The Turbo-Power Laser Atherectomy Catheter is indicated for laser atherectomy of de novo restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).
Device Story
Turbo-Power Laser Atherectomy Catheters are sterile, single-use devices used with the CVX-300 Excimer Laser System for peripheral atherectomy. The catheter features optical fibers arranged eccentrically around a 0.018" guidewire lumen, a hydrophilic-coated working length, and a radiopaque marker band. The device connects to a Motor Drive Unit (MDU). During operation, the system transmits ultraviolet energy from the CVX-300 laser to the distal tip to photoablate arterial obstructions (atheroma, fibrosis, calcium, thrombus) via molecular bond disruption. The device supports a 'jogging' mode and a continuous rotation mode (simultaneous lasing and rotation). Used by physicians in clinical settings to recanalize diseased vessels, the device helps restore blood flow. The output is the physical removal of obstructive material, which aids clinical decision-making by facilitating vessel recanalization and improving patient outcomes in peripheral artery disease.
Clinical Evidence
No clinical data was required. Evidence consists of bench testing (simulated use, functional, physical, and laser testing) and pre-clinical GLP studies comparing usability and procedural safety between the 2.0mm and 2.3mm catheter models.
Technological Characteristics
Catheter subassembly with optical fibers, 0.018" guidewire lumen, PTFE lumen tip, stainless steel torque wire, and hydrophilic coating. Radiopaque marker band for visibility. Energy source: CVX-300 Excimer Laser System (ultraviolet). Operation: simultaneous lasing and rotation or jogging mode. Sterilization: AAMI TIR:28-2009 product adoption. Dimensions: 2.0mm (6Fr) and 2.3mm (7Fr).
Indications for Use
Indicated for laser atherectomy of de novo or restenotic lesions in native infrainguinal arteries and femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive PTA, in patients requiring peripheral atherectomy.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2017
Spectranetics, Inc. Mr. Christopher McLellan Regulatory Affairs Manager 9965 Federal Drive Colorado Springs, Colorado 80921
Re: K172687
Trade/Device Name: Turbo-Power Laser Atherectomy Catheters Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: September 5, 2017 Received: September 6, 2017
Dear Mr. McLellan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Mr. McLellan
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172687
#### Device Name
Turbo-Power Laser Atherectomy Catheters
Indications for Use (Describe)
The Turbo-Power Laser Atherectomy Catheter is indicated for laser atherectomy of de novo restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <div style="float: left; margin-right: 5px;"> <img alt="Checked" height="12" src="checked.png" width="12"/> </div>Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;"> <div style="float: left; margin-right: 5px;"> <img alt="Unchecked" height="12" src="unchecked.png" width="12"/> </div>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Spectranetics logo. The logo consists of a blue diamond shape on the left, followed by the word "Spectranetics" in blue, bold font. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller, lighter blue font. The logo is clean and modern, and the tagline suggests that the company is always striving to improve.
# 510(k) Summary
This 510(k) summary was prepared in accordance with 21 CFR 807.92(c) Prepared on September 06, 2017
| 510(k) Submitter / Holder: | The Spectranetics Corporation |
|----------------------------|-------------------------------------------|
| | 9965 Federal Drive |
| | Colorado Springs, CO 80921.3617 |
| | Establishment Registration No: 3007284006 |
| Contact: | Ms. Stephanie Byrum |
| | Regulatory Affairs Specialist |
| | Office: 719.447.2671 |
| | Mobile: 719.482.4765 |
| | Fax: 719.447.2070 |
| | Email: Stephanie.Byrum@spnc.com |
## Subiect Device
Device Trade Name:
| Device Common Name: | |
|----------------------------|--|
| Device Class: | |
| Classification Regulation: | |
| Regulation Description: | |
| Product Code: | |
| 510(k) Type: | |
| Model Numbers: | |
Turbo-Power™ (2.0mm) and (2.3mm) Laser Atherectomy Catheters Laser Atherectomy Catheter = 21 CFR 870.4875, Intraluminal Artery Stripper Cardiovascular MCW Traditional 420-050, 423 .- 050
#### Predicate Device
The Turbo-Power (2.0mm) and (2.3mm) Laser Atherectomy Catheters are being compared to the following legally marketed predicate device:
| 510(k) Number: | K152181 (cleared November 12, 2015) and K162561 (cleared January 5, 2017) |
|---------------------|---------------------------------------------------------------------------|
| Manufacturer: | The Spectranetics Corporation |
| Trade Name: | Turbo-Power Laser Atherectomy Catheter |
| Device Common Name: | Laser Atherectomy Catheter |
| Model Number: | 420-050, 423-050 |
## Intended and Indications for Use
Turbo-Power™ (2.0mm) and (2.3mm)Laser Atherectomy Catheter is indicated for laser atherectomy of de novo or restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Spectranetics logo. The logo consists of a blue diamond shape with a white circle in the center, followed by the word "Spectranetics" in blue, bold font. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller, regular font.
