MEGASORBTM Absorbable Polyglycolic Acid Surgical Sutures are indicated for use in soft tissue approximation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological procedures. MITSUTM suture is intended for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological tissues. MITSU FST™ suture is indicated only for use in superficial general soft tissue approximation of the skin and mucosa, where only short term wound support (7-10 days) is required. MITSU FST™ suture is not intended for use in ligation, ophthalmic, cardiovascular or neurological procedures. FILAXYN™ sutures are indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardio vascular and neurological procedures. FILAPRON™ dyed / undyed sutures are intended for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
Device Story
Synthetic absorbable surgical sutures; used for soft tissue approximation and ligation; available in braided (Polyglycolic Acid, Polyglactin) or monofilament (Polydioxanone, Poly(glycolide-co-caprolactone)) forms; attached to stainless steel needles; used by surgeons in clinical settings; provides temporary wound support during healing; absorbed by body over time; non-pyrogenic; biocompatible; reduces tissue reaction.
Clinical Evidence
Bench testing only. Performance evaluated per USP requirements: diameter (USP <861>), tensile strength (USP <881>), needle attachment (USP <871>), resorption profile, and suture length. Biocompatibility testing included in vitro cytotoxicity, skin sensitization, intracutaneous reactivity, acute/sub-chronic systemic toxicity, intramuscular implantation, bacterial reverse mutation, mammalian erythrocyte micronucleus, in vitro hemolysis, and pyrogen testing.
Technological Characteristics
Synthetic absorbable polymers: Polyglycolic Acid, Poly(glycolide/l-lactide), Poly(p-dioxanone), Poly(glycolide-co-caprolactone). Braided sutures coated with polycaprolactone/calcium stearate. Monofilament options available. Sizes 6-0 to 8-0 (USP). Attached to stainless steel needles. Sterilized for single use. Non-pyrogenic. No software or electronic components.
Indications for Use
Indicated for soft tissue approximation and/or ligation in general, ophthalmic, or superficial skin/mucosa procedures depending on specific suture type. Contraindicated for cardiovascular and neurological procedures; specific exclusions apply for microsurgery and ophthalmic surgery based on suture variant.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure in profile, with three overlapping faces representing the department's focus on health and human well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 28, 2017
Meril Endo Surgery Private Limited % Mr. Umesh Sharma Deputy General Manager - Quality Assurance Third Floor, E1-E3, Meril Park Survey No. 135/2/B & 174/2 Muktanand Marg, Chala, Vapi, India, Valsad Gujarat, 396191, IN
Re: K172659
Trade/Device Name: MEGASORB - Natural absorbable Polyglycolic Acid Surgical Suture, MITSU - Absorbable Polyglactin 910 Surgical Sutures, MITSU FST -Absorbable Polyglactin 910 Surgical Suture, FILAXYN - Absorbable Polydioxanone Surgical (PDS), FILAPRON - Absorbable poly(glycolide-co-caprolactone) Surgical Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM, NEW Dated: August 28, 2017 Received: September 5, 2017
Dear Mr. Sharma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K172659
Device Name MEGASORB, MITSU, MITSU FST, FILAXYN, FILAPRON
Indications for Use (Describe)
MEGASORB Absorbable Polyglycolic Acid Surgical Sutures are indicated for use in soft tissue approximation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological procedures.
MITSU suture is intended for use in general soft tissue approximation including use in ophthalmic procedures, but not for use in cardiovascular and neurological tissues.
MITSU FST suture is indicated only for use in superficial general soft tissue approximation of the skin and mucosa, where only short term wound support (7-10 days) is required. MITSU FST suture is not intended for use in ligation, ophthalmic, cardiovascular or neurological procedures.
FILAXYN sutures are indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological procedures.
