Classic Cotton Cannulatome®, CannulaTome II® Double Lumen Sphincterotome, Cotton CannulaTome II® PC, UTS® Ultra Taper Sphincterotome, Tri-Tome pc® Triple Lumen Sphincterotome, Howell D.A.S.H.® Sphincterotome with DomeTip®, Billroth II Sphincterotome, Soehendra® BII Sphincterotome, Models ACU-1 and ACU-1-VL active cords
K172665 · Wilson-Cook Medical Inc./Cook Endoscopy · KNS · May 24, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K172665
Device Name
Classic Cotton Cannulatome®, CannulaTome II® Double Lumen Sphincterotome, Cotton CannulaTome II® PC, UTS® Ultra Taper Sphincterotome, Tri-Tome pc® Triple Lumen Sphincterotome, Howell D.A.S.H.® Sphincterotome with DomeTip®, Billroth II Sphincterotome, Soehendra® BII Sphincterotome, Models ACU-1 and ACU-1-VL active cords
Applicant
Wilson-Cook Medical Inc./Cook Endoscopy
Product Code
KNS · Gastroenterology, Urology
Decision Date
May 24, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Howell D.A.S.H.® Sphincterotome with DomeTip® (Cook Reference Part Numbers DASH-21, DASH-21-480, DASH-1, DASH-260, DASH-480, DASH-ACRO-25-450, DASH-35, DASH-ACRO-35-260, DASH-35-480, DASHACRO-35-450): This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP. Tri-Tome pc® Triple Lumen Sphincterotome (Cook Reference Part Numbers TRI-20, TRI-20M, TRI-25, TRI-25M, TRI-25M-P, TRI-25M-SLT, TRI-30, TRI-30M): This device is used for cannulation of the ductal system and for sphincterotomy. Classic Cotton® CannulaTome® (Cook Reference Part Numbers CCPT-25, CCPT-25-MONO, CCPT-25ME), CannulaTome II® Double Lumen Sphincterotome (Cook Reference Part Numbers CT-20, CT-20M, CT-30, CT-30M), Cotton Cannulatome II PreCurved Double Lumen Sphincterotome (Cook Reference Part Numbers CT-25, CT-25M, CT-25M-P UTS®), and Ultra Taper Sphincterotome (Cook Reference Part Numbers UTS-15, UTS-20, UTS-20M, UTS-25, UTS-25M, UTS-30, UTS-30M): This device is used for cannulation of the ductal system and for sphincterotomy. Billroth II Sphincterotome (Cook Reference Part Numbers PTG-20-6-BII-NG) and Soehendra® BII Sphincterotome (Cook Reference Part Numbers PT-5.5-BII-SOEHENDRA): This device is used for the cannulation of the ductal system and for sphincterotomy. Models ACU-1 and ACU-1-VL active cords: This device is used to connect Cook monopolar electrosurgical accessories to compatible electrosurgical generators.
Device Story
Sterile, single-use endoscopic sphincterotomes; long plastic cannula with internal wire; distal wire segment exposed. Used during ERCP; physician inserts device through endoscope accessory channel; high-frequency electrosurgical current applied via wire to cut papilla roof; enables access to biliary/pancreatic orifices. Active cords connect sphincterotomes to electrosurgical generators. Benefits: facilitates ductal cannulation and sphincterotomy; preloaded versions assist in bridging strictures. Used in clinical/hospital settings by gastroenterologists.
Clinical Evidence
Bench testing only. Includes biocompatibility (cytotoxicity, irritation, sensitization), electrical safety (IEC 60601-1, 60601-2-2, 60601-2-18), mechanical integrity (joint strength, force to bow), functional testing (visualization, fluoroscopic visibility, functional age), and sterilization/packaging validation.
Technological Characteristics
Materials: PTFE (catheter), 303/304 stainless steel (wire), Parylene C (coating). Design: Single/double/triple lumen, 4-7 Fr diameter, 196-200 cm length. Energy: Monopolar electrosurgical. Connectivity: Active cords connect to electrosurgical generators. Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for patients requiring cannulation of the biliary or pancreatic ductal systems and sphincterotomy during ERCP procedures. Includes bridging of difficult strictures for preloaded devices. Active cords indicated for connecting electrosurgical accessories to generators.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized symbol. To the right is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' and 'ADMINISTRATION' in blue text.