#### Device Description
The 2.0mm and 2.3mm Turbo-Power Laser Atherectomy Catheters are laser atherectomy devices designed for use with the CVX-300™ Excimer Laser System. The 2.0mm and 2.3mmTurbo-Power Laser Atherectomy Catheters are sterile, single use, prescription only devices used for peripheral atherectomy.
The 2.0mm and 2.3mm Turbo-Power Laser Atherectomy Catheters are used to ablate lesions with reference vessel diameters of ≥3.0mm. 2.0mm and 2.3mm Turbo-Power Laser Atherectomy Catheters are comprised of 2 subassemblies:
- 1. Catheter Subassembly
- 2. Motor Drive Unit (MDU) Subassembly
The working length of the Turbo Power Laser Atherectomy Catheter is constructed of multiple optical fibers arranged eccentrically around a 0.018" (0.46 mm) guidewirecompatible lumen. The PTFE quidewire lumen tip is attached to a stainless steel torque wire which is connected to the MDU at the proximal end of the working length. The multifiber laser catheter transmits ultraviolet energy from the Spectranetics CVX-300 Excimer Laser System through the tip of the laser to an obstruction in the patient's artery. The outer surface of the laser catheter working length is hydrophilic-coated, and the distal tip of the catheter contains a radiopaque marker band for in situ visibility. The ultraviolet energy transmitted from the CVX-300 laser system is used to photoablate multiple morphology lesions which mav be comprised of atheroma, fibrosis, calcium, and thrombus, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.
#### Technological Characteristics
The 2.0mm (6Fr) and 2.3mm (7Fr) Turbo-Power catheters introduce a new mode of operation of simultaneously pressing the two (2) jogging buttons for continuous distal tip rotation and allows for simultaneous lasing and rotation. The subject devices also contain the jogging mode feature, which currently exists for both predicate devices. This device iteration does not affect the fundamental scientific technology used in the Turbo-Power family of devices. The mechanism of action, and intended use, remain unchanged from the predicate devices, 2.0mm (6Fr) Turbo-Power and 2.3mm (7Fr) Turbo-Power Laser Atherectomy Systems.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Spectranetics logo. The logo consists of a blue diamond shape with a white circle in the center, followed by the word "Spectranetics" in blue, bold letters. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller, lighter font.
#### Performance Data1
The following testing was conducted to validate and verify that the subject device met all acceptance criteria as required by the risk analysis that was performed:
### Design Verification and Validation Testing
- Simulated Use Testing ●
- . Functional Testing
- Physical Testing ●
- . Laser Testing
#### Sterilization
- Product adoption equivalency per AAMI TIR:28-2009* ●
#### Pre-clinical and Clinical Data:
In addition to the design verification and validation tests, pre-clinical GLP studies were conducted to compare the usability and procedural safety of the 2.0mm (6Fr) Turbo-Power and the 2.3mm (7Fr) Turbo-Power laser catheters, and support this premarket notification.
New clinical data was not required to demonstrate substantial equivalence.
#### Substantial Equivalence
Based on the similarities in design between the subject devices and predicate devices which are currently in use, and the performance and pre-clinical data, the use of the 2.0mm (6Fr) Turbo-Power and the 2.3mm (7Fr) Laser Atherectomy Catheters for the proposed indication does not raise new questions related to safety and effectiveness compared with the predicate. Therefore the 2.0mm (6Fr) and 2.3mm (7Fr) Turbo-Power Laser Atherectomy Catheters with the continuous rotation and jogging mode features are substantially equivalent to the predicate 2.0mm (6Fr; K162561) and 2.3mm (7Fr; K152181) Turbo-Power Laser Atherectomy Catheters with the jogging mode feature.
<sup>1</sup> All testing marked with an * is leveraged from the predicate Turbo-Power (2.3mm). Additional testing required is summarized within this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.