← Product Code [JTN](/productcode/JTN) · K172621

# HardyDisk Meropenem/Vaborbactam 20/10µg (MEV30) (K172621)

_Hardy Diagnostics · JTN · Oct 3, 2017 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172621

## Device Facts

- **Applicant:** Hardy Diagnostics
- **Product Code:** [JTN](/productcode/JTN.md)
- **Decision Date:** Oct 3, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1620
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

Use of HardyDisk™ Meropenem/Vaborbactam 20/10ug (MEV30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of bacteria to Meropenem/Vaborbactam. Meropenem/Vaborbactam has been shown to be active against susceptible isolates of the following microorganisms both in vitro and in clinical infections: Enterobacter cloacae species complex Escherichia coli Klebsiella pneumoniae Meropenem/Vaborbactam has been shown to be active in vitro against susceptible isolates of the following microorganisms: Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Klebsiella oxytoca Morganella morganii Proteus mirabilis Providencia spp. Pseudomonas aeruginosa Serratia marcescens. HardyDisk™ AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae.

## Device Story

HardyDisk Meropenem/Vaborbactam 20/10ug (MEV30) is an antimicrobial susceptibility test (AST) disk used for semi-quantitative in vitro testing via the Kirby-Bauer agar diffusion method. The device consists of paper disks impregnated with the antibiotic combination Meropenem/Vaborbactam. In a clinical laboratory setting, the disk is placed on an agar plate inoculated with a bacterial isolate. As the antibiotic diffuses into the agar, it creates a concentration gradient. After incubation, the zone of inhibition around the disk is measured. The size of this zone correlates with the susceptibility of the bacteria to the antibiotic. Healthcare providers use these results to guide antibiotic therapy decisions for patients with infections caused by susceptible organisms, such as Enterobacter cloacae, E. coli, and K. pneumoniae. The device aids in selecting appropriate antimicrobial treatment, potentially improving patient outcomes by ensuring targeted therapy.

## Clinical Evidence

No clinical data. Performance characteristics rely on breakpoints and quality control ranges established by CDER for the pharmaceutical drug.

## Technological Characteristics

Paper-based AST disk impregnated with 20/10ug Meropenem/Vaborbactam. Operates via passive diffusion in agar medium. Class II device, product code JTN. No software or electronic components.

## Regulatory Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
K172621

B. Purpose for Submission:
Addition of Meropenem/Vaborbactam Antimicrobial Susceptibility Test Disk

C. Measurand:
Meropenem/Vaborbactam, 20/10 µg

D. Type of Test:
Antimicrobial Susceptibility Test Disks

E. Applicant:
Hardy Diagnostics

F. Proprietary and Established Names:
HardyDisk Meropenem/Vaborbactam 20/10 µg (MEV30)

G. Regulatory Information:

1. Regulation section:
21 CFR 866.1620 Antimicrobial Susceptibility Test Disc

2. Classification:
Class II

3. Product code:
JTN – Susceptibility Test Disc, Antimicrobial

4. Panel:
83, Microbiology

{1}

H. Intended Use:

1. Intended use(s):

HardyDisk AST Disks are used for semi-quantitative *in vitro* susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for *Enterobacteriaceae*, *Staphylococcus* spp., *Pseudomonas* spp., *Acinetobacter* spp., *Listeria monocytogenes*, *Enterococcus* spp., and by modified procedures, *Haemophilus* spp., *Neisseria gonorrhoeae*, *N. meningitidis* and *Streptococcus* spp., including *Streptococcus pneumoniae*.

2. Indication(s) for use:

Use of HardyDisk Meropenem/Vaborbactam 20/10 µg (MEV30) for *in vitro* agar diffusion susceptibility testing is indicated when there is need to determine the susceptibility of bacteria to Meropenem/Vaborbactam.

Meropenem/Vaborbactam has been shown to be active against susceptible isolates of the following microorganisms both *in vitro* and in clinical infections:

- *Enterobacter cloacae* species complex
- *Escherichia coli*
- *Klebsiella pneumoniae*

Meropenem/Vaborbactam has been shown to be active *in vitro* against susceptible isolates of the following microorganisms:

- *Citrobacter freundii*
- *Citrobacter koseri*
- *Enterobacter aerogenes*
- *Klebsiella oxytoca*
- *Morganella morganii*
- *Proteus mirabilis*
- *Providencia* spp.
- *Pseudomonas aeruginosa*
- *Serratia marcescens*

HardyDisk AST Disks are used for semi-quantitative *in vitro* susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for *Enterobacteriaceae*, *Staphylococcus* spp., *Pseudomonas* spp., *Acinetobacter* spp., *Listeria monocytogenes*, *Enterococcus* spp., and by modified procedures, *Haemophilus* spp., *Neisseria gonorrhoeae*, *N. meningitidis* and *Streptococcus* spp., including *Streptococcus pneumoniae*.

2

{2}

3. Special conditions for use statement(s):

For prescription use only.

The following footnote was added to the labeling:

K. pneumoniae ATCC 1705 and K. pneumoniae ATCC BAA-2814 are included for the QC of the vaborbactam activity. Routine testing of meropenem/vaborbactam should include P. aeruginosa ATCC 27853 and K. pneumoniae ATCC BAA-2814.

