← Product Code [FEB](/productcode/FEB) · K172677

# Advantage Plus Pass-Thru (K172677)

_Medivators, Inc. · FEB · Dec 19, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172677

## Device Facts

- **Applicant:** Medivators, Inc.
- **Product Code:** [FEB](/productcode/FEB.md)
- **Decision Date:** Dec 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Advantage Plus Pass-Thru Endoscope Reprocessing System tests, washes, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus Pass-Thru system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is required prior to disinfection in the Advantage Plus Pass-Thru System. The Advantage Plus Pass-Thru Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Plus Detergent in its washing cycle at a concentration of 0.5%. Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used under the following contact conditions -Contact Time: 5 minutes Temperature: 30°C MRC: 850ppm

## Device Story

Electro-mechanical Automatic Endoscope Reprocessor (AER) for high-level disinfection of pre-cleaned endoscopes. Features 'pass-thru' wall-mounted configuration separating 'dirty' (loading) and 'clean' (unloading) sides to prevent cross-contamination. Input: contaminated endoscopes. Process: automated channel connectivity, leak, and blockage testing; washing with Intercept Plus detergent; high-level disinfection using Rapicide PA (peracetic acid-based) at 30°C for 5 minutes. Output: disinfected endoscopes. Operated by clinical staff. System monitors cycle parameters, performs MRC testing via test strips, and provides optional alcohol/air purge. Benefits: standardized reprocessing, reduced cross-contamination risk, and verified disinfection efficacy.

## Clinical Evidence

Bench testing only. Includes simulated-use testing (achieved 6-log reduction of test organism), clinical in-use testing (demonstrated no viable organisms post-reprocessing), water-line/self-disinfection cycle testing (survivors ≤ 10 CFU/100 mL, no biofilm), rinse residual testing (residuals below LD50 exposure limits), material compatibility, and critical parameter functionality testing. Electrical safety/EMC testing per IEC 61010 and IEC 61326; software validation per IEC 62304; human factors evaluation per ISO 62366.

## Technological Characteristics

Electro-mechanical AER. Materials: metals and polymers identical to predicate. Energy: electrical. Connectivity: standalone system with integrated monitoring. Disinfection: peracetic acid-based (Rapicide PA). Cleaning: Intercept Plus detergent. Features: leak testing, channel blockage/connection testing, automated cycle monitoring, pass-thru door configuration. Software: validated per IEC 62304.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Medivators Advantage Plus Endoscope Reprocessing System ([K082988](/device/K082988.md))
- Medivators Advantage Plus Endoscope Reprocessing System ([K102996](/device/K102996.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized symbol that resembles a human figure in profile, with three overlapping faces and flowing lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 19, 2017

Medivators Inc. Megan Skaar Regulatory Affairs Specialist 14605 28th Avenue North Minneapolis, Minnesota 55447

Re: K172677

Trade/Device Name: Advantage Plus Pass Thru Automated Endoscope Reprocessing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: FEB Dated: November 20, 2017 Received: November 21, 2017

Dear Megan Skaar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172677

Device Name

Advantage Plus Pass Thru Endoscope Reprocessing System

#### Indications for Use (Describe)

Advantage Plus Pass-Thru Endoscope Reprocessing System tests, washes, disinfect and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus Pass-Thru system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is required prior to disinfection in the Advantage Plus Pass-Thru System.

The Advantage Plus Pass-Thru Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Plus Detergent in its washing cycle at a concentration of 0.5%.

Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used under the following contact conditions -Contact Time: 5 minutes Temperature: 30°C MRC: 850ppm

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span></span>   |
|----------------------------------------------|-----------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span></span> ✓ |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular icon with several curved lines inside, followed by the word "MEDIVATORS" in large, blue, sans-serif font. Below the company name is the text "A Cantel Medical Company" in a smaller, blue, sans-serif font.

