K172555 · Omm Imports, Inc. D/B/A Zero Gravity · OLP · May 1, 2018 · General, Plastic Surgery
Device Facts
Record ID
K172555
Device Name
SAPPHIRE, ELEVARE SAPPHIRE
Applicant
Omm Imports, Inc. D/B/A Zero Gravity
Product Code
OLP · General, Plastic Surgery
Decision Date
May 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SAPPHIRE is an over -the -counter hand held, battery operated, light therapy device that uses light emitting diodes (LEDs) that emit a specific wavelength of 415nm (Blue Light) that is intended for use in the treatment of mild to moderate inflammatory acne.
Device Story
Sapphire is an OTC, handheld, battery-operated light therapy device for treating mild to moderate inflammatory acne. It utilizes blue light-emitting diodes (LEDs) at a 415nm wavelength (+/- 5nm) to deliver 50 mW/cm² of light energy. The device is applied directly to the skin; it operates via a simple on/off mechanism with no user adjustments required. Powered by rechargeable Li-Ion batteries, it delivers a standard dose of 12 J/cm² per treatment. The device is intended for home use by the patient. By applying blue light to the skin, the device aims to treat inflammatory acne lesions. It is a non-invasive, cordless system designed for consistent light administration without user-serviceable components.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing, including functional performance (power output), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), battery safety (IEC 62133), lamp safety (IEC 62471), and biocompatibility (ISO 10993). A usability/label comprehension study with 35 participants demonstrated that 96% of participants could comprehend labeling and use the device successfully.
Technological Characteristics
Handheld, battery-operated (Li-Ion) light therapy device. Emits blue light (415nm +/- 5nm) via LEDs. Output: 50 mW/cm². Dose: 12 J/cm². Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 62133 (battery), IEC 62471 (lamp safety), and ISO 10993 (biocompatibility). Simple on/off operation; no user settings. Cordless design.
Indications for Use
Indicated for the treatment of mild to moderate inflammatory acne in patients using an over-the-counter, handheld, battery-operated light therapy device.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 1, 2018
Omm Imports, Inc. d/b/a Zero Gravity % Ms. Susan Anthoney-Dewet Aegis, Inc 2424 Dempster Drive Coralville, Iowa 52241
Re: K172555
Trade/Device Name: Sapphire, Elevare Sapphire Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: April 19, 2018 Received: April 20, 2018
Dear Ms. Anthoney-Dewet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -ਟਤ
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172555
Device Name SAPPHIRE
Indications for Use (Describe)
The SAPPHIRE is an over -the -counter hand held, battery operated, light therapy device that uses light emitting diodes (LEDs) that emit a specific wavelength of 415nm (Blue Light) that is intended for use in the treatment of mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K172555
This summary of 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 807.92
Submission Date: August 23, 2017
| 1. Submitter Information: | AEGIS Regulatory, Inc. – Susan Anthoney-DeWet | |
|---------------------------|-----------------------------------------------|--|
| | 2424 Dempster Drive | |
| | Coralville, IA 52241 | |
| | Tel.: 865-982-5552 | |
| | Email: sue@fdalistingconsultants.com | |
| On behalf of Sponsor: | Omm Imports, Inc. d/b/a Zero Gravity |
|-----------------------|--------------------------------------|
| | Attn: Moti Shenfarber |
| | 9737 NW 41st Street STE 234 |
| | Doral, FL 33178 |
| | Email: moti@zerogravityskin.com |
#### 2. General Information
- 2.1 Classification Name: Light Based Over-The-Counter Powered Light Based Laser For Acne
- 2.2 Common/Usual Name: Blue Light Therapy for Acne
- 2.3 Proprietary Names: Sapphire
- 2.4 Classification: Class II
- 2.5 Classification Number: 878.4810
2.6 Product Code: OLP
#### 3. Predicate Device(s):
Primary Predicate Device: K121435 – Silk'n Blue (Home Skinovations LTD)
Secondary Predicate Device: K131461 – LightStim for Acne Mini (LED Intellectual Properties LLC)
{4}------------------------------------------------
#### 4. Device Description:
The Sapphire device is an over-the-counter, battery powered, hand-held device that emits light energy in the blue spectrum for the treatment of mild to moderate inflammatory acne by emitting 50 mW/cm² of blue (415 nm +/- 5nm) light via an electric light emitting diode [LEDs] energy source. The Sapphire device operates by a cordless system drawing upon its rechargeable Li-Ion batteries to deliver the treatment. The unit is applied directly to the skin to ensure consistent administration of light during each treatment. There are no user settings or adjustments required. The device does not contain any user serviceable components. The device is sold as Over the Counter (OTC).
## 5. Indications / Intended Use:
The SAPPHIRE is an over –the -counter hand held, battery operated, light therapy device that uses light emitting diodes (LEDs) that emit a specific wavelength of 415nm (Blue Light) that is intended for use in the treatment of mild to moderate inflammatory acne.
