TENS device-HeadaTerm, eEspress

K172450 · Wat Medical Technology (Ningbo) Co., Ltd. · PCC · Sep 13, 2018 · Neurology

Device Facts

Record IDK172450
Device NameTENS device-HeadaTerm, eEspress
ApplicantWat Medical Technology (Ningbo) Co., Ltd.
Product CodePCC · Neurology
Decision DateSep 13, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5891
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TENS device-HeadaTerm is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older.

Device Story

TENS device-HeadaTerm is a transcutaneous electrical nerve stimulator for migraine prophylaxis. Device adheres to patient's forehead via conductive paste; delivers low-frequency electrical impulses to supraorbital and supratrochlear nerves to modulate pain signaling. User operates device via host key to ramp up intensity to comfort level; device locks voltage. Powered by 3V lithium coin cell. Used in home setting by patient. Output provides drug-free migraine prevention. Biocompatible materials used for skin contact. Conductive paste requires replacement after 7 uses.

Clinical Evidence

No clinical data provided. Substantial equivalence established via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance of nerve/muscle stimulators (IEC 60601-2-10), home use standards (IEC 60601-1-11), and biocompatibility (cytotoxicity, irritation, sensitization).

Technological Characteristics

Transcutaneous electrical nerve stimulator; AC symmetric square wave; 60Hz frequency; 250µs phase duration. Powered by 3V lithium coin cell. Dimensions 127x35x12mm; weight 11g. Biocompatible patient contact materials. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11.

Indications for Use

Indicated for prophylactic treatment of episodic migraine in patients 18 years of age or older.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety. (3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm 2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted. (5) Appropriate software verification, validation, and hazard analysis must be performed. (6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population. (7) Labeling must include the following: (i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator. (ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery. (iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown. (iv) A summary of the expected risks and benefits of using the device. (v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications. (vi) Information on how the device operates and the typical sensations experienced during treatment. (vii) A detailed summary of the device technical parameters. (viii) An expiration date/shelf life for the electrodes and the number of times they can be reused. (ix) Disposal instructions.

In combination with the general controls of the Food Drug & Cosmetic Act, the Transcutaneous Electrical Nerve Stimulator to Treat Headache is subject to the following special controls:

