K172381 · Digitimer , Ltd. · IPF · Mar 6, 2018 · Physical Medicine
Device Facts
Record ID
K172381
Device Name
Digitimer DS7AP Constant Current Stimulator
Applicant
Digitimer , Ltd.
Product Code
IPF · Physical Medicine
Decision Date
Mar 6, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Digitimer DS7AP Constant Current Stimulator is indicated for use as an adjunct in the treatment of high and low anorectal malformations by helping in the identification of the striated muscles to be used in anal reconstructions.
Device Story
The DS7AP is a constant current stimulator used by surgeons or neurophysiological scientists in clinical settings (e.g., neurosurgery operating theaters) to identify striated muscles during posterior sagittal anorectoplasty (PSARP) or 'pull-through' procedures. The device delivers pulsed mono-phasic, rectangular electrical stimuli to tissue via a bipolar stimulation probe. The operator controls pulse duration (200µs or 500µs) and current intensity (0-100mA) via front-panel dials and switches. The device provides visual feedback via LEDs for power, trigger status, compliance limits, and fault conditions. By stimulating tissue and observing muscle response, the clinician locates the target muscles for reconstruction. The device is mains-powered and includes an automatic overload trip for safety. It benefits patients by facilitating precise muscle identification during complex anorectal surgeries.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical safety and performance testing per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10 standards.
Technological Characteristics
Mains-powered (115/230V) constant current stimulator. Output: pulsed mono-phasic rectangular waveform, 50Hz frequency, 0-100mA current, 200/500µs pulse width. Housing: UL94 V-0 flame retardant thermoplastic with metal panels. Connectivity: 4mm touch-proof sockets, 3.5mm jack for optional footswitch. Safety: Type BF applied part, automatic overload trip, compliance with 21 CFR 898. Dimensions: ~255x100x225mm. Weight: ~2.1kg.
Indications for Use
Indicated for use as an adjunct in the treatment of high and low anorectal malformations in pediatric patients (typically <3 years) to assist in the identification of striated muscles for anal reconstruction.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 6, 2018
Digitimer Ltd. John Smale Managing Director 37 Hydeway Welwyn Garden City, Hertfordshire, AL7 3BE, UK
Re: K172381
Trade/Device Name: Digitimer DS7AP Constant Current Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF Dated: January 3, 2018 Received: January 8, 2018
Dear John Smale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172381
Device Name
Digitimer DS7AP Constant Current Stimulator
Indications for Use (Describe)
The Digitimer DS7AP Constant Current Stimulator is indicated for use as an adjunct in the treatment of high and low anorectal malformations by helping in the identification of the striated muscles to be used in anal reconstructions.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter: | Digitimer Ltd |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address: | 37 Hydeway,<br>Welwyn Garden City,<br>Hertfordshire,<br>AL7 3BE, UK |
| Contact person: | John Smale<br>Managing Director<br>+44 (0)1707 328347 |
| Date Summary Prepared: | December 31, 2017 |
| 510(k) Number: | k172381 |
| Device Name: | DS7AP Constant Current Stimulator |
| Device Common Name | Powered Muscle Stimulator; CFR890.5850 |
| Device Classification: | Class II |
| Predicate Device: | Radionics Pena Muscle Stimulator PS-1 - 510(k) K980448 |
