← Product Code [NKB](/productcode/NKB) · K172469

# TiLock Modular Spinal System (K172469)

_Genesys Spine · NKB · Nov 30, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172469

## Device Facts

- **Applicant:** Genesys Spine
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Nov 30, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Genesys Spine TiLock Modular Spinal System is intended for posterior, non-cervical (T1-S2/ilium) pedicle and nonpedicle spinal fixation, to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the racic, lumbar, and sacral spine: degenerative disc disease (DDD); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. When used for posterior non-cervical screw fixation in pediatric patients, Genesys Spine TiLock Modular Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the Genesys Spine TiLock Modular Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis / spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft. The Genesys Spine TiLock Modular Spinal System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion. When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the Genesys Spine TiLock Modular Spinal System is also intended for the following indications: degenerative disc disease (DDD); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion

## Device Story

Modular spinal fixation system; includes tulip heads, polyaxial screws (solid/cannulated), lock screws, rods. Used for posterior or anterolateral spinal stabilization. Implantation via open or over-the-wire technique using k-wires and fluoroscopy. Surgeon-operated in clinical/OR setting. Provides rigid immobilization of spinal segments to facilitate fusion. Benefits patient by stabilizing unstable spinal segments due to trauma, deformity, or degenerative disease.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance demonstrated via mechanical testing per ASTM F1717 (static/dynamic compression bending, static torsion) and ASTM F1798 (tulip pull-off).

## Technological Characteristics

Modular system; Ti-6Al-4V ELI titanium alloy (ASTM F136) and cobalt-chromium-molybdenum alloy (ASTM F1537). Polyaxial screw design; cannulated and solid options. Manual instrumentation for open or over-the-wire (fluoroscopy-guided) implantation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- TiLock² Spinal System ([K161914](/device/K161914.md))
- TiLock² Spinal System ([K171838](/device/K171838.md))
- TiLock Pedicle Screw System ([K100757](/device/K100757.md) / [K103671](/device/K103671.md))
- MAS PLIF Fixation System (NuVasive - [K142737](/device/K142737.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health & Human Services logo, which is a stylized depiction of a human figure embracing a globe. To the right of this is the FDA logo, which features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" written below in a smaller font size.

November 30, 2017

Genesys Spine Mr. Brian J. Bergeron Vice President of Engineering 1250 Capital of Texas Highway South Building 3 Suite 600 Austin, Texas 78746

Re: K172469

Trade/Device Name: TiLock Modular Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWP, KWQ Dated: November 15, 2017 Received: November 21, 2017

Dear Mr. Bergeron:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'neill -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172469

Device Name Genesys Spine TiLock Modular Spinal System

#### Indications for Use (Describe)

The Genesys Spine TiLock Modular Spinal System is intended for the following indications:

The Genesys Spine TiLock Modular Spinal System is intended for posterior, non-cervical (T1-S2/ilium) pedicle and nonpedicle spinal fixation, to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the racic, lumbar, and sacral spine: degenerative disc disease (DDD); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

When used for posterior non-cervical screw fixation in pediatric patients, Genesys Spine TiLock Modular Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the Genesys Spine TiLock Modular Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis / spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

The Genesys Spine TiLock Modular Spinal System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the Genesys Spine TiLock Modular Spinal System is also intended for the following indications: degenerative disc disease (DDD); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion

Type of Use (*Select one or both, as applicable*)

| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

# 510(K) SUMMARY

|                                                                                             | Primary                                                                                                                                                                                                                                                                     | Secondary                                        |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Submitter's Name:                                                                           | Genesys Spine                                                                                                                                                                                                                                                               | Genesys Spine                                    |
| Submitter's Address:                                                                        | 1250 Capital of Texas<br>Highway South                                                                                                                                                                                                                                      | 1250 Capital of Texas<br>Highway South           |
|                                                                                             | Building Three, Suite 600<br>Austin, Texas 78746                                                                                                                                                                                                                            | Building Three, Suite 600<br>Austin, Texas 78746 |
| Submitter's Telephone:                                                                      | 512-381-7071                                                                                                                                                                                                                                                                | 512-381-7080                                     |
| Submitter's Fax:                                                                            | 800-817-4938                                                                                                                                                                                                                                                                | 800-817-4938                                     |
| Contact Name:                                                                               | Brian J. Bergeron                                                                                                                                                                                                                                                           | William W. Sowers                                |
| Date Summary was<br>Prepared:                                                               | November 15, 2017                                                                                                                                                                                                                                                           |                                                  |
| Trade or Proprietary Name:                                                                  | Genesys Spine TiLock Modular Spinal System                                                                                                                                                                                                                                  |                                                  |
| Common or Usual Name:                                                                       | Spinal Fixation System                                                                                                                                                                                                                                                      |                                                  |
| Classification:                                                                             | Class II per:<br>21 CFR 888.3050 – Spinal interlaminal fixation orthosis<br>21 CFR 888.3060 – Spinal intervertebral body fixation<br>orthosis<br>21 CFR 888.3070 – Thoracolumbosacral pedicle screw<br>system                                                               |                                                  |
| Product Codes:                                                                              | NKB, KWP, KWQ                                                                                                                                                                                                                                                               |                                                  |
| Classification Panel:                                                                       | Orthopedic Devices Panel                                                                                                                                                                                                                                                    |                                                  |
| Legally Marketed<br>(unmodified) devices to<br>Which Substantial<br>Equivalence is Claimed: | Primary Predicate:<br>TiLock² Spinal System (Genesys Spine - K161914)<br>Additional Predicate(s):<br>TiLock² Spinal System (Genesys Spine - K171838)<br>TiLock Pedicle Screw System (Genesys Spine - K100757<br>/ K103671)<br>MAS PLIF Fixation System (NuVasive - K142737) |                                                  |

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The Genesys Spine TiLock Modular Spinal System is a spinal system that consists of a variety screws, lock screws, rods and associated general instruments. Implant components are available in a variety sizes and can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. The TiLock Modular Spinal System implants are manufactured from titanium alloy per ASTM F136 and F1472 and Cobalt Chrome per ASTM F1537.

{4}------------------------------------------------

## INDICATIONS FOR USE

The Genesys Spine TiLock Modular Spinal System is intended for the following indications:

The Genesys Spine TiLock Modular Spinal System is intended for posterior, noncervical (T1-S2/ilium) pedicle and non-pedicle spinal fixation, to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine: degenerative disc disease (DDD); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

When used for posterior non-cervical screw fixation in pediatric patients, Genesys Spine TiLock Modular Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally the Genesys Spine TiLock Modular Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis / spondylolysis, and fracture caused by tumor and/or trauma. Pediatric pedicle screw fixation is limited to a posterior approach and is intended to be used with autograft and/or allograft.

The Genesys Spine TiLock Modular Spinal System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.

When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the Genesys Spine TiLock Modular Spinal System is also intended for the following indications: degenerative disc disease (DDD); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

## TECHNICAL CHARACTERISTICS

The Genesys Spine TiLock Modular Spinal System is comprised of inter-changeable Tulip heads, polyaxial screws (solid and cannulated) in various lengths and diameters, lock screws, and rods in various lengths. The TiLock Modular cannulated polyaxial screws may be implanted via a conventional (open) technique or with an over-the-wire technique. Manual instrumentation for implantation of the system is available for both techniques. The over-the-wire procedure is performed using k-wires and fluoroscopy. The implantable components are manufactured from medical grade Ti-6Al-4V ELI titanium alloy per ASTM F136 and cobalt-chromium-molybdenum alloy per ASTM F1537.

{5}------------------------------------------------

## PERFORMANCE DATA

Nonclinical testing was performed to demonstrate that the subject Genesys Spine TiLock Modular Spinal System is substantially equivalent to other predicate device. The following testing was performed:

- Static Compression Bending per ASTM F1717 ●
- o Dynamic Compression Bending per ASTM F1717
- Static Torsion per ASTM F1717 ●
- Tulip Pull-off per ASTM F1798 ●

The results demonstrate that the subject Genesys Spine TiLock Modular Spinal System is substantially equivalent to the predicate.

## CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that Genesys Spine TiLock Modular Spinal System is substantially equivalent to legally marketed predicate devices for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K172469](https://fda-staging.innolitics.com/device/K172469)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
