The Patient Status Engine is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This includes heart rate, heart rate variability (R-R interval), ECG derived respiration rate data (EDR), skin temperature, activity, posture and optional SpO2 and non-invasive Blood Pressure (BP). Data is transmitted wirelessly from the sensors to the Patient Gateways and from the Patient Gateways to a central Server where it is stored for analysis. The Patient Status Engine can include the ability to notify healthcare professionals when physiological data fall outside selected parameters. The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information. The data from the Patient Status Engine is intended for use by healthcare professionals as an aid to diagnosis and treatment. It is not intended for use on critical care patients nor replace standard monitoring and/or routine care.
Device Story
Wireless remote monitoring system; collects physiological data (HR, HRV, EDR, skin temp, activity, posture, optional SpO2/BP) via wearable sensors (Lifetouch Blue, Lifetemp). Data transmits to Patient Gateway (Samsung tablet) then to central Lifeguard Server for storage/analysis. Used in home/healthcare settings by healthcare professionals. Provides physiological information as aid to diagnosis/treatment; notifies clinicians of out-of-parameter data. Benefits patients through continuous monitoring outside critical care environments.
Clinical Evidence
No clinical data; bench testing only. Compliance with BS EN 60601-1, 60601-1-2, 60601-1-11, 60601-2-27, ISO 10993-1, ISO 14971, IEC 62304, and IEC 62366-1.
Technological Characteristics
Wireless physiological monitoring system. Components: Lifetouch Blue sensor, Lifetemp sensor, Samsung tablet Patient Gateway, Lifeguard Server. Connectivity: Wireless sensor-to-gateway, gateway-to-server. Standards: BS EN 60601-1, 60601-1-2, 60601-1-11, 60601-2-27, ISO 10993-1, ISO 14971, IEC 62304, IEC 62366-1. Battery-operated sensors; single-use.
Indications for Use
Indicated for general care patients 18+ years old in home and healthcare settings for continuous physiological monitoring (HR, HRV, EDR, skin temp, activity, posture, optional SpO2/BP). Not for critical care patients.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 23, 2018
Isansys Lifecare Ltd % André Kindsvater Senior Consultant RA & OA Emergo Global Consulting, LLC 2500 Bee Cave Road Austin, Texas 78746
Re: K172329
Trade/Device Name: Patient Status Engine Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II Product Code: DRG Dated: March 26, 2018 Received: March 27, 2018
Dear André Kindsvater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - André Kindsvater
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Wilhelm
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172329
Device Name Patient Status Engine
#### Indications for Use (Describe)
The Patient Status Engine is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This includes heart rate variability (R-R interval), ECG derived respiration rate data (EDR), skin temperature, activity, posture and optional SpO2 and noninvasive Blood Pressure (BP). Data is transmitted wirelessly from the Patient Gateways and from the Patient Gateways to a central Server where it is stored for analysis. The Patient Status Engine can include the ability to notify healthcare professionals when physiological data fall outside selected parameters.
The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information.
The data from the Patient Status Engine is intended for use by healthcare professionals as an aid to diagnosis and treatment. It is not intended for use on critical care patients nor replace standard monitoring and/or routine care.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Patient Status Engine
K 172329
#### 1. Submission Sponsor
Isansys Lifecare Ltd 8C Park Square Milton Park Abingdon Oxfordshire OX14 4RR United Kingdom Phone number: +44 (0) 1235 436225 Contact: Julian Jones Title: Chief Compliance Officer
## 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746 Cell Phone: +41 78 843 7077 Office Phone: (512) 327 9997 Fax: (512) 327 9998 Contact: Andre Kindsvater, Senior Consultant, Regulatory Email: project.management@emergogroup.com
#### 3. Date Prepared
July 28, 2017
#### 4. Device Identification
| Trade/Proprietary Name: | Patient Status Engine |
|-------------------------|------------------------------------------------------------------|
| Common/Usual Name: | Wireless Remote Monitoring System |
| Classification Name: | Transmitters and Receivers, Physiological Signal, Radiofrequency |
| Regulation Number: | 870.2910 |
| Product Code: | DRG |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
#### 5. Legally Marketed Predicate Device
K152139 - Vital Connect Platform/VitalPatch/HealthPatch MD by Vital Connect Inc.
{4}------------------------------------------------
### 6. Device Description
The Patient Status Engine (PSE2) is a wireless data collection system that monitors physiological data and consists of the following components / subsystems and optional third party devices as listed below:
| Device Name: | Patient Status Engine |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Components and Subsystems: | Patient Gateway (Samsung Tablet) – Manufactured by Samsung<br>Patient Gateway Software – Manufactured by Isansys Lifecare Ltd<br>Lifetouch Blue Sensor – Manufactured by Isansys Lifecare Ltd<br>Lifetemp Sensor– Manufactured by Isansys Lifecare Ltd<br>Lifeguard Server Software – Manufactured by Isansys Lifecare Ltd |
| Additional third party Accessories: | Pulse Oximeter 3150– Manufactured by Nonin<br>Blood Pressure Monitor UA767– Manufactured by A&D Medical<br>Skintact ECG Electrodes – Manufactured by Leonhard Lang<br>Ambu White Sensors – Manufactured by Ambu Inc. |
At the subsystem level the Patient Gateway, Lifetouch Blue Sensor and Lifetemp Sensor are collectively identified as the Patient Digitisation Engine (PDE2). The inclusion of the Lifeguard Server completes the Patient Status Engine (PSE2).
