K172662 · Led Technologies, Inc. · OHS · Jan 9, 2018 · General, Plastic Surgery
Device Facts
Record ID
K172662
Device Name
reVive Perioral
Applicant
Led Technologies, Inc.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jan 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The reVive® Perioral LED Light Therapy system is an Over-the-Counter (OTC) device intended for the treatment fine lines and wrinkles, and increase in circulation within the perioral region.
Device Story
Over-the-counter LED light therapy device; treats fine lines/wrinkles and increases perioral circulation. Device emits light via four LED types (605nm amber, 630nm red, 660nm red, 880nm infrared). User applies device directly to skin; no user settings or adjustments required. Powered by internal Li-Ion battery, rechargeable via USB. Intended for home use by consumers. Output provides narrow-band light energy to skin. Benefits include non-invasive wrinkle reduction and improved local circulation.
Clinical Evidence
Bench testing only. Includes functional performance testing, software validation, and biocompatibility testing (ISO 10993-5, ISO 10993-10). Usability study with 15 participants showed 100% success rate in performing light sensitivity tests and device operation.
Technological Characteristics
LED-based light therapy; wavelengths 605nm, 630nm, 660nm, 880nm. Irradiance 67.7 mW/cm². Battery-powered (Li-Ion 3.7V, 5V USB charging). Form factor: handheld LED module. Standards: IEC 60601-1, IEC 60601-1-2, UL 1642, UN 38.3. Biocompatibility: ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for OTC treatment of fine lines and wrinkles and increased circulation in the perioral region.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 9, 2018
LED Technologies, Inc. Ms. Jelena Barbaric Compliance Manager 6000 Greenwood Plaza Blvd., Suite 110 Greenwood Village, Colorado 80111
Re: K172662
Trade/Device Name: reVive Perioral LED Light Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS. ILY Dated: October 12, 2017 Received: October 12, 2017
Dear Ms. Barbaric:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Ms. Jelena Barbaric
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172662
Device Name
The reVive® Perioral LED Light Therapy system
Indications for Use (Describe)
Indications for Use (Describe)
The reVive® Perioral LED Light Therapy system is an Over-the-Counter (OTC) device intended for treatment of fine lines and wrinkles, and increase in circulation within the perioral region.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete time to review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510 (k) Summary
This summary of 510 (k) information is being submitted in accordance with the requirements of 21 CFR § 878.4810.
Submission Date: August 31st, 2017
Preparation Date: January 9th, 2018
| 1. Submitter Information: | LED Technologies, Inc. – Jelena Barbaric | |
|---------------------------|------------------------------------------|--|
| | 6000 Greenwood Plaza Blvd., Suite 110 | |
| | Greenwood Village, CO 80111 | |
| | Tel: 303-407-6882 | |
| | Email: jbarbaric@ledtechnologies.com | |
| For Specification Developer: | LED Technologies, Inc. |
|------------------------------|--------------------------------------|
| | Attn: Lloyd Nelson |
| | 6000 Greenwood Plaza Blvd., Suite 11 |
| | Greenwood Village, CO, 80111 |
| | Tel.: 303-407-6884 |
| | Email: Inelson@ledtechnologies.com |
#### 2. General Information
- 2.1 Classification Name: Light Based Over-The-Counter Wrinkle Reduction device
- 2.2 Common/usual name: reVive® Perioral LED Light Therapy system
- 2.3 Proprietary Names: reVive® Perioral LED Light Therapy system
- 2.4 Classification: Class II
- 2.5 Classification Number: 878.4810
- 2.6 Product Code OHS, ILY
- 2.7 Review panel: General & Plastic Surgery
{4}------------------------------------------------
# 3. Device Description
The reVive® Perioral LED Light Therapy system is an over-the counter light emitting diode (LED) device that emits energy for use in dermatology for treatment fine lines and wrinkles, and increase in circulation within the perioral region. The device uses four types of LEDs: 605nm amber, 630nm red, 660nm red, and 880nm infrared. The treatment time is controlled by the user. There are no user settings or adjustments required.
The reVive® Perioral LED Light Therapy system components include the device containing the LED module, USB charging cord, and a storage bag.
The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components. The device is sold as Over The Counter (OTC).
### 4. Indications/Intended Use:
The reVive® Perioral LED Light Therapy system is an Over-the-Counter (OTC) device intended for the treatment fine lines and wrinkles, and increase in circulation within the perioral region.
