XD880A Ultrasonic Osteotomy Surgical System is an ultrasonic surgical system that includes a handpiece and associated cutting tips intended for osteotomy, osteoplasty and drilling in a variety of surgical procedures, including but not limited to: - Otolaryngology - - Oral/maxillofacial - -Hand and foot - Neurosurgery (bone only) - Spine - Plastic/reconstructive. -
Device Story
Ultrasonic surgical system for bone cutting, osteotomy, osteoplasty, and drilling; comprises console with integrated peristaltic pump, handpiece, tip inserts, torque wrench, footswitch, and irrigation set. Console houses ultrasonic generator, power supply, and microprocessor; provides electrical energy to piezoelectric transducer in handpiece. Transducer converts electrical voltage into mechanical energy, inducing longitudinal micro-vibrations (<0.12mm amplitude) in tip inserts at resonant frequency. Used in OR by surgeons; console LCD interface allows selection/visualization of parameters. Irrigation fluid supplied to surgical site via peristaltic pump. System fragments/reshapes bone while sparing soft tissues. Clinical benefit: precise bone cutting with reduced risk of soft tissue/nerve damage. Output power and irrigation flow activated via footswitch.
Clinical Evidence
GLP animal study on 10 beagles undergoing spine surgery (T11-T13). Evaluated 15 tip types for efficiency (cutting time, bleeding volume) and safety (SEP/MEP neurophysiological monitoring, clinical observation, CT scan, histopathology). Results: average cutting time 10-12 min; low bleeding; no accidental nerve/vessel damage. SEP/MEP values normal; histopathology showed no spinal cord damage. Bench testing confirmed electrical safety (IEC 60601-1) and EMC (IEC 60601-1-2) compliance.
Technological Characteristics
Ultrasonic surgical system; piezoelectric transducer; longitudinal vibration <0.12mm amplitude. Materials: Titanium Alloy TC4 (ISO 5832-2/3) for tips; Medical Silicone Rubber (JIS K 6249) for irrigation components. Console includes peristaltic pump, LCD touch interface, microprocessor control. Power: electrical to mechanical energy conversion. Connectivity: standalone. Sterilization: not specified. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-1.
Indications for Use
Indicated for patients requiring bone cutting, osteotomy, osteoplasty, or drilling in otolaryngology, oral/maxillofacial, hand/foot, neurosurgery (bone only), spine, and plastic/reconstructive surgical procedures.
Regulatory Classification
Identification
A sonic surgical instrument is a hand-held device with various accessories or attachments, such as a cutting tip that vibrates at high frequencies, and is intended for medical purposes to cut bone or other materials, such as acrylic.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 888.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 24, 2018
Morley Research Consortium Dr. Suzan Davis, Pharm. D. RAC, EMBA, CQA Head of Regulatory Affairs and Compliance Four Concourse Parkway Suite 215 Atlanta, Georgia 30328
Re: K172464
Trade/Device Name: XD880A Ultrasonic Osteotomy Surgical System Regulation Number: 21 CFR 888.4580 Regulation Name: Sonic Surgical Instrument and Accessories/Attachments Regulatory Class: Class II Product Code: JDX, DZI, ERL, HBE, HWE Dated: March 15, 2018 Received: March 22, 2018
Dear Dr. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K172464
Device Name
XD880A Ultrasonic Osteotomy Surgical System
#### Indications for Use (Describe)
XD880A Ultrasonic Osteotomy Surgical System is an ultrasonic surgical system consisting of a handpiece and associated tips for cutting bone, osteotomy, osteoplasty and drilling in a variety of surgical procedures, including but not limited to: - Otolaryngology
- Oral/maxillofacial
- Hand and foot
- Neurosurgery
- Spine
- Plastic/reconstructive.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary in accordance with 21 CFR 807.92
# 1. Administrative Information
| Type of 510(k) submission: | Traditional |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Date: | July 31st, 2017 |
| 510(k) Submitter: | SMTP Technology Co., Ltd.<br>1F, 4F Building A, Emerging Industry Incubation Center<br>Zhangjiagang FTZ, Jiangsu, 215600, P.R. China<br>Phone: +86-512-82508100<br>Fax: +86-512-82508100 |
| 510(k) Contact Person: | Dr. Suzan Davis Pharm.D, RAC, EMBA, CQA<br>Head of Regulatory Affairs and Compliance<br>Morley Research Consortium<br>Four Concourse Parkway Suite 215<br>Georgia, Atlanta, 30328, USA<br>Phone: 781-672-4200<br>Fax:781-672-4201<br>sdavis@morleyrc.com |
| Manufacturer: | SMTP Technology Co., Ltd.<br>1F, 4F Building A, Emerging Industry Incubation Center<br>Zhangjiagang FTZ, Jiangsu, 215600, P.R. China<br>SMTP Technology Co., Ltd. |
| 2. Device | |
| Proprietary Name: | XD880A Ultrasonic Osteotomy Surgical System |
| Regulation Number: | 21 CFR 888.4580 |
| Device Classification Name: | Sonic Surgical instrument and accessories/attachments |