#### 7. Indication for Use Statement
The Patient Status Engine is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This includes heart rate, heart rate variability (R-R interval), ECG derived respiration rate data (EDR), skin temperature, activity, posture and optional SpO2 and non-invasive Blood Pressure (BP). Data is transmitted wirelessly from the sensors to the Patient Gateways and from the Patient Gateways to a central Server where it is stored for analysis. The Patient Status Engine can include the ability to notify healthcare professionals when physiological data fall outside selected parameters.
The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information.
The data from the Patient Status Engine is intended for use by healthcare professionals as an aid to diagnosis and treatment. It is not intended for use on critical care patients nor replace standard monitoring and/or routine care.
#### 8. Substantial Equivalence Discussion
The following table compares the Patient Status Engine to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| Manufacturer | Isansys Lifecare Ltd | Vital Connect Inc. | Variances / Equivalence |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Patient Status Engine | Vital Connect Platform,<br>Healthpatch MD,<br>VitalPatch | |
| 510(k) Number | | K152139 | N/A |
| Product Code | DRG | DRG | Same |
| Manufacturer | Isansys Lifecare Ltd | Vital Connect Inc. | Variances / Equivalence |
| Trade Name | Patient Status Engine | Vital Connect Platform,<br>Healthpatch MD,<br>VitalPatch | |
| Regulation Number<br>and Regulation<br>Name | 870.2910<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | 870.2910<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Same |
| Indications for Use | The Patient Status Engine<br>is a wireless remote<br>monitoring system<br>intended for use by<br>healthcare professionals<br>for continuous collection<br>of physiological data in<br>home and healthcare<br>settings.<br>This includes heart rate,<br>heart rate variability (R-R<br>interval), ECG derived<br>respiration rate data<br>(EDR), skin temperature,<br>activity, posture and<br>optional SpO2 and non-<br>invasive Blood Pressure<br>(BP). | The Vital Connect Platform<br>is a wireless remote<br>monitoring system<br>intended for use by<br>healthcare professionals<br>for continuous collection of<br>physiological data in home<br>and healthcare settings.<br>This can include heart rate,<br>electrocardiography (ECG),<br>heart rate variability, R-R<br>interval, respiratory rate,<br>skin temperature, activity<br>(including step count), and<br>posture (body position<br>relative to gravity including<br>fall). | Equivalent - the<br>differences not affecting<br>safety or effectiveness<br><br>The Patient Status<br>Engine does not have<br>electrocardiography<br>ECG.<br>The Patient Status<br>Engine does include<br>optional SpO2 and non-<br>invasive Blood Pressure,<br>supplied as optional<br>cleared third party<br>accessories. |
| | Data is transmitted<br>wirelessly from the<br>sensors to the Patient<br>Gateways and from the<br>Patient Gateways to a<br>central Server where it is<br>stored for analysis | Data are transmitted<br>wirelessly from the Vital<br>Connect Sensor for storage<br>and analysis. | Equivalent |
| | The Patient Status Engine<br>can include the ability to<br>notify healthcare<br>professionals when<br>physiological data fall<br>outside selected<br>parameters | The Vital Connect Platform<br>can include the ability to<br>notify healthcare<br>professionals when<br>physiological data fall<br>outside selected<br>parameters | Equivalent |
| | The device is intended for<br>use on general care<br>patients who are 18 years<br>of age or older as a<br>general patient monitor, | The device is intended for<br>use on general care<br>patients who are 18 years<br>of age or older as a general<br>patient monitor, to provide<br>physiological information. | Equivalent |
| Manufacturer | Isansys Lifecare Ltd | Vital Connect Inc. | Variances / Equivalence |
| Trade Name | Patient Status Engine | Vital Connect Platform,<br>Healthpatch MD,<br>VitalPatch | |
| | The data from the Patient<br>Status Engine is intended<br>for use by healthcare<br>professionals as an aid to<br>diagnosis and treatment.<br>It is not intended for use<br>on critical care patients<br>nor replace standard<br>monitoring and/or<br>routine care. | The data from the Vital<br>Connect Platform are<br>intended for use by<br>healthcare professionals as<br>an aid to diagnosis and<br>treatment. The device is<br>not intended for use on<br>critical care patients. | Equivalent |
| Mechanism of Action | Physiological data<br>acquisition (wireless) and<br>display | Physiological data<br>acquisition (wireless) and<br>display | Equivalent |
| Technology<br>Overview | PEMS and Software,<br>wireless communication | PEMS and Software,<br>wireless communication | Equivalent |
| Touchscreen<br>with<br>Graphical<br>User<br>Interface | Touchscreen to operate<br>Graphical User Interface<br>(GUI) on the Patient<br>Gateway. | The Software Library is run<br>in third party devices.<br>No other information<br>publicly available. | Equivalent - the<br>differences does not<br>affect safety or<br>effectiveness, as<br>determined by<br>verification and<br>validation testing |
