The intended use of the ETG-4100 is the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin.
Device Story
Optical Topography System ETG-4100 measures relative changes in cerebral oxy-hemoglobin, deoxy-hemoglobin, and total hemoglobin. Device emits near-infrared light (695nm and 830nm) via fiber optics attached to subject's head via flexible caps. Reflected light is detected by avalanche photodiodes. System uses frequency-modulated light intensity and lock-in amplifier to assign signals to measurement channels. Data processed to display time-course graphs or 2D/3D topographic images. Used in clinical settings; operated by trained personnel. Provides real-time hemoglobin concentration data to assist in monitoring cerebral hemodynamics. Benefits include non-invasive brain imaging with reduced patient restraint and discomfort compared to other modalities.
Clinical Evidence
Bench testing only. Includes signal stability testing comparing subject and predicate, software verification and validation, and compliance with electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and biocompatibility (ISO 10993) standards.
Technological Characteristics
Near-infrared spectroscopy (695nm, 830nm); laser diode source; avalanche photodiode (APD) detector; frequency-modulated signal processing; lock-in amplifier; fiber optic probes with spring-loaded tips; Windows 7 (64-bit) OS; connectivity via LAN; IEC 60825-1 Class 1M laser safety; biocompatible materials per ISO 10993.
Indications for Use
Indicated for the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin in patients.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the agency's acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 17, 2018
Hitachi Healthcare Americas Doug Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park Twinsburg, Ohio 44087
Re: K172492
Trade/Device Name: Optical Topography System ETG-4100 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: July 13, 2018 Received: July 17, 2018
Dear Doug Thistlethwaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Matthew C. Krueger -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K172492
Device Name Optical Topography System ETG-4100
Indications for Use (Describe) The intended use of the ETG-4100 is the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K172492 510(k) Summary
### Submitter Information
| Submitter: | Hitachi Healthcare Americas |
|-------------------|---------------------------------------|
| | 1959 Summit Commerce Park |
| | Twinsburg, Ohio 44087-2371 |
| Contact: | Douglas J. Thistlethwaite |
| Telephone number: | 330-425-1313 |
| Telephone number: | 330-963-0749 |
| E-mail: | thistlethwaited@hitachihealthcare.com |
| Date: | August 17, 2018 |
## Subject Device
| Trade/Proprietary Name: | Optical Topography System ETG-4100 |
|-------------------------|------------------------------------|
| Regulation Number: | 21 CFR 870.2700 |
| Regulation Name: | Oximeter |
| Product Code | DQA |
| Class | II |
| 510(k) Review Panel | Neurology |
#### Predicate Device
| Trade/Proprietary Name: | K042501 Optical Topography System ETG-4000 |
|-------------------------|--------------------------------------------|
| Regulation Number: | 21 CFR 870.2700 |
| Regulation Name: | Oximeter |
| Product Code | DQA |
| Class | II |
| 510(k) Review Panel | Anesthesiology |
#### Device Intended Use
The intended use of the ETG-4100 is the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin based on the amount of reflected or scattered radiation following transmission of radiation at known wavelengths through blood.
#### Indications for Use
The intended use of the ETG-4100 is the measurement of relative levels of cerebral deoxyhemoglobin and oxyhemoglobin. (Rx Use)
### Device Description
The Optical Topography System ETG-4100 is a device that measures and displays relative changes of oxy-hemoglobin, deoxy-hemoglobin and total hemoglobin in the cerebral cortex at multiple points on the head. The ETG-4100 system emits near-infrared light (670-1300mm) of two different wavelengths (695nm and 830nm) that can efficiently penetrate the skull where the light is scattered, absorbed by hemoglobin and unabsorbed light is reflected back.
#### Physical and Performance Characteristics
Light-emitting and detecting optical fibers are arranged in fiber holders in specific configurations which can be attached in numerous positions to the subject's head via flexible head caps. The ETG-4100 system detects the reflected light and calculates the relative changes of oxyhemoglobin, deoxy-hemoglobin and total hemoglobin based on the difference between injected and reflected light.
The technology is non-invasive and less restraining for the subject, reducing discomfort considerably in comparison to other brain imaging technologies. The spring-loaded fiber tips provide efficient contact between fiber tip and scalp. Contact between fiber tip and scalp is indicated by a shaft that protrudes from the top of the fiber tip touches the scalp; the firmer the contact, the more the shaft protrudes.
The ETG-4100 is equipped with a highly sensitive avalanche photodiode allowing the use of
{4}------------------------------------------------
relatively low light intensities, reducing potential harm to the subject and extending the lifetime of the laser diodes.
Simultaneous measurement of all measurement channels is achieved by frequency-modulating the light intensity of each individual light source. The frequency of reflected near infrared light is analyzed by a lock-in amplifier which assigns the incoming signals to their respective measurement channel based on their frequencies.
The individual frequencies of the emitted near infrared light are different from that of daylight or artificial indoor light, the ETG-4100 can be operated under daylight conditions. Measured changes in hemoglobin concentrations can be displayed as time-course graphs or as 2D or 3D topographic images.
