K172409 · Tzora Active Systems , Ltd. · INI · May 2, 2018 · Physical Medicine
Device Facts
Record ID
K172409
Device Name
Lite
Applicant
Tzora Active Systems , Ltd.
Product Code
INI · Physical Medicine
Decision Date
May 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lite is a mobility assistive device for indoor use on pavements or mild terrain. It is not used as a transportation vehicle on roads and freeways used by cars.
Device Story
Electrically powered, foldable scooter for individuals with mobility limitations; enables indoor and outdoor travel on mild terrain. User operates via control panel featuring key switch, forward/reverse lever, speed adjustment knob, battery gauge, and LED status indicator. Transaxle motor drives rear wheels; electromagnetic brake provides stopping. Foldable frame, steering column, and removable battery pack facilitate storage and portability. User must possess sufficient arm/leg strength for independent entry/exit and steering. LED indicator provides status feedback (steady = normal; blinking = error code). Device benefits include increased mobility and independence for users capable of walking but needing assistance.
Clinical Evidence
Bench testing only. No clinical data. Device performance verified against ISO 7176 series standards, including static/dynamic stability, braking effectiveness, energy consumption, obstacle climbing, and electromagnetic compatibility.
Technological Characteristics
Foldable frame (stainless steel), seat (polypropylene), upholstery (vinyl-covered foam), armrests (foam). Transaxle motor drive; 24V system (2x 12V 12Ah sealed lead acid or lithium batteries). 4-wheel configuration (flat-free PU 8x2"). Compliant with ISO 7176-1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, 14, 16, 21, 25.
Indications for Use
Indicated for individuals able to walk but suffering from minor mobility limitations requiring a mobility assistive device for indoor and outdoor use on mild terrain. Not for use on roads or freeways.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is blue and white.
May 2, 2018
Tzora Active Systems Ltd. % Moshe Rosenberg Regulatory Consultant A. Stein Regulatory Affairs Consulting Ltd. 20 Hataas St., Suite 102 Kfar Saba, 4442520 Il
Re: K172409 Trade/Device Name: Lite Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: March 28, 2018 Received: April 2, 2018
Dear Moshe Rosenberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172409
Device Name Lite
Indications for Use (Describe)
The Lite is a mobility assistive device for indoor use on pavements or mild terrain. It is not used as a transportation vehicle on roads and freeways used by cars.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
K172409 (Premarket Notification [510(k)] Number)
#### 1. Submitter Information
| Manufacturer Name and Address | Tzora Active Systems Ltd. |
|-------------------------------|-----------------------------------------------|
| | Kibbutz Tzora, |
| | 9980300, |
| | Israel |
| Official Correspondent | Moshe Rosenberg |
| | A. Stein - Regulatory Affairs Consulting Ltd. |
| | 20 Hata'as St. (Beit Hapaamon, Suite 102) |
| | Kfar Saba 4442520, |
| | Israel |
#### Date Prepared: May 2, 2018 2.
#### 3. Device Name Lite
| Proprietary Name: | Lite |
|-----------------------------|-----------------------------------------------|
| Common Name: | Vehicle, Motorized 3-Wheeled |
| FDA Classification<br>Name: | 21 CFR 890.3800; Vehicle, Motorized 3-Wheeled |
| FDA Classification: | Class II, Product Code INI |
#### Predicate Devices 4.
The Lite is substantially equivalent to the following devices:
| Manufacturer | Device | 510(k) | Date Cleared |
|---------------------------|--------|---------|-------------------|
| Tzora Active Systems Ltd. | Elite | K052204 | February 15, 2006 |
#### Device Description 5.
The Lite scooter is an electrically powered scooter. It is intended to be used by individuals that are able to walk, but suffer from mobility limitations. The user must have sufficient arm and leg strength to get on and off the Lite alone and to safely steer under all driving conditions.
The Lite is intended for indoor use and restricted outdoor use on pavements or paved footpaths only.
{4}------------------------------------------------
The Lite can be folded. This allows for easy storage and enables portability of the Lite.
The control panel houses all the controls for operating the device. The control panel has a key switch for turning the device on and off, a lever for forward/reverse driving and a knob for speed adjustment. The control panel also contains the control indicator for the status of the device and a battery gauge.
The LED control indicator shows the status of the scooter: steady light mean that all is well, blinking indicates an issue. The number of flashes indicate the type of issue. The User Manual contains a troubleshooting section where each type of error and its solution is specified.
The steering column can be adjusted and put it in the position which is most comfortable for the operator.
The armrests can be lifted to enable easy entry and exit from the scooter.
#### 6. Indications for Use
The Lite is a mobility assistive device for indoor and outdoor use on mild terrain. It is not used as a transportation vehicle on roads and freeways used by cars.
These indications for use and intended use are identical to those of the predicate device.
#### 7. Performance Testing
Testing was performed with the Lite to be in conformance with the following standards:
- ISO 7176-1:2014 Wheelchairs Part 1: Determination of static stability .