| Ambulatory<br>monitoring sensors | Yes | Yes | Equivalent |
| Heart Rate, | Yes | Yes | Equivalent |
| R-R interval | Yes | Yes | Equivalent - the<br>differences do not<br>affect safety or<br>effectiveness, as<br>demonstrated by V&V<br>testing. |
| Manufacturer | Isansys Lifecare Ltd | Vital Connect Inc. | Variances / Equivalence |
| Trade Name | Patient Status Engine | Vital Connect Platform,<br>Healthpatch MD,<br>VitalPatch | |
| Electrocardiography<br>(ECG) | No | Yes | Different - The Patient<br>Status Engine does not<br>include ECG in the<br>indications for use and<br>make no claims of such.<br>This does not raise new<br>questions of safety and<br>effectiveness for its<br>intended use. |
| Respiration Rate, | Yes | Yes | Equivalent - the<br>differences do not<br>affect safety or<br>effectiveness, as<br>demonstrated by V&V<br>testing. |
| Thermistor Probe to<br>detect skin<br>Temperature | Yes | Yes | Equivalent - the<br>differences do not<br>affect safety or<br>effectiveness, as<br>demonstrated by V&V<br>testing. |
| Skin Temperature | Yes | Yes | Equivalent - the<br>differences not affecting<br>safety and<br>effectiveness. |
| Activity | Yes | Yes | Equivalent |
| Posture | Yes | Yes | Equivalent |
| SpO2 | Yes | No | Different - the Patient<br>Status Engine may be<br>used with an FDA<br>cleared SpO2 oximeter.<br>This is an addition to<br>the indications for use<br>as compared to the<br>predicate, but this has<br>been demonstrated as<br>safe and effective<br>through V&V testing. |
| Manufacturer | Isansys Lifecare Ltd | Vital Connect Inc. | Variances / Equivalence |
| Trade Name | Patient Status Engine | Vital Connect Platform,<br>Healthpatch MD,<br>VitalPatch | |
| Non-invasive Blood<br>Pressure (BP) | Yes | No | Different - the Patient<br>Status Engine may be<br>used with an FDA<br>cleared BP monitor.<br>This is an addition to<br>the indications for use<br>as compared to the<br>predicate, but this has<br>been demonstrated as<br>safe and effectives<br>through V&V testing. |
| Sterile | No | No | Equivalent |
| Single-Use | Yes | Yes | Equivalent |
| Battery Operated | Yes | Yes | Equivalent - the<br>differences not affecting<br>safety and<br>effectiveness. |
| AC Powered | No | No | Equivalent |
| Central Server | Yes | Yes | Equivalent - the<br>difference does not<br>affect safety or<br>effectiveness.. Both<br>servers offer a secure<br>means of managing the<br>upload, processing,<br>storage and display of<br>sensor derived data. |
| Complies with BS EN<br>ISO 10993-1 | Yes | Yes | Equivalent. |
| Electrical Safety<br>Testing | Yes | Yes | Equivalent. |
#### Table 5A - Comparison of Characteristics
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the Patient Status Engine and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Isansys Lifecare Ltd completed a number of non-clinical performance tests against applicable standards.
| Table 3B - Performance Standards Testing Summary | | | |
|--------------------------------------------------|----------------------------------|-------------------------------------------|---------|
| | Test | Pass / fail criteria | Results |
| 1 | Electrical safety | Compliance to BS EN 60601-1:2006+A12:2014 | Passed |
| 2 | Electromagnetic<br>compatibility | Compliance to BS EN 60601-1-2: 2007 | Passed |
Table 5B – Performance Standards Testing Summary
{9}------------------------------------------------
| Test | | Pass / fail criteria | Results |
|------|----------------------------------------------|----------------------------------------|---------|
| | | | |
| 3 | Homecare environment | Compliance to BS EN 60601-1-11:2010 | Passed |
| 4 | Electrocardiographic<br>monitoring equipment | Compliance to BS EN 60601-2-27:2014 | Passed |
| 5 | Biocompatibility | Compliance to BS EN ISO 10993-1:2009 | Passed |
| 6 | Risk Management | Compliance to BS EN ISO 14971:2012 | Passed |
| 7 | Software | Compliance to BS EN 62304:2006+A1:2015 | Passed |
| 8 | Usability | Compliance to BS EN 62366-1:2015 | Passed |
The Patient Status Engine passed all the testing in accordance with internal requirements, national standards, and international standards to support substantial equivalence of the subject device:
#### 10. Clinical Performance Data
There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics, but can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device.
lt has been shown through the documents provided in this 510(k) submission that the minor differences between the Patient Status Engine and the predicate devices Vital Connect Platform/Healthpatch MD/VitalPatch do not raise any new questions regarding its safety and performance.
The Patient Status Engine has the same intended use and the same or similar indications for use and technological characteristics as the previously cleared predicate device, Vital Connect Platform/Healthpatch MD/VitalPatch.
The Patient Status Engine as designed and manufactured, is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.