### Comparison of Technological Characteristics
The technological characteristics of the Optical Topography System ETG-4100 as compared to the predicate are listed in the following table.
Where applicable, differences are discussed in detail in the comment table that follows the comparison table below to demonstrate how these characteristics compare to the predicate.
| | K172492 | K042501 | |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------|
| ITEM | OPTICAL TOPOGARAPHY SYSTEM ETG-4100 | OPTICAL TOPOGRAPHY SYSTEM ETG-4000 | DIFFERENCE |
| System Configuration | | | See 01 |
| Illustration | Image: OPTICAL TOPOGARAPHY SYSTEM ETG-4100 | Image: OPTICAL TOPOGRAPHY SYSTEM ETG-4000 | |
| Indications for Use | The intended use of the ETG-4100 is the<br>measurement of relative levels of cerebral<br>deoxyhemoglobin and oxyhemoglobin. (Rx<br>Use) | The intended use of the ETG-4000 is the<br>measurement of relative levels of cerebral<br>deoxyhemoglobin and oxyhemoglobin. (Rx use) | No |
| Dimensions | $582(W)×933(D)×1,430(H)mm$ | $560(W)×933(D)×1,470(H)mm$ | See 01 |
| Weight | 165kg | 130kg | See 01 |
| Power Source | AC100/120V 500VA<br>AC220/230/240V 700VA | AC100/120V 500VA<br>AC220/230/240V 700VA | No |
| Holder and Probe Configurations | | | |
| 3x3 holder | Yes | Yes | No |
| 4x4 holder | Yes | Yes | No |
| 3x5 holder | Yes | Yes | No |
| 3x11 holder | Yes | Yes | No |
| 2x11 holder | Yes | No | See 02 |
| 4x4DD holder | Yes | No | See 02 |
| 3x5DD holder | Yes | No | See 02 |
| 3x5MD holder | Yes | No | See 02 |
| 3x5 Neonate/ Infant probe | Yes | Yes | No |
| 3x3 Neonate/ Infant probe | Yes | Yes | No |
| Non-magnetic Probe Type T | Yes | No | See 03 |
| Non-magnetic probe type M<br>(length: 7.5m) | Yes | No | See 03 |
| Non-magnetic probe type M<br>(length: 10.5m) | Yes | No | See 03 |
| Fiber Configurations | | | |
| Removable Infant Fiber | Yes | No | See 04 |
| Long Fiber(length:5.5m) | Yes | No | See 05 |
| Long Fiber(length:6.5m) | Yes | No | See 05 |
| Measurement Parameters | | | |
| Oxyhemoglobin concentration | Yes | Yes | No |
| Deoxyhemoglobin concentration | Yes | Yes | No |
| ITEM | K172492 | K042501 | DIFFERENCE |
| | OPTICAL TOPOGARAPHY SYSTEM ETG-4100 | OPTICAL TOPOGRAPHY SYSTEM ETG-4000 | |
| Total hemoglobin concentration | Yes | Yes | No |
| Light Source Configurations | | | |
| Light component | Laser diode | Laser diode | No |
| Wavelength | 695, 830nm | 695, 830nm | No |
| Output | 4mW/wavelength | 4mW/wavelength | No |
| Device Model | L9135-41 (695nm), Hamamatsu<br>HL8837MG-A (830nm), Ushio Opt | L9135-41 (695nm), Hamamatsu<br>HL8837MG-A (830nm), Ushio Opt | No |
| Detection method | Frequency modulated locked-in amplifier<br>(Frequency range: 12.0 to 23.8kHz) | Frequency modulated locked-in amplifier (Frequency range:<br>12.0 to 23.8kHz) | No |
| Sampling rate | 10Hz | 10Hz | No |
| Emitters | 18 | 18 | No |
| Laser safety | IEC 60825-1, Class 1M | IEC 60825-1, Class 1M | No |
| Source detector separation<br>distance | 30 mm for adult and child<br>20 mm for neonate, infant and child | 30 mm for adult and child<br>20 mm for neonate, infant and child | No |
| Detector | | | |
| Detector component | APD | APD | No |
| Detectors | 16 | 16 | No |
| Model | C8546, Hamamatsu | C8546, Hamamatsu | No |
| Maximum No. of Channels | 52 | 52 | No |
| Computer | | | |
| Operating system | Windows7 (64bit) | Windows2000 (32bit) | See 06 |
| Features | | | |
| Interface for the video recording | Yes | Yes | No |
| system | | | |
| 3D Positioning Unit | Yes | Yes | No |
| 3D Composite Display Unit | Yes | Yes | No |
| Task Presentation System<br>VFT | Yes | No | See 07 |
| Task Presentation System<br>Custom | Yes | No | See 07 |
| Report Display System VFT | Yes | No | See 08 |
| Report Display System Custom | Yes | No | See 08 |
| Double Density Function | Yes | No | See 09 |
| Multi-Distance ICA Function | Yes | No | See 010 |
| PCA Analysis | Yes | No | See 011 |
| Multi Data Analysis | Yes | No | See 011 |
| Waveform Analysis | Yes | No | See 011 |
| Real-time Output | Yes | No | See 012 |
{5}------------------------------------------------
#### Discussion of differences between the predicate device and the Optical Topography System ETG-4100 are explained in the table below.