FILAPRON dyed / undyed sutures are intended for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
Type of Use (Select one or both, as applicable)
| 区 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|---------------------------------------------|
|------------------------------------------------|---------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K172659 Page 1 of 7
### 510(k) Summary
#### l. Submitter
M/s. Meril Endo Surgery Private Limited Third Floor, E1 – E3, Meril Park, Survey No. 135/2/B & 174/2, Muktanand Marg, Chala, Vapi – 396191 Gujarat, India. Tel. No: +91-260-3052100, Fax: +91-260-3052125 Web site: www.merillife.com E-mail: umesh.sharma@merillife.com
Contact Person: Umesh Sharma Date Prepared: August 26, 2017
### II. Device
| Sr.<br>No. | Trade / Proprietary<br>Name | Common<br>Name | Regulator<br>y Class | Product<br>Code | Regulation Number | Review<br>Panel |
|------------|----------------------------------------------------------------------------------|-----------------------------------------------|----------------------|-----------------|-------------------|---------------------------------|
| 1. | MEGASORBTM<br>absorbable Polyglycolic<br>Acid Surgical Suture | Braided coated<br>Polyglycolic<br>Acid Suture | II | GAM | 878.4493 | General &<br>Plastic<br>Surgery |
| 2. | MITSUTM Absorbable<br>poly(glycolide/l-lactide)<br>Surgical Sutures | Braided coated<br>Polyglactin<br>Suture | II | GAM | 878.4493 | General &<br>Plastic<br>Surgery |
| 3. | MITSU FSTTM<br>Absorbable poly<br>(glycolide / l-lactide) Surgical Sutures | Braided coated<br>Polyglactin<br>Suture | II | GAM | 878.4493 | General &<br>Plastic<br>Surgery |
| 4. | FILAXYNTM Absorbable<br>poly (p-dioxanone)<br>Surgical Suture | Monofilament<br>Polydioxanone<br>Suture | II | NEW | 878.4840 | General &<br>Plastic<br>Surgery |
| 5. | FILAPRONTM<br>Absorbable poly<br>(glycolide -co caprolactone) Surgical<br>Suture | Monofilament<br>Polyglecaprone<br>Suture | II | GAM | 878.4493 | General &<br>Plastic<br>Surgery |
{4}------------------------------------------------
### III. Predicate Device
| Sr. No. | Meril's Suture<br>Proprietary Name | Suture Trade /<br>Predicate Device Name | Predicate Device<br>510(k) No. |
|---------|------------------------------------------------------------------|-----------------------------------------|--------------------------------|
| 1. | MEGASORBTM<br>Polyglycolic Acid Surgical Suture | absorbable<br>MEGASORBTM | K150717 |
| 2. | MITSUTM<br>poly(glycolide/l-lactide) Surgical<br>Sutures | Absorbable<br>MITSUTM | K150717 |
| 3. | MITSU FSTTM<br>poly(glycolide/l-lactide) Surgical<br>Sutures | Absorbable<br>MITSU FSTTM | K150717 |
| 4. | FILAXYNTM Absorbable poly (p-<br>dioxanone) Surgical Suture | FILAPRONTM | K150717 |
| 5. | FILAPRONTM<br>poly(glycolide-co-caprolactone)<br>Surgical Suture | Absorbable<br>FILAXYNTM | K150717 |
### IV. Device Description
## A. MEGASORB™ absorbable Polyglycolic Acid Surgical Suture
MEGASORB™ suture is a synthetic absorbable sterile surgical suture composed of Polyglycolic Acid. Braided MEGASORB™ sutures are coated with polycaprolactone and calcium stearate. Poly Glycolic Acid coated with polycaprolactone and calcium stearate has been found to be non-pyrogenic and eliciting only slight tissue reaction during absorption. The empirical formula of the Polyglycolic Acid, polycaprolactone and calcium stearate is (C2H2O2)n, (C6H10O2)n and C36H70Q4Ca respectively.MEGASORB™ sutures are available either undyed or dyed (D and C Violet No. 2, Colour Index number: 60725) form. MEGASORB™ is available in a range of gauge sizes and lengths, attached to standard stainless steel needles of varying types and sizes.
{5}------------------------------------------------
MEGASORB™ suture is available in U.S.P. size range 8-0 to 2, undyed/dyed form in pack size of 6/12/24/36.
### B. MITSUT™ Absorbable poly(glycolide/l-lactide) Surgical Sutures
MITSUTM is a braided coated synthetic absorbable sterile poly(glycolide/llactide) surgical suture. It is composed of a copolymer made of glycolide and L-lactide. MITSUTM sutures are coated with a mixture containing equal parts of copolymer of glycolide and lactide and calcium stearate. The empirical formula of the copolymer is (C2H2O2)m (C3HzO2)n and calcium stearate is C36H704Ca. Poly(glycolide/l-lactide) copolymer and poly(glycolide/l-lactide) with calcium stearate exhibit non-pyrogenic properties. MITSU™ sutures are available in undyed and dyed (D and C violet No. 2, Colour Index No. 60725) form. Available in a broad range of suture sizes and lengths, MITSU™ comes with standard stainless steel needles of varying types and sizes.
MITSUTM suture is available in size range 8-0 to 2, undyed/dyed form, and pack size of 6/12/24/36.