| Regulation Identification: | A sonic surgical instrument is a hand-held device<br>with various accessories or attachments, such as a<br>cutting tip, that vibrate at high frequencies, and it is<br>intended for medical purposes to cut bone or other<br>materials, such as acrylic. |
| Regulation Medical Specialty: | Orthopedic |
| Classification product codes: | JDX<br>DZI, ERL, HBE, HWE |
{4}------------------------------------------------
### Device Class:
Class II
### 3. Predicate Devices
The substantial equivalence of the subject device is based on the following legally marketed predicate devices
| Trade Name | Manufacturer | Product Code | 510(k) Number | Decision Date |
|----------------------|---------------------------|-------------------------------|---------------|---------------------|
| Piezosurgery Flex | MECTRON<br>Spa | JDX, DZI,<br>ERL, HBE,<br>HWE | K132848 | 29 November<br>2013 |
| Piezoelectric System | SATELECT-<br>Acteon Group | JDX, DZI,<br>ERL, HBE,<br>HWE | K100410 | 28 April 2010 |
## 4. Device Description
XD880A Ultrasonic Osteotomy Surgical System consists of a console (control unit) with an integrated peristaltic pump, a handpiece with a connecting cord, a range of tip inserts, a torque wrench, a footswitch and an irrigation set (liquid-flow tube and liquid-flow sleeve).
The console has a color LCD touch screen user interface for the selection/visualization of device functional parameters. The console activates and controls the ultrasound vibration, controls irrigation flow and displays system condition. Inside the console are located the ultrasonic generator, the electrical power supply module and the micro-processor electronic board that controls and supervises the functional parameters of the device.
The console is connected to the main power by an electrical cord. It includes connectors for the handpiece and for the footswitch. The console incorporates a peristaltic pump which provides, through the irrigation tubing set, a sterile fluid supply to the surgical site. Ultrasonic power and irrigation flow to the handpiece are simultaneously activated by pressing the footswitch. The handpiece contains a piezoelectric ultrasonic transducer which attaches to the generator (inside the console) by a cable at one end of the handpiece. Tip inserts are attached to the other end of the handpiece.
XD880A Ultrasonic Osteotomy Surgery System uses ultrasonic technology to generate mechanical micro-vibrations of the tip insert connected to the handpiece, the piezoelectric transducer converting the electrical voltage supplied by the ultrasonic generator into mechanical energy that induces vibration of the tip insert at the resonant frequency of the tip insert.
The tips are used to fragment and reshape bone tissue through longitudinal vibration at high frequency and small amplitude (less than 0.12mm), while keeping the soft tissues with elastic properties and free of damage.
{5}------------------------------------------------
## 5. Intended Use
XD880A Ultrasonic Osteotomy Surgical System is an ultrasonic surgical system that includes a handpiece and associated cutting tips intended for osteotomy, osteoplasty and drilling in a variety of surgical procedures, including but not limited to:
- Otolaryngology -
- Oral/maxillofacial -
- -Hand and foot
- Neurosurgery (bone only) -
- Spine -
- Plastic/reconstructive. -
## 6. Performance Testing
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the XD880A Ultrasonic Osteotomy Surgical System. The system complies with the IEC 60601-1:2006 standards for safety and the IEC 60601-1-2:2007/2010 standard for EMC.
### Biocompatibility
XD880A Ultrasonic Osteotomy Surgical System components that come into direct contact with patients during surgical procedures are the handpiece and tips inserts. According to ISO 10993-1. these components are classified as 'external communicating devices'. as they get in contact with 'tissue/bone/indirect blood path' for a duration less than 24 hours (< 24 h). The cutting tips are made of Titanium Alloy TC4 which meets the requirements of ISO 5832-2:1999 and ISO 5832-3:1996).
The other system components that can have indirect contact with the surface of patient's body during surgery are liquid-flow sleeve and liquid-flow tube, and the duration of their contact is less than 24 hours (≤24 hours). Liquid-flow sleeve and liquid-flow tube are made of Medical Silicone Rubber that meets the requirements of JISK 6249:2003.
Biocompatibility studies have been conducted on Titanium Alloy TC4 and Medical Silicone Rubber and studies results showed that both materials are biocompatible.
### Software Verification and Validation
XD880A Ultrasonic Osteotomy Surgery System Software verification and validation testing were conducted in line with the requirements of FDA Guidance 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices', 11 May 2005. The software for this device was considered as a "moderate level of concern " and software verification and validation reports have been included in this submission.