- ISO 7176-2:2001 Wheelchairs Part 2: Determination of dynamic stability of electric . wheelchairs
- ISO 7176-3:2012 Wheelchairs Part 3: Determination of effectiveness of brakes .
- ISO 7176-4:2008 Wheelchairs Part 4: Energy consumption of electric wheelchairs . and scooters for determination of theoretical distance range
- . ISO 7176-5:2008 Wheelchairs – Part 5: Determination of overall dimensions, mass and manoeuvring space
- ISO 7176-6:2001 Wheelchairs Part 6: Determination of maximum speed, . acceleration and deceleration of electric wheelchairs
{5}------------------------------------------------
- ISO 7176-7:1998 Wheelchairs Part 7: Measurement of seating and wheel dimensions .
- ISO 7176-8:2014 Wheelchairs Part 8: Requirements and test methods for static, . impact and fatigue strengths
- . ISO 7176-9:2009 Wheelchairs – Part 9: Climatic tests for electric wheelchairs
- ISO 7176-10:2008 Wheelchairs Part 10: Determination of obstacle-climbing ability . of electrically powered wheelchairs
- ISO 7176-11:2012 Wheelchairs Test dummies .
- ISO 7176-13:1989 Wheelchairs Part 13: Determination of coefficient of friction of . test surfaces
- . ISO 7176-14:2008 Wheelchairs – Part 14: Power and control systems for electrically powered wheelchairs and scooters - requirements and test methods
- ISO 7176-15:1996 Wheelchairs Part 15: Requirements for information disclosure, . documentation and labeling
- ISO 7176-16:2012 Wheelchairs Part 16: Resistance to ignition of postural support . devices
- ISO 7176-21:2009 Wheelchairs Part Requirements and test methods for ● electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
- . ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs
#### Technological Characteristics Compared to Predicate Device 8.
The following table shows a comparison between the Lite and the predicate Elite.
{6}------------------------------------------------
# COMPARISON TABLE
| Technological<br>Characteristic | Lite<br>Tzora Active Systems Ltd.<br>(K172409) | Elite<br>Tzora Active Systems Ltd.<br>(K052204) |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code,<br>Class | INI,<br>Class II | INI,<br>Class II |
| Indications for Use | The Lite is a mobility assistive device for indoor use<br>and outdoor use on mild terrain.<br>It is not used as a transportation vehicle on roads and<br>freeways used by cars. | The Elite is a mobility assistive device for indoor use<br>and outdoor use on mild terrain.<br>It is not used as a transportation vehicle on roads and<br>freeways used by cars. |
| Target Population | Individuals that are able to walk but suffer from minor<br>mobility limitations. | Individuals that are able to walk but suffer from minor<br>mobility limitations. |
| Environment Used | Indoor and outdoor | Indoor and outdoor |
| Energy Used /<br>Delivered | No energy delivered | No energy delivered |
| Design: | The Lite consists of a foldable frame, foldable steering<br>column and removable battery pack. | The Elite consists of a foldable frame, foldable and<br>detachable steering column and removable battery<br>pack. |
| - Mechanism of<br>Action | Transaxle motor drive unit, propelling both rear<br>wheels. | Pancake hub motor forming the front wheel. |
| - Components | - 2 Front Wheels - Flat Free PU 8x2"<br>- 2 Rear Wheels - Flat Free PU 8x2"<br>- Transaxle Motor<br>- Controller<br>- Frame<br><br>- 2x 12V 12A/h Sealed Lead Acid Batteries<br>- 2x 12V 12 A/h Lithium Batteries (optional) | - 1 Front Wheel - Flat Free PU 10x2"<br>- 2 Rear Wheels - Flat Free PU 8x2"<br>- Pancake hub motor<br>- Controller<br>- Front Frame<br>- Rear Frame<br><br>- 2x 12V 12A/h<br>- 2x 12V 7A/h (*discontinued) |
| Technological | Lite | Elite |
| Characteristic | Tzora Active Systems Ltd. | Tzora Active Systems Ltd. |
| | (K172409) | (K052204) |
| - Features | Maximum Speed 3.7 mph | Maximum Speed: 3.7 mph |
| | Turning Radius: 35" | Turning Radius: 37" |
| | Carrying Capacity: 250 lbs | Carrying Capacity: 250 lbs |
| | Climbing Slope: 6° | Climbing Slope: 6° |
| | Curb Clearance: 1.6" | Curb Clearance: 2" |
| - Dimensions | 39x22x35" | 43x22x35" |
| - Weight | Total: 66 lbs (Lead Acid), 58 lbs (Lithium) | Total: 64 lbs |
| | Frame: 48 lbs | Front Column: 24 lbs |
| | | Rear Frame: 27 lbs |
| | Battery Pack: 18 lbs (12A/h Lead Acid) | Battery Pack: 18 lbs (12A/h Lead Acid) |
| | 10 lbs (12A/h Lithium) | |
| - Performance | Range: up to 8 mi (Lead Acid) | Range: up to 8 mi (Lead Acid) |
| | up to 9.8 mi (Lithium) | |
| - Human Factors | Control Panel consists of: | Control Panel consists of: |
| | - On/Off Key Switch | - On/Off Key Switch |
| | - Control Lever Forward/Reverse | - Control Lever Forward/Reverse |