### 5. Predicate Devices:
This device is substantially equivalent to the following predicate, which is currently legally marketed under product code OHS:
- K141181 dpl® Nuve for wrinkles (LED Technologies, Inc.) ●
- K171390 dpl® II Panel (LED Technologies, Inc.) ●
- K171386 dpl® SpectraLite (LED Tchnologies, Inc.) ●
# Predicate Chart
| Device | reVive® Perioral<br>LED Light Therapy<br>system<br>LED Technologies,<br>Inc.<br>KXXXXXXXXX | dpl® Nüve for<br>wrinkles<br>LED Technologies,<br>Inc.<br>K141181 | dpl® II Panel<br>LED Technologies,<br>Inc.<br>K171390 | dpl® SpectraLite<br>LED Technologies,<br>Inc.<br>K171386 |
|-------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------|----------------------------------------------------------|
| Wavelengths | 605nm, 630nm,<br>660nm, 880nm | 605nm, 630nm,<br>660nm, 880nm | 605nm, 630nm,<br>660nm, 880nm | 605nm, 630nm,<br>660nm, 880nm |
| Irradiance source | LED | LED | LED | LED |
| Treatment Area | 24 cm² | 30 cm² | 415 cm² | 28 cm² |
| Energy Level | 67.7 mW/cm² | 62.07 mW/cm² | 70.16 mW/cm² | 61.59 mW/cm² |
{5}------------------------------------------------
| Treatment Time | 3 minutes per treatment | 3 minutes per treatment | 3 minutes per treatment | 3 minutes per treatment |
|----------------|-------------------------|-------------------------|-------------------------|-------------------------|
| Type/Class | OTC | OTC | OTC | OTC |
#### Summary of the technological characteristics of the device compared to predicate device:
- 1. Has the same/similar intended use as the predicate device (i.e., treatment of wrinkles).
- 2. Has the same/similar output (i.e., 67.7 mW/cm²) as the predicate devices;
- 3. Utilizes the same number of wavelengths (i.e., 4 wavelengths between 605 nm 880 nm) as the predicate devices;
- 4. Utilizes the same treatment duration (i.e., 180 seconds) as the predicate devices;
- 5. Utilizes the same treatment regimen of five days a week for eight weeks.
The reVive® Perioral LED Light Therapy system and the above referenced predicate devices are Over the Counter Devices used to treat wrinkles as defined in 21 CFR § 878.4810. These devices utilize red and IR diodes between 605 nm to 880 nm to provide narrow bands of light energy to treat wrinkles. The performance achieved by these devices is similar with equal power output. Based upon comparison to the predicate devices, the reVive® Perioral LED Light Therapy system has the same intended uses, and similar technological characteristics as the predicate devices.
The devices are intended to be placed directly on the skin. The materials used for the device are similar to what is used in the dpl® Nüve & dpl® II Panel, and same as used in dpl® SpectraLite.
The reVive® Perioral LED Light Therapy system is battery powered (LI-Ion Battery 5V USB & 3.7 V Battery). Battery is charged via Universal USB charger cord. The change in energy (the addition of battery) is usually part of a redesign to provide a portable version of a predicate device. The indication for use does not change. Operation of device does not change. The system performs as intended and does not raise any new safety or effectiveness issues.
#### 6. Performance Testing and Standards:
Testing of the reVive® Perioral LED Light Therapy system, included functional performance testing, and software validation testing.
Safety and functionality testing demonstrates that the reVive® Perioral LED Light Therapy system conforms to various international consensus standards.
{6}------------------------------------------------
IEC 60601-1: 2012: Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance.
IEC 60601-1-2: 2014: Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance: Collateral Standard: Electromagnetic Compatibility.
UL 1642 Battery Certification UN 38.3 Battery Certification.
#### Biocompatibility
ISO 10993-5:2009 – Cytotoxicity Test ISO 10993-10:2010 – Intracutaneous reactivity test ISO 10993-10:2010 – Skin Sensitization Test
The reVive® Perioral LED Light Therapy system software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices".
A Usability Study was conducted with 15 participants.
The results of the study found that:
100% of the participants were able to demonstrate the light sensitivity test.
100% of the participants were able to use the device successfully.
The conclusions drawn from nonclinical tests demonstrate that the device is safe, as effective, and performs as well as the legally marketed devices.
#### 7. Substantial Equivalence Discussion
After analysis of, indications, intended uses, dose rates, performance, features, design materials, power output, technological properties, treatment areas, treatment regimens and methods of operation, the manufacturer asserts that no significant differences exist between the subject device and predicate, and no different questions of safety and effectiveness arise. Therefore, the subject device is substantial equivalence to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.