# FMP Extended Liners (K172651)

_Encore Medical L.P. · OQG · Dec 13, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172651

## Device Facts

- **Applicant:** Encore Medical L.P.
- **Product Code:** [OQG](/productcode/OQG.md)
- **Decision Date:** Dec 13, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; - rheumatoid arthritis; - correction of functional deformity; - femoral fracture. This device may also be indicated in the salvage of previously failed surgical attempts. This device is to be used for uncemented applications.

## Device Story

FMP Extended Liners are acetabular liners designed as a line extension to the X-alt Highly Cross Linked Acetabular Liner with Vitamin E system. The device expands femoral head compatibility by allowing larger femoral heads to be used with smaller acetabular cups. Used in orthopedic hip replacement surgery; implanted by surgeons in a clinical/hospital setting. The device functions as a bearing surface within a hip prosthesis system. It is intended for uncemented applications. The device is comparable to predicate liners in design, material, and surgical technique, with modifications to collar height and liner thickness to accommodate larger femoral heads. Benefits include improved surgical flexibility for surgeons to address patient-specific anatomical needs during hip arthroplasty.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing: wear (ISO 14242-2/3), impingement (ASTM F2582-14), push-out, lever-out, torsion (ASTM F1820-13), and range of motion (ISO 21535:2009). Bacterial endotoxin testing met USP <161> limits.

## Technological Characteristics

Acetabular liner for hip prosthesis. Materials and design are consistent with predicate X-alt liners. Testing standards: ISO 14242-2/3 (wear), ASTM F2582-14 (impingement), ASTM F1820-13 (torsion/push-out/lever-out), ISO 21535:2009 (range of motion). Bacterial endotoxin levels monitored per USP <161>.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- X-alt Highly Cross Linked Acetabular Liner with Vitamin E ([K130365](/device/K130365.md))
- Stryker Trident Large Diameter Acetabular Insert ([K062419](/device/K062419.md))

## Reference Devices

- Foundation® Porous Coated Hemispherical Acetabular Cup ([K974093](/device/K974093.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 13, 2017

Encore Medical, L.P. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Blvd. Austin, Texas 78758

Re: K172651

Trade/Device Name: FMPTM Extended Liners Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: OQG, LPH Dated: November 28, 2017 Received: November 29, 2017

Dear Teffany Hutto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

# Page 2 - Teffany Hutto

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| 510(k) Number (if known) | K172651               |
|--------------------------|-----------------------|
| Device Name              | FMPT™ Extended Liners |

**Indications for Use (Describe)**

Joint replacement is indicated for patients suffering from disability due to:

- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head;
- rheumatoid arthritis;
- correction of functional deformity;
- femoral fracture.

This device may also be indicated in the salvage of previously failed surgical attempts.

This device is to be used for uncemented applications.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
|-----------------------------------------------------------------------------------------------|---------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

င္ခင္မွ ခ်စ္ေ႕

{3}------------------------------------------------

# 510(k) Summary

Date: September 1 , 2017

Manufacturer: DJO Surgical (Legal Name: Encore Medical, L.P.) 9800 Metric Blvd Austin, TX 78758

Contact Person: Teffany Hutto Manager, Regulatory Affairs Phone: (512) 834-6255 Fax: (512) 834-6313 Email: teffany.hutto@djoglobal.com

| Product                | Classification | Product Code                                                                                                                                                                                                                                      |
|------------------------|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FMP Extended<br>Liners | Class II       | LPH - Hip joint metal/polymer/metal semi-constrained porous-<br>coated uncemented prosthesis per 21 CFR 888.3358<br><br>OQG - Hip prosthesis, semi-constrained, cemented,<br>metal/polymer, + additive, porous, uncemented per 21 CFR<br>888.3358 |

#### Description:

This Traditional 510(k) is a line extension to the X-alt Highly Cross Linked Acetabular Liner with Vitamin E product line to expand femoral head compatibility within the system by making larger femoral heads compatible with smaller acetabular cups.

#### Indications for Use:

Joint replacement is indicated for patients suffering from disability due to:

- noninflammatory degenerative joint disease including osteoarthritis and avascular ● necrosis of the natural femoral head;
- rheumatoid arthritis;
- correction of functional deformity; ●
- . femoral fracture.

This device may also be indicated in the salvage of previously failed surgical attempts.

This device is to be used for uncemented applications.

#### Predicate Devices:

| Device                                                       | Manufacturer         | 510(k) Number | 510(k) Type |
|--------------------------------------------------------------|----------------------|---------------|-------------|
| X-alt Highly Cross Linked<br>Acetabular Liner with Vitamin E | Encore Medical, L.P. | K130365       | Primary     |
| Stryker Trident Large Diameter<br>Acetabular Insert          | Stryker              | K062419       | Additional  |
| Foundation® Porous Coated<br>Hemispherical Acetabular Cup    | Encore Medical, L.P. | K974093       | Reference   |

{4}------------------------------------------------

# Comparable Features to Predicate Device(s):

These components are comparable to the predicate devices in indications, material, design features, manufacturing methods, surgical implantation technique, intended use, packaging, and sterilization.

### Key Differences in Subject Device to Predicate:

Collar height and liner thickness compared to K130365.

Bacterial endotoxin testing conducted based on DJO Surgical's Endotoxin Assessment: alternative to batch testing plan met the bacterial endotoxin limits defined in USP <161> as evidenced through ongoing monitoring conducted concurrently with dose audits. Testing conducted with the most recent dose audit using the worst case product demonstrated an acceptable BET level. The subject FMP Extended Liners require fewer manufacturing operations and less handling than the worst case product (FMP Shell). Refer to NDPCAW-00085 for new product review.

Non-Clinical Testing: Testing performed includes wear per ISO 14242-2/3, impingement per ASTM F2582-14, push-out, lever-out, torsion per ASTM F1820-13, and range of motion per ISO 21535:2009. All evaluations determined that the devices are substantially equivalent to the applicable predicate devices.

Clinical Testing: Clinical testing is not required.

---

**Source:** [https://fda-staging.innolitics.com/device/K172651](https://fda-staging.innolitics.com/device/K172651)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).