# 510(k) Summary

| 510(k) Number:                                                                                          | K172677                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:                                                                                           | Medivators Inc., a Cantel Medical Company                                                                                                                                                                                                                                                                                                                                                               |
| Address:                                                                                                | 14605 28th Avenue North<br>Minneapolis, MN 55447<br>(763) 553-3300                                                                                                                                                                                                                                                                                                                                      |
| Official Contact:                                                                                       | Megan Skaar<br>Regulatory Affairs Specialist, Medivators Inc.                                                                                                                                                                                                                                                                                                                                           |
| Date:                                                                                                   | 19 December 2017                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code:<br>Device Class:<br>Regulation No: | Advantage Plus Pass Thru Endoscope Reprocessing System<br>Accessories, Cleaning, For Endoscope<br>Endoscope and accessories<br>FEB<br>II<br>Subject Device - Advantage Plus Pass Thru Endoscope Reprocessing System,<br>876.1500<br>Predicate Device Trade Name - Medivators Advantage Plus Endoscope<br>Reprocessing System and Rapicide PA High Level Disinfectant (K082988 and<br>K102996), 876.1500 |

#### 1. Indications for Use

Advantage Plus Pass-Thru Endoscope Reprocessing System tests, washes, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes, between patient uses. The Advantage Plus Pass-Thru system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is required prior to disinfection in the Advantage Plus Pass-Thru System.

The Advantage Plus Pass-Thru Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Plus Detergent in its washing cycle at a concentration of 0.5%.

Rapicide PA Test Strips are used after the disinfection cycle to ensure that the disinfectant is above the minimum recommended concentration (MRC) of 850 ppm peracetic acid when used; this confirms that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used under the following contact conditions -

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, followed by the word "MEDIVATORS" in blue, sans-serif font. Below the word "MEDIVATORS" is the text "A Cantel Medical Company" in a smaller, sans-serif font, also in blue. The overall design is clean and corporate, conveying a sense of professionalism and reliability.

Contact Time: 5 minutes Temperature: 30°C MRC: 850ppm

# 2. Device Description

The Advantage Plus Pass-Thru is an electro-mechanical Automatic Endoscope Reprocessor (AER) intended to high-level disinfect pre-cleaned endoscopes and their related accessories by exposing them to a peracetic acid based disinfectant – Rapicide PA disinfectant solution (cleared under 510k – K082988).

The subject device is closely related to its predicate the Advantage Plus Endoscope Reprocessing System cleared in 510k – K082988 and K102996 but differs in that it is designed in a "pass-thru" configuration. The subject device is intended to be built into a wall which separates 'clean' and 'dirty' sides. It provides the ability to load an endoscope from the dirty side and unload the disinfected endoscope from the clean side, thus preventing potential mix-up between high level disinfected and contaminated endoscopes.

Contaminated endoscopes are placed in the subject device from the 'dirty' side and appropriate connections to the endoscope channels are made. The system is designed to test for proper channel connectivity, channel blockages and leaks in the outer endoscope sheath. After the endoscope passes these tests, the system proceeds to the washing cycle with Intercept Plus detergent, and the highlevel disinfection cycle. In the high-level disinfection cycle, Rapicide PA HLD is diluted by the incoming water to a minimum concentration of 850 ppm of peracetic acid. The system measures and maintains the temperature of this disinfectant solution to ensure that it is above the minimum recommended temperature of 30℃ for Rapicide PA for effective High Level Disinfection to occur. Following the 5 minute disinfectant contact time, the user takes a sample to test for MRC (850 ppm of peracetic acid). Once the MRC test passes, the disinfectant is emptied from basin and drained out of the AER. The endoscope is then rinsed, dried and removed from the 'clean side' of the AER.

## 3. Comparison and Summary of Technological Characteristics for the Subject and Predicate Devices

The subject device – Advantage Plus Pass-Thru Endoscope Reprocessing System and its predicate device – Advantage Plus Endoscope Reprocessing System are very closely related and have the same intended use and fundamental technology.

## Similarities Between Subject and Predicate Device

The Advantage Plus Pass-Thru and its predicate device have the same scientific technology including principle of operation, features and materials. Both devices have high level disinfection cycle. They use the same disinfectant chemistry – Rapicide PA under the same use conditions (30°C for 5 minutes at MRC of 850 ppm PAA) to high level disinfect endoscopes. They have identical chemistry dosing mechanisms to produce equivalent concentration of disinfection solution. Additionally, both

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, which has several curved lines inside. To the right of the graphic is the word "MEDIVATORS" in a bold, blue sans-serif font, with the text "A Cantel Medical Company" in a smaller font below it.

the devices have the same fluid flow rate and pathway. The Advantage Plus Pass-Thru System is constructed of the same materials, metals and polymers, as its predicate device.

