Codman EDS 3 CSF External Drainage System, Codman EDS 3 CSF External Drainage System Collection Bag
K172537 · Codman & Shurtleff, Inc. · JXG · Oct 16, 2017 · Neurology
Device Facts
Record ID
K172537
Device Name
Codman EDS 3 CSF External Drainage System, Codman EDS 3 CSF External Drainage System Collection Bag
Applicant
Codman & Shurtleff, Inc.
Product Code
JXG · Neurology
Decision Date
Oct 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Use of the CODMAN EDS 3 CSF External Drainage System (EDS 3) is indicated for draining cerebrospinal fluid (CSF) from the cerebral ventricles or the lumbar subarachnoid space as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated. Use the External Drainage System Collection Bag with the CODMAN EDS 3 External Drainage System to measure and collect cerebrospinal fluid (CSF).
Device Story
System drains CSF from cerebral ventricles or lumbar subarachnoid space to reduce intracranial pressure; utilizes ventricular catheter, patient drain line, 100 mL graduated drip chamber, and 700 mL collection bag. Principle of operation: gravity-based drainage controlled by differential pressure; clinician adjusts drip chamber height along base frame to regulate flow rate. Used in clinical settings; operated by healthcare professionals. Benefits patients by managing elevated intracranial pressure post-trauma. Output is collected CSF volume measured via graduated markings on drip chamber and bag. Device is disposable and sterile.
Clinical Evidence
No clinical data; bench testing only. Verification and validation testing included IV pole clamp design, collection bag design, drainage time, and simulated use. All tests met acceptance criteria.
Technological Characteristics
System components: ventricular catheter, patient drain line, drip chamber, collection bag. Materials: non-patient contacting components updated. Sterilization: Ethylene Oxide (EtO) per ISO 11135-1:2014. Biocompatibility: ISO 10993-1:2009. Gravity-fed drainage system; no software or electronic components.
Indications for Use
Indicated for patients requiring CSF drainage from cerebral ventricles or lumbar subarachnoid space to reduce intracranial pressure and CSF volume when permanent shunting is not indicated.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Codman EDS 3 CSF External Drainage System (K162437)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 16, 2017
Codman & Shurtleff, Inc. Christopher Garete Regulatory Affairs Specialist II 325 Paramount Drive Raynham, Massachusetts 02767
Re: K172537
Trade/Device Name: CODMAN EDS 3 CSF External Drainage System: External Drainage System Collection Bag Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: August 21, 2017 Received: August 22, 2017
Dear Mr. Christopher Garete:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Carlos L. Pena -S
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172537
#### Device Name
CODMAN EDS 3 CSF External Drainage System External Drainage System Collection Bag
#### Indications for Use (Describe)
Use of the CODMAN EDS 3 CSF External Drainage System (EDS 3) is indicated for draining cerebrospinal fluid (CSF) from the cerebral ventricles or the lumbar subarachnoid space as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated.
Use the External Drainage System Collection Bag with the CODMAN EDS 3 External Drainage System to measure and collect cerebrospinal fluid (CSF).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K172537 - 510(k) Summary
#### I. Submitter Codman & Shurtleff, Inc. 325 Paramount Drive Raynham, MA 02767
Fax:
Establishment Registration Number: 1226348
| Primary Contact: | Christopher Garete |
|--------------------|--------------------|
| Phone: | (508) 977-3869 |
| Fax: | (508) 977-6979 |
| Secondary Contact: | Megan Palumbo |
| Phone: | (508) 828-3571 |
(508) 977-6979
Date of Preparation: August 21, 2017
#### II. Device
| Device Proprietary Name | CODMAN EDS 3 CSF External Drainage System<br>External Drainage System Collection Bag |
|---------------------------|--------------------------------------------------------------------------------------|
| Common Name | External Drainage System |
| Classification Name | Central Nervous System Fluid Shunt and<br>Components<br>(21 CFR 882.5550) |
| Regulatory Classification | II |
| Product Code | JXG |
III. Predicate Device
The predicate device for this submission is:
Codman EDS 3 CSF External Drainage System (K162437), which was cleared on October 28, 2016.
