K172520 · Boston Scientific Corporation · FGE · Oct 4, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K172520
Device Name
CRE RX Biliary Balloon Dilatation Catheter
Applicant
Boston Scientific Corporation
Product Code
FGE · Gastroenterology, Urology
Decision Date
Oct 4, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CRE™ RX Biliary Balloon Dilatation Catheter is indicated for use in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. The CRF™ RX Biliary Balloon Dilatation Catheters may be used for injection of contrast medium for fluoroscopic visualization of the bile ducts.
Device Story
CRE™ RX Biliary Balloon Dilatation Catheter; multi-lumen catheter (balloon inflation, guidewire, contrast injection); designed for use with duodenoscope (3.7 mm+ working channel) and 0.035 in guidewire. Features controlled radial expansion (CRE) technology; balloon inflates to 3 distinct, progressively larger diameters at/above 3 atm. Includes FloSwitch™ Adapter for contrast injection. Used by physicians in endoscopic procedures to dilate Sphincter of Oddi; enables fluoroscopic visualization of bile ducts. Rapid exchange design allows guidewire locking. Benefits include precise, multi-stage dilatation and integrated contrast delivery in single device.
Clinical Evidence
Bench testing only. Tests included: inflation/deflation, shaft kink resistance, scope insertion/removal, shaft/tunnel tensile strength, lumen mandrel removal, contrast injection, and hub luer lock fittings. Biocompatibility testing performed per ISO 10993-1, ISO 11135-1, and ISO 10993-7. All results met acceptance criteria.
Technological Characteristics
Multi-lumen catheter; balloon inflation, guidewire, and contrast injection lumens. Controlled radial expansion (CRE) mechanism; 3 cm balloon length. Rapid exchange design. Compatible with 0.035 in guidewire. Sterilized via Ethylene Oxide (ISO 11135-1). Biocompatibility per ISO 10993-1/7.
Indications for Use
Indicated for adults requiring endoscopic dilatation of the Sphincter of Oddi, with or without prior sphincterotomy, and for fluoroscopic visualization of bile ducts via contrast medium injection.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
CRE™ PRO Wireguided Balloon Dilatation Balloon (K112994)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
December 17, 2018
Boston Scientific Corporation Jia Huang Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K172520
Trade/Device Name: CRE™ RX Biliary Balloon Dilatation Catheter Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: August 18, 2017 Received: August 21, 2017
Dear Jia Huang:
This letter corrects our substantially equivalent letter of October 4, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Charles Viviano -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172520
Device Name
CRE™ RX Biliary Balloon Dilatation Catheter
#### Indications for Use (Describe)
CRE™ RX Biliary Balloon Dilatation Catheter is indicated for use in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. The CRF™ RX Biliary Balloon Dilatation Catheters may be used for injection of contrast medium for fluoroscopic visualization of the bile ducts.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
## 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4360 Fax: 508-683-5939
| Contact: | Jia Huang |
|----------------|---------------------------------------|
| | Senior Specialist, Regulatory Affairs |
| Date Prepared: | August 17, 2017 |
#### 2. Proposed Device:
| Trade Name: | CRE™ RX Biliary Balloon Dilatation Catheter |
|----------------------|---------------------------------------------|
| Classification Name: | Biliary Catheter and Accessories |
| Regulation Number: | 876.5010 |
| Product Code: | FGE |
| Classification: | Class II |
# 3. Predicate Device:
| Trade Name: | CRE™ PRO Wireguided Balloon Dilatation Balloon |
|----------------------|---------------------------------------------------|
| 510(k) Number: | K112994 |
| Classification Name: | Biliary Catheter and Accessories |
| Regulation Number: | 876.5010 |
| Product Code: | FGE |
| Classification: | Class II |
| Trade Name: | Hurricane™ RX Biliary Balloon Dilatation Catheter |
| Trade Name: | Hurricane™ RX Biliary Balloon Dilatation Catheter |
| 510(k) Number: | K130484 |
| Classification Name: | Biliary Catheter and Accessories |
| Regulation Number: | 876.5010 |
| Product Code: | FGE |
| Classification: | Class II |
# Reference Device:
| Trade Name: | CRET™ Pulmonary Balloon Dilatation Catheter |
|----------------------|--------------------------------------------------|
| 510(k) Number: | K170759 |
| Classification Name: | Bronchoscope (flexible or rigid) and Accessories |
| Regulation Number: | 874.4680 |
| Product Code: | KTI |
| Classification: | Class II |
# 4. Proposed Device Description:
The CRE™ RX Biliary Balloon Dilatation Catheter is capable of inflating balloon to 3 distinct and progressively larger size diameters via controlled radial expansion. The specific balloon sizes are printed on each package and flag label.
{4}------------------------------------------------
The CRE™ RX Biliary Balloon Dilatation Catheter is designed to pass through a 3.7 mm or greater working channel of a duodenoscope and accept a 0.035 in (0.89 mm) guidewire through its guidewire lumen.