| Indications for Use: | The Digitimer DS7AP Constant Current Stimulator is indicated as an<br>adjunct in the treatment of high and low anorectal malformations by<br>helping the identification of striated muscle to be used in anal<br>reconstructions. |
| Device Description: | The DS7AP Constant Current Stimulator has been specially designed<br>to meet the requirements of pediatric surgeons who need a reliable<br>and easy to use muscle locating stimulator when carrying out the<br>Peña, "Pull-through" or posterior sagittal anorectoplasty (PSARP)<br>anorectal reconstruction technique. |
| Intended Patient Population | The reconstruction surgeries, for which the DS7AP is intended, are<br>typically carried out in young children (<3 years). |
| Intended Part of the Body to be Interacted With | The DS7AP is used in the treatment of anorectal malformations by<br>helping the identification of striated muscles to be used in anal<br>reconstructions. |
| Intended User Profile | The intended User is a specialist clinician trained in<br>Neurophysiological studies, or a surgeon trained in the specialist<br>reconstruction surgeries for which the DS7AP is intended. They are<br>likely to be either a doctor with neurology training or a Clinical<br>Neurophysiological Scientist for which the training requirement<br>recommendations (in the UK) is 4 years of undergraduate study. |
{4}------------------------------------------------
| | Intended Conditions of Use | | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | The DS7AP is intended for static table top use in a clinical | | |
| | environment, typically a neurosurgery operating theatre. | | |
| Electrical Performance: | Power: 100-120V or 200-240V (externally selected), 47-63Hz, 12VA.<br>Classification: IEC 60601-1 Class 1 with Type BF applied part.<br>Power: Push rod On/Off switch.<br>Pulse duration: LO (200µs) or HI (500µs), two position rotary switch.<br>Current: 0-100mA, continuously adjustable dial.<br>Compliance voltage: 100V.<br>Output frequency: 50Hz.<br>Output enable: Toggle Switch (On is Up, Off/Reset is down).<br>Output connections: 4mm shrouded, touch-proof sockets (red and<br>black) on 3/4" centers.<br>Footswitch connection rear panel 3.5mm mono jack socket (for<br>optional D185-FS1 contact closure footswitch). | | |
| Mechanical Performance: | Size: 10" x 4" x 10" (255 x 100 x 225 mm) - d x h x w<br>10.6" x 4.33" x 9" (270 x 110 x 225 mm) - d x h x w - over controls<br>and feet.<br>Weight: 4.5 pounds (2.1 kg) approx. | | |
| Temperature: | Operating Range 50°F to 104°F (10°C to 40°C) 30 to 75%, non-<br>condensing<br>Storage Range -40°F to 158°F (-40°C to 70°C) 10 to 100%, non-<br>condensing<br>Transport Range -40°F to 158°F (-40°C to 70°C) 10 to 100%, non-<br>condensing | | |
| Other: | Power: 115V or 230V @ 50/60 Hz<br>Rating: <30 VA<br>Case Material: UL94 V-0 Flame Retardant | | |
| Front Panel Indicators: | Power: Green LED, illuminated for power ON.<br>Trigger: Amber LED, illuminated to indicate stimulus trigger is active.<br>Compliance: Amber LED, illuminated if the compliance limit is<br>reached i.e. the current requested cannot be delivered (e.g. due to<br>high resistance between probe tips).<br>Reset/Fault: Amber LED illuminated and latched for sensed over<br>current and at power on. | | |
| Installation | The DS7AP does not need installation by Digitimer personnel as they<br>are supplied with comprehensive instructions and are easy to set up<br>& operate. | | |
{5}------------------------------------------------
Footswitch: The footswitch is a medical grade foot switch that is sealed to IP68 so as to meet all the requirements of IEC 60601-1 for use within operating theatres.