The Codman EDS 3 CSF External Drainage System (Codman EDS 3 IV. Device Description System) is designed to drain cerebral spinal fluid (CSF) at a controlled rate based on differential pressure between the device and the patient. Collecting CSF from the patient is performed in efforts to reduce elevated intracranial pressure (ICP) post trauma. The EDS 3 device is comprised of four (4) main parts: ventricular catheter, patient drain line, base frame drip chamber assembly, and a collection bag. Note: The collection bag is sold as part of the Codman EDS 3 System, as well as sold separately.
> The principle of operation of the proposed Codman EDS 3 system is identical to the currently marketed Codman EDS 3 system. The ventricular catheter is placed into one of the ventricles in the brain or in the subarachnoid space and is then connected to the patient drainage line. CSF flows from the brain or lumbar region through the patient line and enters into the 100 mL graduated drip chamber assembly, where it is collected
{4}------------------------------------------------
over a period of time to calculate a flow rate. The drip chamber assembly can then be raised or lowered along the base frame, thereby adjusting the differential pressure to achieve the appropriate flow rate. Once the drip chamber height is set, the collected CSF is then drained into the attached 700 mL collection bag.
The Codman EDS 3 CSF External Drainage System Collection Bag is a sterile, vented 700 mL capacity bag that is graduated in 50 mL increments for accurate measurement. A microbial-retentive atmospheric vent facilitates CSF flow into the bag. One bag is provided with the system and replacement bags are sold separately as an accessory to the drain.
The EDS 3 system is a complete, disposable unit that is provided sterile and is available with or without a ventricular catheter.
The Indications for Use statement of the proposed devices remain identical
to those of the predicate devices.
| Equivalence<br>Comparison | Codman EDS 3 CSF<br>External Drainage System<br>(Predicate: K162437) | Codman EDS 3 CSF<br>External Drainage System<br>(Subject of This<br>Submission) |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| EDS 3 System<br>Indications for Use | Use of the CODMAN EDS 3<br>CSF External Drainage<br>System (EDS 3) is indicated<br>for draining cerebrospinal<br>fluid (CSF) from the cerebral<br>ventricles or the lumbar<br>subarachnoid space as a<br>means of reducing<br>intracranial pressure and<br>CSF volume when the<br>insertion of a permanent,<br>internal shunt is not<br>indicated.<br>Use the External Drainage<br>System Collection Bag with<br>the CODMAN EDS 3<br>External<br>Drainage System to measure<br>and collect<br>cerebrospinal fluid (CSF). | Same as predicate |
## V. Indications for Use
VI. Comparison to Predicate Device
The proposed Codman EDS 3 CSF External Drainage System is identical to the currently marketed Codman EDS 3 CSF External Drainage System (K162437) except for the following changes proposed in this submission:
{5}------------------------------------------------
- a new filter membrane material in the burette cap .
- a wider pole clamp design with a shorter thumb screw
- a straightened and wider diameter base frame tube ●
- a new collection bag design .
The indications for use, principle of operation, clinical utility, packaging, and sterilization remain identical to the predicate devices. The proposed design is similar to that of the predicate device.