#### Differences Between Subject and Predicate Device

The subject device provides the ability to load an endoscope into the AER from the dirty, contaminated side and unload the disinfected scope from the clean side - creating the pass-thru system. To facilitate this system configuration, the basin door arrangement was altered from the arrangement of the predicate device. There were no changes in critical parameters to achieve high level disinfection with the alteration of a pass-thru door configuration. Verification of software and performance testing on the subject device pass-thru configuration demonstrates the configuration difference between the subject device and the predicate device does not impact the safety and effectiveness profile of the devices. The pass-thru configuration of the subject AER physically separates the 'clean' and 'dirty' sides of the reprocessing unit and hence reduces the risk of cross contamination as demonstrated in the performance testing outlined in Table 2 below.

Performance bench testing, included in Section 4 of this summary, supports substantial equivalence of the pass-thru system configuration. A device comparison table which supports substantial equivalence of the subject device to the predicate device is provided below –

| Important Parameters                                                            | Subject Device – Advantage Plus<br>Pass-Thru                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device - Advantage Plus<br>(cleared under 510k - K082988<br>and K102996)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                           |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                                                    | Advantage Plus Pass-Thru<br>Endoscope Reprocessing System<br>tests, washes, disinfects and rinses<br>endoscopes, such as fiberoptic and<br>video endoscopes, between<br>patient uses. The Advantage Plus<br>Pass-Thru system is indicated to<br>provide high level disinfection of<br>heat sensitive semi-critical<br>endoscopes and related<br>accessories. Manual cleaning of<br>endoscopes is required prior to<br>disinfection in the Advantage Plus<br>Pass-Thru System.<br>The Advantage Plus Pass-Thru<br>Endoscope Reprocessing System<br>uses Rapicide PA High Level<br>Disinfectant to provide high level | Medivators Advantage Plus<br>Endoscope Reprocessing System<br>tests, washes, disinfects and rinses<br>endoscopes, such as fiberoptic and<br>video endoscopes, between<br>patient uses. The Advantage Plus<br>system is indicated to provide high<br>level disinfection of heat sensitive<br>semi-critical endoscopes and<br>related accessories. Manual<br>cleaning of endoscopes is not<br>required prior to disinfection in the<br>Advantage Plus Pass-Thru System.<br>The Advantage Plus Endoscope<br>Reprocessing System uses Rapicide<br>PA High Level Disinfectant to<br>provide high level disinfection of |                                                                                                                           |
| disinfection of endoscopes when<br>used according to the directions<br>for use. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | endoscopes when used according<br>to the directions for use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                           |
| Product Code                                                                    | FEB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | FEB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                           |
| FDA Regulation                                                                  | 21 CFR§ 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 21 CFR§ 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                           |
| Pass-Thru Configuration/<br>AER separates Clean and<br>Dirty Side?              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                           |
| Type of Machine Function                                                        | Electro-mechanical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Electro-mechanical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                           |
| System Technology Features                                                      | The system performs endoscope<br>connection test, leak and channel<br>blockage test, washing (cleaning),<br>disinfection and rinsing. The cycle<br>steps for each basin are<br>monitored. Built in safety features<br>and alarms.                                                                                                                                                                                                                                                                                                                                                                                   | The system performs endoscope<br>connection test, leak and channel<br>blockage test, washing (cleaning),<br>disinfection and rinsing. The cycle<br>steps for each basin are<br>monitored. Built in safety features<br>and alarms.                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                           |
|                                                                                 | Disinfection time,<br>temperature and minimum<br>recommended<br>concentration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Rapicide PA - 5 minutes at 30°C;<br>850ppm Peracetic acid MRC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Rapicide PA - 5 minutes at 30°C;<br>850ppm Peracetic acid MRC                                                             |
|                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Detergent used in cleaning<br>cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Intercept Plus Detergent                                                                                                  |
|                                                                                 | Disinfection efficacy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >106 log reduction of<br><i>Mycobacterium terrae</i>                                                                      |
| Manual cleaning required<br>before use?                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes (per K082988) however if the<br>automated cleaning cycle is used,<br>manual cleaning is not required<br>(per K102988) |
| Performs leak test?                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                           |
| Performs channel<br>connection and blockage<br>test?                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                           |
| Post-disinfection treatment                                                     | Optional alcohol purge followed by<br>air purge of endoscope channels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Optional alcohol purge followed by<br>air purge of endoscope channels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                           |

#### Table 1 – Predicate Comparison Table

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, followed by the word "MEDIVATORS" in a bold, blue sans-serif font. Below the company name, in a smaller font, is the text "A Cantel Medical Company," also in blue. The overall design is clean and corporate, conveying a sense of professionalism and reliability.