## Disposable Precision Bipolar Stimulation Probe:
Available in packs of 10 probes, the Technomed Neurosign Precision Bipolar Probe is an ideal single-use stimulation probe. The probes are fitted with 1.5mm DIN 42 802 connectors and need to be used in conjunction with our 2m long D185-HB4-2m Output Extension Cable The Precision Bipolar Probe has been cleared for marketing by the FDA under number k050325. This included an evaluation of biocompatibility
### Technological Characteristics:
| Characteristic: | Subject Device: DS7AP | Predicate Device: PS-1 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K172381 | K980448 |
| Intended Use | Used to identify striated muscle in<br>the treatment of high and low<br>anorectal malformations | Used to identify striated muscle in<br>the treatment of high and low<br>anorectal malformations |
| Indications for Use | The DS7AP is indicated for use of<br>as an adjunct in the treatment of<br>high and low anorectal<br>malformations by helping in<br>helping in identification of the<br>striated muscle to be used in anal<br>reconstructions | The PS-1 is indicated for use of as<br>an adjunct in the treatment of<br>high and low anorectal<br>malformations by helping in<br>helping in identification of the<br>striated muscle to be used in anal<br>reconstructions |
| Classification | Class II | Class II |
| Product Code | IPF | IPF |
| Regulation Number | CFR 890.5850 | CFR 890.5850 |
| Power Source | Supply Mains (115/230V; 50/60Hz) | Primary battery (PP3 / 1604) |
| Number of Output<br>Channels | One | One |
| Output Control | Constant Current | Constant Current |
| Firmware Control? | No | No |
| Automatic Overload<br>Trip? | Yes | No |
| Compliance with 21<br>CFR 898? | Yes | No |
| Weight | 4.85lb | 1.52lb |
| Dimensions (wxhxd) | 8.6"x4.4"x 9.9" | 3.75"x3.5"x6.25" |
| Housing Materials | Thermo-plastic with metal front<br>and rear panels | Thermo-plastic with metal front<br>and rear panels |
| Waveform | Pulsed mono-phasic | Pulsed mono-phasic |
| Waveform Shape | Rectangular | Rectangular |
| Maximum Output<br>Voltage1 | 56V @ 500Ω<br>126V @ 2kΩ | 122V @ 500Ω<br>220V @ 2kΩ |
{6}------------------------------------------------
| Characteristic: | Subject Device: DS7AP | Predicate Device: PS-1 |
|--------------------------------------------------------------------------------------------|--------------------------------------------|---------------------------------------------|
| Maximum Output<br>Current² | 112mA @ 500Ω<br>63mA @ 2kΩ<br>13mA @ 10k Ω | 244mA @ 500Ω<br>110mA @ 2kΩ<br>22mA @ 10k Ω |
| Pulse Width | 200µs / 500µs | 175µs |
| Pulse Frequency | 50Hz | 65Hz |
| Maximum Phase<br>Charge | 56µC @ 500Ω | 42.2µC @ 500Ω |
| Maximum Current<br>Density³ | 0.127mA/cm² @ 500Ω | 0.117mA/cm² @ 500Ω |
| Maximum Power<br>Density³ | 0.191W/cm² @ 500Ω | 0.161W/cm² @ 500Ω |
| Notes: | | |
| 1. Peak voltage, dependent upon the current control setting and the fixed load resistance. | | |
2. Peak current, limited by the maximum compliance voltage available.
3. Figures assume use of the G3604-00 Disposable Probe for both devices.
# Different Technological Characteristics:
The subject device and the predicate differ in the power source used, with the subject using mains power and the predicate using batteries.
The power supply of the subject device is identical to that of the Digitimer DS7A/AH Constant Current Stimulators, which are approved for marketing under number k051357.
Nonclinical Tests:
Nonclinical tests have been based on the compliance with the following international standards.
| Standard Utilized | FDA Consensus Standard | Comments |
|---------------------------------------------------------------------|------------------------------------------------------|----------------------------------------------------|
| EN 60601-1:2006 | AAMI ANSI ES60601-<br>1:2005/(R) 2012 and<br>A1:2012 | Differences<br>covered by<br>additional<br>testing |
| EN 60601-1-2:2007 | IEC 60601-1-2:2007 | Same |
| IEC 62366-1:2015 | AAMI ANSI IEC 62366-<br>1:2015 | Same |
| ISO 14971:2007 | ISO 14971:2007 | - |
| EN 60601-2-10:2015<br>+ A1:2016 | IEC 60601-2-10:2012 | - |
| FDA Guidance<br>Document for<br>Powered Muscle<br>Stimulator 510(k) | None | Comparison<br>of DS7AP and<br>predicate<br>device |
{7}------------------------------------------------
Clinical Tests:
No clinical tests are required for this submission
#### Conclusions of Nonclinical & Clinical Tests:
The DS7AP Constant Current Stimulator meets the requirements of the non-clinical testing defined above.
The intended use and the technological characteristics have also been compared in detail with the predicate device and found to be substantially equivalent.
No clinical tests were required to support this 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.