh, NC 27615 Phone: (919) 582-8107 Fax: (919) 882-8962
Date Prepared: September 5, 2017 Trade Name: DESCRIBE® PFD Patch Common or Usual Name: DESCRIBE Patch Classification Name: Accessory to laser surgical instrument for use in general and plastic surgery and in dermatology Regulation: 21 C.F.R. § 878.4810 Product Code: PKO Classification Panel: General and Plastic Surgery Predicate Device: DeScribe Transparent PFD Patch (K150212) Indications for Use:
The DESCRIBE PFD Patch is indicated for use as an accessory to laser tattoo removal procedures using 532, 694, 755, and 1064 nm standard Q-Switched (QS) lasers and 532, 755, 785, and 1064 nm standard picosecond lasers in Fitzpatrick Skin Type I-III patients.
# Technological Characteristics
The DESCRIBE® PFD Patch consists of two components: a dual-layer medicalgrade silicone transparent film, and an optical clearing agent (OCA). The dual-layer silicone film is composed of a thin low-friction silicone membrane and a tacky silicone
{4}------------------------------------------------
polymer layer. The low-friction silicone membrane forms the upper layer of the film. The tacky silicone polymer layer (i.e., the lower layer) contacts the patient's skin during the device's use. The OCA in the Device is sterile, high-purity perfluorodecalin (PFD), an inert, non-toxic liquid (FluoroMed, LLC).
#### Performance Data
Bench performance testing was conducted in support of the additional wavelength indications and is described below. It was our expectation that after exposure to the additional lasers and wavelengths, no optical changes to either the patch or the OCA (PFD) would be observed, and that the PFD would remain chemically unchanged after said exposure. In all instances, the DESCRIBE® PFD Patch functioned as intended and results observed were as expected.
# Performance Testing- Bench
Bench performance testing was conducted in order to confirm that the optical transparency of the DESCRIBE® PFD Patch and its optical clearing agent PFD are unchanged after exposure to standard, commercially available lasers. Briefly, the patch and PFD were exposed to the highest available laser intensities at multiple wavelengths (Q-switched 532, 694, 755 and 1064 nm lasers, as well as 532, 755, 785, and 1064 nm picosecond lasers). The exposed patches and their unexposed controls, as well as the respective associated PFD, were then analyzed using UV/V/S/NIR spectroscopy. No optical changes were observed within the precision of the instrument, demonstrating optical stability of the patch and the OCA after laser exposure. This testing verified that the Patch is optically suitable for use with Q-switched 532, 694, 755 and 1064 nm lasers and 532, 755, 785, and 1064 nm picosecond lasers.
Gas chromatography and mass spectrometry (GC/MS) testing was also conducted on the OCA after exposure to the same set of lasers outlined above. This testing confirmed that no chemical changes to the PFD occurred after laser exposure within the precision of the instrument.
# Stability Data
# Three-year stability tests
Merz wishes to amend the declared shelf life from 2 years to 3 years based on the attached real-time aging test protocol and reports. Briefly, twenty samples of the Patch from identical lots manufactured in 2014 were placed in storage; after 3 years,
{5}------------------------------------------------
fourteen patches were tested for package burst testing and bubble emission leak testing (Bubble and Burst Nonporous Packing Validation GLP Test as per ISO 11607, Nelson Laboratory Standard Test Protocol STP07170). For burst testing, test articles were conditioned for a minimum of 24 hours at 23 ± 2℃ and 50 ± 5% relative humidity (RH). The packages were then attached to the T.M. Electronics package tester by adhesive package ports and pressurized at a constant rate until they burst. For bubble emission leak testing, a probe was attached to a package then the test article was completely immersed in the fluid. Next, the package was pressurized to a minimum of 10 inches H1,0 and inspected for evidence of bubble emission originating from the package or seals. All Nonporous Packaging Validation test method acceptance criteria were met.
Additionally, a representative sample Patch from the above batch was used to test three parameters to confirm stability after 3 years. Visual verification was conducted to ensure the product remained properly sealed. Additionally, the heat seal was visually verified via measurement to ensure a width of no less than approximately 5 mm. Finally, the product weight was measured both before and after evaporative removal of the perfluorodecalin to verify that 1.8 mL + 30% - 10% had remained in the pouch. All acceptance criteria were met and thus the product was deemed to have successfully passed 3-year real time aging testing.
# Substantial Equivalence
The DESCRIBE PFD Patch is as safe and effective as the predicate device (DeScribe Patch, ON Light Science, K150212) as the two devices are identical. They have the same intended uses and similar indications, technological characteristics, and principles of operation. This submission contains bench studies in which the Patch was exposed to the highest energy fluence at a number of laser wavelengths (Q-switched 532, 694, 755 and 1064 nm lasers and 532, 755, 785 and 1064 nm picosecond lasers), including the wavelength currently contained in the indication (specifically, a 755nm Q-Switched Alexandrite laser).
A key feature of the Patch is that it is completely optically transparent to all dermatological lasers. Hence, it does not absorb or attenuate laser light any more than a sheet of glass. The bench studies contained within this application show that upon measuring the optical absorption spectra of both the patch and liquid PFD with a UV/Visible Spectrometer, the patch is transparent at all of the wavelengths tested. We
{6}------------------------------------------------
conclude that since no light is absorbed at any of the new wavelengths of interest, the Patch has been demonstrated to be formally clinically suitable for use at these laser wavelengths (Q-switched 532, 694, 755 and 1064 nm lasers and 532, 755, 785 and 1064 nm picosecond lasers). Thus, the DESCRIBE PFD Patch, with the amended indication including additional laser wavelengths, is substantially equivalent to the original cleared device.
# Conclusions
The DESCRIBE PFD Patch with its revised indication statement is as safe and effective as the predicate device. Performance testing as described above has been conducted to support these claims.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.