| System Configuration | |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 01 | The differences in appearance, weight, and size have no effect on the safety and effectiveness of the device. |
| Holder and Probe Configurations | |
| 02 | The differences in size and/or optical fiber layout have not effect on the safety, effectiveness and performance of the device. |
| 03 | Changing the holder material to non-magnetic material while retaining the other specifications of the predicate holders does not effect on the safety, effectiveness and performance of the device. |
| Fiber Configurations | |
| 04 | This fiber is for Infant use with shape of tip smaller than previous fiber for adult. However, the other specifications are same as previous neonate/infant probe. Therefore, they have no effect on the safety, effectiveness and performance of the device. |
| 05 | These fibers are longer than previous fiber (normally 3.5m). The longer fibers have a smaller transparent which reduces the power to the patient. Therefore safety is improved with no effect on effectiveness. |
| Computer | |
| 06 | Changing from Windows2000 (32bit) to Windows7 (64bit) has no effect on the safety and effectiveness of the device. |
| Features | |
| 07 | These features present tasks to the patient during the measurement. Therefore, they have no effect on the safety, effectiveness and performance of the device. |
| 08 | These features display the measurement results in the above task presentation system. Therefore, they have no effect on the safety, effectiveness and performance of the device. |
| 09 | The Double Density option overlays a standard 3x5 or 4x4 fiber configuration with a second set of detectors and emitters resulting in a spatial resolution twice that of the standard configuration. Therefore, they have no effect on the safety and effectiveness and minimal effect on performance of the device. |
| 010 | Multi-Distance Measurement Function performs two different SD distance measurements and separates hemoglobin data into deep components and shallow blood flow components. Therefore, they have no effect on the safety and effectiveness and minimal effect on performance of the device. |
| 011 | These features provide additional data analysis techniques which no not affect the measurement function. Therefore, they have no effect on the safety, effectiveness and performance of the device. |
| 012 | This feature allows hemoglobin data to be sent in real time to the PC through the LAN during measurement. Therefore, they have no effect on the safety, effectiveness and performance of the device. |
{6}------------------------------------------------
Therefore, based on a thorough analysis and comparison of the Optical Topography System ETG-4100 and the predicate device, the technological characteristics do not impact safety and effectiveness.
## Substantial Equivalence
Substantial Equivalence was based on analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance testing which are summarized in the following table:
| ITEM | OVERALL RATIONALE ANALYSIS |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System Configuration | Based on that there are no significant differences from the predicate device, Hitachi judges that the |
| Holder and Probe Configurations | Optical Topography System ETG-4100 has no additional issues with safety and effectiveness. |
| Fiber Configurations | |
| Measurement Parameters | |
| Light Source Configurations | |
| Detector | |
| Computer | |
| Features | Based on results of design inspection as confirmed in the Declaration of Conformity, Hitachi judges<br>that the Optical Topography System ETG-4100 device has no additional issues with safety and<br>effectiveness. |
# Summary of Non-Clinical Testing
The Optical Topography System ETG-4100 is in conformance with the applicable parts of the following standards:
- 1. Signal stability testing for the subject device and predicate device were submitted to show that the new characteristics produce matching results and output.
- 2. Software verification and validation reports were submitted to account for the new features proposed in the subject device. These reports demonstrate that the subject device with these new features is substantially equivalent to the predicate since the methods acceptably demonstrate that the algorithm that achieves the intended use (measurement of the relative levels of cerebral deoxyhemoglobin and oxyhemoglobin) as compared to the predicate.
- 3. AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 Edition 3: 2007 4. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- 5. IEC 60825-1 Edition 2.0 2007-03
Safety of laser products - part 1: equipment classification, and requirements
- 6. IEC 62304:2006 Medical Device Software - Software Life Cycle Processes.
- 7. IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - part 1-6: general requirements for basic safety and essential performance - collateral standard: usability
- ISO 10993-1 Fourth edition 2009-10-15 8. Biological evaluation of medical devices - part 1: evaluation and testing within a risk management process [including: technical corrigendum
- 9. ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - part 5: tests for in vitro cytotoxicity. (Biocompatibility)
{7}------------------------------------------------
- 10. ISO 10993-10 Third Edition 2010-08-01
Biological evaluation of medical devices - part 10: tests for irritation and skin sensitization. (Biocompatibility)
#### Conclusions
The proposed Optical Topography System ETG-4100 is considered substantially equivalent to the currently marketed predicate device (Optical Topography System ETG-4000 (K042501) in terms of design features, fundamental scientific technology, intended use, and safety and effectiveness.
Based on the technological characteristics and performance testing to account for any differences in those characteristics, Optical Topography System ETG-4100 is substantially equivalent to the Optical Topography System ETG-4000 (K042501).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.