4. Special instrument requirements:

N/A

I. Device Description:

HardyDisk AST Disks utilize 6-mm diameter white filter paper disks. The disks are prepared by impregnating absorbent paper with a known concentration of 20 µg Meropenem and 10 µg Vaborbactam. The disks are marked with the code MEV30 on both sides.

HardyDisk AST Disks are supplied in plastic cartridges containing 50 disks each. They are also packaged as one cartridge per vial with desiccant or five cartridges per vial with desiccant.

J. Substantial Equivalence Information:

1. Predicate device name(s):

HardyDisk Tigecycline 15 µg

2. Predicate 510(k) number(s):

K062245

3. Comparison with predicate:

Table 1. Comparison with the Predicate

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device K172621 HardyDisk Meropenem/Vaborbactam 20/10 µg | Predicate K062245 HardyDisk Tigecycline 15 µg  |
|  Test Method | Antimicrobial Susceptibility Testing using paper disks impregnated with an antimicrobial agent | Same  |

{3}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device K172621 HardyDisk Meropenem/Vaborbactam 20/10 μg | Predicate K062245 HardyDisk Tigecycline 15 μg  |
|  Intended Use | Antimicrobial susceptibility Test Disks are used for in vitro susceptibility test by standardized agar diffusion test procedures | Same  |
|  Methodology | Kirby-Bauer Disk Diffusion Susceptibility Test Protocol requires the user to determine categorical interpretations (S/I/R) using the measured zone diameters. | Same  |
|  Inoculum | Prepared from pure isolated colonies to match the turbidity equivalent of a 0.5 McFarland in Tryptic Soy Broth. | Same  |
|  Inoculation Method | Dip a sterile swab into the prepared inoculum and streak an appropriate agar plate's surface three times. Add the disks impregnated with the antimicrobial agent to the surface of the plate. Incubate the agar plate agar side up in a 35 ± 2°C incubator for 16-18 hours. | Same  |
|  Reading Method | The user will interpret the zone diameters according established interpretive criteria for the drug. | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Product Name | HardyDisk Meropenem/Vaborbactam 20/10 μg (MEV30) | HardyDisk Tigecycline  |
|  Antimicrobial Agent | Meropenem/Vaborbactam | Tigecycline  |
|  Concentration | 20/10 μg | 15 μg  |

# K. Standard/Guidance Document Referenced (if applicable):

CLSI M02-A12, Performance Standards for Antimicrobial Disk Susceptibility Tests; Approved Standard- Twelfth Edition

CLSI M100-S27, Performance Standards for Antimicrobial Susceptibility Testing; Twenty-Seventh Informational Supplement

{4}

L. Test Principle:

The HardyDisk AST Disk is based on the agar diffusion (Kirby-Bauer) methodology. It utilizes dried filter paper disks impregnated with a known concentration of an antimicrobial agent that are placed onto the test medium surface. Mueller Hinton agar is recommended for agar diffusion testing of non-fastidious organisms and Mueller Hinton with 5% Sheep Blood for Streptococcus spp. Three to five similar colonies are transferred to 4-5 mL of a suitable broth medium. The broth is incubated at 35°C for 2-6 hours to develop a turbidity that exceeds or is equivalent to a 0.5 McFarland standard. Alternatively, a direct broth or saline suspension of colonies may be prepared from an overnight culture. The final inoculum density should be equivalent to a 0.5 McFarland turbidity standard. The inoculum density may also be standardized photometrically.

Within 15 minutes of inoculum preparation, the Mueller Hinton agar plate is streaked with an inoculated swab to obtain an even inoculation of organism. Disks are aseptically placed onto the agar surface with a disk dispenser and the disks are pressed down with a sterile needle or forceps to make contact with the agar surface. Agar plates are incubated in an ambient air incubator at 35±2°C for 16 - 18 hours. Fastidious organisms are tested using appropriate media incubated in an atmosphere enriched with 5% CO₂, as recommended in the CLSI M02 approved standard document.

After incubation the agar medium is examined for zone of inhibition around the disks. The zones of inhibition are measured to the nearest millimeter and compared against recognized zone size ranges for the antimicrobial agent being tested.

M. Performance Characteristics (if/when applicable):

Descriptive characteristics were sufficient for this disk because the drug studies evaluated by CDER generated the breakpoints and quality control (QC) expected ranges used for review of this device.

1. Analytical performance:

a. Precision/Reproducibility:
Not applicable

b. Linearity/assay reportable range:
Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable

{5}

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
The interpretive criteria, QC isolates and expected ranges are the same as recommended by FDA/CDER in the approved pharmaceutical drug package insert. The interpretive criteria are shown in Table 2.

{6}

7

Table 2. FDA Breakpoints for Meropenem/Vaborbactam

|  Indications For Use Organism(s) | Interpretative Criteria  |   |   |
| --- | --- | --- | --- |
|   |  Zone Diameter (mm)  |   |   |
|   |  R | I | S  |
|  Enterobacteriaceae | ≤13 | 14-16 | ≥17  |

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K172621](https://fda-staging.innolitics.com/device/K172621)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com