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. September 13, 2018 WAT Medical Technology Inc. % Rodney Zhang CTO WAT Medical Technology Inc. Room703-711, No.2 North Taoyuan Road Ningbo, 315600 China Re: K172450 Trade/Device Name: TENS device-HeadaTerm, eEspresso Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous electrical nerve stimulator to treat headache Regulatory Class: Class II Product Code: PCC Dated: August 1, 2017 Received: August 14, 2017 Dear Rodney Zhang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Timothy A. Marjenin -S For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K172450 #### Device Name TENS device-HeadaTerm, eEspresso Indications for Use (Describe) The TENS device-HeadaTerm is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" in green and pink, followed by the text "WAT Med" in black. The text is in a bold, sans-serif font. # SECTION 1 # SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 1 - 1 of 7 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for WAT Med. The logo features a stylized checkmark symbol in blue, green, and pink. To the right of the checkmark is the text "WAT Med" in a bold, sans-serif font. ### 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS (As Required by 21 CFR 807.92) ### 1. Date Prepared [21 CFR 807.92(a)(1)] 09/11/2018 #### 2. Submitter's Information [21 CFR 807.92(a)(1)] | Company Name: | WAT Medical Technology Inc. | |------------------|-----------------------------------------------------------------------------------| | Company Address: | Room703-711, No.2 North Taoyuan Road, 315600,<br>Ningbo, Zhejiang Province, P.R.C | | Contact Person: | Dr. Rodney Zhang | | Phone: | 86-574-6506-0811 | | Email: | 13500799711@139.com | #### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] | Trade Name: | TENS device-HeadaTerm, eEspresso | |--------------------|--------------------------------------------------------------| | Common Name: | Transcutaneous electrical nerve stimulator to treat headache | | Model No. | YF-HT-W1. | | Product Code: | PCC | | Regulation Number: | 21 CFR 882.5891 | | Device Class: | II | #### 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[ The identification of predicates within this submission is as follow: | Predicate I | | |----------------------|-----------------------------------------------------------------| | Manufacturer: | STX-Med SPRL | | Trade Name: | Cefaly | | Product Code: | PCC | | Classification Name: | Transcutaneous electrical nerve stimulator to treat<br>headache | | Regulation Number: | 21 CFR 882.5891 | | Classification: | Class II | | FDA 510 (k) #: | K122566 (Primary), K160237 | #### 5. Description of the Device [21 CFR 807.92(a)(4)] TENS device-HeadaTerm offers a drug-free scheme, which introduces electric impulses to act on the supraorbital nerve and supratrochlear nerve, the very nerves that transmit migraine pain, to treat and prevent migraine headaches. TENS device-HeadaTerm enables an electronic feedback mechanism to adjust the electrical ### SECTION 1 - 2 of 7 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" made up of blue, green, and pink elements, followed by the text "WAT Med" in bold, black letters. The logo is simple and modern, with a focus on the company name. impulses to the specific requirements of the individual user for personalized efficacy and comfort. The device could adhere to patients forehead with conductive paste, the current passing through the electrodes and release current stimulate the nerve underneath patients' forehead skin; the stimulation would alleviate migraine headaches, and during ramping up, patients could choose comfortable intensity level by pressing the host key to lock up the current voltage, and the patient contact material is biocompatibility safe. The conductive paste could be replaced once the paste lost its adhesiveness and patients are recommended to replace the conductive paste after 7 times of treatment uses. #### 6. Previous Device Names and Company: Previous Device Name: Other names appears in various test reports including but not limit to Biocompatibility reports, electrical safety, EMC are recorded as HeadaTerm-wet, HeadaTerm-wet(shell), HeadaTerm-wet device for preventing and treating primary headache, these names are the previous names being used for proposed products. Previous Company: The company appears in various test reports is WAT Medical Technology (Ningbo) Co., Ltd, which is the same company as the current WAT Medical Technology Inc. #### 7. Indications for Use [21 CFR 807.92(a)(5)] The TENS device-HeadaTerm is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older. #### 8. Technological Characteristic [21 CFR 807.92(a)(6)] By releasing the low frequency pulse with a particular frequency and reaching the advanced nerve center of cerebral cortex via nervus supraorbitalis, they can adjust all signals causing headache which come from the biological and physical channels, so as to