| | - Speed Adjustment Knob | - Speed Adjustment Knob |
| | - Battery Gauge | - Battery Gauge |
| | - Status Indicator LED | - Status Indicator LED |
| - Standards Met | ISO 7176 series safety standards | ISO 7176 series safety |
| - Materials | Frame: Stainless Steel | Frame: Stainless Steel |
| | Seat: Polypropylene | Seat: Polypropylene |
| | Upholstery: Vinyl Covered Foam | Upholstery: Vinyl Covered Foam |
| | Armrests: Foam | Armrests: Foam |
| - Biocompatibility | Materials are biocompatible | Materials are biocompatible |
| Technological<br>Characteristic | Lite<br>Tzora Active Systems Ltd.<br>(K172409) | Elite<br>Tzora Active Systems Ltd.<br>(K052204) |
| Compatibility<br>With the<br>Environment and<br>Other Devices | The Lite is compliant with the ISO 7176-21 standard. | The Elite is compliant with the IEC 60601-1-2 (EMC<br>Compatibility) standard. |
| Electrical Safety | Power Requirements:<br>110-120 VAC / 60 Hz<br>220-240 VAC / 50 Hz<br>The Lite is compliant with the ISO 7176 series safety<br>standards. | Power Requirements:<br>110-120 VAC / 60 Hz<br>220-240 VAC / 50 Hz<br>The Elite is compliant with the EN 12184 (Electrical<br>Safety) standard. |
| Mechanical Safety | The Lite is compliant with the ISO 7176 series safety<br>standards. | The Elite is compliant with the EN 12184 (Electrical<br>Safety) standard. |
| Chemical Safety | Not Applicable | Not Applicable |
| Thermal Safety | The Lite is compliant with the ISO 7176 series safety<br>standards. | The Elite is compliant with the EN 12184 (Electrical<br>Safety) standard. |
| Radiation Safety | The Lite is compliant with the ISO 7176-21 standard. | The Elite is compliant with the IEC 60601-1-2 (EMC<br>Compatibility) standard. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
## Control Panel
The Lite has the same control panel as the predicate device.
### Maximum speed
The Lite can reach the same maximum speed of 3.7 mph as the predicate device (K052204). Also the minimum braking distance remained the same: 0.7 m.
## Batteries
The Lite batteries are identical to the batteries of the predicate device: two 12V 12A/h sealed lead acid batteries.
The Lite can also be provided with two 12V 12A/h lithium batteries.
## Seat
The Lite has the same seat and armrests as the predicate device.
### Frame
The Lite has a foldable frame, where the seat and the steering column can be folded down. The Elite has a foldable and detachable frame, where the steering column can be folded with the frame or it can be detached from the frame.
### Wheels
The Lite has 4 wheels (2 front, 2 rear wheels), whereas the predicate device has 3 wheels (1 front, 2 rear wheels). The 4-wheel design is identical to the previously cleared EasyTravel (K012592).
### Motor
The Lite is propelled by a transaxle motor with electromagnetic brake. This motor unit is attached to the rear of the frame and has the rear wheels attached to it. The predicate device is propelled by a pancake motor placed, with the electromagnetic brake, in the front drive wheel.
{10}------------------------------------------------
The differences in the above specifications do not adversely affect the safety and effectiveness and performance of the Lite. The Lite was tested to be in conformance with the in section 8 aforementioned standards and found compliant.
| Standard | Subject: Lite (K172409) | Predicate: Elite (K052204) | Equivalent? |
|-------------|-------------------------|----------------------------|-------------|
| ISO 7176-1 | Passed | Passed | Yes |
| ISO 7176-2 | Passed | Passed | Yes |
| ISO 7176-3 | Passed | Passed | Yes |
| ISO 7176-4 | Passed | Passed | Yes |
| ISO 7176-5 | Passed | Passed | Yes |
| ISO 7176-6 | Passed | Passed | Yes |
| ISO 7176-7 | Passed | Passed | Yes |
| ISO 7176-8 | Passed | Passed | Yes |
| ISO 7176-9 | Passed | Passed | Yes |
| ISO 7176-10 | Passed | Passed | Yes |
| ISO 7176-11 | Passed | Passed | Yes |
| ISO 7176-13 | Passed | Passed | Yes |
| ISO 7176-14 | Passed | Passed | Yes |
| ISO 7176-16 | Passed | Passed | Yes |
| ISO 7176-21 | Passed | Passed | Yes |
| ISO 7176-25 | Passed | N/A | N/A |
The technological characteristics, e.g., overall design, materials, mechanism of action, mode of operation, performance characteristics, etc., and the indications for use of the Lite are substantially equivalent to the predicate device cited above.
#### 9. Conclusion
The performance testing and comparison to the predicate device demonstrate that the Lite is as safe, as effective and performs as well as the legally marketed Elite predicate device. Therefore, the Lite is substantially equivalent to the Elite and may be legally marketed in the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.