Mecta-C with titanium markers and Mecta-C TiPEEK with titanium markers
Applicant
Medacta International S.A.
Product Code
ODP · Orthopedic
Decision Date
Sep 18, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Mecta-C intervertebral body fusion device is indicated for anterior cervical interbody fusion procedures in sketally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. The implants are intended to be used with supplemental spinal fixation. The Mecta-C device is intended for use at one level in the cervical spine, from C2-T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment prior to treatment with the device.
Device Story
Intervertebral body fusion device; PEEK cage with titanium markers; TiPEEK version includes commercially pure titanium coating. Used in anterior cervical interbody fusion procedures; inserted between cervical vertebrae to provide support/correction; packed with autogenous bone graft to promote fusion. Markers allow radiological assessment of implant position/orientation. Titanium markers replace tantalum to reduce imaging artifacts; marker spikes modified (90° to 50° conical edge) for improved penetration/contact. Used in clinical setting by surgeons; requires supplemental spinal fixation. Benefits include stabilization of cervical spine and promotion of bone fusion during healing.
Clinical Evidence
Bench testing only. No clinical data provided. Verification activities conducted on worst-case implants per pre-defined acceptance criteria. Performance tests (shear, tension, fatigue, abrasion, subsidence, wear, imaging artifact reduction) previously established for predicate devices.
Technological Characteristics
Materials: PEEK (ASTM F2026-16), Titanium (ISO 5832-3:1996), and CPTi coating (ASTM F1580-12). Form factor: Intervertebral cage with various widths, heights, and lordosis (5°, 7°). Sterilization: Gamma radiation. Single-use. Markers: Titanium with 50° conical edge.
Indications for Use
Indicated for anterior cervical interbody fusion in skeletally mature patients with cervical disc disease (neck pain of discogenic origin with degeneration confirmed by history/radiography) at one level from C2-T1, following six weeks of failed non-operative treatment. Requires use with autogenous bone graft and supplemental spinal fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of what appears to be an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2017
Medacta International SA % Elizabeth Rose, MST, RAC (US) Manager, Medical Devices Mapi USA. Inc. 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504
Re: K172587
Trade/Device Name: Mecta-C with titanium markers and Mecta-C TiPEEK with titanium markers Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: August 25, 2017 Received: August 28, 2017
Dear Ms. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Vincent J. Devlin -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K172587
### Device Name
Mecta-C with titanium markers and Mecta-C TiPEEK with titanium markers
### Indications for Use (Describe)
The Mecta-C intervertebral body fusion device is indicated for anterior cervical interbody fusion procedures in sketally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. The implants are intended to be used with supplemental spinal fixation.
The Mecta-C device is intended for use at one level in the cervical spine, from C2-T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment prior to treatment with the device.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 3.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Manager, Medacta International SA Date Prepared: August 25, 2017
#### II. Device
| Device Proprietary Name: | Mecta-C with titanium markers and Mecta-C TiPEEK<br>with titanium markers |
|--------------------------|---------------------------------------------------------------------------|
| Common or Usual Name: | Intervertebral Body Fusion Device, Cervical |
| Classification Name: | Intervertebral Body Fusion Device, Cervical |
| Primary Product Code: | ODP |
| Regulation Number: | 21 CFR 888.3080 |
| Device Classification | 2 |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
- Mecta-C, K112862, Medacta International SA ●
- Mecta-C TiPEEK, K142744, Medacta International SA ●
#### IV. Device Description
The purpose of this submission is to add the Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers to Medacta's Mecta-C product offering. The Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers are intervertebral body fusion devices intended for stabilization and to promote bone fusion during the normal healing process following surgical correction of disorders of the cervical spine. The Mecta-C and the Mecta-C TiPEEK intervertebral body fusion devices are indicated for the treatment of degenerative diseases of the cervical disc and can be used for cervical fusion from C2-T1.
The Mecta-C with titanium markers are intervertebral body fusion devices manufactured with an implant grade polyetheretherketone (PEEK) body and titanium markers. The Mecta-C
{4}------------------------------------------------
TiPEEK with titanium markers are intervertebral body fusion devices manufactured with a PEEK body, commercially pure titanium (CPTi) coating, and titanium markers' material is changing from tantalum to titanium to reduce artefact during imaging scans.
The titanium marker's spikes are also being modified to reduce the conical edge from 90° to 50°. The purpose of the conical edge change is to create a sharper tip design which allows for better penetration of the titanium pin and a more uniform contact between the PEEK cage and vertebral body.
The markers are placed in the implant on each end of the TiPEEK cages to allow easier radiological assessment of the position and orientation of the radiolucent TiPEEK cages.
The cages are offered in various widths, heights, footprint geometries, and lordosis which can be inserted between two cervical vertebra bodies to give support and correction during cervical interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft.
