Fixed Patient Leads ECG Cable (Model 3153)

K172591 · Boston Scientific Corporation · DSA · Nov 3, 2017 · Cardiovascular

Device Facts

Record IDK172591
Device NameFixed Patient Leads ECG Cable (Model 3153)
ApplicantBoston Scientific Corporation
Product CodeDSA · Cardiovascular
Decision DateNov 3, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2900
Device ClassClass 2

Indications for Use

The Fixed Patient Leads ECG Cable is used to transmit signals from patient-connected electrodes or transducers to electrocardiograph recorders/monitors for both diagnostic and monitoring purposes. ECG cables are supplied non-sterilizable; they are reusable devices and are intended for use by trained operators in a medical environment. Usage is limited to the indications for use of the connected monitoring or dagnostic equipment.

Device Story

Fixed Patient Leads ECG Cable (Model 3153) transmits electrical ECG signals from patient-affixed electrodes to BSC programming system display monitors; device consists of 5 leadwires with clips at one end and a trunk cable with a 6-pin DIN AAMI connector at the other; leadwires are permanently molded into the trunk cable. Used in medical environments by trained operators; facilitates diagnostic and monitoring functions by providing signal path between patient and external recording equipment. Benefits include reliable signal transmission for cardiac monitoring/diagnostics. Device is reusable and non-sterile.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including component qualification, electrical device verification (EMI/EMC, electrical safety), biocompatibility assessment (ISO 10993-5, ISO 10993-10), and packaging assessment.

Technological Characteristics

5-leadwire ECG cable; 6-pin DIN AAMI connector; fixed leadwire-to-trunk joint. Materials are biocompatible per ISO 10993-1. Standards: ANSI/AAMI EC53:2013, IEC 60601-1:2005, IEC 60601-1-2:2014. Non-sterile, reusable. Passive electrical transmission device.

Indications for Use

Indicated for patients requiring ECG signal transmission from electrodes or transducers to diagnostic or monitoring equipment. Intended for use by trained operators in medical environments.