stop or postpone the transmission of headache signal to cerebral center. | Clinical Testing | Not Applicable | |----------------------|---------------------------------------------| | Non-Clinical Testing | Electrical Safety | | | Electromagnetic Safety | | | Performance of nerve and muscle stimulators | | | Performance of Device for Home Use | #### 9. Clinical Testing and Non-Clinical Testing ### SECTION 1 - 3 of 7 {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for WAT Med. The logo features a stylized "W" in blue, green, and pink, followed by the text "WAT Med" in bold, black letters. The logo is simple and modern, and the colors are bright and eye-catching. | | Biocompatibility: Cytotoxicity | |--|---------------------------------| | | Biocompatibility: Irritation | | | Biocompatibility: Sensitization | The non-clinical testing conducted are listed above, and all tests come back with "Pass" results which demonstrating the same safety and effectiveness with the predicate device. #### 10. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)] | | | Proposed Device | Predicate Device | |---------------------------------------------------------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------| | Product Name | | TENS device-HeadaTerm | CEFALY | | Product Code | | PCC | PCC | | Regulation No. | | 21 CFR 882.5891 | 21 CFR 882.5891 | | Classification | | Class II | Class II | | Intended Use | | The TENS device-HeadaTerm is<br>indicated for the prophylactic<br>treatment of episodic migraine in<br>patients 18 years of age or older. | The Cefaly® device is indicated for<br>the prophylactic treatment of<br>episodic migraine in patients 18<br>years of age or older. | | 510 (k) Number | | N/A | K160237, K122566 | | Device Feature | | | | | Device Name | | TENS device-HeadaTerm | CEFALY | | Power Source | | One 3V lithium coin cells | One lithium battery | | Follow Current | | Yes | Yes | | Voltage Overload<br>Detection | | Yes | Yes | | Adjustable Intensity | | ON/OFF button on front<br>of device | ON/OFF button on front<br>of device | | Channel | | 1 | 1 | | | Start | Yes | Yes | | Operatio<br>n Tips | Low<br>Battery | No | Yes | | | Work | Yes | Yes | | Software-controlled | | No | Yes | | Time Set | | No - Manually Adjust | No - Manually Adjust | | Constant Current | | No | No | | Automatic overload<br>trip voltage level | | Yes | Yes | | Patient override<br>control method | | On/Off button on front of device | On/Off button on front of device | | Indicator displays | | Unit functioning<br>Electrical connection | Unit functioning<br>Low battery<br>Electrical connection | | Timer Setting | | Yes | Yes | | Weight | | 11g | 30g | | Dimensions | | 127x35x12mm | 160 x 170 x 40 mm | | Waveform | | AC Symmetric Square<br>Wave | AC Symmetric Square<br>Wave | | Phase Duration<br>(µsec) | | 250us(±20) | 250us(±0.5%) | | Phase Interval | | 10us(±0.5%) | 10us(±0.5%) | | Pulse Period | | 500us(±0.5%) | 505us(±0.5%) | | Frequency (Hz) | | 60Hz(±1) | 60Hz(±0.5%) | | Net Charge (µC) per<br>pulse | | 0 | 0 | | Maximum output<br>voltage (V): | | 8.00V(±0.5%) | 8.00V(±0.5%) | | @500 ohms | | | | | Maximum output<br>voltage (V):<br>@2000 ohms | 32V(±0.5%) | 32V(±0.5%) | | | Maximum output<br>voltage (V):<br>@10000 ohms | 60V(±3) | 60V(±0.5%) | | | Maximum output<br>current (mA):<br>@500 ohms | 16mA(±0.5%) | 16mA(±0.5%) | | | Maximum output<br>current (mA):<br>@2000 ohms | 16mA(±0.5%) | 16mA(±0.5%) | | | Maximum output<br>current (mA):<br>@10000 ohms | 6mA(±0.5%) | 6mA(±0.5%) | | | Maximum phase<br>charge (μC) @500Ω | 3.95 | 4 | | | Maximum Current<br>Density, (mA/cm2,<br>r.m.s.) @500Ω | 2.35 | 2.37 | | | Maximum Average<br>Power Density,<br>(W/cm2) @500Ω | 0.000017 | 0.000017 | | | Maximum Average<br>Current (average<br>absolute value, mA)<br>@500Ω | 0.48 | 0.48 | | | IEC60601-1 | Yes | Yes | | | IEC60601-1-2 | Yes | Yes | | | IEC60601-2-10 | Yes | Yes | | | IEC 60601-1-11 | Yes | No | | ### SECTION 1 - 4 of 7 {7}------------------------------------------------ # W. WAT Med ## SECTION 1 - 5 of 7 {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" in blue, green, and pink, followed by the text "WAT Med" in black. The text is in a bold, sans-serif font. ## SECTION 1 - 6 of 7 {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for WAT Med. The logo consists of a stylized "W" symbol in blue, green, and pink, followed by the text "WAT Med" in bold, black letters. The "W" symbol appears to be made up of three curved lines, with the left line in blue, the middle line in green, and the right line in pink. #### 11. Conclusion [21 CFR 807.92(b)(3)] TENS device-HeadaTerm are substantially equivalent to predicate device because both devices use identical technology and same intended use, also testing standards are identical with the predicates. The differences between both devices are insignificant in terms of safety and effectiveness. ## SECTION 1 - 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...