← Product Code [JWH](/productcode/JWH) · K172336

# Smith & Nephew Patient Matched Cutting Blocks (K172336)

_Smith & Nephew, Inc. · JWH · Jan 10, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172336

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Jan 10, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Indications for Use

Smith & Nephew's VISIONAIRE Patient Matched Cutting Blocks are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use: - Genesis II Knee System - Legion Knee System - Journey BCS Knee System - Journey II Knee System The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for single use only.

## Device Story

VISIONAIRE Patient Matched Cutting Blocks are patient-specific surgical instruments used during total knee replacement. The device utilizes patient imaging scans to identify anatomic landmarks for precise implant alignment and bone marking. This submission introduces a software component update for the design and manufacturing process and extends the device shelf life. The blocks are used intra-operatively by surgeons to guide bone cuts and component positioning. The system integrates with existing Smith & Nephew knee systems (Genesis II, Legion, Journey BCS, Journey II). The device is single-use. The software component facilitates the translation of patient-specific imaging data into the physical design of the cutting blocks, ensuring alignment with the patient's unique anatomy. This improves surgical accuracy and implant positioning, potentially enhancing patient outcomes.

## Clinical Evidence

No clinical data was needed or provided to support the safety and effectiveness of the subject device. Evidence consists of software verification and validation testing performed in accordance with FDA guidance.

## Technological Characteristics

Patient-specific surgical instrumentation. No changes to materials, design, or manufacturing processes. Software-driven design process. Single-use. Connectivity/integration with Smith & Nephew knee systems.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Patient Matched Cutting Blocks ([K082358](/device/K082358.md))
- Patient Matched Cutting Blocks ([K130708](/device/K130708.md))
- Patient Matched Cutting Blocks ([K170282](/device/K170282.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 10, 2018

Smith & Nephew, Inc. Brad Sheals Principal Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova, Tennessee 38016

Re: K172336

Trade/Device Name: Smith & Nephew Patient Matched Cutting Blocks Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH, OOG Dated: July 26, 2017 Received: August 2, 2017

Dear Brad Sheals:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES         | Form Approved: OMB No. 0910-0120                                                                                                                                                                                                                                                                                                                                                                         |                          |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Food and Drug Administration                    | Expiration Date: January 31, 2017                                                                                                                                                                                                                                                                                                                                                                        |                          |
| Indications for Use                             |                                                                                                                                                                                                                                                                                                                                                                                                          | See PRA Statement below. |
| 510(k) Number (if known)                        | K172336                                                                                                                                                                                                                                                                                                                                                                                                  |                          |
| Device Name                                     | Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks                                                                                                                                                                                                                                                                                                                                                 |                          |
| Indications for Use (Describe)                  | Smith & Nephew's VISIONAIRE Patient Matched Cutting Blocks are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. |                          |
|                                                 | The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use:                                                                                                                                                                                                                   |                          |
|                                                 | - Genesis II Knee System                                                                                                                                                                                                                                                                                                                                                                                 |                          |
|                                                 | - Legion Knee System                                                                                                                                                                                                                                                                                                                                                                                     |                          |
|                                                 | - Journey BCS Knee System                                                                                                                                                                                                                                                                                                                                                                                |                          |
|                                                 | - Journey II Knee System                                                                                                                                                                                                                                                                                                                                                                                 |                          |
|                                                 | The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for single use only.                                                                                                                                                                                                                                                                                                           |                          |
| Type of Use (Select one or both, as applicable) | <span style="text-decoration: overline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span>                                                                                                                                                                                                                                                    |                          |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

|  | Department of Health and Human Services |
|--|-----------------------------------------|
|  | Food and Drug Administration            |
|  | Office of Chief Information Officer     |
|  | Paperwork Reduction Act (PRA) Staff     |
|  | PRAStaff@fda.hhs.gov                    |

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# ~We are smith&nephew

| Submitted by:                                | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116               |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Date of Summary:                             | July 26, 2017                                                                                                   |
| Contact Person and Address:                  | Brad Sheals<br>Principal Regulatory Affairs Specialist<br>T 901-399-6897<br>F 901-566-7911                      |
| Name of Device:                              | Smith & Nephew VISIONAIRE Patient Matched Cutting<br>Blocks                                                     |
| Common Name:                                 | Knee Prosthesis                                                                                                 |
| Device Classification Name and<br>Reference: | 21 CFR 888.3560- Knee joint patellofemorotibial<br>polymer/metal/polymer non-constrained cemented<br>prosthesis |
| Device Class:                                | Class II                                                                                                        |
| Panel Code:                                  | Orthopaedics/87                                                                                                 |
| Product Code:                                | JWH, MBH, OOG                                                                                                   |

#### Device Description

The subject of this premarket notification is to seek FDA clearance of a software component to be used in the design and manufacture of the VISIONAIRE Patient Matched Cutting Blocks. Also within this notification is an extension of the subject device's shelf life. Patient Matched Cutting Blocks were previously cleared for market via premarket notifications- K082358, K130708 and K170282.

#### Intended Use

Smith & Nephew's VISIONAIRE Patient Matched Cutting Blocks are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use:

- Genesis II Knee System
- Legion Knee System
- Journey BCS Knee System
- Journey II Knee System

The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for single use only.

{4}------------------------------------------------

### Technological Characteristics

No new mechanical testing was performed. No implants or new blocks are being introduced in this premarket notification. There are no changes to the block design, packaging, material composition or manufacturing of Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks as a result of these chanqes described in the premarket notification. Clinical data was not needed to support the safety and effectiveness of the subject device(s).

Software verification and validation testing was completed in line with FDA's guidance document entitled, "General Principles of Software Validation; Final Guidance for Industry and FDA Staff," dated January 11, 2002. A review of this testing has demonstrated that there are no new issues related to the safety and effectiveness of the subject device and the software will perform as intended.

Based on the documentation within this premarket notification, there are no new issues related to the safety and effectiveness of the subject device. Clinical data was not needed to support the safety and effectiveness of the subject device.

# Substantial Equivalence Information

The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are identical in function, equivalent design features, intended use, indications for use, operational principles, manufacturing processes, and materials as the predicate device- Patient Matched Cutting Blocks (K082358, 11/25/2008, K130708, S.E. 02/28/2014 and K170282, S.E. 05/22/2017).

## Conclusion

This premarket notification is being submitted to request clearance of the subject Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks. Based on the similarities to the predicate cutting blocks and a review of the testing performed, the subject device is substantially equivalent to the predicate device- Patient Matched Cutting Blocks (K082358, 11/25/2008; K130708, S.E. 02/28/2014 and K170282, S.E. 05/22/2017).

---

**Source:** [https://fda-staging.innolitics.com/device/K172336](https://fda-staging.innolitics.com/device/K172336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com