May 24, 2018
Wilson-Cook Medical, Inc. / Cook Endoscopy Theresa De Prat Global Regulatory Affairs Specialist 4900 Bethania Station Road Winston-Salem, NC 27105
#### Re: K172665
Trade/Device Name: Classic Cotton Cannulatome® CannulaTome II® Double Lumen Sphincterotome Cotton CannulaTome II® PC UTS® Ultra Taper Sphincterotome Tri-Tome pc® Triple Lumen Sphincterotome Howell D.A.S.H.® Sphincterotome with DomeTip® Billroth II Sphincterotome Soehendra® BII Sphincterotome Models ACU-1 and ACU-1-VL active cords Regulation Number: 21 CFR& 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS Dated: April 23, 2018 Received: April 24, 2018
Dear Theresa De Prat:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172665
#### Device Name
Classic Cotton Cannulatome®, CannulaTome II® Double Lumen Sphincterotome, Cotton CannulaTome II® PC, UTS® Ultra Taper Sphincterotome, Tri-Tome pc® Triple Lumen Sphincterotome, Howell D.A.S.H.® Sphincterotome with DomeTip®, Billroth II Sphincterotome, Soehendra® BII Sphincterotome, and Models ACU-1 and ACU-1-VL active cords
#### Indications for Use (Describe)
Howell D.A.S.H.® Sphincterotome with Dome Tip® (Cook Reference Part Numbers DASH-21, DASH-1, DASH-260, DASH-480, DASH-ACRO-25-450, DASH-35, DASH-ACRO-35-260, DASH-35-480, DASH-ACRO-35-450): This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP.
Tri-Tome po® Triple Lumen Sphincterotome (Cook Reference Part Numbers TRI-20, TRI-25, TRI-25M, TRI-25M-P, TRI-25M-SLT, TRI-30, TRI-30M): This device is used for cannulation of the ductal system and for sphincterotomy.
Classic Cotton® CannulaTome® (Cook Reference Part Numbers CCPT-25-MONO, CCPT-25ME), CannulaTome II® Double Lumen Sphincterotome (Cook Reference Part Numbers CT-20, CT-30M), Cotton Cannulatome II® PreCurved Double Lumen Sphincterotome (Cook Reference Part Numbers CT-25, CT-25M, CT-25M-P UTS®), and Ultra Taper Sphincterotome (Cook Reference Part Numbers UTS-15, UTS-20M. UTS-25, UTS-30, UTS-30, UTS-30M): This device is used for cannulation of the ductal system and for sphincterotomy.
Billroth II Sphincterotome (Cook Reference Part Numbers PTG-20-6-BII-NG) and Soehendra® BII Sphincterotome (Cook Reference Part Numbers PT-5.5-BII-SOEHENDRA): This device is used for the cannulation of the ductal system and for sphincterotomy.
Models ACU-1 and ACU-1-VL active cords: This device is used to connect Cook monopolar electrosurgical accessories to compatible electrosurgical generators.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K172665 Page 1 of 4
#### 510(k) Summary
#### Sphincterotomes and Active Cords
Traditional 510(k) Premarket Notification
April 23, 2018
### Applicant Information
| Applicant: | Wilson-Cook Medical, Inc. /Cook Endoscopy |
|------------|---------------------------------------------------|
| | 4900 Bethania Station Road |
| | Winston-Salem, North Carolina 27105 |
| Contact: | Theresa de Prat, Regulatory Affairs Specialist II |
| Phone: | (336) 744-0157 ext. 396518 |
| Fax: | (336) 201-5994 |
#### Device Information
| Trade Names: | Classic Cotton Cannulatome®<br>CannulaTome II® Double Lumen Sphincterotome<br>Cotton CannulaTome II® PC<br>UTS® Ultra Taper Sphincterotome<br>Tri-Tome pc® Triple Lumen Sphincterotome<br>Howell D.A.S.H.® Sphincterotome with DomeTip®<br>Billroth II Sphincterotome<br>Soehendra® BII Sphincterotome<br>Models ACU-1 and ACU-1-VL active cords |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Names: | Papillotome, sphincterotome, active cord |
| Regulation Name: | Endoscopic electrosurgical unit and accessories |
| Regulation Number: | 21 CFR 876.4300 |
| Product Code: | KNS |
| Device Class: | Class II |
| Review Panel: | Gastroenterology-Urology |
### Predicate Device
| Name: | Zimmon® Papillotome |
|----------------|-----------------------|
| 510(k) Number: | K901443 |
| Date: | Cleared June 18, 1990 |
#### Device Description
The sphincterotomes described in this submission are a sterile, single use devices compatible with the accessory channel of endoscope. The device consists of a long, thin plastic tube (cannula) with a wire running the length of its interior. A small portion of
{4}------------------------------------------------
that wire is exposed at its distal end. The roof of the papilla is opened by passing highfrequency current through the wire, exposing the biliary or pancreatic orifices for selective cannulation. The active cords are accessories to sphincterotomes so that a sphincterotome can be connected to an electrosurgical unit.
### Intended Use
Howell D.A.S.H.® Sphincterotome with DomeTip® (Cook Reference Part Numbers DASH-21, DASH-21-480, DASH-1, DASH-260, DASH-480, DASH-ACRO-25-450, DASH-35, DASH-ACRO-35-260, DASH-35-480, DASHACRO-35-450): This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP.
