K172395 · Sofradim Production · FTL · Feb 1, 2018 · General, Plastic Surgery
Device Facts
Record ID
K172395
Device Name
Duatene bilayer mesh
Applicant
Sofradim Production
Product Code
FTL · General, Plastic Surgery
Decision Date
Feb 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Duatene™ Bilayer Mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists. It is indicated for the open repair of groin hernia defects.
Device Story
Duatene™ Bilayer Mesh is a three-dimensional, one-piece knitted polypropylene monofilament surgical mesh. It consists of a circular/elliptic posterior layer for preperitoneal placement and an oblong anterior layer for onlay placement, linked by crossing threads. The device is used in an extraperitoneal site via an open surgical approach to reinforce the groin region. It provides long-term soft tissue reinforcement for hernia repair. The device is intended for use by surgeons in an operating room setting. It functions as a permanent implant to support tissue integrity.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via bench testing and animal models. Bench tests evaluated pore size, thickness, surface density, bursting strength/distension, breaking strength, elongation at break, tear strength, and suture pull-out strength. Animal studies confirmed equivalent tissue integration compared to the predicate. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Three-dimensional polypropylene monofilament textile; knitted one-piece construction. Bilayer design with differentiated knitting patterns. Sterilization method not specified. Compliant with ISO 10993-1 for biocompatibility. Mechanical properties assessed via bench testing.
Indications for Use
Indicated for patients requiring reinforcement of abdominal wall soft tissue where a weakness exists, specifically for the open repair of groin hernia defects.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.
## February 1, 2018
Covidien (Sofradim Production) % Ms. Mary Mellows Regulatory Affairs Manager 60 Middletown Avenue North Haven, Connecticut 06473
# Re: K172395
Trade/Device Name: Duatene™ Bilayer mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: January 2, 2018 Received: January 3, 2018
# Dear Ms. Mellows:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172395
Device Name Duatene™ Bilayer Mesh
Indications for Use (Describe)
Duatene™ Bilayer Mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists. It is indicated for the open repair of groin hernia defects.
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared: | August 4, 2017 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Sofradim Production (subsidiary of Covidien llc)<br>116, avenue du Formans<br>01600 Trevoux, France<br>Telephone: +33 (0)4 74 08 90 00<br>Fax: +33 (0) 4 74 08 90 02 |
| Contact: | Jose Marquez<br>Regulatory Affairs Manager<br>Covidien<br>60 Middletown Avenue<br>North Haven, CT 06473, USA<br>Telephone: (203) 492-7290<br>Fax: (203) 492-5029<br>Email: jose.a.marquez@medtronic.com |
| Name of device: | Duatene™ Bilayer Mesh<br>Trade/Proprietary name: Surgical Mesh<br>Common name: Mesh, Surgical, Polymeric<br>Classification name: Panel number and product code: 79 FTL<br>Regulation number: 21 CFR 878.3300 |
| Predicate Device: | Prolene™ (Polypropylene) Hernia System<br>Trade/Proprietary name: Surgical Mesh<br>Common name: Mesh, Surgical, Polymeric<br>Classification name: Panel number and product code: 79 FTL<br>Regulation number: 21 CFR 878.3300<br>510(k) Number: K984220<br>Manufacturer: Ethicon |
| Reference Device: | Parietene™ Macroporous Mesh<br>Trade/Proprietary name: Surgical Mesh<br>Common name: Mesh, Surgical, Polymeric<br>Classification name: Panel number and product code: 79 FTL<br>Regulation number: 21 CFR 878.3300<br>510(k) Number: K142091<br>Manufacturer: Sofradim Production |
{4}------------------------------------------------
| Reference Device: Trade/Proprietary name: | Ultrapro™ Hernia System |
|------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name: | Surgical Mesh |
| Classification name: | Mesh, Surgical, Polymeric |
| 510(k) Number: | Panel number and product code: 79 FTL<br>Regulation number: 21 CFR 878.3300 |
| Manufacturer: | K071249<br>Ethicon |
| Reference Device: | Parietene™ Polypropylene Mesh (modified into Parietene™ Flat Sheet Mesh) |
| Trade/Proprietary name: | Surgical Mesh |
| Common name: | Mesh, Surgical, Polymeric |
| Classification name: | Panel number and product code: 79 FTL<br>Regulation number: 21 CFR 878.3300 |
| 510(k) Numbers: | K991400, K140941 |
| Manufacturer: | Sofradim Production |
| Reason for 510(k) Submission: | To obtain market clearance for the Duatene™ Bilayer Mesh. |