| Substantial Equivalence Comparison | | | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--|
| Characteristic | Codman EDS 3 CSF<br>External Drainage System<br>and Bags (Predicate:<br>K162347) | Codman EDS 3 CSF<br>External Drainage System<br>and Bags (Subject of This<br>Submission) | |
| Manufacturer | Codman & Shurtleff, Inc. | Same as predicate | |
| Classification Panel | Neurology | Same as predicate | |
| Classification<br>Name | Central Nervous System Fluid<br>Shunt and Components<br>(21 CFR 882.5550) | Same as predicate | |
| Indications for Use | Use of the CODMAN EDS 3<br>CSF External Drainage<br>System (EDS 3) is indicated<br>for draining cerebrospinal<br>fluid (CSF) from the cerebral<br>ventricles or the lumbar<br>subarachnoid space as a<br>means of reducing<br>intracranial pressure and<br>CSF volume when the<br>insertion of a permanent,<br>internal shunt is not<br>indicated.<br>Use the External Drainage<br>System Collection Bag with<br>the CODMAN EDS 3<br>External<br>Drainage System to measure<br>and collect<br>cerebrospinal fluid (CSF). | Same as predicates | |
| Operating Principle | Drainage of CSF based on a<br>differential pressure between<br>the patient and the device. | Same as predicate | |
| Single Use Only | Yes | Same as predicate | |
| Shelf Life | 3 years (EDS3 System)<br>5 years (Collection Bags) | 3 years (EDS3 System and<br>Collection Bags) | |
| Sterilization<br>Method | EtO | Same as predicate | |
| Sterility Assurance<br>Level (SAL) | 10-6 | Same as predicate | |
| Non-Pyrogenic | Yes | Same as predicate | |
| Packaging | EDS 3 System: product is<br>placed in a blister tray, sealed<br>with a lid, and then placed | Same as predicate | |
{6}------------------------------------------------
| into a unit box (EDS3<br>System) |
|---------------------------------------------------------------------------------------------------------------------------------------------|
| Collection Bags: the bags are<br>placed in blister tray sealed<br>with a lid, placed into a unit<br>box. Five (5) blisters per unit<br>box. |
VII. Performance Data
The following performance data has been provided in support of the substantial equivalence determination.
### Bench Testing:
Design Verification and Design Validation were performed to verify that the performance of the proposed device is substantially equivalent to that of the predicate device. Please see the Summary of Bench Testing Table below.
| Summary of Bench Testing | | |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Test Method Summary | Result |
| IV Pole Clamp Design<br>Verification | Ensure new IV Pole Clamp<br>can hold at least the same<br>amount of weight as the<br>predicate device. | Pass - proposed device<br>design met the acceptance<br>criteria and is therefore<br>substantially equivalent to<br>the predicate device. |
| Collection Bag Design<br>Verification | Ensure new Collection Bag<br>connects to the EDS 3<br>system, is leak free, and<br>exhibits at least the same<br>tensile strength as the<br>predicate device. | Pass - proposed device<br>design met the acceptance<br>criteria and is therefore<br>substantially equivalent to<br>the predicate device. |
| Drainage Time Design<br>Verification | Ensure drainage time for the<br>new system meets design<br>inputs specifications and is at<br>least as fast as the predicate<br>device. | Pass - proposed device<br>design met the acceptance<br>criteria and is therefore<br>substantially equivalent to<br>the predicate device. |
| Design Validation /<br>Simulated Use of EDS<br>3 System and Bag | Ensure finished product<br>meets all user inputs and<br>needs and inputs of the<br>predicate device. | Pass - proposed device<br>design met the acceptance<br>criteria and is therefore<br>substantially equivalent to<br>the predicate device. |
### Sterilization Testing:
The Codman EDS 3 CSF External Drainage System and Collection Bags are sterilized using a validated ethylene oxide sterilization cycles. The sterilization cycle has been validated to ensure a sterility assurance level (SAL) of 10-6 in accordance with ISO 11135-1:2014, "Sterilization of health care products - Ethylene Oxide Sterilization - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices."
## Shelf Life Testing:
The Codman EDS 3 CSF External Drainage System and Collection Bags were subjected to accelerated aging. The aging studies established that the
{7}------------------------------------------------
device and packaging remain functional and maintain sterility for up to 3 years.
#### Biocompatibility Testing:
Biocompatibility assessments were performed for the material changes (filter, tubing, thumbscrew, and collection bag). All material changes were made to non-patient contacting device components. The proposed materials were determined to be acceptable for their intended use and the final device continues to meet the requirements of ISO 10993-1:2009, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
#### Animal Testing:
No animal studies were performed as appropriate verification and validation of the proposed device was achieved based on the comparison to the predicate device and from the results of the bench testing.
## Clinical Testing:
No clinical studies were performed as appropriate verification and validation of the proposed device was achieved based on the comparison to the predicate device and from the results of the bench testing.
#### Conclusion:
Based on the intended use, fundamental scientific technology, comparison to the predicate device, and testing conducted, it is concluded that the subject device, the Codman EDS 3 CSF External Drainage System and Collection Bags are substantially equivalent to the predicate device and therefore do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.