#### V. Indications for Use
The Mecta-C intervertebral body fusion device is indicated for anterior cervical interbody fusion procedures in skeletally mature patients. The device systems are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. The implants are intended to be used with supplemental spinal fixation.
The Mecta-C device is intended for use at one level in the cervical spine, from C2-T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment prior to treatment with the device.
#### VI. Comparison of Technological Characteristics
The Mecta-C with titanium markers, the Mecta-C TiPEEK with titanium markers, and the predicate devices share the following characteristics:
- cage and coating materials of construction;
- cage design: ●
- sterility:
- diameter:
- shelf life; and
- packaging.
{5}------------------------------------------------
The Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers are technologically different from the predicate devices as the markers' material of construction has changed and the conical edge has been reduced to 50°.
The Mecta-C with titanium markers are manufactured using PEEK Implant Grade Polyetheretherketone (ASTM F2026-16 Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications) and titanium (ISO 5832-3:1996 Implants for Surgery - Metallic Materials - Part 3: Wrought Titanium 6-Aluminum 4-Vanadium Alloy). The Mecta-C TiPEEK with titanium markers are manufactured using PEEK Implant Grade Polyetheretherketone (ASTM F2026-16) and titanium (ISO 5832-3:1996) with a commercially pure titanium (ASTM F1580-12 Standard Specification for Titanium and Titanium-6 Aluminum-4 Vanadium Alloy Powders for Coatings of Surgical Implants) coating.
These materials have a long history of use in implantable orthopedic devices and material information has been provided in previous Medacta 510(k) submissions for the Mecta-C (K112862) and Mecta-C TiPEEK (K142744).
Due to the extensive history of use in currently marketed medical devices, as well as identical manufacturing processes between the subject and predicate devices the Mecta-C (K112862) and the Mecta-C TiPEEK (K142744), additional biocompatibility testing was deemed unnecessary for the Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers.
A comparison of the subject and predicate devices is provided in the table below.
| Parameters | Mecta-C with titanium markers<br>and Mecta-C TiPEEK with<br>titanium markers | Mecta-C PEEK and<br>Mecta-C TiPEEK<br>(K112862 and K142744) |
|-----------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------|
| Height | 4 mm - 9 mm | 4 mm - 9 mm |
| Lordosis | 5° and 7° | 5° and 7° |
| Width | 14 mm, 16 mm, and 18 mm | 14 mm, 16 mm, and 18 mm |
| Depth | 12 mm, 14 mm, and 15 mm | 12 mm, 14 mm, and 15 mm |
| Material of<br>Construction | Cage: PEEK and PEEK w/CPTi<br>coating<br>Markers: titanium | Cage: PEEK and PEEK w/CPTi<br>coating<br>Markers: tantalum |
| Conical Edge | 50° | 90° |
| Device Usage | Single Use | Single Use |
# Technological comparison
{6}------------------------------------------------
| Parameters | Mecta-C with titanium markers<br>and Mecta-C TiPEEK with<br>titanium markers | Mecta-C PEEK and<br>Mecta-C TiPEEK<br>(K112862 and K142744) |
|------------|------------------------------------------------------------------------------|-------------------------------------------------------------|
| Sterility | Gamma Radiation | Gamma Radiation |
| Shelf Life | 5 Years | 5 Years |
| Packaging | Individual packaging | Individual packaging |
# Discussion
The technological differences between the subject and predicate devices do not raise new questions of safety and effectiveness. As seen above, the Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers are identical to the predicate devices in terms of materials of construction (except markers material), design, diameter, sterility, and packaging. Performance data provided within this submission supports that the Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers are substantially equivalent to the identified predicate devices.
#### VII. Performance Data
As the subject devices are a line extension to previously cleared devices, verification activities, as identified through risk analysis, were previously conducted on the worst-case implants to written protocols with pre-defined acceptance criteria. Engineering rationales determined that the proposed implants did not represent a new worst-case for mechanical testing.
The following performance tests were conducted on the predicate devices and reviewed as part of the Mecta-C (K112862) and the Mecta-C TiPEEK (K142744) submissions:
- shear testing;
- tension testing; ●
- shear and bonding fatigue; ●
- coating characterization; ●
- abrasion resistance;
- mechanical testing;
- subsidence testing;
- wear testing; and
- implant imaging properties artefacts reduction. ●
Pyrogenicity
- Medacta uses both the Bacterial Endotoxin Test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and the Pyrogen Test according to USP chapter <151> for pyrogenicity determination.
{7}------------------------------------------------
- Medacta does not intend to label the subject devices as non-pyrogenic or pyrogen free. .
# VIII. Conclusion
Based on the above information, the Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers can be considered substantially equivalent to the identified predicate devices.
Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations. The Mecta-C with titanium markers and the Mecta-C TiPEEK with titanium markers are as safe and effective as the predicate devices, Mecta-C (K112862) and Mecta-C TiPEEK (K142744).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.