| Device Description: | Duatene™ Bilayer Mesh is designed to be placed in an extraperitoneal site by open approach. Duatene™ Bilayer Mesh is a three-dimensional device made out of polypropylene monofilament textile knitted in one-piece. The three-dimensional textile is composed of: <ul><li>a posterior textile layer which is circular or elliptic in shape (for preperitoneal placement)</li><li>an anterior textile layer which is oblong in shape (for onlay placement)</li></ul> The two textile layers are designed with differentiated knitting patterns, adapted to the function of each layer. The two layers are linked by crossing threads from both textiles. The mesh is designed to be placed over the groin region to ensure long term reinforcement of soft tissues. |
| Intended Use: | Duatene™ Bilayer Mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists. |
| Indications for use: | Duatene™ Bilayer Mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists. It is indicated for the open repair of groin hernia defects. |
| Summary comparing the technological characteristics of the subject and predicate device: | The subject Duatene™ Bilayer Mesh is substantially equivalent to the predicate device Prolene™ (Polypropylene) Hernia System (K984220) in terms of: <ul><li>Indications</li><li>Design: bilayer mesh linked with connector and equivalent in regards of shapes and sizes of the reference devices Parietene™ Macroporous Mesh (K142091) and Ultrapro™ Hernia System (K071249)</li></ul> |
{5}------------------------------------------------
- Polypropylene textile performance, including in regards of mechanical performance of the reference device Parietene™ Macroporous Mesh (K142091)
Duatene™ Bilayer Mesh has been evaluated and found compliant with ISO Standard 10993-1.
The following performance data is provided in support of the substantial equivalence determination:
- In vitro (bench) tests have been performed in accordance with . the "Guidance for the Preparation of a Premarket Notification Application of a Surgical Mesh" issued March 2. 1999 to evaluate the performance characteristics of the subject Duatene™ Bilayer Mesh in comparison with the predicate Prolene™ (Polypropylene) Hernia System (K984220).The following mesh characteristics were assessed:
- Pore size -
Materials:
Performance data:
- -Thickness
- -Surface density
- -Bursting strength
- -Bursting distension
- Breaking strength -
- -Elongation at break
- -Tear strength
- Suture pull-out strength -
The reference device Parietene™ Macroporous Mesh (K142091) is introduced in this submission to confirm that there are no issues associated with the mechanical differences identified in terms of bursting strength, bursting distension, breaking strength, elongation at break, tear strength and suture pull-out strength between the subject Duatene™ Bilayer Mesh and the predicate Prolene™ (Polypropylene) Hernia System (K984220) (refer to section 21 Performance Testing - Bench).
Connector dimensions and breaking strength were compared between the subject device Duatene™ Bilayer Mesh and the predicate device Prolene™ (Polypropylene) Hernia System (K984220) and found to be substantially equivalent.
- In vivo pre-clinical tests on representative animal model were conducted in comparison with the predicate Prolene™ (Polypropylene) Hernia System (K984220) to evaluate the tissue integration. Results demonstrate that no difference was observed between the subject Duatene™ Bilayer Mesh and the predicate Prolene™ (Polypropylene) Hernia System (K984220) in terms of tissue integration.
- Stability study was conducted and the proposed device shelf ●
{6}------------------------------------------------
life was demonstrated.
- Biocompatibility evaluation was performed and confirmed that . Duatene™ Bilayer Mesh and its constitutive components are compliant with ISO Standard 10993-1 for their intended patient contact profile. A chemical and toxicological assessment was performed. The biocompatibility data of the reference device Parietene™ Macroporous Mesh (K142091) has been leveraged for genotoxicity and systemic toxicity. Cytotoxicity, irritation, sensitization, acute systemic toxicity, pyrogen and implantation studies have been performed on the subject device.
- . Usability tests were conducted and demonstrate that the subject Duatene™ Bilayer Mesh is safe and effective for the intended users, uses and use environments.
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
### Conclusion:
All testing demonstrate that the subject Duatene™ Bilayer Mesh is substantially equivalent to the predicate device, Prolene™ (Polypropylene) Hernia System (K984220).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.