The CRE™ RX Biliary Balloon Dilatation Catheter has three lumens: balloon inflation lumen, guidewire lumen and contrast injection lumen. A FloSwitch™ Adapter is connected to the hub of the "Injection" lumen. The FloSwitch Adapter is to remain closed until distal contrast medium iniections are required.
# 5. Indications for Use:
CRE™ RX Biliary Balloon Dilatation Catheter is indicated for use in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. The CRE™ RX Biliary Balloon Dilatation Catheters may be used for injection of contrast medium for fluoroscopic visualization of the bile ducts.
The indications for use are similar when compared to those of the predicates CRE™ PRO Wireguided Balloon Dilatation Catheter (K112994) and Hurricane™ RX Biliary Balloon Dilatation Catheter (K130484).
## 6. Technological Characteristics:
The proposed CRE™ RX Biliary Balloon Dilatation Catheter has identical intended use as the predicates CRE™ PRO Wireguided Balloon Dilatation Catheter (K112994) and Hurricane™ RX Biliary Balloon Dilatation Catheter (K130484). The proposed CRE™ RX Biliary Balloon Dilatation Catheter has similar technological characteristics as the predicates CRE™ PRO Wireguided Balloon Dilatation Catheter (K112994) and Hurricane™ RX Biliary Balloon Dilatation Catheter (K130484) and reference CRE™ Pulmonary Balloon Dilatation Catheter (K170759).
Although the technological characteristics of the CRE™ RX Biliary Balloon Dilatation Catheter are not completely identical to either of the predicate devices, the technological characteristics of the proposed CRE™ RX Biliary Balloon Dilatation Catheter are the combination of the technological characteristics of the two predicates and can fulfil the identical intended use as that of the two predicate devices.
The proposed CRE™ RX Biliary Balloon Dilatation Catheter has the same design as predicate CRE™ PRO Wireguided Balloon Dilatation Catheter (K112994) and reference CRETM Pulmonary Balloon Dilatation Catheter (K170759) in terms of balloon controlled radial expansion that the balloon will dilate to three distinct and progressively larger size diameters at or above 3 atm (304 kPa).
The proposed CRE™ RX Biliary Balloon Dilatation Catheter has the same design as predicate Hurricane™ RX Biliary Balloon Dilatation Catheter (K130484) in terms of guidewire rapid exchange and contrast medium injection that the guidewire can exit the proposed CRE™ RX Biliary Balloon Dilatation Catheter at the guidewire exit port in order to be locked in place and the distal bile duct can be viewed fluoroscopically by contrast medium injection through the injection lumen of the proposed CRE™ RX Biliary Balloon Dilatation Catheter.
The proposed CRE™ RX Biliary Balloon Dilatation Catheter has the identical 3 cm length balloon as the 3 cm balloon of the reference CRE™ Pulmonary Balloon Dilatation Catheter (K170759) that will provide physician another length option to dilate the Sphincter of Oddi using the proposed CRE™ RX Biliary Balloon Dilatation Catheter.
{5}------------------------------------------------
The proposed CRE™ RX Biliary Balloon Dilatation Catheter has similar material as predicates CRE™ PRO Wireguided Balloon Dilatation Catheter (K112994) and Hurricane™ RX Biliary Balloon Dilatation Catheter (K130484) and reference CRE™ Pulmonary Balloon Dilatation Catheter (K170759).
### 7. Performance Data:
The proposed CRE™ RX Biliary Balloon Dilatation Catheter meets the requirements of ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing", ISO 11135-1 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices" and ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals".
The following bench tests were performed on the proposed CRE™ RX Biliary Balloon Dilatation Catheter: Inflate, Deflate, Shaft Kink Resistance, Scope Insertion, Scope Removal, Shaft Tensile, Tunnel Tensile. Lumen Mandrel Removal, Contrast Injection, Hub Luer Lock Fittings.
The testing performed demonstrated that the proposed device is substantially equivalent to the predicate devices because the acceptance criteria of these product specifications are identical or similar to those of the predicate devices.
#### 8. Conclusion:
All biocompatibility tests conducted on the proposed CRE™ RX Biliary Balloon Dilatation Catheter passed. Therefore, the proposed device is considered biocompatible for its intended use.
All device bench test results were acceptable. The data demonstrate that the proposed CRE™ RX Biliary Balloon Dilatation Catheter sufficiently meets the design specifications and is suitable for the intended use.
Boston Scientific Corporation has demonstrated that the proposed CRE™ RX Biliary Balloon Dilatation Catheter is substantially equivalent to the currently marketed CRE™ PRO Wireguided Balloon Dilatation Catheter (K112994) and Hurricane™ RX Biliary Balloon Dilatation Catheter (K130484) and can be safely and effectively used for its proposed indications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.