# 4. Summary of Non-Clinical Performance Data

Medivators has conducted the following testing to demonstrate the substantial equivalence of Advantage Plus Pass Thru AER –

- Performance Testing Identical performance testing has been conducted on the subject and . predicate devices to ensure the Advantage Plus Pass-Thru meets equivalent acceptance criteria as the Advantage Plus. This testing confirms the subject device performs with the same safety

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for MEDIVATORS, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, featuring several curved lines. To the right of the graphic is the word "MEDIVATORS" in a bold, blue sans-serif font. Below "MEDIVATORS" is the text "A Cantel Medical Company" in a smaller, lighter blue font.

and effectiveness as the predicate device, please refer to the table below for a brief discussion of the performance testing.

|  | Table 2 – Comparison of Non-Clinical Performance Tests for the Predicate and Subject Devices |
|--|----------------------------------------------------------------------------------------------|
|--|----------------------------------------------------------------------------------------------|

| Performance Testing                                                                                                                                                                                                                                  | Acceptance Criteria                                                                                                                               | Results |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Simulated-Use Testing -<br>Demonstrates that when used as indicated, the<br>system high level disinfects endoscopes using<br>Rapicide PA HLD and Intercept Plus detergent.                                                                           | Achieve 6-log reduction of the test<br>organism in each endoscope<br>channel and endoscope exterior.                                              | Pass    |
| Clinical In-Use Testing -<br>Demonstrates that when used as indicated, the<br>system high level disinfects nominal endoscopes<br>using Rapicide PA HLD after patient/clinical use.                                                                   | Demonstrate no viable organisms<br>remain after reprocessing.                                                                                     | Pass    |
| Water-Line/Self-Disinfection Cycle Testing -<br>Evaluation of the bactericidal effectiveness of<br>the self-disinfection cycle (SDC) and additional<br>subject device biofilm analysis by quantitative<br>Polymerase Chain Reaction gene expression. | Survivors after the SDC must be ≤<br>10 CFU of test organism per 100<br>mL.<br>Absence of biofilm formation<br>following self-disinfection cycle. | Pass    |
| Rinse Residual Testing -<br>Evaluation of residual levels of Rapicide PA and<br>Intercept Plus to demonstrate that endoscopes<br>reprocessed in the AER are safe for both<br>patients and users.                                                     | Any chemistry residuals must be<br>equal to or lower than the lowest<br>reports LD50 exposure with a 10-<br>fold safety margin.                   | Pass    |
| Material Compatibility -<br>Demonstrates the system compatibility with<br>Rapicide PA High Level Disinfectant and subject<br>device compatibility with Intercept Plus<br>Detergent.                                                                  | Acceptance criteria included<br>dimensional and weight changes<br><5%, visual observations and FTIR<br>spectra.                                   | Pass    |
| Critical Parameter/Component Functionality<br>Testing -<br>Demonstrates the system consistently and<br>reliably delivers the specified physical process<br>parameters.                                                                               | All critical process parameters must<br>achieve and maintain required<br>specifications.                                                          | Pass    |

- Electrical Safety and Electromagnetic Compatibility – Electrical safety and electromagnetic compatibility testing has been conducted to demonstrate compliance with IEC 61010 and IEC 61326.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a blue circular graphic on the left, followed by the word "MEDIVATORS" in blue, block letters. Below the company name is the text "A Cantel Medical Company" in a smaller font, also in blue. The logo is simple and professional, conveying a sense of trust and reliability.

- Software Validation Software validation testing has been conducted to demonstrate ● compliance with IEC 62304.
- Human Factors/Usability Evaluation Human factors testing has been conducted to ● demonstrate compliance with ISO 62366.

## 5. Conclusion

The Advantage Plus Pass-Thru intended use and fundamental scientific technology is the same as the predicate device, the Advantage Plus AER. Performance testing, software verification and validation testing and human factors validation demonstrate that the subject device pass-thru configuration is substantial equivalent to the predicate device. Therefore, based on the intended use, technology and performance data, the subject device is as safe and as effective as the legally marketed predicate device. The Advantage Plus Pass-Thru AER is substantially equivalent to predicate device Advantage Plus AER cleared in 510(k)s – K082988 and K102996.

---

**Source:** [https://fda-staging.innolitics.com/device/K172677](https://fda-staging.innolitics.com/device/K172677)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