### C. MITSU FST™ Absorbable poly(glycolide/l-lactide) Surgical Sutures
MITSU FST™ is a braided coated synthetic absorbable sterile poly(glycolide/llactide) surgical suture. It is composed of a copolymer made from glycolide and L-lactide. MITSU FST™ sutures are coated with a mixture containing equal parts of copolymer of glycolide and calcium stearate. The empirical formula of the copolymer is (C2H2O2)m (C3H4O2)n and calcium stearate is C36H704Ca. The rapid loss of strength is achieved by using polymer material with lower molecular weight than that of regular MITSUTM suture. Poly(glycolide/l-lactide) copolymer and poly(glycolide/l-lactide) with calcium stearate exhibit non-pyrogenic properties and elicit only slight tissue reaction during absorption. MITSU FST™ sutures are available in undyed form. Available in a broad range of suture sizes and lengths, MITSU FST™ comes with standard stainless steel needles of varying types and sizes.
{6}------------------------------------------------
MITSU FST™ suture is available in U.S.P. size range 6-0 to 1, undyed form in pack size of 6/12/24/36.
### D. FILAXYN™ Absorbable poly (p-dioxanone) Surgical Suture
FILAXYN™ is a sterile synthetic absorbable monofilament suture composed of Poly (p-dioxanone). The empirical molecular formula of the polymer is (C4HgO3), Poly (p-dioxanone) polymer has non-pyrogenic properties. FILAXYN™ sutures are available in undyed or dyed (D and C Violet No.2, Color Index number: 60725) form. FILAXYN™ is available in a range of gauge sizes and lengths, attached to standard stainless steel needles of varying types and sizes.
FILAXYN™ suture is available in size range 7-0 to 2, undyed/dyed form in pack size of 6/12/24/36.
## E. FILAPRON™ Absorbable poly(glycolide-co-caprolactone) Surgical Suture
FILAPRON™ is a sterile synthetic absorbable monofilament suture is composed of poly(glycolide-co-caprolactone). The empirical molecular formula of the polymer is (C2H2O2)m (C6H10O2)n poly(glycolide-cocaprolactone) has non-pyrogenic properties. FILAPRON™ sutures are available either in undyed or dyed (D and C violet No. 2, Colour Index No. 60725) form. FILAPRON™ is available in a range of gauge sizes and lengths and attached to standard stainless steel needles of varying types and sizes. FILAPRON™ suture is available in size range 6-0 to 2, undyed/dyed form in pack size of 6/12/24/36.
{7}------------------------------------------------
### V. Intended Use
- 1. MEGASORBTM Absorbable Polyglycolic Acid Surgical Sutures are indicated for use in soft tissue approximation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological procedures.
- 2. MITSUTM suture is intended for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neurological tissues.
- 3. MITSU FST™ suture is indicated only for use in superficial general soft tissue approximation of the skin and mucosa, where only short term wound support (7-10 days) is required. MITSU FST™ suture is not intended for use in ligation, ophthalmic, cardiovascular or neurological procedures.
- 4. FILAXYN™ sutures are indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardio vascular and neurological procedures.
- 5. FILAPRON™ dyed / undyed sutures are intended for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
{8}------------------------------------------------
### VI. Substantial Equivalence
The device design, material of construction, performance, packaging and intended uses are similar to the predicate device. Substantial equivalence is conducted based on the following parameters:
- 1. Product description
- 2. Intended use
- 3. Suture Size
- 4. Colorant Used
- 5. Suture Coating
- 6. Single use
- 7. Sterilisation method
- 8. Packaging
- 9. Label Claim
- 10. Performance
- a. Diameter USP <861>
- b. Tensile strength USP <88|>
- Needle attachment USP <871> ﻥ ﻧ
- d. Resorption profile
- e. Suture Length
11. Labelling and Instructions for use (IFU)
### VII. Performance Data
The Surgical Suture was subjected to the performance testing as per USP requirements. The safety and effectiveness of the Surgical Suture has been evaluated for the following performance and safety requirements.
- 1. Diameter USP <861>
- 2. Tensile strength USP <881>
- 3. Needle attachment USP <871>
- 4. Resorption Profile
- 5. Suture Length
- 6. Biocompatibility
- a. In Vitro Cytotoxicity Study
{9}------------------------------------------------
- b. Skin Sensitization Study
- c. Intracutaneous Reactivity Test
- d. Acute Systemic Toxicity Study
- e. Sub Chronic Toxicity Study
- Intramuscular Implantation Test f.
- Bacterial Reverse Mutation Test g.
- h. Mammalian Erythrocyte Micronucleus Test
- In Vitro Hemolysis Test İ.
- j. Pyrogen Test
#### Conclusion VIII.
Meril's Absorbable surgical Sutures are composed of the same materials, as are the predicate devices and has the same design, as do the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.