Regulatory Classification

Identification

A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 3, 2017 Boston Scientific Corporation Elizabeth Lisowski Principal Regulatory Affairs Specialist 4100 Hamline Avenue North St. Paul, Minnesota 55112 Re: K172591 Trade/Device Name: Fixed Patient Leads ECG Cable (Model 3153) Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer and Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: October 4, 2017 Received: October 5, 2017 Dear Elizabeth Lisowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, M.A. Willemen for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K172591 Device Name Fixed Patient Leads ECG Cable (Model 3153) Indications for Use (Describe) The Fixed Patient Leads ECG Cable is used to transmit signals from patient-connected electrodes or transducers to electrocardiograph recorders/monitors for both diagnostic and monitoring purposes. ECG cables are supplied non-sterilizable; they are reusable devices and are intended for use by trained operators in a medical environment. Usage is limited to the indications for use of the connected monitoring or dagnostic equipment. | Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------------| | <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # 1. GENERAL INFORMATION | Device Common / Usual<br>Generic Name: | ECG Cable | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name: | Fixed Patient Leads ECG Cable (Model 3153) | | Applicant's Name and<br>Address: | Boston Scientific Corporation<br>4100 Hamline Avenue North<br>St. Paul, Minnesota 55112-5798<br>FDA Registration Establishment No. 2124215 | | Product Code /<br>Regulation Citation: | DSA, Cardiovascular / 21 CFR Part §870.2900<br>The ECG Cable is a Class II device as defined in 21 CFR Part<br>§870.2900 and has a product code of DSA<br>Panel: 74 Cardiovascular | | Contact Name and<br>Information: | Elizabeth A. Lisowski<br>Principal Regulatory Specialist<br>Phone: 651-532-5131<br>Fax: 651-582-5134<br>E-mail: beth.lisowski@bsci.com | | Owner / Operator: | Boston Scientific Corporation<br>One Boston Scientific Place<br>Natick, Massachusetts 01760<br>FDA Registration Establishment No: 9912058 | | Manufacturing Facility: | Affinity Medical Technologies<br>3545 Harbor Boulevard<br>Costa Mesa, California 92626<br>FDA Registration Establishment No: 2031044 | #### 2. DEVICE Fixed Patient Leads ECG Cable (Model 3153) #### 3. PREDICATE DEVICE Removable Patient Leads ECG Cable (K132253; cleared December 16, 2013) #### DEVICE DESCRIPTION 4. The BSC Model 3153 ECG Cable is an external device used to transmit ECG signals from electrodes that are affixed to the patient's body to an external BSC programming system display monitor. One end of each leadwire clip is attached to an ECG patient electrode; the other end is affixed / molded into one end of the trunk cable. The trunk cable then plugs into the BSC programming system. K172591 Pg. 1 of 4 {4}------------------------------------------------ # 5. INDICATIONS FOR USE The Fixed Patient Leads ECG Cable is used to transmit signals from patient-connected electrodes or transducers to electrocardiograph recorders / monitors for both diagnostic and monitoring purposes. ECG cables are supplied non-sterile and are not sterilizable; they are reusable devices and are intended for use by trained operators in a medical environment. Usage is limited to the indications for use of the connected monitoring or diagnostic equipment. ### DEVICE COMPARISON TO PREDICATE 6. The BSC ECG cable incorporates substantially equivalent design, packaging, fundamental technology, and Indications for Use as the predicate ECG Cable (Removable Patient Leads ECG Cable; K132253). Refer to Table 1 for a comparison between the BSC Fixed Patient Leads ECG Cable and the BSC Predicate Removable Patient Leads ECG Cable. | Characteristic | Proposed BSC Fixed Patient Leads<br>ECG Cable | Predicate BSC Removable Patient<br>Leads ECG Cable | |----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | ECG Cable is used to transmit signals<br>from patient-connected electrodes or<br>transducers to electrocardiograph<br>recorders / monitors for both<br>diagnostic and monitoring purposes. | ECG Cable is used to transmit signals<br>from patient-connected electrodes or<br>transducers to electrocardiograph<br>recorders / monitors for both diagnostic<br>and monitoring purposes. | | Patient Usage | Reusable | Reusable | | Sterility | Supplied non-sterile; cannot be<br>sterilized | Supplied non-sterile; cannot be<br>sterilized | | Anatomical Sites | Attached to electrodes placed at<br>standard specified locations on chest<br>wall and extremities | Attached to electrodes placed at<br>standard specified locations on chest<br>wall and extremities | | Design / Appearance / Functionality | | | | Functionality | Transmits signals from 5 patient-<br>connected leadwires | Transmits signals from 5 patient-<br>connected leadwires | | No. of Leadwires<br>(Patient Wire) | 5 | 5 | | Leadwire / Trunk<br>Cable Joint | Fixed | Removable | | Trunk Cable Length<br>(from edge of yoke to<br>edge of patient clip) | $80\pm2"$ | $120\pm6"$ | | Leadwire Length<br>(each) | $30\pm2"$ | $40\pm1.5"$ | | Connector - Patient<br>End | Patient Clip | Patient Clip | | Connector - Monitor<br>End | 6 pin DIN AAMI | 6 pin DIN AAMI | | Table 1: Comparison between BSC Fixed Patient Leads ECG Cable and BSC Predicate | | | |---------------------------------------------------------------------------------|--|--| | Removable Patient Leads ECG Cable. | | | {5}------------------------------------------------ | Characteristic | Proposed BSC Fixed Patient Leads<br>ECG Cable | Predicate BSC Removable Patient<br>Leads ECG Cable | |-------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------| | Materials | Biocompatible per 10993-1 | Biocompatible per 10993-1 | | Leadwire Colors<br>(IEC 60601-2-27) | White    Red<br>Green   Black<br>Brown | White    Red<br>Green   Black<br>Brown | | Packaging | Outer labeled box<br>Inner labeled bubble pouch<br>Non-sterile | Outer labeled box<br>Inner labeled bubble pouch<br>Non-sterile | # 7. PERFORMANCE DATA Component Qualification, Electrical Device Verification Testing (DVT) (including Electromagnetic Immunity / Electromagnetic Compatibility (EMI / EMC), Electrical Safety, and general functional testing), Biocompatibility Assessment, and a Packaging Assessment were performed to support a determination of substantial equivalence. Design Verification Testing conducted to support this submission demonstrated that all requirements were met. Component Qualification verified the ability for Affinity to provide ECG Cables that meet BSC mechanical and electrical specifications, as well as functionality / performance requirements. Electrical DVT verified that the Model 3300 Programmer and associated cables, including the proposed ECG Cable, meet system electrical design specifications. Design Validation Testing conducted on the proposed ECG Cable demonstrated that the cable meets user needs and its intended use. Results of Biocompatibility Testing demonstrated that the ECG Cable is biocompatible for its intended use. Design Analysis and Evaluation Testing evaluated device robustness and correct implementation of the design intent. Results of the Packaging Assessment demonstrated that Affinity-manufactured ECG Cable box labels meet adhesion, legibility, and indelibility requirements. Standards applicable to the proposed ECG Cable: - ANSI/AAMI EC53:2013: ECG trunk cables and patient leadwires ● - IEC 60601-1:2005: Medical electrical equipment Part 1: General requirements for ● basic safety and essential performance - IEC 60601-1-2:2014: Medical electrical equipment Part 1-2: General requirements . for safety - Collateral standard: Electromagnetic compatibility - requirements and tests Standards used to perform Biocompatibility Testing: - . EN ISO 10993-5:2009: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity {6}------------------------------------------------ - EN ISO 10993-10:2013: Biological evaluation of medical devices Part 10: Tests for ● irritation and skin sensitization Results of this testing provide reasonable assurance that the proposed device design meets the performance requirements and performs as intended. ### CONCLUSION 8. Based on the Indications for Use, technological characteristics, and safety and performance testing, the BSC Fixed Patient Leads ECG Cable has been shown to be appropriate for its Indications for Use and is considered to be substantially equivalent to the predicate BSC Removable Patient Leads ECG Cable cleared by FDA on December 16, 2013 (K132253).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...