#### Mechanical and Acoustic Testing
{6}------------------------------------------------
Bench testing and acoustic testing were conducted on the subject device. Testing results demonstrated that the subject device performs within its specifications.
### Animal Study
A GLP animal study has been conducted with the 15 types of tips on 10 beagles. The objective of the study was to evaluate the efficiency, convenience and safety of the 15 tips during spine surgery. Animals were split into two groups of 5 animals each. The 15 types of tips were split between the two groups of animals. All animals were anesthetized with pentobarbital sodium and were cut at T11-T13 thoracic vertebral posterior and vertebral plate to reduce the pressure. Each animal was cut at T12 spinous process at the mid-line and cut through the subcutaneous tissue. After separating the attached dorsal muscles from the spinous process, the vertebral plate and T11-T13 flat joints, the vertebral plate was cut with XD880A Ultrasonic Osteotomy Surgical Tips.
All the operating procedures were performed in the same conditions and with same system parameters (power: level 5. pulse: level 6. wash flow: level 4). The set-up parameters were at their maximum value to evaluate the efficiency and safety of the tips when operating at the maximum conditions.
The tips' efficiency was evaluated by analyzing the cutting time of the bone and bleeding volume when using each tip on the animals. The tips' safety was evaluated by monitoring and recording neurophysiological parameters of each animal prior, during and after the surgery. The Sensory Evoked Potential (SEP) latent period and fluctuation, and Motor Evoked Potential (MEP) were used to evaluate the neurophysiological parameters. SEP and MEP values were recorded for both back legs and vertebral plate before the operation (Phase I), at the start of the cutting process (Phase II), after exposure of spinal dura mater (Phase III) and after the end of operation (Phase IV). Additionally, a clinical observation of animals' mental state and behavior was performed on all animals on day 1, day 3 and day 7 after operation. CT scan was performed on day 7 and histopathology was performed on day 8 after operation.
Study results demonstrated that all tips were efficient and safe when used at the recommended conditions. The average cutting time was 12min and 2 second and 10min and 24 second in the first and second group of animals respectively. The average bleeding volume was 31.4mL and 28.6mL in the first and second group of animals respectively. The was normal and the bleeding was low during surgeries, except for one animal, where the cutting time was longer than usual, but the delay was not related to the device. The cutting effect of the tips on hard bone was good. No accidental damage was caused to nerves, vessels or soft tissues during surgeries. The bone cutting was performed from inside the spinal canal to the outside, which reduced the surgery risk significantly.
SEP latent period and fluctuation and MEP values were normal for all animals. This means that no damage was caused to animals' motor and sensory reflection during surgeries. Except for two
{7}------------------------------------------------
animals from the first group, who showed respectively an SEP with latent period longer than 10% for both sides before and after cutting, and an SEP fluctuation greater than 50% on the left sides after cutting, when compared to baseline values. These abnormal SEP values were probably related to the anesthesia as the animal returned to the normal state after completion of the surgery.
Clinical evaluation using the Tarlov rating standard process on day 1, day 3 and day 7 after surgery recorded a score of 4 for all animals, except for one animal who had a poor mental state during the first two days following surgery, but he returned to the normal state on the third day after surgery. No any other abnormal clinical symptoms were observed on the other animals.
CT scan performed on all animals on day 7 after surgery showed that the spine was in order and the curvature was normal. The posterior part of the vertebral plate of T11-T13 was missing, and an edema of soft tissue was observed. The soft tissue edema was a postoperative complication and wasn't caused by the tips themselves. The spinal cord near the spinal canal was not swelling, its density was normal and no obvious anomalies were seen in the spinal canal.
Histopathology examination performed on day 8 after surgery, showed no histopathological damage of the spinal cord tissue. Some inflammatory cells were observed in the specimens of the spinal cord membrane of 5 animals, and they were related to normal tissue reaction to surgery and not to the device.
# 7. Substantial Equivalence
Many of the features and technical characteristics of XD880A Ultrasonic Osteotomy Surgical System (subject device) are identical to those of the primary or secondary predicate devices (PD1 and PD2), and where there are differences, such differences do not have an impact on the safety or effectiveness of the subject device.
The differences are mainly in the dimensions of the systems components or number of operatorusable handpieces. Other differences, such as user interface design and minimum/maximum irrigation flow rates, provide more flexibility for users. Justification is provided in this submission for substantial equivalence for mentioned differences, including indications for use/intended use, and output power.
# 8. Conclusion
Based on the information contained within this submission, it is concluded that the XD880A Ultrasonic Osteotomy Surgical System is substantially equivalent to the identified predicate devices which are already in commercial distribution in the United States.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.