Tri-Tome pc® Triple Lumen Sphincterotome (Cook Reference Part Numbers TRI-20, TRI-20M, TRI-25, TRI-25M, TRI-25M-P, TRI-25M-SLT, TRI-30, TRI-30M): This device is used for cannulation of the ductal system and for sphincterotomy.
Classic Cotton® CannulaTome® (Cook Reference Part Numbers CCPT-25, CCPT-25-MONO, CCPT-25ME), CannulaTome II® Double Lumen Sphincterotome (Cook Reference Part Numbers CT-20, CT-20M, CT-30, CT-30M), Cotton Cannulatome II PreCurved Double Lumen Sphincterotome (Cook Reference Part Numbers CT-25, CT-25M, CT-25M-P UTS®), and Ultra Taper Sphincterotome (Cook Reference Part Numbers UTS-15, UTS-20, UTS-20M, UTS-25, UTS-25M, UTS-30, UTS-30M): This device is used for cannulation of the ductal system and for sphincterotomy.
Billroth II Sphincterotome (Cook Reference Part Numbers PTG-20-6-BII-NG) and Soehendra® BII Sphincterotome (Cook Reference Part Numbers PT-5.5-BII-SOEHENDRA): This device is used for the cannulation of the ductal system and for sphincterotomy.
Models ACU-1 and ACU-1-VL active cords: This device is used to connect Cook monopolar electrosurgical accessories to compatible electrosurgical generators.
### Comparison to Predicate Device
The subject device and predicate device have the same intended use and different technological characteristics.
{5}------------------------------------------------
| Characteristic | Zimmon® Papillotome (K901443) | Cook Sphincterotomes and Active Cords (K172665) |
|------------------------------------------|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Used for endoscopic papillotomy/sphincterotomy. | This device is used for cannulation of the ductal system and for sphincterotomy. If preloaded, also aids in bridging difficult strictures during ERCP. |
| Sterile/Non-Sterile | Sterile, EO | Same |
| Disposable/Reusable | Disposable | Same |
| Cutting Wire Length | 15-30 mm | 15-30 mm |
| Cutting Wire Diameter | 0.012" (braided) | 0.012" (braided)<br>0.0075-0.010" (monofilament) |
| Handle configuration | Three Ring Handle | Same |
| Tip configuration | Straight or Tapered | Tapered or DomeTip |
| Wire Guide / Injection Port<br>Extension | No | Yes |
| Wire Control Port (WCP) with<br>Slide | No | Yes (on TriTome pc® Select) |
| Cutting Wire Material | 304 Stainless Steel | 303 stainless steel or 304 stainless steel; Parylene C added for "Protector" versions |
| Ink Marker Material | Gem Black | Black, Green, Blue, Silver,<br>Copper/Bronze, Gold |
| Packaging | PTEG Tray with Tyvek Lid | Tyvek and Mylar pouch |
| Catheter Design | Single lumen | Single, double or triple lumen |
| Catheter Design/Material | Polytetrafluoroethylene (PTFE) | Same |
| Catheter Length | 200 cm | 196 or 200 cm |
| Catheter Diameter | 5 Fr | 4-7 Fr |
| Wire Guide Compatibility | Not applicable | 0.021-0.035" |
# Substantial Equivalence Comparison
None of the differences in technological characteristics raise different questions of safety and effectiveness. Furthermore, performance data from acceptable scientific testing methods provide evidence that the subject device is substantially equivalent to the predicate device.
# Performance Data
Performance testing consisting of sterilization, shelf life, biocompatibility, and nonclinical bench testing, as follows, demonstrate that the subject device meets the performance requirements to fulfill the intended use of the device.
{6}------------------------------------------------
- GLP Cytotoxicity 1X MEM ISO elution .
- GLP ISO Intracutaneous irritation - 2 extracts
- GLP ISO Guinea Pig Maximization, 2 extracts ●
- IEC 60601-1: 2006 + A12: 2014 (3.1 edition)- Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- . IEC 60601-2-2: 2017 (6.0 edition) - Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories
- IEC 60601-2-18: 2009 (3.0 edition) Medical electrical equipment Part 2-18: . Particular requirements for the basic safety and essential performance of endoscopic equipment
- Ink visualization
- Visualization of Cutting Wire ●
- Orientation of Cutting Wire
- Fluoroscopic Visibility ●
- Drive Wire to Handle Joint Strength
- Brass Insert to Handle Joint Strength ●
- Active Cord to Pin Detachment ●
- Force to Bow .
- Functional Age Testing
- Sterile Barrier and Functional Testing After Shipping ●
- Active Cord: Wire to Universal Insert Solder Joint Strength
- Active Cord: Wire to Banana Plug Solder Joint Strength ●
- Active Cord: Wire to Valley Lab Plug Solder Joint Strength
- Active Cord: Functional Testing at Time Zero ●
- Active Cord: Functional Testing after Shipping ●
- Sterilization validation ●
- ETO residual testing
- Packaging visual inspection ●
- . Burst testing
- Dye leak testing ●
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.