The High-Flo Silver Polyethylene Catheter is intended for the delivery of contrast media and therapeutic agents to the peripheral, carotid, and coronary vasculature, not including the neurovasculature.
Device Story
Sterile, single-use angiographic catheter; 5.5 French size; 65-100 cm lengths. Features luer lock adapter, connecting cap, and single lumen braided shaft. Used by clinicians in angiographic procedures to deliver contrast media or therapeutic agents to peripheral, carotid, and coronary vasculature. Operates via manual insertion into the vascular system. Output is the targeted delivery of fluids/agents, facilitating diagnostic imaging or therapy. Benefits include precise agent delivery to specific vascular sites.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, hemocompatibility, complement activation, thrombogenicity, PTT) met ISO 10993-1:2009(R)2013 criteria. Mechanical testing (tensile strength of hub-to-shaft and tip-to-shaft bonds, liquid/air leakage, static burst pressure, dimensional verification, and hub pressure) met BS EN ISO 10555-1:2013 standards.
Technological Characteristics
Single lumen braided polyethylene catheter; 5.5 French; 65-100 cm length. Includes luer lock adapter and connecting cap. Tested per BS EN ISO 10555-1:2013 (Annexes B, C, D, F) for mechanical integrity and leakage. Biocompatibility per ISO 10993-1:2009(R)2013. Sterile, single-use.
Indications for Use
Indicated for delivery of contrast media and therapeutic agents to peripheral, carotid, and coronary vasculature; excludes neurovasculature.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 3, 2018
Cook Incorporated Jessica Swafford Regulatory Affairs Specialist 750 Daniels Wav Bloomington, Indiana 47402
Re: K172635
Trade/Device Name: High-Flo Silver Polyethylene Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO Dated: March 5, 2018 Received: March 6, 2018
Dear Jessica Swafford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K172635
### Device Name High-Flo Silver Polyethylene Catheter
### Indications for Use (Describe)
The High-Flo Silver Polyethylene Catheter is intended for the delivery of contrast media and therapeutic agents to the peripheral, carotid, and coronary vasculature, not including the neurovasculature.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, block letters in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" in white, block letters.
COOK INCORPORATED OOMINGTON. IN 47404 USA N.COOKME
# 510(k) Summary -- K172635
# High-Flo Silver Polyethylene Catheter Traditional 510(k) Summary 21 CFR §807.92
## Submitter Information
| Applicant: | Cook Incorporated |
|-----------------------|------------------------------------------|
| Address: | 750 Daniels Way<br>Bloomington, IN 47404 |
| Contact: | Jessica P. Swafford |
| Email: | RegSubmissions@cookmedical.com |
| Contact Phone Number: | 812-335-3575 ext. 104260 |
| Contact Fax Number: | 812-332-0281 |
| Date Prepared: | 05 March 2018 |
| Device Information | |
| Trade Name: | High-Flo Silver Polyethylene Catheter |
| Common Name: | Angiographic Catheters |
### Predicate Device
Classification Name:
The predicate device of the subject High-Flo Silver Polyethylene Catheter is the Slip-Cath® Beacon® Tip Catheter cleared under 510(k) number K122937. The predicate Slip-Cath® Beacon® Tip Catheter is visually identified by a distal radiopaque tip bonded onto a stainless steel braided catheter shaft. The predicate device is manufactured in lengths of 60 to 150 centimeters and in sizes of 4.0 to 6.5 French. The shaft of these catheters have an inner lumen that tapers to a 0.035 or 0.038 inch endhole diameter and are manufactured in a variety of distal tip configurations.
Catheter, Intravascular, Diagnostic
DQO (21 CFR §870.1200)
{4}------------------------------------------------
## Comparison to Predicate(s)
It has been demonstrated that the subject High-Flo Silver Polyethylene Catheter and the predicate device are substantially equivalent in terms of intended use, duration of use, principles of operation, fundamental technological characteristics, and insertion method. The design, dimensions, manufacture, and materials of the subject device are either similar to the materials of the predicate device(s) or have been used in other cleared devices. The differences between the subject device and the predicate device(s) do not raise new questions of safety and effectiveness as demonstrated by performance and biocompatibility testing.
## Device Description
The High-Flo Silver Polyethylene Catheter, subject of this submission, is a sterile, single use device designed for use in angiographic procedures. The High-Flo Silver Polyethylene Catheter is available in a 5.5 French size and is manufactured in lengths of 65 to 100 centimeters. Each configuration includes a luer lock adapter, connecting cap, and a single lumen braided shaft.
## Intended Use
The High-Flo Silver Polyethylene Catheter is intended for the delivery of contrast media and therapeutic agents to the peripheral, carotid, and coronary vasculature, not including the neurovasculature.
## Test Data
The High-Flo Silver Polyethylene Catheter, subject of this submission, was subjected to applicable testing to assure reliable design and performance under the testing parameters. The following tests were conducted to ensure reliable design and performance:
- . Biocompatibility testing - Testing (i.e., cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, hemocompatibility, complement activation, in vivo thrombogenicity, and partial thromboplastin time) demonstrated that the device is biocompatible for the intended use. In conformance with the applicable sections of ANSI AAMI ISO 10993-1:2009(R)2013, the predetermined acceptance criteria were met.
{5}------------------------------------------------
- . Tensile Testing of the Hub-to-Shaft Bond - Testing verified that under proper clinical use of the catheter, the peak load value of the hub-to-shaft connection is in accordance with the methods of BS EN ISO 10555-1:2013. Annex B. The predetermined acceptance criterion was met.
- Tensile Testing of the Tip-to-Shaft Bond Testing verified that under proper . clinical use of the catheter, the peak load value of the tip-to-shaft connection is in accordance with the methods of BS EN ISO 10555-1:2013, Annex B. The predetermined acceptance criterion was met.
- Liquid Leakage Testing Testing verified that under proper clinical use of the . catheter, there will be no liquid leakage when tested in accordance with BS EN ISO 10555-1:2013, Annex C. The predetermined acceptance criterion was met.
- Air Leakage Testing Testing verified that under proper clinical use of the . catheter. there will be no air leakage when tested in accordance with BS EN ISO 10555-1:2013, Annex D. The predetermined acceptance criterion was met.
- Static Burst Testing Testing successfully characterized the catastrophic failure . pressure for the catheter is accordance with BS EN ISO 10555-1:2013. Annex F.
- . Dimensional Verification Testing - Testing verified that the dimensional requirements of the subject device are within a specified tolerance. The predetermined acceptance criteria were met.
- Hub Pressure Testing Testing successfully characterized the hub pressure, when . tested at maximum flow rate, and to verify that it does not exceed the static burst pressure.
In conclusion, the results of these tests support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.