## 6. Technological Characteristics & Substantial Equivalency:
Summary of the technological characteristics of the device compared to predicate device:
| Device | Silk 'N Blue | LightStim for Acne Mini | Sapphire |
|-------------|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Home Skinovations,<br>LTD.<br>K121435<br>A Predicate Device | LED Intellectual<br>Properties, LLC<br>K131461<br>A Predicate Device | Zero Gravity, LLC<br>K172555<br>This Submission |
| Indications | The Silkn Blue is<br>indicated as an over<br>the counter<br>phototherapy device<br>for the treatment of<br>mild to moderate acne | LightStim for Acne<br>Mini is an over-the-<br>counter, hand held<br>device intended for the<br>use in the treatment of<br>mild to moderate acne. | The SAPPHIRE is an<br>over -the -counter<br>hand held, battery<br>operated, light therapy<br>device that uses light<br>emitting diodes<br>(LEDs) that emit a<br>specific wavelength of<br>415nm (Blue Light)<br>that is intended for use<br>in the treatment of<br>mild to<br>moderate<br>inflammatory acne. |
| Handheld | Yes | Yes | Yes |
| Wavelength | 415 nm +/- 15nm | 415 nm +/- 6nm | 415 nm +/- 5nm |
{5}------------------------------------------------
| Device | Silk 'N Blue | LightStim for Acne Mini | Sapphire |
|-------------------------------------|------------------------------------------------------------------------------|-------------------------------------|------------------------------------------------------------------------------|
| | Home Skinovations,<br>LTD. | LED Intellectual<br>Properties, LLC | Zero Gravity, LLC |
| | K121435 | K131461 | K172555 |
| | A Predicate Device | A Predicate Device | This Submission |
| Modes | On/Off | On/Off | On/Off |
| Energy Level | 50 mW/cm2 | 50 mW/cm2 | 50 mW/cm2 |
| Treatment<br>Time | 4 minutes per area, twice<br>per week for 4 weeks<br>(total of 8 treatments) | 3-4 minutes per area | 4 minutes per area, twice<br>per week for 4 weeks<br>(total of 8 treatments) |
| Total Energy<br>delivered per<br>Tx | 12 J/cm2 | 9-12 J/cm2 | 12 J/cm2 |
| Location for<br>Use | OTC | OTC | OTC |
1. Has the same intended use as the predicate(s) (i.e., Treatment of mild to moderate inflammatory acne ;
2. Has the same output (i.e., 50 mW/cm2 ) as the predicate(s);
3. Utilizes the same wavelength (i.e., 415 nm +/- 5nm) as the predicate device(s);
4. Utilizes the same standard dose (i.e., 121/cm2) as the predicate device(s);
The Sapphire device and the above referenced predicate devices are Over the Counter devices used to treat mild to moderate inflammatory acne as defined in 21 CFR § 878.4810. These devices utilize blue diodes at 415nm to provide narrow bands of light energy to treat mild to moderate inflammatory acne vulgaris. The performance achieved by these devices is similar with equal power output. The devices are handheld, and intended to be placed directly on the skin. They are manufactured out of similar materials. Based upon comparison to the predicate device, the Sapphire device has the same intended use, with similar technological characteristics as the predicate device. The system performs as intended and does not raise any new safety or effectiveness issues.
{6}------------------------------------------------
# 7. Performance Data/Non-Clinical Testing:
Testing of the Sapphire device included safety testing to various international standards, bench testing--functional performance testing (power output), software validation testing and user safety testing.
The results of this testing are as follows:
Conforms to international consensus standards:
# ELECTRICAL SAFETY: Recognition Number 19-4:
- IEC60601-1 Medical Electrical Equipment Part 1: General Requirements For Basic ● Safety And Essential Performance
## EMC:
# Recognition Number 19-8:
- IEC 60601-1-2 Edition 4.0 2014-02, Medical Electrical Equipment Part 1-2: . General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
# BATTERY SAFETY:
# Recognition Number 19-13:
- IEC 62133;2013:Secondary Cells And Batteries Containing Alkaline Or Other Non-. Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications
# HOME HEALTHCARE ENVIRONMENT:
# Recognition Number 19-14:
- IEC 60601-1-11 Medical Electrical Equipment Part 1-11: General Requirements . For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used in Home Healthcare Environment
## LAMP SAFETY:
## Recognition Number 12-249:
- IEC 62471 First Edition 2006-07, Photobiological Safety Of Lamps And Lamp Systems. (Radiology)
## BIOCOMPATIBILITY:
## Recognition Number 2-156:
- o ISO 10993-1:2009/(R) 2013. Biological Evaluation Of Medical Devices -- Part 1: Evaluation And Testing Within A Risk Management Process.
## Recognition Number 2-153:
- ISO 10993-5:2009/(R) 2014, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity.
## Recognition Number 2-174:
- ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices -● Part 10: Tests For Irritation And Skin Sensitization.
{7}------------------------------------------------
The Sapphire device's software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices."
A Usability/Label Comprehension Study was conducted with 35 participants. The results of the study found that: 96% of the participants were able to comprehend the labeling. 96% of the participants were able to use the device successfully.
User Safety testing reflects device can be used in a safe and effective manner.
# 9. Substantial Equivalence Conclusion
After an analysis of the safety, indications, intended uses, performance, design materials, power output, technological properties, treatment areas, and treatment regimes the manufacturer believes that no significant differences exist between the device and the predicate device. Therefore substantial equivalency is requested.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.